BS-120 Chemistry Analyzer Preventive Maintenance Manual
[email protected]
Dear Service Engineer, It is well known fact that instrumentation which is well taken care of will out perform and out last than which is not. The Mindray BS-120 Automatic Chemistry Analyzer is no exception to this rule. Chemistry systems that are well taken care of by their users and service personnel will have many fewer problems than those which are not. Fewer problems mean fewer service calls, and happier customers. The preventive maintenance (PM) manual is offered as a standard guide to follow. You will find that, depending on the quality of care which an instrument receives, 12 months preventive maintenance visits will require varying degrees of thoroughness. The important point here is that the reliability of the Mindray BS-120 is somewhat dependant on the quality and quantity of attention it receives during its 12 months preventive maintenance visit. The service engineer has a substantial amount of influence on the call rate of each system.
[email protected]
Catalog 1
Part list of Preventive Maintenance Kit .................................................................... 4
2
Replacement ................................................................................................................ 5
3
2.1
Replacing Halogen Lamp(20W) ........................................................................ 5
2.2
Replacing Sample Probe And Washer&Uncloging Probe . ............................ 7
2.3
Replacing Syringe Plunger .................................................................................11
2.4
Replacing Mixing Bar and Retaining Nut ....................................................... 12
2.5
Replacing Filter Assembly ................................................................................. 14
2.6
Replacing Check Valve ..................................................................................... 16
Chemistry Maintenance schedule………………………………………….18
[email protected]
1 Part list of Preventive Maintenance Kit Order Number
Definition
BS-120 801-BA10-00150-00
Preventive Maintenance Kit
Part No
Name
BA10-30-78192
Halogen Lamp(20W)
2
BA40-30-61525
Sample probe
1
0040-10-32307
Washer for Probe
4
115-003182-00
Probe Unclog Tool
1
0000-10-10916
Acupuncture Needle
10
082-000731-00
Syringe Plunger (500ul, zero dead volume)
Qty
1
0040-10-32303
Washer for Syringe
2
BA31-20-41651
Mixing Bar
1
BA30-20-06774
Retaining Nut
1
BA31-30-56750
Filter Assembly (2 Pcs,
1
[email protected]
2 Replacement 2.1 Replacing Halogen Lamp(20W) 2.1.1 Part Number BA10-30-78192
Halogen Lamp
2.1.2 Purpose For sufficient lamp intensity and reliable testing results 2.1.3 Interval Every 6 months Every 1000 hours When the machine reminds you to do so 2.1.4 Procedures
CAUTION
[email protected]
2.1.4.3 Unplug the connector of the light source cable 2.1.4.4 Loosen the retaining screws with hands, and then remove the lamp
[email protected]
2.2 Replacing
Sample
Probe
And
Washer&
Uncloging Probe 2.2.1 Part Number BA40-30-61525
Sample Probe
0040-10-32307
Washer for Probe
0000-10-10916
Acupuncture Needle
115-003182-00
Probe Unclog Tool
2.2.2
Purpose
If the probe is bent or damaged, it must be replaced immediately. Follow the procedure given below to replace the damaged or bent probe.
2.2.3
Interval
Every year
2.2.4 Procedures 2.2.4.1 Place the Power to OFF.
[email protected]
clean place 2.2.4.6 Press the circuit board with one hand and disconnect the probe’s circuit connector from the board with the other hand. Exercise caution when disconnecting the connector. Excessive force may damage the connector and/or the circuit board.
2.2.4.7 Use a slotted head screwdriver to remove the retaining screw on the probe and take out the spring. Store the removed probe in a safe place where it will neither endanger people working around the area nor be damaged. Exercise caution when pulling the probe away from the arm
2.2.4.8 Slowly pull the probe away from the probe arm. Exercise caution so that the gasket inside the probe does not drop out and if it does, store it in a clean place for later installation. When the probe is clogged, the fluid flow will become abnormal. Follow the steps given below to remove, unclog and install the probe.
[email protected]
2.2-1.3.5.Leave the sample probe soaked with wash solution for about 10 minutes. Push and pull the syringe plunger for several times until liquid comes out of the probe tip evenly. 2.2-1.3.6.Use the syringe to aspirate deionized water and rinse the sample probe for at
[email protected]
2.2-2.3.2 Remove the old washer with tweezers and put a new one inside the probe. Dispose the old one in proper way.
The next step is installing the probe 2.2.4.9 Insert the probe back into the hole on probe arm, and align the hole on probe 2.2.4.10 plate to the rotor inside the arm. 2.2.4.11 Sleeve the spring on the rotor and screw the retaining screws to secure. 2.2.4.12 Pinch the probe by the part near the probe arm. Gently push the probe 2.2.4.13 upward and then release the probe to see if the spring can move freely. If yes, proceed to the next step. If not, check for errors and try again after removing the errors. 2.2.4.14 Connect the probe’ s circuit connector back to the circuit board
2.2.4.15 Ensure the gasket is inside the probe
[email protected]
2.2.5 Verification Run the machine if no leakage around the connector ,the replacement is OK.
2.3 Replacing Syringe Plunger 2.3.1 Part Number 082-000731-00
Syringe Plunger (500ul, zero dead volume)
0040-10-32303
Washer for Syringe
2.3.2 Purpose For reliable aspiration 2.3.3 Interval Every year Every 100000 movements Replace it if necessary 2.3.4 Procedures
[email protected]
2.3.4.4 Unscrew (counter-clockwise) the lower retaining screw. 2.3.4.5 Unscrew (counter-clockwise) the four retaining screws, remove the screws and space bars, and remove the syringe from the holder.
