Dear Service Engineer, It is well known fact that instrumentation which is well taken care of will out perform and out last than which is not. The Mindray BS-380 Automatic Chemistry Analyzer is no exception to this rule. Chemistry systems that are well taken care of by their users and service personnel will have many fewer problems than those which are not. Fewer problems mean fewer service calls, and happier customers. The preventive maintenance (PM) manual is offered as a standard guide to follow. You will find that, depending on the quality of care which an instrument receives, 12 months preventive maintenance maintenance visits will require varying degrees of thoroughness. thoroughness. The important point here is that the reliability of the Mindray BS-380 is somewhat dependant on the quality and quantity of attention it receives during its 12 months preventive maintenance visit. The service engineer has a substantial amount of influence on the call rate of each system. Taking a little extra time to educate users as to the proper way of caring for an instrument will, on most cases, increase the instruments reliability, reliability, and the customer’s customer’s satisfaction.
[email protected]
Catalog
3
1
Part list of Preventive Maintenance Kit ................... ......... ................... .................. ................... ................... ............. .... 4
2
Replace Replacement ment................................ ................................................. .................................. ................................. ................................. ................. 5 2.1
Replacing Halogen Lamp(20W) ........................................................................ 5
2.2
Replacing Sample Probe And Washer&Uncloging Probe ............................. 7
2.3
Replacing Replacing Syringe Plunger & Syringe
2.4
Replacing Mixing Bar&Retaining Nut ............................................................. 14
2.5
Replacing Water Inlet Filter …………………………………………………..16 ………………………………………………… ..16
2.6
Replacing DI Water Filter ................................................................................. ................................................................................. 18
2.7 2.8
Replacing First and Second Phase Washing Tube on the Wash Unit …..19 Replacing Cuvette Spring…………………………………………………… Spring…………………………………………………….20 .20
2.9
Replacing Check Valve ................... .......... .................. ................... ................... ................... ................... ................. ........ 21
……………… ……………………1 …………………… 11
Chemistry Maintenance schedule…………………………………………… schedule……………………………………………23 23
[email protected]
1 Part list of Preventive Maintenance Kit FRU
Definition
NewBS-380 801-BA38-00063-00
Preventive Maintenance Kit
Part No
Name
115-008803-00
Halogen Lamp(20W)
2
BA40-30-61525
Sample probe
1
0040-10-32307
Washer for Probe
4
115-003182-00
Probe Unclog Tool
1
0000-10-10916
Acupuncture Needle
10
0040-10-32316
500ul Syringe Plunger Tip
Qty
1
3100-10-49438
Syringe(100ul)
1
0040-10-32303
Washer for Syringe
4
BA31-20-41651 +
Mixing Bar + Retaining
BA30-20-06774 BA40-30-61993 BA38-30-88299
Nut Water Inlet Filter Diluted Detergent filter assembly
1 2 2
[email protected]
2 Replacement 2.1 Replacing Halogen Lamp(20W) 2.1.1 Part Number 115-008803-00
Halogen Lamp(20W)
2.1.2 Purpose For sufficient lamp intensity and reliable testing results 2.1.3 Interval Every 6 months Every 2000 hours When the system reminds you to do so 2.1.4 Procedures 2.1.4.1 Place the Main Power to OFF. Wait at least 10 minutes for the lamp and its
housing to cool down. 2.1.4.2 The light source assembly is at the right rear part of the instrument. Loosen the
screws on the back cover manually or using a screw driver. Pull the cover
[email protected]
2.1.4.4 Unscrew the retaining screws on the lamp base manually. Pull the lamp out.
[email protected]
2.2 Replacing
Sample
Probe
And
Washer&
Uncloging Probe 2.2.1 Part Part Number
2.2.2
BA40-30-61525
Sample probe
0040-10-32307
Washer for Probe
115-003182-00
Probe Unclog Tool
0000-10-10916
Acupuncture Needle
Purpose
If the probe is bent or damaged, it must be replaced immediately. Follow the procedure given below to replace the damaged or bent probe.
2.2.3
Interval
Every year
2.2.4
Procedures
2.2.4.1 Place the Power to OFF. 2.2.4.2 Remove the sample/reagent disk. 2.2.4.3 Pull the probe arm to the highest point by hand. Rotate the probe arm to move
[email protected]
clean place 2.2.4.6 Press the circuit board with one hand and disconnect the probe’s circuit connector from the board with the other hand. Exercise caution when disconnecting the connector. Excessive force may damage the connector and/or the circuit board.
