XT-2000i /XT-1800 /XT-1800i ASTM Host Interface Format Revision 2.1
Note : The standard specification is under discussion in NCCLS. This specification may may be modified modified to match the standard when it will be published.
Revised on March 23, 2007
SYSMEX CORPORATION
Revision History Revision
Date
Major contents of changes
1.0
12 th March, 2002
Initial version of Revision 1.0 is created
1.1
25 th May, 2002
Revision 1.1 to add XT-1800 i descriptions
1.2
25th March, 2005
Revision 1.2 to correct typing mistakes (No software updates)
1.3
19th April, 2006
Revision 1.3 to change description of the inquiry sample number and rack number (No software updates)
2.0
5th September, 2007
·Compatibility with the XT IG master and XT RET master is additionally described. · “Products and Additional Software” is added, and the description on XT IG master and XT RET master is added. · “IG#”, “IG%” and “RET-H e” are added to the measurement item ID in request records. · “IG#”, “IG%” and “RET-H e” are added to the measurement item list and QC measurement item list in result records. · “IG Present” is added to the IP message (abnormal) list in result records.
2.1
23th March, 2007
· Compatibility with the ASTM E1381-02 is additionally described. · Erroneous descriptions are corrected in transmission and reception examples in comment records.
XT-2000i/XT-1800i ASTM Host Interface Format. Revision 2.1
2 /43
Revision History Revision
Date
Major contents of changes
1.0
12 th March, 2002
Initial version of Revision 1.0 is created
1.1
25 th May, 2002
Revision 1.1 to add XT-1800 i descriptions
1.2
25th March, 2005
Revision 1.2 to correct typing mistakes (No software updates)
1.3
19th April, 2006
Revision 1.3 to change description of the inquiry sample number and rack number (No software updates)
2.0
5th September, 2007
·Compatibility with the XT IG master and XT RET master is additionally described. · “Products and Additional Software” is added, and the description on XT IG master and XT RET master is added. · “IG#”, “IG%” and “RET-H e” are added to the measurement item ID in request records. · “IG#”, “IG%” and “RET-H e” are added to the measurement item list and QC measurement item list in result records. · “IG Present” is added to the IP message (abnormal) list in result records.
2.1
23th March, 2007
· Compatibility with the ASTM E1381-02 is additionally described. · Erroneous descriptions are corrected in transmission and reception examples in comment records.
XT-2000i/XT-1800i ASTM Host Interface Format. Revision 2.1
2 /43
i
i
XT-2000i/XT-1800i ASTM Host Interface Format. Revision 2.1
3 /43
1. Products and Additional Software 1.1 XT-2000i and and XT-1800i Because the XT-1800 i is not equipped with the RET channel, it does not output the following items. Refer to the instruction manual for explanation of the following items. • RET#, RET%, HFR, MFR, LFR and IRF The following items are not handled in accuracy control. • RBC-O, PLT-O, RBC-X, RBC-Y, d-RBC, d-PLT, Dw/X and Dw/Y
1.2 Additional software 1.2.1 Upgrade Software for Automated Hematology Analyzer XT IG IG master master Installing Upgrade Software for Automated Hematology Analyzer XT IG master enables output of the following items. Contact a dealer or agency near you for the XT IG master. • IG# • IG% • IG Present (IP message) Note:
• When “IG#” and “IG%” are reliable (that is, when “*” is not attached), the XT judges “IG Present”. Present”. When the condition condition for “IG Present” is satisfied satisfied as a result of judgment, judgment, the XT outputs “IG Present”. (In this case, the XT does not output “Immature Gran?”.) • When “IG#” and “IG%” are not reliable (that is, when “*” is attached), the XT cannot judge “IG Present”. It judges “Immature Gran?” Gran?” instead. When When the condition for “Immature Gran?” is satisfied as a result of judgment, the XT outputs “Immature Gran?”. (In this case, the XT does does not output “IG Present”.)
1.2.2 Upgrade Software for Automated Hematology Analyzer XT RET master Installing Upgrade Software for Automated Hematology Analyzer XT RET master enables enables output of the following item. However, this upgrade software is not applicable to the XT-1800i. Contact a dealer or agency near you for the XT RET master. • RET-He
1.3 XT for f or North America The XT for North America does not output the following items in measurement result output and real-time accuracy control data output. • PDW • P-LCR • PCT • HFR • MFR • LFR
XT-2000i/XT-1800i ASTM Host Interface Format. Revision 2.1
4 /43
1.4 Holland SI units When the XT is set to Holland SI units, it outputs data items “HGB”, “MCH”, “MCHC” and “RET-He” with the following units. Item
Holland SI unit
HGB
10-1mmol/L
MCH
amol
MCHC
10-1mmol/L
RET-He amol
2. Scope This document describes the Data Communication Specifications for XT-2000 i/XT-1800i using ASTM E1394-97, E1381-02. [Note] ASTM (America Society for Testing and Materials), one of the world largest volunteer non-profit organizations, founded in 1898 for the purpose of creating standard regulations for materials, products and system services. This specification conforms to the following two standards: ASTM E 1381-02: Specifications for low-level protocols to transfer data between clinical laboratory instruments and computer systems ASTM E 1394-97: Standard specifications for transferring data between clinical instruments and computer systems [Note] The Information Processing Unit (IPU) of the XT-2000 i/XT-1800i supports connection by the RS-232C and Ethernet. For the Ethernet connection, only the presentation layer conforms to ASTM E 1394-97. All other layers conform to the IEEE 802.3.
XT-2000i/XT-1800i ASTM Host Interface Format. Revision 2.1
5 /43