This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
L3 Security and Detection Systems Monoblock® Xray Generator Specification The XRB150P&N135X3408 Monoblock® is used in a Baggage Inspection Application. This Monoblock® consists of an integrated X-ray tube, a dual output high voltage power supply and a filament supply with control circuitry. This Monoblock® is programmed and monitored via an RS232 Interface. The unit is built in accordance to the following specifications: 1. INPUT POWER: Input Line Voltage Range:
100VAC – 264VAC, 50Hz / 60Hz, 3 Amps RMS maximum.
2. OUTPUT VOLTAGE PERFORMANCE: 2.1
Tube Voltage Operational Range: The nominal high voltage output range shall be 80kV to 150 kV (+/40kV to +/-75kV), with an absolute maximum output voltage of 153kV (+/- 76.5kV).
2.2
kV Accuracy: The High Voltage measured at the X-ray tube shall be within +/-2% of the selected value.
2.3
kV Ripple: The low frequency ripple (<1kHz) shall be 1% pk-pk of the maximum kV, measured across the X-Ray tube.
2.4
Voltage Regulation: < 0.05% for ±10% of nominal input line change < 0.05% for 150uA to 900uA load change.
2.5
mA/kV Rise Time: The mA and kV rise time shall be controlled by a ramp up circuit. Ramp time less than 200 mS from 10% to 90% of the output voltage and current.
2.6
kV Overshoot: The kV Overshoot will return within 5% of full output voltage in less than 10ms.
3.0 X-RAY TUBE, FILTER AND EMISSION EMISSION CURRENT PERFORMANCE: PERFORMANCE: SHV p/n 105739-033
3.1 X-Ray Tube:
3.2 Beam Filter: 5052 aluminum, 0.080” thick, ±0.01” inch window, 6061 aluminum filter, .016 thick +/-0.005, secured but not obstructed by RTV. 3.3 Aperture Cover:
Mylar® Tape .002” thick X 2.0” wide, adhered over aperture opening
3.4 Tube Current:
150uA to 900uA @ 150kV, 135 Watts maximum
3.5 mA Regulation:
<0.5% at 80 -150 kV, 150uA to 900uA.
3.6 Tube Envelope Thickness: Spellman shall receive and maintain glass window thickness measurements for each SHV p/n 105739-033 tube, traceable by the tube serial number. Average glass thickness shall be 1.7mm +/- 10%. UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 \\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
1
of 11 Sheet 2 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
4.0 PROTECTION CIRCUITRY: 4.1 Current protection: The OVER CURRENT (High mA) trip point is set at 18% over programmed output current value. The UNDER CURRENT (Low mA) trip point is set at 18% under programmed output current value. 4.2 Voltage protection: The OVER VOLTAGE (High kV) trip point is when the kV output exceeds 165kV for more than 10ms. The UNDER VOLTAGE (Low kV) trip point is set at 10% under the programmed output voltage value. 4.3 Over temperature protection: Over temperature trip point shall be set within 65°C to ± 5°C 5.0 FILAMENT POWER SUPPLY: Filament Current is provided to the X-ray tube f ilament via a two-wire system, referenced to Cathode (-) High Voltage Output. The wires are defined as Filament and Filament common. 5.1 Filament Current Output:
3.2 ± 0.2 Amps RMS
5.2 Filament Voltage Output:
3.5 Volts RMS
6.0 SAFETY AND QUALITY DOCUMENTATION REQUIREMENTS: 6.1 Data Items: To be provided prior to any L3 approved change to any part of this documented specification. 6.1.1
X-Ray Leakage Diagram Not to be greater than 0.5mR/hr at 5cm outside the external surface per FDA 21 CFR 1020.40 and OSHA 29 CFR 1020.96. 6.1.1.1 Indicate date, model #, and serial # of the unit. 6.1.1.2 Indicate model # and serial # of the X-ray tube. 6.1.1.3 Indicate kV and mA of the unit during the test. 6.1.1.4 All the equipment used for production units shall be calibrated per the requirements of the ISO 9000 standard. In addition, the type, model, serial # and calibration due date of any radiation survey meters used must be recorded.
6.1.2
Process Data Sheets Indicating test and burn-in data. 6.1.2.1 Indicate date, model #, and serial # of the unit. 6.1.2.2 Indicate model # and serial # of the X-ray tube. 6.1.2.3 Indicate kV, mA, duration (in hours), cycle (in seconds), date/time started, and date/time stopped for each step of the current ramp. 6.1.2.4 Indicate pass/fail of radiation leakage test. 6.1.2.5 All the equipment used for production units shall be calibrated per the requirements of the ISO 9000 standard. In addition, the type, model, serial # and calibration due date of any radiation survey meters used must be recorded.
