INSTRUCTION MANUAL
Hitachi Automatic Blood Cell Washing Centrifuge
Model MC450 Thank you very much for purchasing a Hitachi Automatic Blood Cell Washing Centrifuge Model MC450. Before using this centrifuge, carefully read through this instruction manual to ensure efficient and safe operation. Keep this instruction manual available as an important reference when using this centrifuge.
June 2008 S99972708
Contents Page
Safety Reminder 1. Operation Principle 2. Specifications 3. List of Standard Accessories and Consumable Part 4. Location and Function of Parts 5. Control Panel 6. Condition-setting Panel 7. Preparation for Operation 7.1 Preparation 7.2 Injection Volume Adjustment and Pump Deaeration 7.3 Setting of Bowl, Rotor and Distributor 7.4 Setting the Operating Conditions 7.5 Preparation of the sample 8. Operation 8.1 Operating Procedure in Automatic Mode 8.2 Operating Procedure in Manual Mode 8.3 Operating Procedure of Agitation-Centrifugation 9. Maintenance 9.1 Maintenance of Pump, Tank and Tubing 9.2 Maintenance of Rotor, Distributor and Bowl 9.3 Maintenance of Chamber, Inner Chamber and Drain Cover 9.3.1 Removing and Mounting the Drain Cover 9.3.2 Removing and Mounting the Inner Chamber 10. Troubleshooting 10.1 Error Code 10.2 Troubleshooting when no Error Code is indicated 11. Installation 11.1 Location 11.2 Installation Installati on Procedure 11.3 Removing the Motor Holder and Mounting the Motor Guard Plate (2) 11.4 Connection of the Power Cord 11.5 Removing the Motor Packing 12. Opening the Door during Power Failure Appendix: Decontaminati Decontamination on Sheet Warranty After-sales Service Service
i 1 2 3 4 6 8 10 10 11 12 13 13 14 14 14 15 16 17 17 17 18 18 19 19 20 21 21 21 21 21 21 22
Safety Reminder Common safety precautions Carefully read the following safety precautions for a thorough understanding. Follow the instructions and procedures described in this Instruction Manual to operate this centrifuge
・
safely. Be sure to observe the precautions in this Instruction Manual and on your equipment.
・
Safety messages are indicated as follows.
They are in combination with signal words of
・
“WARNING” and “CAUTION” with the safety alert symbol to call your attention to items or operations that could be dangerous to you or other persons using this equipment.
The definitions of signal
words are as follows:
WARNING:Warning notes indicate any condition or practice, which if not strictly observed, could result in serious injury or possible death.
CAUTION:Caution notes indicate any condition or practice, which if not strictly observed or remedied, could result in personal injury, damage or destruction of the equipment.
NOTE:Notes indicate an area or subject of special merit, emphasizing either the product’s capabilities or common errors in operation or maintenance.
Do not operate this centrifuge in any manner not described in this Instruction Manual.
・
Should you
have any troubles with this centrifuge, call a Hitachi Koki authorized sales/service representative. The precautions described in this Instruction Manual are carefully developed in an attempt to cover
・
all the possible risks.
However, it is also important that you are alert for unexpected incidents. Be
careful when operating this centrifuge.
i
Safety Reminder WARNING 1. This centrifuge is not explosion-proof. Never use explosive or flammable samples. 2. Do not install the centrifuge in or near places where flammable gases are generated or chemicals are stored. 3. Take all necessary safety measures before using samples that are toxic or radioactive, or blood samples that are pathogenic or infectious. You use such samples at your own responsibility. 1) If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic or radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods. 2) If there is a possibility that the centrifuge, rotor, or an accessory is contaminated by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor, or the accessory properly before requesting repairs from a Hitachi Koki authorized sales/service representative. 3) It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or parts properly before returning them to a Hitachi Koki authorized sales/service representative. 4. Because the liquid samples, the saline, etc. are used in this centrifuge, your centrifuge must be grounded properly. 5. To avoid electrical shocks, do not handle the power cord or turn on or off the POWER switch with wet hands. 6. For safety purposes, do not go within 30 cm of this centrifuge while it is in operation. 7. While the rotor is rotating, never forcedly unlock the door. 8. Repairs, disassembly, and other modifications to the centrifuge are strictly prohibited unless performed by a Hitachi Koki authorized sales/service representative.
