Shantou Institute of Ultrasonic Instruments Co., Ltd.
Note
This operation manual shall apply to the CTS-49/CTS-59 ultrasonic thickness gauge. The major differences of the two models on functions and specifications are listed below:
I
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Model CTS-49
Item Dual-element
delay
CTS-59
type
(with
Dual-element
delay
probe cable):
type(with probe cable):
TG5-10L (standard configuration)
TG5-10L
TG7.5-6L
configuration)
TG5-10HL
TG7.5-6L
TG2-12L
TG5-10HL
TG5-6L
TG2-12L
TG5-8L
TG5-6L
(standard
TG5-8L Single-element delay type :
Compatible Probes
TG5-6DL
Single-element
TG7.5-6DL
type :
TG10-6DL
TG5-6DL
TG7.5-3DMD
TG7.5-6DL
TG10-3DMD
TG10-6DL
TG15-3DMD
TG7.5-3DMD
delay
TG10-3DMD Single-element contact type:
TG15-3DMD
TG0.5-25NL TG0.5-30NL
Single-element contact
TG1-20NL
type:
TG2.5-20NL
TG0.5-25NL TG0.5-30NL TG1-20NL TG2.5-20NL
Dialog probe series are supported and can be automatically identified.
II
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Standard (R-B1, transmit pulse to
Standard
the first echo);
transmit pulse to the
Thru-coating
measurement
EchoToEchoMode
or
measurement
(B1-B2, or Bm-Bn)
(R-B1,
first echo); Thru-coating measurement
or
EchoToEchoMode
Measurement mode
measurement(B1-B2, or Bm-Bn); If velocity of coating is known, the thickness of coating
can
be
measured. User-defined V-path
√
TDG functio n
√
Temperature
√
compensation Multilayer
√
measurement
√
Time-based B-scan
Store up to 20,000 sets
Storage capacity
Store up to 10,000 sets of
of measurement data
measurement data (including
(including
measurement value, parameter
measurement value,
setup and waveform data)
parameter setup and waveform data)
Standard configured
√
external fast charger CD-B010
【Note】: The mark √ in the table above indicates the corresponding model has such function or specification.
The complication of this operation manual is based on the model CTS-49, and the parameters and the operation shall apply to the CTS-49. As the operation of the CTS-59 is compatible with the CTS-49, for additional functions and operation on the CTS-59, see Appendix D.
III
Shantou Institute of Ultrasonic Instruments Co., Ltd.
DCY2.781.49/59 SS/V1.2/A-E
TAB LE OF CONTENTS
PREFACE
..........................................................................................................................................VII
OPERATION CAUTIONS AND SAFETY GUIDE ................................................................................. IX CHAPTER 1
SYSTEM .............................................................................................................................1
1.1
APPEARANCE ..................................................................................................................................1
1.2
PORT AND SOCKET .........................................................................................................................3
1.3
KEY ..................................................................................................................................................4
1.4
MENU............................................................................................................................................... 4
1.5
SCREEN INTERFACE ......................................................................................................................8
1.6
W AVEFORM DISPLAY AREA ...........................................................................................................10
1.7
MEASUREMENT D ATA AREA ........................................................................................................10
1.8
SYMBOL AND INFORMATION PROMPT AREA .................................................................................10
CHAPTER 2
SYSTEM OPERATION ...................................................................................................12
2.1
POWER SUPPLY ............................................................................................................................ 12
2.2
POWER ON .................................................................................................................................... 12
2.3
POWER OFF .................................................................................................................................. 13
2.4
OPERATION INTERFACE SELECTION............................................................................................. 13
2.5
M AIN MENU SELECTION ............................................................................................................... 13
2.6
SUBMENU SELECTION .................................................................................................................. 13
2.7
TURNING P AGE.............................................................................................................................. 14
2.8
SUBMENU ADJUSTMENT ............................................................................................................... 14
2.9
SHORTCUT OPERATION ................................................................................................................ 14
2.10
C ALIBRATION FUNCTION ............................................................................................................... 15
2.11 AUTOSEARCH AND FLOATINGG AIN FUNCTIONS .......................................................................... 16 2.12
WORKMODE .................................................................................................................................. 17
2.13 VELOCITY MEASUREMENT ............................................................................................................. 20 2.14 PARAM STORAGE......................................................................................................................... 20 2.14.1
Save....................................................................................................................................... 20
2.14.2
Recall..................................................................................................................................... 21
2.14.3
Delete .................................................................................................................................... 21
2.15
DATA STORAGE ............................................................................................................................ 21
2.15.1
Save....................................................................................................................................... 22
2.15.2
Recall..................................................................................................................................... 22
2.15.3
Delete .................................................................................................................................... 23
CHAPTER 3 3.1
APPL ICATION OF THICKNESS MEASUREMENT ..................................................24
GENERAL WORK FLOW................................................................................................................. 24
3.2 APPLICATION KEYNOTE OF THICKNESS MEASUREMENT ................................................................ 25 CHAPTER 4
PERIPHERAL DEVICES, PORTS AND ACCESSORIES ........................................26 IV
Shantou Institute of Ultrasonic Instruments Co., Ltd.
4.1
PROBE SOCKET ............................................................................................................................ 26
4.2
MINIUSB PORT ............................................................................................................................. 26
4.2.1
Connect to a PC.................................................................................................................. 26
4.2.2
Disconnect from a PC ........................................................................................................ 26
4.3
MICROSD C ARD HOLDER ............................................................................................................ 26
4.3.1
Insert microSD card ............................................................................................................ 26
4.3.2
Remove microSD card ....................................................................................................... 27
4.4
B ATTERY........................................................................................................................................ 27
4.4.1
Battery Installation and Removal...................................................................................... 27
4.4.2
Charging the Battery .......................................................................................................... 27
4.4.3
Using the Battery ................................................................................................................ 28
4.4.4
Battery Maintenance .......................................................................................................... 28
4.4.5
Airlift Requirements for battery ......................................................................................... 29
4.4.6
Cautions on using the battery: .......................................................................................... 29
4.5
POWER ADAPTER.......................................................................................................................... 30
4.6
F AST CHARGER CD-B010 (OPTION) ........................................................................................... 30
CHAPTER 5
TRAINING, SYSTEM MAINTENANCE AND SERVICE ............................................32
4.7
TRAINING ........................................................................................................................................32
4.8
SYSTEM M AINTENANCE .................................................................................................................32
4.9
SERVICE .........................................................................................................................................32
APPENDIX A
COMMON FAILURES AND TROUBLESHOOTING .................................................34
A.1
THE SYSTEM CANNOT POWER ON ................................................................................................. 34
A.2
THE SYSTEM CANNOT POWER OFF ................................................................................................ 34
A.3
C ANNOT CHARGE ........................................................................................................................... 34
A.4
NO ECHO ........................................................................................................................................ 34
A.5
ABNORMAL ERROR OF THICKNESS MEASUREMENT READING....................................................... 35
APPENDIX B
SPECIFICATIONS...........................................................................................................36
APPENDIX C SUBMENU DEFAULT VALUE AND RANGE OR OPTION ......................................39 APPENDIX D AVAIL AB LE PROB ES SPECIFICATION ....................................................................48 APPENDIX E AB OUT CTS-59 ...............................................................................................................52 E.1 APPEARANCE .................................................................................................................................... 52 E.2 MENU STRUCTURE ........................................................................................................................... 52 E.3 CTS-59 SYMBOL AND INFORMATION PROMPT AREA ...................................................................... 57 E.4 FUNCTIONAL OPERATION OF ADDITIONAL M AIN MENU ................................................................... 58 E.4.1 PROBE ....................................................................................................................................... 58 E.4.2 V-PATH........................................................................................................................................ 59 E.4.3 TDG ............................................................................................................................................. 61 E.4.4 TEMP .......................................................................................................................................... 62 E.4.5 COAT........................................................................................................................................... 62 E.4.6 B-SCAN ...................................................................................................................................... 63 E.4.7 MULTI.......................................................................................................................................... 68 V
Shantou Institute of Ultrasonic Instruments Co., Ltd.
APPENDIX F ACOUSTIC VEL OCITY OF MATERIAL S ...................................................................71 CONTACT INFORMATION........................................................................................................................71
VI
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Preface The CTS-49 digital ultrasonic thickness gauge is the perfect combination of various advanced high technologies such as ultrasound, computer, electronics, molding and processes. This palmheld thickness gauge is suitable for thickness or acoustic velocity measurement of metal and many other materials. It supports dual and single element probes for erosive and wall-thickness measurement. Outstanding measurement features, such as A-scan display, high sensitivity, precise measurement and through-coating thickness measurement, are available on the system. More features such as multi-language menus, large data storage, computer communication, compact design, as well as left & right handed operation, can be found on the unit. It is suitable for ultrasound thickness measurement in machinery, chemistry, ship building, aeronautical and aerospace industries. Main features of the CTS-49 digital ultrasonic thickness gauge: 1) Small size —— The dimension of th e whole unit is 105mm × 180 mm × 42mm (W×H×D). 2) 5-inch 800×480 pixels high-definition color TFT-LCD monitor, high-brightness LED backlight (visible under sunlight). 3) Light weight —— The total weight of the unit (including battery) is 0.6kg only. 4) Low consumption ——a fully-recharged battery can support the system to work continuously over 6 hours. 5) High performance ——It supports dual and single element probes, A-scan display, thrucoating thickness measurement, auto-search measurement, velocity dynamic measurement, and single/dual point correction functions. 6) Rich functions —— The Landscape/ Portrait screen display good for left/right handed operation; microSD card holder and minUSB slave port for powerful data storage, transmission and post-processing. 7) Easy operation —— Concise buttons, fast button switch between standard/ simple menu interface, user-defined buttons for flexible setting, friendly interface, easy to read. This Operation Manual contains the following chapters: VII
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Chapter 1 gives a brief introduction to the system. Chapter 2 describes system menus and functional operation. Chapter 3 describes general workflow and main points of system application, so as to facilitate the users to operate the system quickly and conveniently. Chapter 4 describes how to use peripheral devices, ports and accessories. Chapter 5 describes operator training and system maintenance. Appendix A describes how to identify system common failures and troubleshoot the problems. Appendix B describes system main performance specifications. Appendix C describes system submenus. Appendix D describes available probe specification. Appendix E describes differences between CTS-59 and CTS-49, and corresponding operation methods. Appendix F lists common material velocity.
