Service manual METTLER TOLEDO MultiRange ID1 Plus / ID1 Plus-A weighing terminals terminals
Bild
M E T T L E R
I D 1 P
l u s
ID1 Plus / ID1 Plus-A weighing terminals
Contents
Contents
Page
1
Checklists .........................................................................................................................................
2
1.1
Maintenance checklist .........................................................................................................................
2
1.2
Service checklist .................................................................................................................................
3
2
Spare parts .......................................................................................................................................
4
2.1
ID1 Plus exploded view diagram, with interfaces ....................................................................................
4
2.2
ID1 Plus-A exploded view diagram, analog connection ...........................................................................
5
2.3
Spare parts list ...................................................................................................................................
6
3
Error displays ....................................................................................................................................
8
3.1
Operating errors..................................................................................................................................
8
3.2
Error messages ..................................................................................................................................
8
4
Troubleshooting ................................................................................................................................
9
5
Corrective maintenance .....................................................................................................................
10
5.1
Opening the terminal ...........................................................................................................................
10
5.2
Checking the voltages .........................................................................................................................
10
5.3
Removing the main board ...................................................................................................................
10
5.4
Changing the microfuses .....................................................................................................................
11
5.5
Changing the program ........................................................................................................................
11
5.6
Changing the main board ....................................................................................................................
11
5.7
Checking and changing the transformer, setting the power supply voltage .................................................
12
5.8
Changing the keypad ..........................................................................................................................
12
5.9
Changing the analog board .................................................................................................................
13
5.10
Installing the main board .....................................................................................................................
13
5.11
Closing the terminal ............................................................................................................................
13
6
Settings ............................................................................................................................................
14
6.1
Configuring terminal and weighing platform ...........................................................................................
14
6.2
Preparing the identcard (ID1 Plus only) .................................................................................................
14
7
Software compatibility .......................................................................................................................
15
8
Pin assignment .................................................................................................................................
15
Service manual 00705863D
08/02
1
Checklists
1
Checklists
1.1
Maintenance checklist
1.1.1
Visual inspection
ID1 Plus / ID1 Plus-A weighing terminals
• Check condition of the following scale components: – – – – –
Housing Keypad Protective hood Weighing platform Peripherals
• Check condition of the following cables: – Power cable – Weighing platform connection cable – Data transfer cable (if used) • Check for protected position of cables. 1.1.2
Function check
• Press power switch. • For verified scales, check the ID code (setting on ID card and display in the terminal must match after pressing the "Test key"; see installation instructions or operating instructions, Installation section). • Carry out calibration and display checks with the "Test key". • Check functions by making entries via the keypad (see operating instructions). • Check the following settings of the weighing platform (see service manual of the connected weighing platform): – – – –
Calibration Cornerload Linearity Hysteresis
• Check the following plug-in connections for firm seating: – – – –
2
Power cable Weighing platform connection cable Data transfer cable (if used) Peripheral connection (if installed)
Service manual 00705863D
08/02
ID1 Plus / ID1 Plus-A weighing terminals
1.2
Checklists
Service checklist
Carry out the following check procedure at the terminal and the weighing platform before troubleshooting and after servicing. 1.2.1
At the terminal
• Check whether an operating error is present. • Check power and weighing platform connection cables. • Check calibration via the "Test key". • Check functions – entries via keypad (see operating instructions). • Carry out display, keypad and printer tests; for this purpose: – – – – – – – – –
Unplug terminal. Unplug weighing platform. Switch terminal on. PLUG IN appears in the display. Press SET TO ZERO key until SERVICE MODE appears in the display. Press Enter key until all display elements illuminate. Press all keys in sequence from left to right. If a GA46 printer is connected, a test printout is printed automatically after the last key is pressed. "END?" appears in the display. The test routine can be repeated by pressing the Tare key. Press the Enter key to return to weighing mode.
• Check all plug-in connections and program EPROM for firm seating. • Check connected peripherals. 1.2.2
At the weighing platform
• Check that transport locks of the weighing platform have been released properly and are in the correct position. • Check supports of the weighing platform. • Check that the load plate is free and has no contact with the surroundings. • Check maximum capacity and linearity of the weighing platform. • Check play of all stops and limits. • Check that the lever system is free on all sides. • Check knife edges, bearings and flexible bearings.