2.2.4.7 Use a slotted head screwdriver to remove the retaining screw on the probe and take out the spring. Store the removed probe in a safe place where it will neither endanger people working around the area nor be damaged. Exercise caution when pulling the probe away from the arm
2.2.4.8 Slowly pull the probe away from the probe arm. Exercise caution so that the gasket inside the probe does not drop out and if it does, store it in a clean place for later installation. When the probe is clogged, the fluid flow will become abnormal. Follow the steps given below to remove, unclog and install the probe.
2.2-1 How to use unclogging tool
[email protected]
2.2-1.3.5.Leave 2.2-1.3.5.Leave the sample probe soaked with wash solution for about 10 minutes. Push and pull the syringe plunger for several times until liquid comes out of the probe tip evenly. 2.2-1.3.6.Use 2.2-1.3.6.Use the syringe to aspirate deionized water and rinse the sample probe for at least 3 times. 2.2-1.3.7.Remove 2.2-1.3.7.Remove the unclogging device and the syringe, and then connect the tubing
[email protected]
2.2-2.3.2 Remove the old washer with tweezers and put a new one inside the probe. Dispose the old one in proper way.
The next step is installing the probe 2.2.4.9 Insert the probe back into the hole on probe arm, and align the hole on probe 2.2.4.10 plate 2.2.4.10 plate to the rotor inside the arm. 2.2.4.11 Sleeve the spring on the rotor and screw the retaining screws to secure. 2.2.4.12 Pinch 2.2.4.12 Pinch the probe by the part near the probe arm. Gently push the probe 2.2.4.13 upward 2.2.4.13 upward and then release the probe to see if the spring can move freely. If yes, proceed to the next step. If not, check for errors and try again after removing the errors. 2.2.4.14 Connect the probe’ s circuit connector back to the circuit board
2.2.4.15 Ensure the gasket is inside the probe
[email protected]
2.2.5 Verification Run the machine if no leakage around the connector ,the replacement is OK.
2.3 Replacing Syringe Plunger & Syringe 2.3.1 Part Number 0040-10-32316
500ul Syringe Plunger Tip
3100-10-49438
Syringe(100ul)
0040-10-32303
Washer for Syringe
2.3.2 Purpose For reliable aspiration 2.3.3 Interval Every year Every 100000 tests The old one is apparently damaged 2.3.4 Procedures 2.3.4.1 Place the ANALYZING UNIT POWER to OFF. 2.3.4.2 Open the middle door and you will see both of the syringes, the sample
[email protected]
[email protected]
2.3.4.10 Grab the T-piece with one hand and the syringe connector with the other
hand. Screw (clockwise) the syringe into the T-piece until secure. 2.3.4.11 Place the syringe on the holder. 2.3.4.12 Install the space bar and two upper retaining screws. Note that do not
tighten the screws now. Adjust the height of syringe to keep the V-shaped groove level le vel with sample syringe scale11 (or reagent syringe 7.5). Then tighten the screws. 2.3.4.13 Rotate syringe assembly motor. When syringe plunger moves
downward till the plunger goes out of the syringe for 15~20mm, gently rotate the plunger retaining screws to push the front end of plunger connector close to the drive part.Gently tighten the plunger retaining screw to avoid plunger rod beat. 2.3.4.14 Place the Power back to ON. Wait for about 30 seconds, and then
execute S ys tem R es et on the Daily Maint. page. Repeat the instruction for several times if necessary, and check if the T-piece is leaking. If not, go to the next step. If
yes, tighten the syringe. If leakage remains, replace the T-piece and
connector. 2.3.4.15 Close the front doors of the analyzer. 2.3.4.16 Place the ANALYZING ANALYZING UNIT POWER back to ON.
[email protected]
2.4 Replacing Mixing Bar & Retaining Nut 2.4.1 Part Number BA31-20-41651 + BA30-20-06774
Mixing Bar + Retaining Nut
2.4.2 Purpose For reliable testing results 2.4.3 Interval Every year If the mixing bar is damaged, it must be replaced immediately. 2.4.4 Procedures 2.4.4.1 Place the ANALYZING UNIT POWER to OFF. 2.4.4.2 Prepare bar it
a
new
mixing
with detergent-dipped dry
bar.
Wash
gauze or
the
flat
part
cotton sticks
of
the
new
and then
wipe
with deionized water-dipped gauze.