To know system features and operation, as well as how to perform ultrasonic thickness measurement with the system, please read this Operation Manual carefully before using the system.
VIII
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Operation Cautions and Safety Guide 1) Please keep this Operation Manual and always operate the system by following the steps and methods described in the Operation Manual. 2) Before operating the system, the operator shall have received formal training on ultrasonic testing. 3) Do not try to dismantle and repair the system unless specifically described in the Operation Manual. 4) The system warranty will be void if the fragile labels on specific positions are damaged. 5) Do not place the system in the place of moisture or near radiator, strong magnetic field or heat source. 6) When the system is not in use for a long time, it should be powered up at lease once every month. 7) Avoid collide any hard object against the display screen. 8) Do not use a damaged or worn-out power cable. 9) For cautions on using the battery, please refer to Chapter 4.
IX
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Chapter 1 System 1.1 Appearance
Direction and confirmation key
Power on/off /
Standard/ simple
Shift/ step key
interface switch key
User-defined
Power/ charging
function key 1/2
indicator
Fig. 1-1
System Front View
1
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Battery fixing clip
Battery
Zero calibration block
Handle
Fig. 1-2
System Back View
Handle
Fig. 1-3
System Side View 2
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Probe Socket T Probe Socket R
Fig. 1-4
System Top View
miniUSB port (slave)
I/O port (reserved)
DC Power In Socket
microSD card holder
Fig. 1-5
System Bottom View
1.2 Port and Sock et Table 1-1
Usage of Ports and Sockets
Name
Usage Compatible with LEMO00, for connecting to probes. T for
Probe Socket T, R transmitting while R for receiving. microSD card holder miniUSB port
To be inserted with a microSD card, for storing measurement data sets. Slave device, for connecting to a PC For main unit power supply when connected to the external
DC Power In Socket
power adapter (CD-92A), or charging the battery (DC-A004) installed on the main unit.
I/O port
Reserved
3
Shantou Institute of Ultrasonic Instruments Co., Ltd.
1.3 Key Table 1-2 Key
,
,
Key Function Description
Key Name ,
Function To select menu, increase/ decrease value and switch
Direction option. Special function in simple menu interface. To enter or exit a submenu, execute a function or Confirm confirm an item. User-defined function key 1, which can be defined as either AutoSearch (default) or WorkMode. User-defined User-defined function key 2, which can be defined as Freeze/MeasVEL/SaveMeasData Switch key between standard/ simple interface, for fast
Function
switch between standard/ simple menu. Power on/off: Press and hold it briefly for about one
Power/
second to turn on/off the system.
Function
Shift/ step key: To run coarse and fine adjustment/ shift/ step function subject to the operation. Pressing the two keys at the same time to switch to the
+
Combination waveform frozen/ unfrozen state in testing.
1.4 Menu Please refer to Table 1-3 for the menu structure. For detailed menu description, see Appendix C. Table 1-3 Menu Structure Main Menu
Submenu
CAL
ProbeType WorkMode ProbeZero^ Vel^ CalOption Ref1 4
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Submenu Ref2 CAL GAStart dB △
Table 1-3 Menu Structure (continued) Main Menu
Submenu Range^ DisplayDelay^ dB △ AutoSearch MaxSearchGain FloatingGain MeasMode
MEAS
DiffRef AVGNum MeasVEL THKValueRef^ LastReading
Table 1-3 Menu Structure (continued) Main Menu
Submenu GAStart GAWidth GAThresh
GATE
GBStart GBWidth GBThresh
5
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Table 1-3 Menu Structure (continued) Main Menu
Submenu MeasCount Rectify ScreenMode F1-Function F2-Function Brightness Color Precision
CONF
Language Unit Date Time AutoPowerOff Default Update About
Table 1-3 Menu Structure (continued) Main Menu
Submenu StoreNum FileName Save Recall
PARAM
Delete DeleteAll ToSDCard AllToSDCard StoreCapability
Table 1-3 Menu Structure (continued) Main Menu
Submenu
DATA
FileNum FileName FileFormat Line Column Point
6
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Submenu Save Recall Delete DeleteFile DeleteAllFiles StoreLocation StoreCapability
Table 1-3 Menu Structure (continued) Main Menu
Submenu MaxAlarmSW MinAlarmSW
ALARM
MaxAlarm MinAlarm BuzzerSW
【Note 1】: The submenu marked with ^ has coarse adjustment/ fine adjustment. The submenu
marked with ^ on the right indicates it is in fine adjustment, while the submenu with mark in coarse adjustment.
【Note 2】: The submenu marked with △ has step adjustment, and the number on the right of
the submenu with △ means the step value.
7
Shantou Institute of Ultrasonic Instruments Co., Ltd.
1.5 Screen Interface Portrait screen standard operation:
Symbol and information prompt area
Waveform display area
Measurement data area Submenu display area
Main menu display area
Portrait screen simple operation:
Measurement value buffer area Simple operation menu area Operation prompt area
8
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Landscape screen standard operation : Symbol and information prompt area Main menu display area Measurement data area
Submenu display area
Waveform display area
Landscape screen simple operation: Measurement value buffer area Simple operation menu area
Operation prompt area
9
Shantou Institute of Ultrasonic Instruments Co., Ltd.
1.6 Waveform Disp lay Area
Measurement point symbol (standard operation)
Display range start and end
1.7 Measurement Data Ar ea Coupling symbol
1.8 Symbol and Inform ation Prompt Area See the figure below for the view of symbol and information prompt area. Whenever a message shows, the prompt will overlay the symbol, or the symbol will always display if no prompt message. See Table 1-4 for the description of symbols and parameters.
10
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Table 1-4
Description of Parameters and Symbols
Parameter & Symbol
Descrip tion Start Au to Searc h or FloatingGain functions. A: start Au to Searc h function. FA: start FloatingGain function of gate A.
A FB: start FloatingGain function of gate B. E: start EchoToEchoMode. If no, the display is blank.
*
Start Freeze function. If no, the display is blank. Measurement mode, symbols for Standard/ Maximum/ Minimum/
STD/MAX/MIN/DIFF/AVG Difference/ Average 2011-10-01 00:59
Date and time
5920m/s
Current velocity
7.921us
Current probe delay
/
Single/ Dual element probe type WorkMode
/
symbol,
for
normal/
through
coating
(or
EchoToEchoMode) Alarm indicator. In alarm state, the color changes and the mark is filled and flashing. In non-alarm state, it is hollow. On its right, H is for over high limit alarm, and L is for over low limit alarm. Battery capacity indicator.
11
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Chapter 2 System Operation 2.1 Power Supply The power of this system can be supplied by either the battery (DC-A004) or the power adapter (CD-92A): a) Battery Operation: Install the battery according to 4.4.1 Battery Installation and Disassembly. b) Adapter Operation: 1) Insert the adapter AC power in plug in the AC socket (AC 100V~240V). 2) Insert the adapter DC power out plug in DC Power In Socket (DC-12V) at the bottom of the system.
2.2 Power on Operation: Press Power Key
to turn on the system, and the power indicator on the panel will turn
green. The system screen will display SIUI logo and the model. After initialization, it will go to standard operation screen. 【Note 1】: The system power-up parameter settings are the same as those before the system
is powered off last time in normal state. 【Note 2】: If the system cannot be powered on, please refer to Appendix A Common Failures
and Troubleshooting.
12
Shantou Institute of Ultrasonic Instruments Co., Ltd.
2.3 Power Off Operation: Press and hold Power Key
for about two seconds, and the power-off screen will be
displayed and the system will be turned off. 【Note】: If the system cannot be powered off, please refer to Appendix A Common Failures and
Troubleshooting.
2.4 Operation Interface Selection Press
on the keypad to switch between the standard and simple menu interface.
2.5 Main Menu Select ion
Provided that no submenu is selected (i.e. the submenu is highlighted in white), press
or
to switch selecting the 7 main menus.
2.6 Submenu Selection If not specified, the submenu shall be selected as below: Operation i n standard interface: Press the direction key
or
highlighted in white. Press
to select the submenu. And the selected submenu is
to enter the selected submenu, and the submenu is highlighted in
red. Then press the direction key
or
to adjust the submenu.
Operation i n simp le interface: Press or
to enter the submenu, and the selected submenu is highlighted in red. Then press to change the menu value. Press
or
to select other submenu. Press
exit the submenu.
13
again to
Shantou Institute of Ultrasonic Instruments Co., Ltd.
2.7 Turning Page 1) Main menu turning page: Press
or
, when the main menu reaches the last menu of
the current page, it will jump to the first menu of the main menu page automatically. 2) Submenu turning page: Press
or
, when the submenu reaches the last menu of
the current page, it will jump to the first menu of the submenu page automatically.
2.8 Submenu Adju stment If not specified, the submenu value or item shall be adjusted by following the steps below. Operation: 1) Select the submenu to be adjusted. Please refer to 2.6 Submenu Selection. 2) Use the direction key
or
to increase/decrease submenu value or toggle among items.
【Note 1】: When adjusting submenu value, hold on the adjust key may speed up the value
increase/decrease. 【Note 2】: For submenu with status switch operation, a symbol is indicated on the right of the
submenu. Switch between the states by repeatedly pressing the
key.