Service manual 00705863D
08/02
3
Spare parts
ID1 Plus / ID1 Plus-A weighing terminals
2
Spare parts
2.1
ID1 Plus exploded view diagram, with interfaces
4
Cable connections for example cable A to connector A Service manual 00705863D
08/02
Spare parts
ID1 Plus / ID1 Plus-A weighing terminals
2.2
ID1 Plus-A exploded view diagram, analog connection
Service manual 00705863D
08/02
5
Spare parts
2.3
ID1 Plus / ID1 Plus-A weighing terminals
Spare parts list
Item
Designation
Part No. ID1 Plus-A
1
Cover, compl., with display window and keypad
00 505 326
00 505 326
2
Keypad
00 203 585
00 203 585
3
Identcard S, short
00 504 468
–
4
Cover seal
00 203 594
00 203 594
5
Power cable 00 504 247 00 504 248 00 504 249 00 504 250 00 504 251 00 504 252 00 504 253
00 504 247 00 504 248 00 504 249 00 504 250 00 504 251 00 504 252 00 504 253
-
6
Part No. ID1 Plus
Euro GB DK CH USA ZA AUS, RA
6
Ring seal
00 203 654
00 203 654
7
Leveling foot
00 504 186
00 504 186
8
O-ring 16 x 1.25
00 203 656
–
9
IDNet connector
00 203 598
–
10
Transformer (set to 230 V)
00 203 587
00 203 587
11
Insulating plate
00 203 655
00 203 655
12
Spacer
00 504 470
00 504 470
13
Main board ID1 Plus
00 203 578
00 203 578
14
Program ID1 Plus
00 505 303
00 505 303
15
Microfuse slow-blow 315 mA - L 250 V
00 087 431
00 087 431
16
Microfuse slow-blow 1.25 A - L 250 V
00 087 805
00 087 805
17
Microfuse slow-blow 2.5 A - H 250 V
00 089 922
00 089 922
18
Interface 101 CL20 mA Interface 102 RS232 Interface 104 RS422/485
00 505 237 00 505 238 00 506 847
00 505 237 00 505 238 00 506 847
19
Interface 103 digital I/O
00 505 239
00 505 239
Service manual 00705863D
08/02
Spare parts
ID1 Plus / ID1 Plus-A weighing terminals
Item
Designation
Part No. ID1 Plus
Part No. ID1 Plus-A
20
Analog board (point ADC board) without Eprom
–
00 507 000
21
Point cable
–
00 507 003
22
Minispreader
–
00 204 130
23
Reducing ring
–
00 506 860
24
Cable gland, heavy gauge 7
–
00 201 887
25
Sealing sleeve
–
00 507 008
26
Grub screw M3 x 6 DIN 913
–
00 205 958
27
Verification set (for updating Identcard bracket)
22 000 386
22 000 386
28
Eprom for analog board
–
00 507 274
Set of labels
–
00 507 010
Service manual 00705863D
08/02
7
Error displays
3
Error displays
3.1
Operating errors
ID1 Plus / ID1 Plus-A weighing terminals
Incorrect operations at the terminal and operating conditions of the weighing platform under which a weight determination is not admissible or not possible are displayed by the terminal in clear text, see operating instructions.
3.2
Error messages
If an operational fault is identified internally by the terminal, the faulty component is shown in the display.
Error messages ID1 Plus / ID1 Plus-A terminals Error messages
Meaning
Testing
Component
ERR ID1
Error terminal
Check all plug-in connections and EPROM for firm seating.
EPROM main board ID1 Plus
ERR I-F
Error interface
Check attached peripherals and connections. Check EPROM.
EPROM main board ID1 Plus
ERR BRU
Error weighing platform
Check connection cable of measuring cell.
Measuring cell
CAL ERR
Calibration error
Perform calibration again.
Measuring cell
Clearing error message: Switch terminal off and then on again. If error is redisplayed: Replace faulty component, see section 5, Corrective maintenance or service manual of attached weighing platform.