2.4.4.3 Gently pull the bar to its highest point and rotate it to a position convenient to operate. 2.4.4.4 Pinch and
the
bar
by
the
knurled
part
with
one
hand
unscrew (counter-clockwise) the retaining nut with the other until the
[email protected]
2.4.4.6 Pinch 2.4.4.6 Pinch the mixing bar by the knurled part and align the hole of the nut to the axis of the mixing bar and push the nut onto the bar until it reaches the end of the bar. Tighten the nut (screw clock-wise) with the other hand. 2.4.4.7 After 2.4.4.7 After replacing the bar, visually check whether the bar is vertical to the bar arm.If yes, proceed to the next step. If not, repeat step 5 to remove the bar and re-install it. 2.4.4.8 Pull the bar arm to its highest point and rotate it back to a position above its wash well. 2.4.4.9 Place the ANALYZING UNIT POWER to ON.
2.4.5
Verification The system will automatically reset the bar and wash it with deionized water.
[email protected]
2.5 Replacing Filter Assembly 2.5.1 Part Number BA38-30-88299
Diluted Detergent filter assembly
2.5.2 Purpose Keep reliable testing results 2.5.3 Interval Every 6 months 2.5.4 Procedures 2.5.4.1 Make sure analyzing unit is not running tests. 2.5.4.2 Wear rubber gloves. Take out a soft pipe (inner diameter: 1/8 ”, length: 50mm) from the supplies bag. Connect one end to new filter assembly, and the other to disposable syringe. Inject water to the filter through the syringe till water comes out of the filter. Take off the soft pipe and syringe. The purpose is to increase the weight of filter assembly, so as to be placed at the bottom of wash solution bucket.
[email protected]
2.5.4.4 Mount new filter assembly to the end of wash solution bottle soft pipe 2.5.4.5 Put wash solution bottle cap assem bly back into wash solution bucket.
2.5.5 Verification Select S ys tem Pr i me in Utilities screen. Click Execute. Then select Pr ime Was Was h Unit to ensure there are wash solution in Phase1 and 2 wash probes, and remove air bubbles.
[email protected]
2.6 Replacing DI Water Water Filter 2.6.1 Part Part Number BA40-30-61993
Water Inlet Filter
2.6.2 Purpose To keep the quality of water and make the flow expedite 2.6.3 Interval Every 6 months When you find the inlet water flow is not expedite When you find the DI water filter f ilter changes color seriously 2.6.4 Procedures 2.6.4.1
Ensure the Power is placed to OFF.
2.6.4.2
Cut off the power supply of the water suppl y module.
2.6.4.3
The DI water filter is placed outside the analyzer, and connected to DI water tubing. Pull out the connecting pipes on both ends of the old filter assembly to take out the old filter assembly.
[email protected]
2.7 Replacing First and Second Phase Washing Tube on the Wash Unit 2.7.1 Part Number 3001-10-07069
Tubing,1/16"X1/8",Tygon
2.7.2 Purpose To keep the quality of Washing 2.7.3 Interval Every year When turn dirty or yellow after a period 2.7.4 Procedures 2.7.4.1 Place the Po wer to OFF. OFF. 2.7.4.2
Open the back cover. Disconnect the dirty tubing of the aspiration wash probe(the longer straight probe). The No. for the first phase washing tubing is 12, and 11 for the second phase washing tubing. Pull the tubing out from the tubing f ixer. Take Take off the pipe m arker for spare use.
[email protected]
2.8 Replacing Cuvette Spring 2.8.1 Part Part Number BA40-20-73118
Cuvette Spring
2.8.2 Purpose For reliable testing results 2.8.3 Interval After the C uvette/L uvette/L amp amp C heck if a cuvette is found to be damaged
2.8.4
Procedures
2.8.4.1
Place the Power to OFF
2.8.4.2
Manually rotate the probes and mixers to a position convenient for cuvette replacement and then remove the reaction disk cover.
[email protected]
2.9 Replacing Check Valve Valve 2.9.1 Part Part Number M6Y-020003--M6Y-020003---
Check Valve Valve
2.9.2 Purpose As consumables and and keep the flow unobstructed unobstructed 2.9.3 Interval Every 6 Months 2.9.4 Procedures 2.9.4.1
Place the Power to OFF
2.9.4.2 Remove right and rear plate. Locate the four check valves as shown in following figure.
[email protected]
ensure reliable connection. 2.9.4.4 Check if tubings are connected properly. Pay attention to the direction of check valves when replacing. After replacing, two check valves should guarantee the flow direction in the tubing as: D7 →D8, D9 → D10, D3→D11, D4→D12. Refer to following figure for check valve flow direction
2.9.4.5 Reinstall the right and rear plate.
2.9.5 Verification Perform startup initialization after replacing, to check the tubing for leak age.If not,the replacement is ok.