2.9 Shortcut Operation 1) User-defined key F1 and F2 Press the keypad
or
key to enable or disable the preset functions, and change the
F1 and F2 preset functions in the main menu setup. 2) In the simple operation interface, before entering the submenu, the direction key has the shortcut operation function, and the user may follow the on-screen prompt to operate. After pressing
to enter the submenu, the direction key returns to its original submenu select or
value adjust function.
14
Shantou Institute of Ultrasonic Instruments Co., Ltd.
2.10 a)
Calibr ation Functio n
Standard Calibration
When WorkMode is Normal, go to the CAL main menu, select 1 Point or 2Points in CalOption. I. 1 Point is for zero calibration only, generally a test block with known acoustic velocity and thickness is used for calibration. Steps are as follows: 1)
Set the thickness of Ref1.
2)
Place the probe on the test block with the corresponding thickness.
3)
Ensure that the coupling between the probe and the test block is good, adjust the gain, range and gate A, and bring gate A to record the first echo of Ref1.
4)
Go to CAL sub-menu, press
5)
If the calibration is finished successfully, a prompt message will be displayed in the
to run calibration.
information prompt area. II. 2Points is for zero and velocity calibration. Usually two test blocks with known but different thickness are used for calibration. Steps are as follows: 1)
Set the thickness of Ref1 and Ref2.
2)
Place the probe on the test block with the thickness of Ref1.
3)
Ensure that the coupling between the probe and the test block is good, adjust the gain, range and gate A, and bring gate A to record the first echo of Ref1.
4)
Go to CAL sub-menu, press
5)
Place the probe on the test block with the thickness of Ref2.
6)
Ensure that the coupling between the probe and the test block is good, adjust the gain,
to confirm the measured value of Ref1.
range and gate A, and bring gate A to record the first echo of Ref2. 7)
Go to CAL sub-menu, press
to confirm the measured value of Ref2 and run
calibration. 8)
If the calibration is finished successfully, a prompt message will be displayed in the information prompt area.
b)
Fast Zero Calibration
When WorkMode is Normal, press
to enter the simple operation interface. Here are the 15
Shantou Institute of Ultrasonic Instruments Co., Ltd.
zero calibration steps: 1)
Adjust the waveform waveform and the gate, and bring gate A to record the first echo of the test block on the system.
2)
Press
Note 1
to run zero calibration.
: Calibration can only be performed when WorkMode is Normal. Normal . When performing
the above calibration, use Au use Au to Searc h for quick adjustment, if necessary turn off Au off Au to Searc h for manual precise adjustment. Note 2
: Quick zero calibration should be performed on the test block on the system.
When calibration is finished, you may begin to measure. See Chapter 3 3 for General thickness measurement workflow. workflow.
2.11 2.11
Aut oSearch oSearch and FloatingGain FloatingGain Functio ns
2.11.1 2.11.1 Aut oSearch When using dual-element probe , with this function quick waveform search and adjustment can be carried out subject to waveform signals received. The AutoSearch function provides a shortcut for quick system adjustment and then performs measurement based on the waveform signals of the measured thickness. Such function includes: adjusting DisplayDelay, Range, Gain and gate measurement. However, AutoSearch cannot replace manual adjustment completely, which is also subject to probe selection, as well as material, thickness and surface status of work piece. If AutoSearch function can not achieve proper capture for measurement, please turn off the AutoSearch function, but adjust manually for measurement. Operation: When using dual-element probe, turn on the Au to Searc h function in the MEAS MEAS main menu, or press
(provided
is defined as Au as Au to Searc h ) to activate the function.
When WorkMode is WorkMode is Normal, Normal , through search by the software, bring gate A to record the first echo, and increase autogain of the first peak of the echo to 80%. When WorkMode is Thru-Coating, Thru-Coating, through search by the software, bring gate A to record the first echo, and increase the autogain of the max. Echo peak to 80%, and gate B will auto record the second echo.
16
Shantou Institute of Ultrasonic Instruments Co., Ltd.
【Note】:
1.
When using single-element probe, AutoSearch probe, AutoSearch function in Normal and Through-Coating Through -Coating work modes can only have gain automatic adjustment within gate A.
2.
When EchoToEchoMode EchoToEchoMode is selected, the AutoSearch function is disabled. The gain automatic adjustment can only work by FloatingGain function.
3.
The AutoSearch function is default to auto search continuously after bootup. It can be set to auto search once, that is to run auto search once only upon each bootup. Select Au to Searc h from the main menu MEAS, MEAS, press the shift button between Continue and One. One . Press
2.11. 2.11. 2
or
, and toggle
to switch between start or stop.
Floatin gGain
When FloatingGain is on, gate A or gate B can perform FloatingGain function (only for ThruCoating mode and EchoToEchoMode). When WorkMode is Normal, the system will automatically bring the first wave peak amplitude within gate A to 80% height. When WorkMode is Thru-Coating mode or EchoToEchoMode, the system will automatically bring the highest wave peak amplitude within selected gate to 80% height. This function can not be activated with AutoSearch function at the same time. 【Note】: When AutoSearch or FloatingGain is on, the maximum value of self-adjusting gain
is
limited by the MaxSearchGain. By adjusting this set value, users can define the
maximum value of self-adjusting gain. If this value is set too high, it may search the clutter signal. If this value is set too low, it may not search the desired signal. Specific settings need to be adjusted based on factors such as probe and workpiece.
2.12
WorkMode
There are three work modes for selection: Normal, Thru-Coating and EchoToEchoMode. EchoToEchoMode. a)
When WorkMode is Normal, Normal , the measurement reading is the measurement value between the display zero and the zero-crossing before the first peak higher than GAThresh (the position is auto tracked by the small upside down triangular symbol). See Fig. a below.
b)
When WorkMode is Thru-Coating, Thru-Coating, the measurement reading is the measurement value 17
Shantou Institute of Ultrasonic Instruments Co., Ltd.
between the zero-crossing before the highest peak higher than GAThresh (the position is auto tracked by the small upside down triangular symbol) and the zero-crossing before the highest peak higher than GBThresh. See Fig. b as below. c)
When WorkMode is EchoToEchoMode is EchoToEchoMode,, the measurement reading is the measurement value between the zero-crossing before the highest peak higher than GAThresh (the position is auto tracked by the small upside down triangular symbol) and the zerocrossing before the highest peak higher than GBThresh. See Fig. c as below. In EchoToEchoMode, gate A is the triggered gate. Adjust gate A to hitch the first bottom wave (It is neccessary to make sure that the measuring position within the gate is the correct highest wave peak position of the first bottom beam. It can be judged by adjusting gate start, gate width and the small upside down triangular symbol.) The system will automatically move the GBStart in proportion to make gate B cover the second bottom wave based on the highest peak position within gate A. The difference between highest peak within gate B and gate A is the thickness of Echo to Echo (It is necessary to make sure that the measuring position within the gate B is the correct highest wave peak position of the second bottom beam. It can be judged by adjusting GBWidth and the small upside down triangular symbol. ). EchoToEchoMode is suitable for quick measurement of different thickness by B1-B2. When it is necessary to arbitrarily adjust two gate positions to have manual measurement, please use Thru-Coating mode to adjust the start and width of both gates and measure required Bm-Bn (e.g. B2-B3, or B3-B4) value.
When WorkMode is Normal zero-crossing before peak
peak of measurement point
(a)
18
Shantou Institute of Ultrasonic Instruments Co., Ltd.
When WorkMode is Thru-Coating Peak symbol of
Zero-crossing before
thickness measurement
peak of measurement
end under coating Peak symbol of
Zero-crossing before peak
thickness measurement
of measurement end
start under coating
(b) When WorkMode is EchoToEchoMode Zero-crossing
Peak symbol of
before
beam measurement point
peak
of
beam measurement
of the second wave
Zero-crossing before Peak symbol of
peak
beam measurement point
of
beam
measurement point of
of the first wave
the second wave
(C)
【Note】:
1.
Adjust the measured waveform amplitude or gate based on the measurement requirements above, and enable the signal at the measured position higher than gate thresh, so as to ensure correct measurement. In case AutoSearch or FloatingGain function does not work, disable auto functions and adjust manually.
2.
When Rectify is set at RF, RF, the default GAThresh is minus, i.e. measuring the negative signal of waveform (below the baseline).The measurement value is corresponding to the first peak of negative signal higher than GAThresh. The zero-crossing before the peak is stable, which is good for stable display of measurement value. When GAThresh is set as positive, positive , i.e. measuring the positive signal of waveform (above the baseline), the 19
Shantou Institute of Ultrasonic Instruments Co., Ltd.
measurement value is corresponding to the first peak of positive signal higher than GAThresh. The zero-crossing before the peak, influenced by baseline noise, is bad in stability, which is not good for stable display of measurement value. 3.
Rectify is recommended to be set as RF (at the same time, set GAThresh as minus) or Negative, so as to get the most stable measurement value display.
2.13
Veloci ty Measurement
Use this function to measure velocity of workpiece, provided that the thickness of the workpiece is known. Operation: 1)
Go to Meas main menu.
2)
Set up THKValueRef .
3)
Turn on the submenu MeasVEL.
4)
Place the probe on the test block with thickness corresponding to THKValueRef . Ensure that the coupling is good and the gate has recorded the echo.
5)
2.14
The data displayed in the measurement display area is the velocity value.
PARAM Stor age
PARAM storage includes saving all menu setup, waveforms and measurement values. Up to 100 parameter sets can be saved. This function is also for saving user-defined probe parameter setting or special testing status. The parameter sets are only saved in main unit and the current or all parameter sets can be transferred to SD card in the microSD card holder. In FileName submenu, the file name can be inputted from the virtual keyboard. The created file name will be corresponded with StoreNum. Only FileNum with created FileName can be stored.
2.14.1
1)
Save
Select StoreNum, and select the store set number without the sign StoreNum with
means the set number has been saved with data. 20
. The
Shantou Institute of Ultrasonic Instruments Co., Ltd.