8
Service manual 00705863D
08/02
Troubleshooting
ID1 Plus / ID1 Plus-A weighing terminals
4
Troubleshooting Fault
Possible causes
Fault rectification
Terminal can not be switched on
• Power cable not plugged in
• Establish connection to power supply • Check power cable • Check microfuses on main board and replace if necessary (see 5.4) • Switch on display • Change main board (see 5.6)
• No line voltage • Fuse blown • Display switched off • Main board faulty • Weighing platform not attached to terminal • Connection cable of weighing platform faulty • Fuse blown
After terminal switched on, - PLUG IN - appears
• Measuring cell faulty
• Ribbon cable for weighing platform connection on main board (item H in exploded view diagram) not attached • Analog board faulty • Point cable faulty After terminal switched on, no weight display
• Transport safeguards of weighing platform not released • Mechanical system touching somewhere • Main board faulty • Analog board faulty • Point cable faulty
No data transfer to measuring cell
• Replace connection cable • Check microfuses on main board and replace if necessary (see 5.4) • Replace measuring cell (see service manual of attached weighing platform) • Plug in weighing platform cable
• Change analog board (see 5.9) • Change point cable • Release transport safeguards (see operating instructions of attached weighing platform) • Adjust frame and lever system (see service manual of attached weighing platform) • Change main board (see 5.6) • Change analog board (see 5.9) • Change point cable
• Analog board faulty • Point cable faulty
• Check microfuses on main board and replace if necessary (see 5.4) • Change main board (see 5.6) • Replace measuring cell (see service manual of attached weighing platform) • Change analog board (see 5.9) • Change point cable
• • • •
• • • •
• Fuse blown • Main board faulty • Measuring cell faulty
No entry via keypad possible
• Attach weighing platform
Keypad faulty Main board faulty Analog board faulty Point cable faulty
Replace keypad (see 5.8) Change main board (see 5.6) Change analog board (see 5.9) Change point cable
Note See also operating instructions of ID1 Plus or ID1 Plus-A, section "What if...?". Service manual 00705863D
08/02
9
Corrective maintenance
5
ID1 Plus / ID1 Plus-A weighing terminals
Corrective maintenance
Caution Before opening the instrument: • Disconnect power plug. • Unplug attached weighing platform and peripherals.
5.1
Opening the terminal
• Unscrew 2 screws on underside of the cover at the front and lift off cover. Caution Do not scratch coating on the inside of the display window, if need be simply wipe carefully using a soft cloth. • Unplug keypad cable (J) at main board.
5.2
Checking the voltages • Open terminal, see section 5.1. • Measure voltages at ribbon cable on main board at bottom right.
STP 10
1 BRC1
Pin
Assignment
1, 2 3 4 5 6 7 8 9 10
GND VDIS VNOR VCC VOPT2 VOPT1 Vheat2 Vheat1 VBA2
Value 47 VDC 16 VDC 5 VDC
} }
12 VAC 7
VAC
3.2 VDC
• If the voltages are not correct, check microfuses and transformer, see sections 5.4 and 5.7.
5.3
Removing the main board
• Open terminal, see section 5.1. • Unplug weighing platform cable (H) on right of main board next to the display. • Unscrew heavy gauge screw fitting and push in power cable approx. 5 cm. • Unscrew 2 screws on main board and lift main board out of guide. • Unplug interface cable (A). • Unplug 6 connectors on underside and lift out board.
10
Service manual 00705863D
08/02
Corrective maintenance
ID1 Plus / ID1 Plus-A weighing terminals
5.4
Changing the microfuses • Remove main board, see section 5.3. • Check microfuses and replace if necessary. Fuse ratings SI2
SI3
SI 1
slow blow 2.5 A - H 250 V
SI 2
slow-blow 1.25 A - L 250 V
SI 3
slow-blow 315 mA - L 250 V
SI 5
slow-blow 315 mA - L 250 V
SI5 SI1
5.5
• Reinstall main board, see section 5.10.
Changing the program
• Open terminal, see section 5.1. • Remove old EPROM program from socket using IC extractor. • Insert new EPROM program in the IC socket paying heed to the marking on the IC and the socket. Carefully bend IC contacts straight. • Close terminal, see section 5.11.
5.6
Changing the main board
• Remove main board, see section 5.3. • Remove 5 spacers; insulating plate, top and bottom are separate parts. • Break off new main board into top and bottom. • Mount insulating plate between top and bottom with the 5 spacers. • Pull off program from old main board socket using IC extractor and insert in the IC socket of the new main board paying heed to the marking on the IC and the socket. Carefully bend IC contacts straight. • Reinstall main board, see section 5.10. Caution The battery must not be changed as improper handling can lead to an explosion hazard. Disposal of the board with battery only by the manufacturer!
Service manual 00705863D
08/02
11
Corrective maintenance
5.7
ID1 Plus / ID1 Plus-A weighing terminals
Checking and changing the transformer, setting the power supply voltage Checking the transformer • Change main board, see section 5.3. • Check voltages on secondary side.
V 0 5 2
7
S1
S9
S2
S10
V 5 1 1
6
5
4
3
2
1
Pin
Rated current
Voltage at rated current
Open-circuit voltage
S1, S2 S4, S5 S5, S6 S7, S9 S8, S10
0.95 A 0.225 A 0.225 A 0.13 A 0.17 A
12 V ± 5 % 3.2 V ± 5 % 3.2 V ± 5 % 32 V ± 5 % 10 V ± 5 %
13.8 V 4.0 V 4.0 V 36.5 V 11.6 V
• If the tolerances are not complied with, change transformer. Changing the transformer, setting the power supply voltage P7
5.8
S1
• Remove transformer.