[email protected]
3. Chemistry Maintenance schedule New BS-380 Maintenance Schedule I: Inspect. Clean, Adjust or Replace if necessary.
C: Clean,
R: Replace,
A: Adjust,
ITEM
L: Lubricate,
P: Perform,
N: No action needed
Related Part
Qty
WHICHEVER COMES FIRST EVERY
DI Water Tank
r e m o t s u C y B
N/A
I
C
C
C
C
Reference
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5.
High Conc entrate Was te tank
N/A
N/A
I
C
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5.
Dilut ed Detergent Tank
N/A
N/A
I
C
C
C
C
Was h with water
Inlet Tubing of Cap assembly of Diluted Detergent Tank
N/A
N/A
N/A
I
I
I
I
I
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
N/A
Outlet tubing of Cap assembly of High concentrate Waste tank
N/A
N/A
I
I
I
I
I
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Reagent /Sample Dis k
N/A
N/A
N
C
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Was h Wells of probe and mix ing bar
N/A
N/A
N
C
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Dus t Proof Net
N/A
N/A
N
N
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Sample/ Reagent Bar Code Reader Windows
N/A
N/A
N
C
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Cleaning P anels of Analy z ing Unit
N/A
N/A
N
C
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Clean the Rotor of Reagent/Sample Probe/Mixer Drive Assembly
N/A
N/A
N
N
C
C
C
Wi Wipe the rotor with c lean gauze.
Wipe Bloc k s
N/A
N/A
N
N
C
C
C
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
Was te Tube
N/A
N/A
I
I
I
R
R
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
IS E daily cleaning
N/A
N/A
P
P
P
P
P
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
IS E pump c alibration
N/A
N/A
P
P
P
P
P
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
IS E Na elec trode
N/A
N/A
10,000 tests or 6 months
I
I
I
I
I
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
IS E K elec trode
N/A
N/A
10,000 tests or 6 months
I
I
I
I
I
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
I
I
I
I
I
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
I
I
I
I
I
BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5 BS-380 Chemistry Analyzer Operator’s Manual(V3.0) Chapter 5
IS E Cl elec trode
N/A
N/A
10,000 tests or 6 months
IS E Li elec trode
N/A
N/A
3,000 tests or 6 months 10,000 tests or 6 months
I
I
I
I
I
2000 hours
N
N
I
R
R R
IS E Ref elec trode
N/A
N/A
Lamp
115-008803-00
Halogen Lamp(20W )
2
Sample probe
BA 40-30-61525
Sample probe
1
I
C
C
C
Probe was her
0040-10-32307
W as her for P robe
4
I
C
C
C
R
Probe unc log tool
115-003182-00
P robe Unc log Tool
1
N
N
P
P
P
Acupuncture Needle
r e m o t s u C y B
N/A
S y h s T y h l l y k t N H l t i n e O o n M o M D a W e 2 M 6 1
0000-10-10916
Acupuncture Needle
10
N
N
P
P
P
Reagent Sy ringe Plunger Tip (500ul)
0040-10-32316
500ul Sy ringe Plunger Tip
1
100,000 t imes
I
I
I
I
R
Sample Sy ringe (100ul)
3100-10-49438
S y ringe(100ul)
1
100,000 t imes
I
I
I
I
R
Sy ringe was her
0040-10-32303
W as her for Sy ringe
4
I
C
C
C
R
1
I
C
C
C
R
Mixing Bar + Retaining Nut
BA31-20-41651 + BA30-20-06774
Mix ing B ar + Retaining Nut
Water Inlet Filt er
BA 40-30-61993
Water Inlet Filter
2
N
N
N
R
R
Was h Solution Filter
BA 38-30-88299
Diluted Detergent filter as s embly
2
N
N
N
R
R
W as h Uni t Tube for P has e 1, 2 and 3
3001-10-07069
Tubi ng, 1/ 16" X1/ 8" ,T ,Ty go gon
15ft
N
N
N
N
R
Cuvette S pring
BA 40-20-73118
Cuvette S pring
10
I
I
I
I
I