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
2
of 11 Sheet 3 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
6.1.3
Sticker on Unit 6.1.3.1 Indicate date, model #, and serial # of the unit 6.1.3.2 Indicate serial # of the X-ray tube
6.1.4
Certificate of Conformance to be kept with the unit. 6.1.4.1. L3 purchase order number 6.1.4.2. Indicate date, model, part #, revision and serial # of the unit 6.1.4.3. Indicate model #, serial #, and manufacturer of the X-ray tube 6.1.4.4. Supplier QA buy off
6.2 Data Delivery Requirements: Items to be created for each unit. 6.2.1 All items in section 6.1 are to be included with each unit delivered to L3 for first article approval. 6.2.2 Items 6.1.3 and 6.1.4 are to be included with each production unit delivered to L3. 6.2.3 Item 6.1.2 is to be maintained at supplier for each unit supplied and furnished on demand to L3. 6.3 Ground fault compliance The construction of the tank shall comply with CDRH requirements of ground fault protection, that is, a ground fault to any tank input cannot result in activating x-rays for a tank that is OFF, or in increasing the kV or mA of a tank that is ON, or in preventing the x-ray output from being de-activated. 7.0
PHYSICAL SPECIFICATIONS 7.1
Environmental: Operating ambient temperature: 0 °C to 40°C. Storage ambient temperature: -40 °C to +70°C
7.2
Humidity: 95% non-condensing
7.3
Dimensions: 18.0”L X 13.5”W X 6.25”H
7.4
Weight: 120 Lbs
7.5
Cooling method: Through natural convection/ customer supplied external cooling fans
7.6
Compliance Standards: Unit shall be designed to comply with Safety standard, EN 61010 –1 and EMC s tandards, EN 50081 –1 & EN 50082 –1. With an external filter, Delta p/n: 03DRCG5 Oil leakage from tank: Oil leakage from the tank is not acceptable . Vibration tolerance: The tank is able to withstand 10 g during the shipping.
7.7 7.8
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
3
of 11 Sheet 4 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
7.9
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
Geometry: 7.9.1
Angular beam spread: The beam angular coverage will be 80 deg with the beam plane perpendicular to the X-ray tube axis. With respect to the face plate angular coverage will range from –20 deg to + 60 deg.
7.9.2
Orientation: The tank shall be operable when placed on any of its sides: Top, Bottom, Left, Right, Front or Back.
7.9.3
Beam filtration Beam filtration consists of the oil, aluminum window and aluminum filter. (see Par. 3.2 for beam filter spec).
8.0
INTERFACE SPECIFICATIONS: 8.1 AC Power Connector (Molex p/n 26-60-4060, pc mtg 0.156 ctr, 06 pos) Voltage: 100VAC – 264VAC, 50Hz / 60Hz, • Current: 3 Amps RMS maximum • Pin 1 – Line Pin 2 – Removed Pin 3 – Neutral Pin 4 – Removed Pin 5 – Spare Pin 6 – Spare Ground stud provided on chassis. 8.2 Analog Signals Interface (local Setting) –– J7(pc mtg rt 07 pos 0.156 ctr): (PC Mtg RT Ag 07 Pos 0.156 Ctr, Molex #: 26-60-5070) + X-ray Signal (Exgate) Input signal: Analog signal • +12VDC = Enable X-ray • Low (or Open) = Disable X-ray • Connections: J7, pin 1 • X-ray Signal Return Connections: J7, pin 2 • Removed Connections: J7, pin 3. •
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
4
of 11 Sheet 5 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
VMTR Signal (KV monitor) Output signal: Analog signal • Voltage: 0 to +10.00 V max • Scale Factor: 0 – 10.00 VDC = 0 - 170 kV • Connections: J7, pin 4 • Signal Ground Connections: J7, pin 5 • IMTR Signal (mA monitor) Output signal: Analog signal • Voltage: 0 to +10.00 V max • Scale Factor: 0 – 10.00 VDC = 0 – 1.10mA • Connections: J7, pin 6 • Fault Signal Output signal: Open Collector • High (Open) = No Fault • Connections: J7, pin 7 8.3 Digital Signals Interface (RS232), 9 Pin D-male – J5: (See RS232 Specification, Par. 10) Transmitted Data Signal: Conforms to EIA RS-232-C. • Connections: J5, pin 2 • Received Data Signal: Conforms to EIA RS-232-C. • Connections: J5, pin 3 • Signal Ground Connections: J5, pin 5 • + X-ray Signal Input signal: Analog signal • +12VDC = Enable X-ray • Low (or Open) = Disable X-ray • Connections: J5, pin 8 • Spares Connections: J5, pin 1, 4, 6, 9. • UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
5
of 11 Sheet 6 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
9.0
LED INDICATORS: LED1 (Error 6) – Illuminated when High kV occurs, OV LED2 (Error 5) – Illuminated when Low kV occurs, UV LED3 (Error 4) – Illuminated when Low mA occurs, UC LED4 (Error 3) – Illuminated when High mA occurs, OC LED5 (Error 2) – Illuminated when Arc F ault occurs, ARC LED6 (Error 1) – Illuminated when Over temperature occurs, OT LED7 – Illuminated when X-rays are enabled, XRAY ON LED8 – Illuminated when power is on, PWR ON
10.0
X3408 COMMUNICATIONS INTERFACE SPECIFICATION 10.1
Scope: This section defines the requirements for the communications interface on the X3408 high voltage power supply. The communications interface will be used to program voltage and current monitor the output voltage and current, control various functions and report status and faults.