2)
Go to FileName, press
, ,
,
selected character. Press 3)
Press
to select characters and press
to confirm the
to delete and backspace one character.
to confirm the inputted FileName. Press
to cancel the input and exit
the FileName input screen. 4)
Enable Save function.
5)
If the save operation is finished, the information prompt area will prompt Completed.
2.14.2
Recall
1)
Select the StoreNum with a
2)
Enable Recall function, and recall all the menu setup and waveforms, and the system
.
is frozen. Unfreeze to quit recall.
2.14.3
Delete
1)
Select the StoreNum with a
2)
Turn on Delete. The file of the selected StoreNum is deleted, and the
. mark
disappears.
2.15 DATA Storag e DATA Storage includes saving all menu setup, waveforms and measurement values. The difference from PARAM (parameter sets) is the storage structure, and the storage content is saved in files. In each file, the user may save the data with certain common features, and each data set has its multiple data format.
Data sets can be saved in main unit or SD card. Before
storage, please confirm the storage location (Local/SDCard) . In FileName submenu, the file name can be inputted from the virtual keyboard. The created file name will be corresponded with StoreNum. Only FileNum with created FileName can be stored.
21
Shantou Institute of Ultrasonic Instruments Co., Ltd.
2.15.1
Save
To save to system disk. 1)
2)
Go to FileName. Press
,
to confirm it. Press
to delete and backspace one character.
Press
,
or
to select the character, and then press
to confirm the inputted file name. Press
to cancel the inputted and
exit the FileName input screen. 3)
Go to FileFormat , and select 1D,2D or 3D.
4)
Select Line, Column or Point with
mark. While
mark means that data has
been saved at this location, and no additional storage is allowed. The actual storage capability for Line, Column or Point is related with the rest storage capability. 5)
Turn on Save. If the save operation is finished, the information prompt area will prompt Completed.
【 Note 1 】 : If the file name for the file number already exists, changing the file name is
equivalent to file rename, with the file content remains unchanged. 【Note 2】: Up to 8 characters only (including spaces) can be inputted as the file name. If the file
name is longer than 8 characters, the last character will be replaced with a new character. 【Note 3】: If the file is already saved with data, it is not allowed to change the file format. 【Note 4】: Only Line can be selected for 1D format; Line and Column can be selected for 2D
format; Line, Column or Point can be selected for 3D format. The options not selectable are displayed with *. 【Note 5】: The actual file number that can be created is related with the rest storage capability.
2.15.2
1)
Recall
Select FileNum.
2) Select Line, Column or Point with
mark.
3) Enable Recall function, and recall all the menu setup and waveforms corresponding with the measured location, and the system is frozen. Unfreeze and the system will refresh
22
Shantou Institute of Ultrasonic Instruments Co., Ltd.
the setup with the recalled menu setup as the current system setup.
2.15.3
a)
Delete
Delete Current File Storage Location Data 1) Select FileNum. 2) Select the data location with a
.
3) Turn on Delete. The location content of the selected FileNum is deleted, and the mark disappears. b)
Delete Current File All Storage Location Data 1) Select FileNum. 2) Turn on DeleteFile, and all storage location data for the current file will be deleted.
c)
Delete All File Alll Storage Location Data 1) Select FileNum. 2) Select DeleteAllFiles , and all storage location data for all files t will be deleted.
23
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Chapter 3 Application of Thick ness Measurement 3.1 General Work Flow
Fig. 3-1
General Work Flow
The general work flow of thickness measurement on the system is as shown in Fig. 3-1 in Normal mode. 24
Shantou Institute of Ultrasonic Instruments Co., Ltd.
3.2
Appl icatio n Keynote of Thickness Measurement
For most materials, such as carbon steel, etc., the general workflow above is practical and effective. On how to use the system, operation of calibration and auto search function in particular, please see Chapter 2 System Operation. When testing a thickness close to the critical measurement range, the auto-search function sometimes needs manual adjustment and decision by the operator, so as to ensure accurate measurement. While for coarse-grained, high-attenuation material (such as ductile iron), special composite material and multi-layer material, please adjust manually for measurement. It would be important that the operator has certain ultrasound theory and testing knowledge.
25
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Chapter 4 Perip heral Devices, Ports and Accesso ries 4.1
Probe Socket
The probe sockets of the system are compatible to LEMO00 standard connectors. When testing with the system, use the cable configured with the system or a cable of the same type to connect to a proper probe. For testing with one single-element probe, either probe socket (in internal parallel connection) can be used. While for testing with a dual (TR) probe, connect one end to the transmitting socket T and the other end to the receiving socket R.
4.2
MiniUSB Port
4.2.1 Connect to a PC Operation: 1) With the system turned on, connect one end of the communication cable to the miniUSB port of the system, and the other end to the USB port of a power-up PC. 2) Upon successful connection, the PC software gives a prompt message.
4.2.2 Disconn ect fr om a PC Remove the USB cable from the miniUSB port of the system directly.
4.3
Micro SD Card Holder
4.3.1 Insert mic roSD card Insert the microSD card in the correction direction to the microSD card holder on the system, push it in till a sound is heard when it reaches the fixing device in the card holder. Restart the system, and the microSD card will get ready. Now the user may perform data storage or software
26
Shantou Institute of Ultrasonic Instruments Co., Ltd.
update.
4.3.2 Remove microSD card Turn off the system, push the microSD card into the system, release it only when a sound is heard from the fixing device in the card holder. Now the microSD card comes out, and you may remove it.
4.4
Battery
4.4.1 Battery Installation and Removal Follow the below operation procedure to install the battery (DC-A004): 1) Align the positioning device at the bottom of the battery to the bottom of the system. 2) Insert the battery, push the battery in place, and press the battery against the back of the system. Fix the battery fixing clip. Follow the operation procedure to remove the battery: 1) Press the battery fixing clip downward and pull it away from the system back, so as to detach the battery from the fixing clip of the system. 2) Take out the battery upwards.
4.4.2 Charging the Battery Operation: 1) Insert the adapter (CD-92A) power plug into the AC socket (AC 100V~240V). 2) Insert the DC Out plug of the adapter in the main unit power in socket. If the battery is installed in the main unit, the power/ charging indicator at the bottom of the main unit keypad will be on (in red) for recharge. 3) When the power/ charging indicator turns green, it means the recharge is finished. Usually it takes 6 hours for one full recharge (with the battery in the system). 4) Disconnect the power adapter from AC power supply and the main unit. 【Note】: 27
Shantou Institute of Ultrasonic Instruments Co., Ltd.
1) The system supports recharge while it is in operation. However, it is not recommended to keeyp on such operating state frequently. Because when the battery is fully discharged and starts recharge, the load on the power adapter is big and the heat is obvious, and the recharge while in operation will keep the main unit at a nearly full load state, which is not good for its operating life. 2) When the battery is recharged while in operation, the power/ charging indicator at the bottom of the system keypad will be always on (in green). 3) When the battery is charging, if the power adapter indicator flashes between red and green briefly, it is normal.
4.4.3 Using the Battery The battery capacity indicator is shown on the top right of symbol information area. More occupied boxes mean higher power capacity.
When the battery box indicator is blank and has a
red frame, it means that the battery power is running out, and you should recharge or replace the battery in time. A new DC-A004 battery, if fully charged, may work continuously over 6 hours. 【Note】: When the battery power is very low, to avoid battery over-discharge, the system will be
turned off automatically.
4.4.4 Batt ery Maintenance The capacity and the service life of the rechargeable battery depend on whether the battery is used properly. Please recharge the battery (DC-A004) through the system main unit with the configured power adapter (CD-92A). On how to recharge the battery, please refer to 4.4.2 Charging the Battery. In the following cases, the battery shall be charged: 1) Before using the battery for the first time; 2) If the battery is stored for or over 3 months; 3) After frequent discharge.
28
Shantou Institute of Ultrasonic Instruments Co., Ltd.
4.4.5 Air lift Requirements for battery
a)
For air transporting the system, the battery should be put in the system and placed in the unchecked baggage. For shipping, insulating terminals of the battery and placing the battery inside the system. (although switch of the system prevents unintentional activation );
b)
For air transporting the battery, be sure to insulate terminals of the battery (such as placing single battery in a separate plastic bag or protective case to prevent short circuit) and put battery in the unchecked baggage. Note
: Battery matching this system has passed the test required in the section 38.3 of Part III of UN Manual of Tests and Criteria. If you have any questions, please contact local aviation authorities for further guidance.
4.4.6 Cauti ons on usi ng the battery: 1) When the system is being operated with an adapter or the battery is being recharged, please connect with stable and reliable mains power supply (AC 100V~240V, 50Hz~60Hz); 2) The system battery can only be recharged with the adapter in the method specified. 3) Please recharge the system battery in a cool and ventilated room, otherwise, the battery charging effect might be affected. 4) When the system is not in use for a long time, the battery should be recharged and discharged at least once every three months. 5) During storage or transportation, the battery shall be stored with half capacity, i.e. its voltage shall be 7.2V~7.7V, so as to avoid any gas generated or gas inflation. 6) When system battery comes to the end of its service life, in order to avoid environmental pollution, do not discard, break or crush the battery at will. Please dispose it properly according to applicable laws and regulations or national standards. 7) It is prohibited to throw the battery into fire, so as to avoid explosion or producing toxic chemicals. 8) Do not charge the battery near fire source or in extremely hot environment. 29
Shantou Institute of Ultrasonic Instruments Co., Ltd.
9) It is prohibited to make electric conductor contact ba ttery pole plate, so as to avoid short circuit. 10) Keep the battery from moisture and never throw it into water. 11) Do not store the battery in the temperature higher than 60℃ or lower than -20℃; 12) Do not use the battery if it is in peculiar smelling, heats, distortion, changed color or any other abnormity.