2.3.4.6 Grab the Tee with one hand and the syringe connector with the other hand and unscrew (counter-clockwise) the syringe. Exercise caution so that the gasket on the syringe does not drop out and if it does, store it in a clean place for later installation.
2.3.4.7 Unscrew (counter-clockwise) the plunger guide cap and pinch the plunger button to gently pull the plunger assembly from the syringe.
[email protected]
2.4.3 Interval Every year Replace it if necessary 2.4.4 Procedures 2.4.4.1 Place the Power to OFF 2.4.4.2 Gently pull the bar arm to its highest point and move the bar to a position convenient to operate.
2.4.4.3 Pinch the bar by the knurled part with one hand and unscrew (counter-clockwise) the retaining nut with the other hand until the mixing bar looses. Pull the bar downward to remove it and remove the nut.
[email protected]
2.5 Replacing Filter Assembly 2.5.1 Part Number BA31-30-56750
Filter Assembly (2 Pcs, 1pkg)
2.5.2 Purpose Keep reliable testing results 2.5.3 Interval Every 6 months 2.5.4 Procedures 2.5.4.1 Remove cap assembly from the DI water tank and place it on a clean desktop. Carefully remove the filter assembly from the cap assembly.
[email protected]
2.5.5.4 Expel air from the filter assemby 1) Connect the DI water tank assembly to the analyzer (pay attention to the color of the adapters). Put the assembly into the water tank and make sure that the filter sink smoothly to the tank bottom and does not twist with the floater connecting rod. Tighten the cap of the DI water until secure.
[email protected]
2.6 Replacing Check Valve 2.6.1 Part Number M6Y-020003---
Check Vavle
2.6.2 Purpose To keep the sealingness of the fluidic path and unobstructed flow
2.6.3 Interval Every 6 months 2.6.4 Procedures 2.6.4.1 Remove the rear panel and find the Check Vavle.As following CV1&CV2.
[email protected]
2.6.5 Verification Run our analyzer and check if the liquid is weeping ;If not,the replacement is OK.
[email protected]
3. Chemistry Maintenance Schedule BS-120 Maintenance Schedule I: Inspect. Clean, Adjust or Replace if necessary. C: Clean, R: Replace, A: Adjust, L: Lubricate, P: Perform, N: No action needed
ITEM
Related Part
Qty
WHICHEVER COMES FIRST EVERY
r e m o t s u C y B
Reference
DI water tank
N/A
N/A
N/A
I
C
C
C
C
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.3.3
Waste tank
N/A
N/A
N/A
Inlet tubing of Cap assembly of DI Water Tank
N/A
N/A
N/A
I
C
C
C
C
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.3.4
I
I
I
I
I
Outlet tubing of Cap assembly of Waste Tank
N/A
N/A
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.3.3
N/A
I
I
I
I
I
Reagent/Sample disk
N/A
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.3.3
N/A
N/A
N
C
C
C
C
Wash Wells of probe and mixing bar
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.3.5
N/A
N/A
N/A
N
C
C
C
C
Clean by customers
Dust Proof Net
N/A
N/A
N/A
N
N
C
C
C
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.5.1
Clean the Rotor of Reagent/Sample Probe/Mixer Drive Assembly
N/A
N/A
N/A
N
N
C
C
C
W ipe the rotor with clean gauze.
Cleaning Panels of Analyzing Unit
N/A
N/A
N/A
N
C
C
C
C
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.3.6
ISE daily cleaning
N/A
N/A
N/A
P
P
P
P
P
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.2.8
ISE pump calibration
N/A
N/A
N/A
P
P
P
P
P
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.2.8
ISE Na electrode
N/A
N/A
N/A
10,000 tests or 6 months
I
I
I
R
R
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.6.7
ISE K electrode
N/A
N/A
N/A
10,000 tests or 6 months
I
I
I
R
R
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.6.7
I
I
I
R
R
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.6.7
ISE Cl electrode
N/A
N/A
N/A
10,000 tests or 6 months
ISE Li electrode
N/A
N/A
N/A
3,000 tests or 6 months
I
I
I
R
R
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.6.7
I
I
I
R
R
BS-120 Chemistry Analyzer Operator’s Manual(V7.0) Chapter 5.6.7
N/A
N/A
N/A
10,000 tests or 6 months
Lamp
BA10-30-78192
Halogen Lamp(20W)
2
1,000 hours
Reagent/Sample probe
BA40-30-61525
Sample probe
1
Probe washer
0040-10-32307
Washer for Probe
Probe unclog tool
115-003182-00
Probe Unclog Tool
Acupuncture Needle
0000-10-10916
Acupuncture Needle
Syringe Plunger (500ul, zero dead volume)
082-000731-00
Syringe washer
0040 10 32303
ISE Ref electrode
r e e n i g n E y B
S H l y t l y h h s T l y e k t i N n a n D W e o M o M O 2 M 6 1
Syringe Plunger (500ul, zero dead volume) Washer for Syringe
N
N
I
R
R
I
C
C
C
R
4
I
C
C
C
R
1
N
N
P
P
P
10
N
N
P
P
I
I
I
I
R
C
C
C
R
1 2
100,000 movements
I
P