10.2
INTERFACE CONNECTIONS: The connector will be a standard 9-pin DB (female) connector. The pin assignments will be as follows: Pin 1 2 3 4 5 6 7 8 9
10.3
Name Spare Transmitted Data Received Data Spare Signal Ground Spare Spare X-Ray Enable Spare
Usage RS-232 RS-232 RS-232
+12 VDC = Enable X-Ray
COMMUNICATION PARAMETERS: Baud Rate 9600 Parity Even Data Bits 8 Stop Bits 1
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
6
of 11 Sheet 7 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
10.4 Command StructureCommands sent from the host computer to the power supply shall be structured as an ASCII string as shown below: CMD ARG; CMD ARG
Start of message. The Hex value 0x02 used to indicate the start of a message. Command. The commands are defined below. Argument. When a command has an argument there will be one ASCII space between the command and the argument.
;
Semi-colon. The ASCII value for Semi-colon (Hex 0x13) will be used as a End of Message terminator Checksum. The Check Sum is calculated by adding all bytes following the to th (and including) the ‘;’. Take the 2’s complement. Set D 7 (the 7 bit of D7 – D0) to a logical zero and D6 to a logical one. Carriage Return. The ASCII value for carriage return (Hex 0x13) will follow the checksum. Line Feed. The ASCII value for Line Feed (Hex 0x10) will follow the Carriage Return.
10.5
Report Structure Reports sent from the power supply to the host computer shall be structured as an ASCII string as shown below: RPT ARG; RPT ARG
Start of message. The Hex value 0x02 used to indicate the start of a message. Report. The reports are defined below. Argument. When a command reports an argument there will be one to four ASCII characters between the start of the message and the semi-colon.
;
Semi-colon. The ASCII value for Semi-colon (Hex 0x13) will be used as a End of Message terminator Checksum. The Check Sum is calculated by adding all bytes following the to th (and including) the ‘;’. Take the 2’s complement. Set D 7 (the 7 bit of D7 – D0) to a logical zero and D6 to a logical one. Carriage Return. The ASCII value for carriage return (Hex 0x13) will follow the checksum. Line Feed. The ASCII value for Line Feed (Hex 0x10) will follow the Carriage Return.