4.5
Power Adapt er
The system configured power adapter (CD-92A) may serve for power supply of system operation, or charging the battery through the system.
Cautions on us ing t he adapter: 1) Do not use the adapter on a plug-loosen socket. 2) Do remember that the total load of the adapter plugged into the socket shall not exceed its rated load.
4.6
Fast Charger CD-B010 (Opti on)
When the battery DC-A004 is removed from the main unit, you may charge the battery outside of the system with the fast charger CD-B010. A complete recharge takes about 3 hours. Operation: 1)
Insert the power plug of the adapter CD-92A to the AC outlet (AC 100V ~ 240V).
2)
Insert the DC power out plug of the adapter to the power in socket (DC 12V) of the fast charger CD-B010. Then insert the charge output plug of CD-B010 to the charge input socket of the battery. Now the charger power / charge indicator is red, which means charging.
3)
When the CD-B010 charger power / charge indicator turns green, it means charging is finished. A full charge usually takes about 3 hours.
4)
When charging is finished, disconnect the power adapter from the AC power and the charger. Unplug the CD-B010 charge output plug connected to the battery DC-A004.
30
Shantou Institute of Ultrasonic Instruments Co., Ltd.
【Note】:
1. If the configured car charging cable is in use, that is one end of the cable is connected to the car power supply (DC 12V) socket, and the other end to the power in socket (DC 12V) of the fast charger CD-B010, such connection is an alternative to the adapter CD-92A. You may charge the DC-A004 battery on the car. 2. In the charging process, you may feel the obvious heat from the housing of the charger, which is normal. When charging the battery, please keep the CD-B010 charger in a wellventilated place, away from flammable and explosive materials.
31
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Chapter 5 Trainin g, System Maintenance and Service 5.1
Training
To ensure a safe and proper operation, the operator shall receive appropriate training on ultrasound testing, and read this operation instruction before operating this system. The lack of relevant knowledge on ultrasound testing may lead to unpredictable wrong testing results. Therefore, to acquire information on ultrasonic testing training and qualification, please contact relevant NDT societies, organizations or SIUI.
5.2
System Maintenance
Please use the factory configured accessories and options, such as the probe cable and the communication cable. Incompatible cables may result in system internal circuit failure or external connector damage. DO NOT drop any liquid into the system. The system housing, glass panel and other parts can be cleaned with a piece of cloth soaked with a little water or neutral household cleanser. 【Note】: DO NOT use any solvent to clean the system. The plastic parts may be damaged or
become fragile if being cleaned with solvent.
5.3
Service
The system adopts the latest technologies and is made of high quality components. The process inspection, the in-process testing and the quality management system are all certified by ISO9001, ensuring that the system performance is optimized. SIUI provide 2 years’ warranty for the system main unit. In case any failure is found in the system, please shut off the system and take out the battery. Contact SIUI service department whenever it is necessary to repair or recalibrate the system. The warranty does not cover any unauthorized repair or disassembly.
32
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Under the following situations, the safe operation of the system is not ensured: 1) The system is obviously damaged (with abnormal display or internal sound). 2) The system is stored in adverse environment for a long time (e.g. abnormal temperature or high humidity, or erosive condition). 3) Severe fall-down or pressure during transportation. 【 Note 】 : Before transportation, the system shall be packed to meet the requirements of
shockproof, waterproof and humidity-proof, in order to avoid system damage due to rain, snow, mechanical collision or strong shock.
33
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Appendix A A.1 a)
b)
Common Failures and Troubleshooting
The sys tem cannot power on If the power supply is from the battery: 1)
Try to replace the battery with full capacity if the capacity is low.
2)
Change the power supply to the adapter.
If the power supply is from the adapter: 1)
Check if the system is connected to the adapter plug, or if the plug comes loose.
2)
Check if the adapter indicator indicates the adapter works properly.
3)
If the battery is not used on the system, try to use the system with a fully charged battery.
c)
If the failure still exists after trying the solutions above, please contact the distributor or SIUI.
A.2
The sys tem cannot power off
a)
Long press the power key
b)
If force shutdown fails, unplug the adapter or remove the battery to cut off power
for 3 seconds to force the system shutdown.
supply. Reboot the system to see whether it is in normal operation. If not, try to contact the distributor or SIUI for repair as soon as possible.
A.3
Cann ot charge
a)
Check if the plug connection of the adapter is good or not.
b)
Check if the indicator of the power adapter indicates normal operation.
A.4 a)
No Echo Check if ProbeType is set up properly. If it is set Dual but connected to a single-
34
Shantou Institute of Ultrasonic Instruments Co., Ltd.
element probe, or the vice versa. b)
Check if the probe cable is in bad contact. Try to replace another probe cable.
c)
Check if the configurations such as gain, range, delay and zero are correct.
d)
Set the system at Default, connect a standard configured probe, and start the AutoSearch function. Couple the probe on the test block of the system. If echoes are displayed, the failure is due to improper setting.
e)
If the failure still exists after trying the solutions above, please contact the distributor or SIUI.
A.5
Abn ormal error of thic kness meas ur emen t readin g
a)
Check if ProbeType and WorkMode are set up properly.
b)
Check if velocity or zero is calibrated.
c)
Check if gain and gate are set up properly, such as the gate does not record the echo to be measured, or the measurement point is not found properly due to AutoSearch function, which can be corrected with manual adjustment.
35
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Appendix B
Specifications Table B-1 Specifications
No. 1
Item
Specifications and Configu ration
Display
800×480 pixels, 5” high-resolution color TFT-LCD monitor, with
Screen
high-brightness LED backlight (visible under sunlight)
2
Standard (R-B1, transmit pulse to the first echo); Measurement
Thru-coating measurement or EchoToEchoMode measurement
Mode
(B1-B2, or Bm-Bn); All measurements using Zero Crossing.
3
4
Measurement
0.5~600mm (subject to probe, material, temperature and selected
Range
configuration)
Display Resolution
0.01mm / 0.1mm (0.001in / 0.01in) Dual-element delay type(with probe cable):
5
TG5-10L (standard configuration) TG7.5-6L TG5-10HL TG2-12L TG5-6L TG5-8L Single-element delay type : Compatible Probes
TG5-6DL TG7.5-6DL TG10-6DL TG7.5-3DMD TG10-3DMD TG15-3DMD Single-element contact type: TG0.5-25NL TG0.5-30NL TG1-20NL TG2.5-20NL
6
Velocity Range
400~15000 m/s
7
Gain
0-110dB manually adjustable/auto (for auto-search or floating gain)
8
Off/On: With this function activated, proper display range and gain Auto Search
can be adjusted automatically based on the measured waveform echo, which improves measurement efficiency.
36
Shantou Institute of Ultrasonic Instruments Co., Ltd.
No. 9 10
11
Item
Specifications and Configu ration
A -scan Rectify
RF /Full/Positive/Negative Negative square wave transmission, with pulse-width and voltage
Pulser
auto matching the probe
Measurement
4/8/16/32
times
When using the standard configured probe (TG5-10L) :
12
0.80mm ~ Display Error
Tube
Wall
Thickness
± (1‰H + 0.04)mm
100.0mm ~400.0mm
± 3‰H mm
15
: H is thickness of the detected material.
With a configured probe, it can measure steel tube with diameter not less than 20mm and wall thickness not less than 2.0mm
Measurement 14
± 0.05mm
10.00mm ~ 99.99mm Note
13
9.99mm
a.Fast zero point calibration with the built-in test block.
Calibration
b.User-defined calibration (1 point/ 2 points calibration)
Measurement
STD/MAX/MIN/DIFF/AVG
Function 16
Interface Mode
Standard /Simple menu measurement interface
17
Other
Velocity dynamic measurement, measurement value over-limit
Functions
symbol, sound alarm and freeze function.
18
Portrait Landscape Screen
/
Portrait /Landscape screen/ auto (gravity-sensing auto switch), suitable for left/ right handed operation Main Unit Storage: Max. 10,000 sets of measurement data (including measurement value, parameter setup and waveform data. Three file formats
19
Storage
available for different applications);
Function
Up to 100 sets of parameter file data (such as measurement value , parameter setup and waveform data) MicroSD card: Max. 40,000 sets of measurement data.
20
21
Data Transmission Measure Unit
The data can be stored to a micro SD card and transferred to a PC via a card reader; It can also be transferred to a PC via the miniUSB port. inch/mm
37
Shantou Institute of Ultrasonic Instruments Co., Ltd.
No.
Item
22
Language
23
Auto Shutoff
24
25
26
27
28
Specifications and Configu ration English/ French/ Spanish/ Portuguese/Polish/German/Chinese/ Russian Off/2/5/10/20/30 minutes for selection
Operation
-10~45℃
Temperature Power Supply
b. ≥6hours’ operation with 7.4V rechargeable lithium batte ry set a. With battery in the system: approx. 6 hours
Battery Charge Time Dimension
a. DC 12V power adapter
&
Weight Main Unit Port
b. With external fast charger (CD-B010 charger, option): approx. 3 hours 105 mm × 180 mm × 42 mm (W×H×D) Approx. 600g (including battery) MiniUSB, micro SD card holder, DC-IN (DC12V input) LEMO 00 compliant (T/R)
38
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Appendix C
Submenu Default Val ue and Range or Option Table C-1 Submenu Default Value and Range or Option
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option To select probe type, and the probe setup will be the last probe setup. Dual-element delay
Select Default will
type(with probe
restore probe(in
cable):
addition to
TG5-10L (standard
custom1~custom10)
configuration)
setup for all the probes
TG7.5-6L
to their factory setting.
TG5-10HL TG2-12L
The last 10 items in the
TG5-6L
probe type list are user-
TG5-8L
defined probes (default names:
CAL
ProbeType
TG5-10L
Single-element
custom1~custom10).
delay type :
Based on the selection
TG5-6DL
of factory built-in probe
TG7.5-6DL
parameters (i.e. for the
TG10-6DL
current selected probe),
TG7.5-3DMD
the probe parameters
TG10-3DMD
can be adjusted
TG15-3DMD
according to the user probe actual state (e.g.