S2 S7
• With new transformer, check connections on primary side:
P5
S9 S10
P3
S8 S4 S5 S6
for 230 V
connections 3 and 7 (factory setting)
for 115 V
connections 3 and 5
• To switch to 115 V, carefully undo and lift off cover, unsolder and resolder connection accordingly. • Install transformer. • Reinstall main board, see section 5.10.
Changing the keypad
• Open terminal, see section 5.1. • Pull off affixed foilsheet keypad from display window. • Remove residual adhesive and dirt from adhesion surface of the display window. • Pull protective film off new keypad. • Lead connection cable through the cutout in the display window. • Affix keypad and smooth down firmly. • Close terminal, see section 5.11.
12
Service manual 00705863D
08/02
Corrective maintenance
ID1 Plus / ID1 Plus-A weighing terminals
5.9
Changing the analog board • Remove main board, see section 5.3. • Disconnect weighing platform cable and point cable. • Undo nut and screw and lift out analog board. • Disengage minispreader from old analog board and lock in place on new board. • Plug in analog board and install with screw and nut. • Plug in weighing platform cable (1) and point cable (2). • Reinstall main board, see section 5.10.
2
5.10
1
Installing the main board • Plug 6 connectors into bottom part of main board. • Plug weighing platform connector and interface connector, if used, into top part of main board. • Route power cable outward and tighten heavy gauge screw fitting. • Insert main board in the guide and fix with 2 screws.
ST9 ST7
ST4 ST6
• Close terminal, see section 5.11.
ST3
Pin assignment
ST5
ST 3 ST 4 ST 5 ST 6 ST 7 ST 9
5.11
L1 (brown) N (blue)
}
transformer shield bonding
Closing the terminal
Caution If necessary, wipe display window on the inside very carefully using a soft cloth! • Plug in keypad cable. • Engage cover at top in screw heads and fasten at front using 2 screws on underside of cover.
Service manual 00705863D
08/02
13
Settings
ID1 Plus / ID1 Plus-A weighing terminals
6
Settings
6.1
Configuring terminal and weighing platform
6.1.1
ID1 Plus-A
All other information on the ID1 Plus-A terminal is contained in the operating instructions for the ID1 Plus-A. 6.1.2
ID1 Plus
For the ID1 Plus terminal, the service mode settings are described in the service manual for the service mode of the connected weighing platform. Service mode operation = YES
Program blocks are selected the same way in service mode as in master mode.
= NO
Entering service mode • Press and hold the Enter key until MASTER appears in the display. • Press the NO key; SERVICE appears in the display. • Press the YES key to enter service mode; the CODE prompt appears. • Enter the following code: YES NO NO YES. The first service mode block, RETURN, appears.
6.2
Preparing the identcard (ID1 Plus only)
6.2.1
New identcard (socket)
In the factory, the identcard is provided with the measurement data plate of the standard configuration of the weighing platform, the identcode shown is "1". For mounting the identcard, see installation information for the identcard. 6.2.2
Old identcard (bracket) Putting into operation after changing the weighing platform configuration
1
• Remove identcard. • Affix appropriate measurement data plate (1).
1 p p p p p p p p p p p p p p p p p p p
p p
p p p p p
p p
p p p p
1
p p p p
p p
p p
p p
p p
p p
p p p
• Adjust identcode disk (2) so that it matches the displayed identcode.
p p p p
p p p p p p p
2
• Assemble identcard. • With certified scales, seal screw (3).
3
6.2.3
Updating from old identcard (bracket) to new identcard (socket)
Therefore you need the calibration set 22 000 386. The procedure is described in the installation information for the identcard. 14
Service manual 00705863D
08/02
Software compatibility / Pin assignment
ID1 Plus / ID1 Plus-A weighing terminals
7
Software compatibility
The ID1 Plus weighing terminal is compatible with all weighing platforms of the F, K, N and D series. With the ID1 Plus-A terminal: see ID1 Plus-A operating instructions, section "Technical data".
8
Pin assignment Weighing platform connection B
A TXD+, transmitting loop of the weighing platform
K
A
C
J L
C VNOR 13 V M H
D
E
B VDIS 40 V
F
G
D RXD+, receiving loop of the weighing platform F
RXD–, receiving loop of the weighing platform
G Protective ground Front view
H Signal ground J TXD–, transmitting loop of the weighing platform The pin assignment of the ID1 Plus-A is described in the ID1 Plus-A operating instructions.