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
7
of 11 Sheet 8 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
10.6
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
COMMAND AND REPORT DEFINITIONS: Note – All commands have a response. Command and Report Definitions Argument Response 0 – XXXX ; 0 – XXXX ; N/A XXXXX; N/A XXXXX; N/A ; N/A XXX; N/A X;
Name Voltage Reference Current Reference Voltage Monitor
Command VREF IREF VMON
Current Monitor
IMON
Fault Clear Report Fault HV Status
CLR FLT STAT
X-Ray Enable
ENBL
X
;
WatchDog Timer
WDTE
X
;
WatchDog Timer Tickle
WDTT
N/A
;
Comments XXXX = 0 - 1530 XXXX = 0 - 900 XXXX = 0 - 1530 XXXX = 0 - 900
XXX = See Below X = 1 is X-Ray On X = 0 is X-Ray Off X = 1 Enable X-Ray X = 0 Disable X-Ray X = 1 Enable Timer X = 0 Disable Timer
10.6.1 VREF COMMAND: The VREF command specifies the tank kV in 0.1 kV increments. Thus to set the voltage to 150 kV, an argument of 1500 is used. The response from the tank for this command is always ;. 10.6.2 IREF COMMAND: The IREF command specifies the tank current in 1-microAmp increments. Thus to set the current to 900uA, an argument of 900 is used. The response from the tank for this command is always ;. 10.6.3 VMON COMMAND: The VMON command requests the tank to transmit its measured tank voltage. The command has no argument. The response from the tank is: XXXX; where XXXX is the voltage in .1 kV increments. 10.6.4 IMON COMMAND: The IMON command requests the tank to transmit its measured tank current. The command has no argument. The response from the tank is: XXXX; where XXXX is the current in microAmps. 10.6.5 CLR COMMAND: The CLR command instructs the tank to clear (reset) all fault conditions. The command has no argument. The response from the tank for this command is always ;. UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
8
of 11 Sheet 9 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
10.6.6 FLT COMMAND: The FLT command requests the tank to transmit its fault status. The command has no arguments. The response from the tank is: XXX; where XXX is a number that represents a certain fault. See fault definitions below. 10.6.7 STAT COMMAND: The response to the STAT command is X; where X is 1 if x-rays are being produced and 0 if no X-rays are being produced. 10.6.8 XRAY COMMAND: The ENBL command enables or disables X-ray production. It has one argument, which is a ‘1’ to enable the production of X-rays and a ‘0’ to disable production of X-rays. The response is always ;. When X-rays are enabled via the ENBL command, we say that the system is in an enabled state. When Xrays are disabled, we say that the system is in a disabled state. 10.6.9 WATCHDOG TIMER COMMAND: The WDTE enables or disables the watchdog timer. If the watchdog is not reset within one second the tank will shutdown X-ray production. It has one argument, which is a ‘1’ to enable the watchdog and a ‘0’ to disable. The response is always ;, when the watchdog is enabled. 10.6.10 WATCHDOG TIMER TICKLE COMMAND: The WDTT resets the watchdog timer. The command has no argument. The response from the tank for this command is always ;. 10.7
COMMUNICATIONS COMMAND TIME-OUT: The tank will never, for more than 100 ms, be in a state where it has received a partial command and is waiting for the remainder of the command. All commands from the host computer will be fully transmitted to the tank in less than 100 ms. If a command is started, and is not terminated within 100 ms, the tank will flush the current command and begin searching for the start of a new command. New command can be sent to the tank as soon as a report message is received.
10.8
REPORT FAULT DEFINITIONS:
FLT # 000 001 002 003 004
Report Fault Definitions FLT # Description 005 Low kV 006 High kV 007 Spare 008 Spare 009 Spare
Description No Fault Present Temp High Tube ARC High mA Low mA
Reference Definitions: 1. Voltage Reference – a) The DSP will set a digital to analog converter output based on a request received via the RS-232 communications interface. UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
9
of 11 Sheet 10 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
2. Current Reference a) The DSP will set a digital to analog converter output based on a request received via the RS-232 communications interface. Monitor Definitions: 1. Voltage Monitor - Voltage Monitoring is available via the RS-232 Voltage Monitor Command. 2. Current Monitor - Current Monitoring is available via the RS-232 Current Monitor Command. Status Definitions: 1. Report Fault – Active Faults can be monitored via the FAULT LED’s and the RS-232 Report Fault Command. In addition, if any fault is active J7-7 will be low impedance to ground. 2. HV Status - HV Status can be monitored via the X-RAY ON LED and the RS232 HV Status Command. Fault Cause and Effect: If a shutdown occurred from a fault, the fault must be cleared before X-Ray On can be enabled. Faults can be cleared with a low to high transition on the enable signal to the DSP, or an RS232 reset command. Once cleared, a ‘High’ on the enable signal to the DSP will turn X-Rays On. Note, if the RS232 X-ray ENBL command is used, and a fault occurs, the RS232 reset command must be used to clear the fault. The enable will not clear faults in this m ode. 10.8.1 Arc Fault – An Arc Fault occurs when one or both of the Arc Detect Signals goes Low. This fault will produce the following results: a) Light the ARCFLT LED for 1 minute. b) Respond to a RS232 Report Fault Command with FLT# 002. c) Change the status of J7-7 to low impedance to ground for 1 minute. This is not a latched fault. However the tank will shut down if 4 arc events are detected within a 10 second period. Each arc event will be debounced in software for 100mS to prevent ringing of the arc detect signal from erroneously tripping a shutdown. 10.8.2 High mA – A High mA Fault occurs when the output current exceeds the current reference set value by 18%. This fault will produce the following results: a) Light the HIGH MA LED b) Respond to an RS232 Report Fault Command with FLT# 003. c) Change the status of J7-7 to low impedance to ground. d) Shutdown the unit. This is a latched fault. 10.8.3 Low mA – A Low mA Fault occurs when the output current is below the current reference set value by 18% after the filament ramp is complete. This fault will produce the following results: a) Light the LOW MA LED. b) Respond to an RS232 Report Fault Command with FLT# 004. c) Change the status of J7-7 to low impedance to ground. This is not a latched fault. The fault will only remain active as long as the output current is below the current reference set value by 18%. UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
10
of 11 Sheet 11 of 12
This information contained herein is proprietary to Spellman High Voltage Electronics Corp. and is to be used only for the purpose intended and is not to be reproduced either in whole or part without the expressed written permission of Spellman High Voltage Electronics Corp.