Single-element
probe delay, test state,
contact type:
dual probe V-path
TG0.5-25NL
correction) and saved.
TG0.5-30NL
The user-defined probe
TG1-20NL
names can be revised.
TG2.5-20NL Operate by following the screen prompt.
39
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option Normal is to use single gate measurement, and the measurement reading is the measurement value between the display zero and the first peak higher than GAThresh; Thru-Coating and Normal/
WorkMode
Normal
EchoToEchoMode are
Thru-Coating/
to use dual gate
EchoToEcho
measurement, and the measurement reading is the measurement value between the highest peak higher than GAThresh and the highest peak higher than GBThresh. See section 2.12 WorkMode for details.
ProbeZero^
us
7.930
0~200
Vel^
m/s
5920
400~15000
1 Point
1 Point /2Points
CalOption Ref1
mm
4
0.1~1000.
Ref2
mm
8
0.1~1000
select 1 Point or 2Points Set the thickness of Ref1 Set the thickness of Ref2 To run calibration, and
CAL
0
0/1/2
the number means operation steps
GAStart
5%
0%~120% Gain adjustment, and
dB △
db
45
0~110
the step value can be changed.
MEAS
Range^
mm
25
10~1000
DisplayDelay^
mm
0
0~500
40
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option Gain adjustment, and
dB △
db
45
0~110
the step value can be changed. The AutoSearch function is default to auto search continuously after bootup. It can be set to auto search once, that is to run auto search once only upon each bootup. Select
off (Continue)/ on (Continue)/
off
AutoSearch
(Continue)
off (One)/ on (One)
Au to Searc h, press shift button
,and
toggle between Continue and One. Press
or
to
switch between start or stop. In EchoToEchoMode, the AutoSearch function is restricted. It is recommended to use FloatingGain. In AutoSearch or FloatingGain, the user can limit the maximum of the automatic adjustment gain. If this value is set too high, it may search the clutter MaxSearchGain
db
78
0~110
signal. If this value is set too low, it may not search the desired signal. Specific settings need to be adjusted based on factors such as probe and workpiece.
41
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option When FloatingGain is on, the user can automatically adjust the gain based on GA or GB (only in ThruCoating/ EchoToEchoMode). In Normal mode, the system automatically brings the first wave peak amplitude within
FloatingGain
off
off/GA/GB
GA to 80% height. If it is in Thru-Coating and EchoToEchoMode, the system will automatically bring the highest wave peak amplitude within selected gate to 80% height. This function can not be activated with AutoSearch function at the same time. To select measurement mode of Standard/ Maximum/ Minimum/ Difference/ Average. If MAX/MIN is selected, the interruption of probe
MeasMode
STD
STD/MAX/MIN/
coupling less than 4
DIFF/AVG
seconds is ignored. Only when the interruption of probe coupling is more than 4 seconds, a new scan will be started to capture the maximum /minimum value.
42
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option If DIFF is selected, the measurement is the
DiffRef
mm
4
0.1~1000
actual measurement minus the difference reference value. If AVG is selected, the
AVGNum
2
2~9
measurement is the average measurement of such number. To enable MeasVEL,
MeasVEL
off
off/on
and the measurement result is the velocity. When measuring
THKValueRef^
mm
4
4~500
velocity, THKValueRef needs to be set up properly. To set up whether to maintain
LastReading
Blank(---)
Hold/Blank(---)
the
last
reading or show blank in
the
measurement
display column if no coupling signal.
GAStart
5%
0%~120%
GAWidth
85%
4%~120%
60%/-60
0~90%/-10%~90%
(RF wave)
(RF wave)
GBStart
30%
0%~120%
GBWidth
85%
4%~120%
65%/-65
0~90%/-10%~90%
(RF wave)
(RF wave)
16
4/8/16/32
GAThresh GATE
GBThresh
CONF
MeasCount
Hz
43
Refresh rate of measurement reading
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option The default is RF. It is recommended to use RF (with gate thresh set
Rectify
RF
RF/Positive/
at minus) or Negative
Negative/Full
for testing, so as to achieve the most stable display of measurement value. See 2.11.
Portrait / Landscape ScreenMode
Portrait
0°/ Landscape 180°/Auto
F1-Function
AutoSearch
F2-Function
Freeze
AutoSearch/ WorkMode
User defined F1 function key
Freeze/MeasVEL/ SaveMeasData Every time when the system starts, the default brightness is 5, which can be changed but not saved when
Brightness
5
1~10
poweroff. If the screen has been in operation at the high brightness, the battery operation time decreases accordingly.
Color
1
1~2
Precision
.0mm/.001in
01mm/.001in/ .1mm/.01in Chinese(CN)/ English(EN)/ French (FR) / Spanish(SPA)/
Language (语言)
EN(英文)
Portuguese(PT)/ Polish(PL)/ German(DEU)/ Russian(RU) Czechic (CZ)
Unit
mm
mm/inch
44
To select display precision.
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option Go to the submenu,
Date
and use the direction key to change the date. Go to the submenu,
Time
and use the direction key to change the time. The poweroff countdown starts when no coupling signal and no button operation is carried out. When the
AutoPowerOff
off/2min/5min/10min
off
/20min/30min
countdown reaches half of the time setting, the screen brightness decreases to the minimum, and the software will shut off the system when time is out. To restore the menu to factory setup, including all the probe setup. The
Default
off
off/on
operation does not affect the saved parameter sets and data sets. DO NOT power down
Update
off
off/on
the system in software update!
About
off
off/on
StoreNum
1
1~100
45
System version information
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option
PARAM
In FileName submenu, file name can be input with the virtual keyboard. The created file name will be
FileName
corresponded with StoreNum. Only FileNum with created FileName can be stored. To save all the menu setup and the current waveform data.
Save
off
off/on
Parameter sets can be saved and recalled as user-defined probe parameter file. To recall the saved
Recall
off
off/on
menu setup and waveform data
Delete
off
off/on
DeleteAll
off
off/on
To SDCard
off
Off/on
Delete current parameter file. Delete all parameter files. Save current parameter file to SD card. Save all current
AllToSDCard
off
Off/on
parameter files to SD card.
StoreCapability
off
Off/on
Check current storage capability. Select FileNum, and the file name will be
DATA
FileNum
1
1~100
renewed. The actual FileNum value is related with the rest storage capability.
46
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main Menu
Ad ju stabl e ran ge Submenu
Unit
Default
Remark or option Enter the submenu, and input the file name from the virtual keyboard. The created
FileName
~
~
file name will be corresponded with StoreNum. Only FileNum with created FileName can be stored.
FileFormat
1D/2D/3D
File format of the data saved
Line
1
1~200
Adjustable range. The
Column
1
1~200
actual storage capability for Line,
Point
1
1~200
Column or Point is related with the rest storage capability. To save valid measure
Save
off
off/on
data to designated storage location To recall and display
Recall
measure data
Delete
off
off/on
DeleteFile
off
off/on
DeleteAllFiles
off
off/on
StorageLocation
local
To delete all data of the current file To delete all data of all files.
Check the current storage capability
MaxAlarmSW
off
off/on
MinAlarmSW
off
off/on
MaxAlarm
mm
5
3~1000
MinAlarm
mm
3
3~1000
off
off/on
BuzzerSW
data
Local/SDCard
StorageCapability
ALARM
To delete the current
47
To set up the maximum alarm threshold. Alarm buzzer switch
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Appendix D
Available Probes Specification Measurement Range
Probe Type
Frequen cy
Ap pl ic ati on Type
(Having Crystal
Contact
measurement
Diameter Diameter on 45# steel in
Remark
Normal mode without special instruction) Dual-element delay type TG5-10L
5MHz
Dual-element
10mm
11.8mm
0.8~400mm
delay type,
With probe cable
standard, commonly use TG7.5-6L
7.5MHz
Dual-element
6mm
9mm
0.8~100mm
delay type,
With probe cable
standard, thin plate TG2-12L
2MHz
Dual-element
12mm
16mm
Steel: 3~600mm
delay type, high
Nodular cast
With probe cable
penetration Iron: 3~40mm TG5-10HL
5MHz
Dual-element
10mm
12mm
3~300mm
delay type
With probe cable
High temperature: 200°C TG5-6L
5MHz
Dual-element
6mm
10mm
0.8~200mm
delay type,
cable
standard
TG5-8L
5MHz
Dual-element
With probe
8mm
delay type,
10mm
0.8~300mm
With probe cable
standard
48
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Single-element delay type TG5-6DL
5MHz
Single-
6mm
8.2mm
Standard: 1~20mm
element delay type, thin plate
Thru-Coating (Bm-Bn): 0.5~10mm
TG7.5-6DL
7.5MHz
Single-
6mm
8.2mm
Standard: 1~15mm
element delay type, thin plate
Thru-Coating (Bm-Bn): 0.5~8mm
TG10-6DL
10MHz
Single-
6mm
8.2mm
Standard: 0.8~15mm
element delay type, thin plate
Thru-Coating (Bm-Bn): 0.5~8mm
TG7.5-3DMD
7.5MHz
Single-
3mm
4mm
Standard: 1~10mm
element delay type, thin plate
Thru-Coating (Bm-Bn): 0.5~5mm
TG10-3DMD
10MHz
Single-
3mm
4mm
Standard: 0.8~10mm
element delay type, thin plate
Thru-Coating (Bm-Bn): 0.5~5mm
49
Delay
line
probe without probe cable
Delay
line
probe without probe cable
Delay
line
probe without probe cable
Delay
line
probe without probe cable
Delay probe without probe cable
line
Shantou Institute of Ultrasonic Instruments Co., Ltd.
TG15-3DMD
15MHz
Single-
3mm
4mm
Standard: 0.8~12mm
element delay type, thin plate
Thru-Coating (Bm-Bn): 0.4~6mm
Delay
line
probe without probe cable
Single-element contact type TG0.5-25NL
0.5MHz
Single-
25mm
28mm
Steel: 10~600mm
element contact type, Low
Nodular cast
frequency
iron:
0.5MHz
Single-
30mm
36mm
Steel: 10~600mm
element contact type, Low
Nodular cast
frequency
iron:
1MHz
Single-
20mm
24mm
Steel: 10~600mm
element contact type, Low
Nodular cast
frequency
iron:
2.5MHz
Single-
Without probe cable
Without probe cable
10~80mm
application TG2.5-20NL
cable
10~80mm
application TG1-20NL
probe
10~50mm
application TG0.5-30NL
Without
20mm
24mm
Steel: 5~600mm
element contact type, High
Nodular cast
penetration
iron:
Without probe cable
5~80mm
【Note】:
a.
If low frequency (less than 2.5MHz) or high temperature probe is adopted, it is recommended to select precision as 0.1mm or 0.01 in.
b. If testing coarse-grain or high attenuation materials, it is recommended to select precision as 0.1mm or 0.01 in. 50
Shantou Institute of Ultrasonic Instruments Co., Ltd.
c.
If the probe of single-element delay type is used for the first time, or it has not been used for a long time but before reuse, please remove the delay line of the probe carefully along the screw thread, add sufficient coupling gel between the delay line and the crystal, and then assemble the delay line back to the probe.
d.
Using other probes (including the user-defined probes Custom1-10) except the thickness probe supported by our company, index like the measurement range and precision can not be predictable and guaranteed.
51
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Appendix E E.1
About CTS-59
Appearance
The appearance, interface and keyboard of the CTS-59 are the same as those of the CTS-49. Please refer to 1.1 Appearance, Chapter 1.
E.2
Menu Struc tu re Table E-1 Menu Structure Main Menu
Submenu ProbeType V-PathSelect WorkMode ProbeZero^ Vel^ CalOption
CAL
Ref1 Ref2 CAL GAStart dB △
Table E-2 Menu Structure (Continued) Main Menu
Submenu Range^ DisplayDelay^ dB △ AutoSearch MaxSearchGain FloatingGain MeasMode
MEAS
DiffRef AVGNum MeasVEL THKValueRef^ LastReading
52
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Table E-3 Menu Structure (Continued) Main Menu
Submenu GAStart GAWidth GAThresh
GATE
GBStart GBWidth GBThresh
Table E-4 Menu Structure (Continued)
Main Menu
Submenu MeasCount Rectify ScreenMode F1-Function F2-Function Brightness Color
CONF
Precision Language (语言) Unit Date Time AutoPowerOff Default Update About
Table E-5 Menu Structure (Continued) Main Menu
Submenu StoreNum FileName Save Recall
PARAM
Delete DeleteAll ToSDCard AllToSDCard StoreCapability
53
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Table E-6 Menu Structure (Continued) Main Menu
Submenu FileNum FileName FileFormat Line Column Point
DATA
Save Recall Delete DeleteFile DeleteAllFiles StoreLocation StoreCapability
Table E-7 Menu Structure (Continued) Main Menu
Submenu MaxAlarmSW MinAlarmSW
ALARM
MaxAlarm MinAlarm BuzzerSW
Table E-8 Menu Structure (Continued) Main Menu
Submenu ProbeType ProbeCode
PROBE
V-PathSelect ProbeZero^ dB △
Table E-9 Menu Structure (Continued) Main Menu V-PATH
Submenu V-PATH Enable TestBlockPointVal TestBlockPoint Calibrate DeletePoint VEL ProbeZero^ 54
Shantou Institute of Ultrasonic Instruments Co., Ltd.
dB △ ApplyToCustom FileNum FileName Save Recall Delete DeleteAll StoreCapability Table E-10 Menu Structure (Continued) Main Menu
Submenu TDGEnable InitGain
TDG
TDGSlope StartBlank MaxGain
Table E-11 Menu Structure (Continued) Main Menu
Submenu TempCompEnable DegreeUnits
TEMP
CalibrationTemp TempCoefficient CurrentTemp
Table E-12 Menu Structure (Continued) Main Menu
Submenu CoatTHK CoatVEL^ GAStart GAWidth
COAT
GAThresh GBStart GBWidth GBThresh
55
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Table E-13 Menu Structure (Continued) Main Menu
Submenu B-ScanEnable FileNum FileName FileFormat Save
B-SCAN
Recall Delete DeleteAll ToSDCard AllToSDCard StoreCapability
B-SCAN (menu in the measurement setup subpage)
B-ScanGridSize B-ScanDirection B-ScanLOSMode B-ScanFreezeMode B-ScanMaxTHK
Table E-14 Menu Structure (Continued) Main Menu
Submenu
MULTI
MultiMeasEnable Meas1 Meas2 Meas3 Meas4 Sum LayerVel1^ LayerVel2^ LayerVel3^ LayerVel4^ GAStart GAWidth GAThresh dB △ Range^ FileNum FileName Save Recall Delete 56
Shantou Institute of Ultrasonic Instruments Co., Ltd.
DeleteAll ToSDCard AllToSDCard StoreCapability
【Note 1】: The submenu marked with ^ has coarse adjustment/fine adjustment. The submenu
marked with ^ on the right indicates it is in fine adjustment, while the submenu with – mark in coarse adjustment. 【Note 2】: The submenu marked with △ has adjustable steps. and the number on the right of
the submenu with △ means the step value.
E.3
CTS-59 Symbo l and Information Promp t Area
See the figure below for the view of symbol and information prompt area. Whenever a message shows, the prompt will overlay the symbol, or the symbol will always display if no prompt message.
Table E-15 Description of Parameters and Symbols Parameter & Symbol
Descrip tion Start Au to Searc h and FloatingGain functions. A: start Au to Searc h function. FA: start FloatingGain function of gate A.
A FB: start FloatingGain function of gate B. E: start EchoToEchoMode. If no, the display is blank.
*
Start Freeze function. If no, the display is blank. Measurement mode, symbols for Standard/ Maximum/
STD/MAX/MIN/DIFF/AVG Minimum/ Difference/ Average 2011-10-01 00:59
Date and time
57
Shantou Institute of Ultrasonic Instruments Co., Ltd.
5920m/s
Current velocity
7.921us
Current probe delay
/
Single/ Dual element probe type WorkMode
/
symbol,
for
standard/
thru-coating
(or
EchoToEcho) Alarm indicator. In alarm state, the color changes and the mark is solid and flashing. In non-alarm state, it is hollow. On its right, H is for over high limit alarm, and L is for over low limit alarm. Battery capacity indicator. Start dedicated function symbol: V: Start V-PATH function G: Start TDG function
V T: Start TEMP function B: Start B-SCAN function M: Start MULTI measurement function
E.4
Functi onal Operation of Addi tion al Main Menu
E.4.1 PROBE The main menu PROBE shall be used with the menu V-PATH. In ProbeType, see the probe models available in Appendix C or Appendix D. The dual (TR) probes above have corresponding dialog dual probe models for option. Dialog dual probes enable auto probe identification, optimize the configuration between the main unit and the probe, and reduce measurement error. The dialog dual probe has an identification signal pin. When connecting a dialog probe, please connect the pin and the T/R plug right into the probe socket on the system carefully. When starting up the system, it can identify the probe type automatically, and use the built-in 58
Shantou Institute of Ultrasonic Instruments Co., Ltd.
parameters of the probe to set up. If probe connected, a
mark will be shown after the
probe type item. If disconnected, the mark will disappear. When a dialog probe is connected, ProbeType in the main menu Probe will include the optional probe model automatically (with the mark shown after the probe model) for selection. 【Note】:
ProbeCode: unique identification of probe type, cannot be changed by the users. V-PathSelect: Select the V-PATH curve for the current dual probe. Four options for selection: of f , Instrument , Probe and Custom. When using dual-element probe, the default setting is Instrument, which means adopting build-in V-path correction curve. When using dialog probe, the system automatically select Probe. For wear probes or probes from other manufactures, select Custom to use user-defined V-PATH curve. It is needed to apply existing or re-produced V-PATH curve to custom. If no need to modify, you can choose to turn off. When using single-element probe, V-PATHSelect is regularly off.
E.4.2 V-PATH A set of V-PATH calibration curve is set up by default in the system, which is for all the factory original dual probes. The users may make a set of V-PATH curve for a specific probe. When making a V-PATH curve, 10 measurement points can be set at most, and the 10 points should be in the actual measurement area where the user prefers to have smaller measurement error. For example, try to be below 10mm (steel longitudinal wave) or in the range of 20-40mm (steel longitudinal wave), rather than selecting any measurement point in the minimum to maximum measurement range. For high-precision application in different test range, the user may make a number of V-PATH curves. a)
How to make a V-PATH 1)
Set V-PathSelect to Off.
2)
Set V-PATH Enable to On.
3)
Select a test point on the test block, set TestBlockPoint at 1, which means to calibrate the first test point. 59
Shantou Institute of Ultrasonic Instruments Co., Ltd.
4)
Set TestBlockPointVal to a known value, e.g. 4.00mm.
5)
When the probe is in good coupling and the thickness reading is stable, set Calibrate to On, so as to record and calibrate the first test point.
6)
For calibrated test points recorded, a
mark will appear in TestBlockPointVal.
7)
Repeat steps 3)-5) to record more test points.
【Note】:
b)
c)
d)
A valid V-PATH curve shall have at least 3 test points recorded.
Up to 10 test points can be recorded.
When making a V-PATH curve, V-PathSelect must be set to Off .
Save a finished V-PATH 1)
Select a blank file number in FileNum from V-PATH menu.