REVISION LTR
DESCRIPTION
DATE
APPROVED
E
REVISED PER ECN NO. 19627
11/9/06
AL, MT, JG, NF
F
REVISED PER ECN NO. 20005
2/28/07
AL,MT,BR,NF
10.8.4 High kV – A High kV Fault occurs when the output voltage exceeds 165kV. This fault will produce the following results: a) Light the HIGH kV LED. b) Respond to an RS232 Report Fault Command with FLT# 006. c) Change the status of J7-7 to low impedance to ground. d) Shutdown the unit. This is a latched fault. 10.8.5
Low kV – A Low kV Fault occurs when the output voltage is below the voltage reference set value by 9% after the kV ramp is complete. This fault will produce the following results: a) Light the LOW KV LED. b) Respond to an RS232 Report Fault Command with FLT# 005. c) Change the status of J7-7 to low impedance to ground. d) Shutdown the unit. This is a latched fault.
High Temp - A High Temp Fault occurs when the Temp Sense Signal goes ‘High’. This fault will produce the following results: a) Light the OVER TEMP LED. b) Respond to an RS232 Report Fault Command with FLT# 001. c) Change the status of J7-7 to low impedance to ground. d) Shutdown the unit. This is a latched fault. X-RAY ENABLE and X-RAY ON/OFF COMMANDS: The tank will power up in the X-rays enabled state with VREF = 0 and IREF = 0. Thus when the tank powers up, X-rays will not actually come on, even if the enable is asserted, until VREF and IREF have been set by the host software. If the host software wants to set VREF and IREF without producing X-rays then it must either de-assert the 12V Xray enable line or send an XRAY 0 command to cause the tank to transition of the X-rays disabled state. Once an Xray command is sent the 12V Xray enable line will act as an interlock. 10.8.6
10.9
When X-rays are enabled via the XRAY command, X-rays will still not be produced unless the 12V X-ray enable line is held high. The X-ray enabled / disabled state can be set and cleared by the XRAY command regardless of the state of the X-ray enable. However X-rays will not go on unless the X-ray enable line is held high while the X-rays are in an enabled state. Control Definitions: X-Ray Enable – A ‘High’ on the enable signal to the DSP or an X-ray command with the enable signal ‘High’ starts the voltage ramp. The voltage ramp shall step the voltage from 0 to 10V in 20mS +/- 10%. When the output voltage exceeds 80kV the filament ramp shall start. If the kV level falls below 80kV the filament will shutdown. The filament ramp shall step the voltage from 0 to 10V in < 1mS.
END of Document
UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES. TOLERANCES ARE:
CONTRACT NO.
APPROVALS
FRACTIONS DECIMALS ANGLES +/.XX +/+/+/.XX +/+/-
DRAWN
MATERIAL
CHECKED EM DESIGN
\\Shvdesign\release\137\137408 -001.DOC DO NOT SCALE DRAWING
JP
DATE
475 Wireless Blvd. Hauppauge, New York 11788
11/19/03
SPECIFICATION CONTROL DRAWING XRB150P&N135X3408
11/19/03
N/A
N/A
PROJ. ENG MT
11/21/03
QA/MFG.
11/26/03
KK, JC
Spellman High Voltage Electronics Corporation
PROPRIETARY INFORMATION
SIZE
A SCALE:
CODE IDENT NO.
DRAWING NO.
Rev F
137408-001
10610 Part Code:
Sheet
0125-10455-00
11
of 11 Sheet 12 of 12