2)
Input a file name which is relevant to the curve and easy to be identified in FileName.
3)
Set Save to On and save the curve.
Recall a saved V-PATH 1)
Select a file number saved with the curve from FileNum.
2)
Set Recall to On and recall the curve.
Change a test point 1)
Select a test point to be deleted. saved curve which to be deleted
2)
Set DeletePoint to On.
【Note】: After being deleted, the test point will become uncalibrated. An uncalibrated test point
can be calibrated again. e)
Apply a curve as the current V-PATH curve
The curve with calibrated test points can be applied as the current V-PATH curve. Or the user may recall a saved V-PATH curve and apply it as the current V-PATH curve. Set Ap po yToCus to m to On, and change V-PathSelect to Custom. 【 Note 】 : In addition to Custom option for V-PathSelect, the other functions are: Off ,
Instrument and Probe. . Off :
To turn off V-PATH Enable function if not required for measurement.
Instrument : To use a set of calibrated curve in instrument de fault for measurement. Probe: To use a set of calibrated curve in the dialog probe for measurement. 60
Shantou Institute of Ultrasonic Instruments Co., Ltd.
E.4.3 TDG The TDG function is to compensate echo amplitude loss due to transmission sound path. Set TDGEnable in TDG menu On, and TDG function is enabled. The examples of figures before and after TDG function enabled are shown below:
Before TDG function enabled
After TDG function enabled 【Note】: When TDG function is enabled, a TDG line will appear on the screen.
InitGain: The initial dB value of TDG curve TDGSlope: the increment of dB value per us StartBlank: The valid start time of TDG curve MaxGain: The maximum gain that TDG reaches
61
Shantou Institute of Ultrasonic Instruments Co., Ltd.
E.4.4 TEMP In general, the higher the temperature of the solid material, the lower the velocity. Test data shows that, if the temperature of thermal material rises by 100°C, the velocity decreases by 1%. When the temperatures of the calibration test block and the tested work piece are different, the user may enable the temperature compensation function. Turn TempCompEnable item to On, enable temperature compensation function. 【Note】:
●DegreeUnits: Select the unit of temperature, ℉ (fahrenheit) or °C (centigrade) ●CalibrationTemp: The temperature of the workpiece when calibrating the system. Adjustable range: -10°C to 200°C or 14℉ to 392 ℉ .
●TempCoefficient : Temperature compensation coefficient. Adjustable range: -1.0 to 1.0. ●CurrentTemp: Temperature of the workpiece tested. Adjustable range: -10°C to 200°C or 14℉ to 392 ℉.
E.4.5 COAT When WorkMode is Thru-Coating, enable coating measurement function. Set up the velocity of the coating. While testing the thickness through coating, the system will test the thickness of coating. Option: 1) Select COAT menu, adjust CoatTHK to On and set up CoatVEL. 2) At this time, the measurement reading under coating is the measurement value between the zero-crossing before the first peak higher than GAThresh and the zero-crossing before the first peak higher than GBThresh. When CoatTHK is set On, the software will measure the zerocrossing of the first highest peak higher than GAThresh, and calculate the thickness of the coating. See the figure below for the corresponding peak position. See 2.11 for Information about zero-crossing. 【 Note 】 : By adjusting parameters like gain and gate position, make sure the wave form
amplitude and gate position meet the measurement requirements above (see the figure below). Otherwise, it will affect the test result. Please try to use manual adjustment in this case. 62
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Measurement end thru- coating Measurement start thru- coating Measurement point of coating + thru-coating
Thickness of thru-coating
Thickness of coating
Coupling mark
E.4.6 B-SCAN B-scan (time base) function measures at a fixed speed and displays the measurement reading in scanning as B-mode images in proper proportion, which is convenient for recording or analysis for measurement. See the table for enable and file management menu. The access operation of B-scan data is similar to that of datasets.
63
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main menu
Submenu
Unit
Default
Range or opti on
Remarks Enable B-scan function. Press to enter B-scan measurement simple interface (see the description of Bscan measurement simple interface for details). The B-scan
B-Scan Enable
off
off/on
menu in the standard menu interface is mainly for file storage and management. Note
with B-Scan
enabled, other B-SCAN
functions may be restricted. Select
the
file
number, and the file name will change. FileNum
1
1~100
The
actual
file
number that can be created is related to the
remainder
storage capability.
64
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main menu
Submenu
Unit
Default
Range or opti on
Remarks Enter FileName submenu, and use the virtual keyboard to input file name. The created file name is associated
FileName
with the FileNum. Only the FileNum with a corresponding FileName can be saved with B-scan data. Save the entire
SaveType
Entire
Entire / Screen
process of B-scan or the current screen only. Save all the menu setup and the current waveform
B-SCAN Save
off
off/on
data. Parameter set files can be saved or recalled as userdefined probe parameter files. Recall the saved B-
Recall
off
off/on
scan waveforms and setup parameters.
Delete
off
off/on
DeleteAll
off
off/on
ToSDCard
off
off/on
AllToSDCard
off
off/on
Delete the current B-scan file. Delete all the Bscan files. Save the current Bscan file to SD card. Save all the B-scan files to SD card. Check the store
StoreCapability
off
off/on
capability of the current B-scan files.
65
Shantou Institute of Ultrasonic Instruments Co., Ltd.
After enabling B-scan function on the standard menu interface, press measurement simple interface, as shown in Fig. a below. Press
to enter B-scan
again to return to standard
menu. In B-Scan measurement simple screen, you may set up the main parameters (Range, Delay, Gain, GAStart, GAWidth, Thickness), and operate by following the on-screen prompt. The operations on B-Scan measurement simple screen:
Press
to start or stop B-Scan.
Press
When B-Scan is stopped and frozen, adjust the left/right cursor key to select the
to freeze or pause B-Scan.
thickness or A-Scan waveform for the playback measurement position.
Press
to enter B-Scan measurement setup screen (Fig. b). Press
again to
return to the simple interface. See the screen prompt for specific operation, and see the table below for functional description.
66
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Main menu
Submenu
Unit
B-ScanGridSize
Default HalfSize
B-ScanDirection
LeftToRight
Range or option
Remarks
HalfSize/ FullSize
See figures c and d below.
LeftToRight /
Indicated
RightToLeft
“→”/“←”
as
under B-scan display. StopOnLOS / ContinueOnLOS
B-ScanLOSMode
When the probe is not coupled
StopOnLOS
properly, stop/continue scanning. ShowMinimum/
When freeze
Menu option
ShowMaximum/
function is
in the B-Scan
ShowCurrent
measurement setup subpage
enabled, the measurement
B-
cursor stays
ShowCurrent
ScanFreezeMode
at the corresponding position of the scanned display. The current thickness
B-ScanMaxTHK
mm
50.00
4~500
range value is displayed under B-scan display.
67
Shantou Institute of Ultrasonic Instruments Co., Ltd.
(a) B-scan Measurement Simple Interface
(a)
FullSize
(b) B-scan Measurement Setup Subpage
(b)
HalfSize
E.4.7 MULTI If the workpiece has multi-layer materials and the layers are bonded firmly, the MULTI function is recommended for measurement, provided that velocity of each material layer is known. The actual measurement range is related with the materials.
68
Shantou Institute of Ultrasonic Instruments Co., Ltd.
Select MULTI menu, adjust MultiMeasEnable to On, and enable multi measurement function. Operation of MULTI funct ion in the standard menu in terface: a)
Adjust LayerVel1, LayerVel2, LayerVel3 and LayerVel4 according to velocity of each layer.
b)
Couple the probe on the surface of the tested workpiece, adjust parameters (range, velocity and gain), till the screen shows echoes from each layer.
c)
Adjust GAStart, GAWidth and GAThresh , so as to hitch the echo of the first layer. Note: the back flank of the gate should not cover the test range of the reflection echo of the next layer.
d)
Select Meas1 item, press OK, at this time, screen prompts Press -> to update measurement area, <- to turn off measurement area. Follow the prompt to press
key,
and create a measurement area for Meas1 (the start of the measurement area is GAStart, and the size of the measurement area is GAWidth). Two vertical lines will appear onscreen to indicate the start and end of the measurement area. Press
key
to turn off the measurement area for Meas1, i.e. not to display the measurement value of the measurement area. e)
Repeat steps c)- d) to create another 3 measurement areas.
Note 1
: Measurement value of each layer and the total thickness value are shown at the
submenu area. The value shown in the measurement reading area (big reading) is the sum of measured values for each layer. Note 2
: When performing step c) above, pressing the key
repeatedly may turn on/ off the
current measurement area (the 2 vertical lines in the measurement area indicate the on/off status). No measurement or statistics are done if the measurement area is turned off. Note 3
: The display of MULTI measurement in standard menu interface is as shown in Fig. a
below. In the standard menu screen, enable MULTI, finish setup of velocity for each layer, then press to enter the simple screen for multi measurement, as shown in Fig. b below. The functions achieved are the same as those functions in the standard menu screen, but the main operations 69
Shantou Institute of Ultrasonic Instruments Co., Ltd.
ca be easily performed in this simple interface. In the multi measurement simple screen, you may set up the key parameters(Range, Delay, Gain, GAStart, GAWidth, GAThresh)by following the screen prompt. Operation of MULTI funct ion i n the simp le menu interface: a)
Press
switch key, and select the current measurement area number for setup.
Press the key repeatedly to switch among the measurement area 1~4. The operation prompt area will show the selected measurement area number. b)
Adjust parameters of gate and gain to have the gate hitch the reflected signal of the
measurement for the layer, and then press
to u update tthe m measur ement a ar ea s setup.
c)
Repeat steps a) and b) to create the other 3 measurement areas.
d)
Either during or after the test process, you may press
to return to the standard
menu interface.
(a) MULTI Measurement Standard Interface
70
(b) MULTI Measurement Simple Interface