ii
Safety Reminder CAUTION 1. Do not move or relocate this centrifuge while the rotor is rotating. 2. Do not pour any solution such as water, detergent, or disinfectant directly into the rotor chamber. If you do so, the bearings of the drive unit might corrode or deteriorate. 3. Connect the saline tank to the pump inlet connector of the centrifuge properly with the hose bands (standard accessories). Otherwise the liquid might leak from the tube and the liquid might seep into the inside of the centrifuge. If the liquid might have seeped into the inside of the centrifuge, contact a Hitachi Koki authorized sales/service representative for cleaning and drying the centrifuge. 4. Before running this centrifuge, remove any dropped objects and tube fragments from inside the rotor chamber. 5. Always check for corrosion and damage on the rotor before using it. Do not use the rotor if you find such a problem. 6. Be sure to use the specified test tubes with a wall thickness of 1 mm or more. When using thin test tubes such as disposable test tubes, do not use them repeatedly. If any of scratches, crack, inside warp, etc. is found on the test tubes, do not use them because they cannot bear a centrifugal force. 7. Be sure to load test tubes to all the holders. When test tubes are not loaded unavoidably to all the holders, load test tubes and empty tubes to all the holders. 8. Use one or two drops (about 50μl) of 3-5% erythrocyte suspension as the sample volume at blood cell washing. When using precipitated erythrocyte layer, use one or two drops (about 50μl or less) per a test tube. Use the 80% or less of the test tube capacity as the sample volume for centrifugation. 9. If you observe some abnormality in this product, stop using it immediately and contact a Hitachi Koki authorized sales/service representative. Notify the service representative of the alarm code if displayed. 10. If the centrifuge will not be used for a long time, remove the power cord from the receptacle. 11. For connection to a different outlet, the power cord might be needed to be replaced. Follow local electrical codes. 12. Depending on the magnitude, an earthquake might damage the centrifuge. If you observe some abnormality, contact a Hitachi Koki authorized sales/service representative.
Note: During operation, it is not abnormality that the control panel and the surface of this centrifuge are warmed by themselves.
iii
1. Operation Principle The MC450 automatically carries out a series of blood cell washing processes of erythrocyte suspension for antiglobulin test (direct Coomb’s test or indirect Coomb’s test). The operational sequence is shown below. ) m p r ( d e e p s r o t o R
3000 Injection of saline 1200 400
①
Process
②
③
④
⑤
1 cycle Time
Blood cell washing operational sequence Description of operational sequence Process
Operation
Description
Reference illustration
①
WASH
A fixed amount of saline is pumped into the distributor when the rotor speed reaches 1200 rpm. By centrifugal force, the saline is jetted from the distributor into test tubes.
At that time, blood cells
in each test tube are sufficiently suspended in the
Saline
saline.
②
CENTRIFUGE
Blood cells are centrifuged. The standard centrifugation time is 35 sec. (selectable).
. s e m i e t l c 4 y c r g o n 3 i h d s t e a a W e p e R
Immediately
before
the
rotor
is
decelerated, injection of saline is resumed to overflow the test tubes. (Overflow is also selectable.)
③
DECANT
The rotor is rotated at a low speed with the rotor holder kept at an angle to slightly open its top end by magnetic force.
By this operation, only the
saline is decanted from the test tubes and the blood cells remain.
④
AGITATE
The rotor repeats rotation and stop at short, quick intervals to disentangle the remained blood cells.
⑤
CENTRIFUGE
The rotor is rotated about 5 seconds to collect the blood cells adhered to the wall surfaces of the test tubes to the bottom.
It is to ensure the reaction
with the Coombs reagent.
This operation is
performed at the end of the washing cycle.
1
Centrifugal force
2. Specifications Number of samples
24 (or 12)
Size of applicable test
12 mm or 10 mm in diameter and 75 mm in length
tubes Rotor speed
WASH
1,200 rpm (190 ×g)
(Centrifugal force at the
CENTRIFUGE
3,000 rpm (1,180 ×g)
bottom of test tubes)
DECANT
350 to 500 rpm
Timer
AUTO
1 to 99 seconds (in 1-second increments)
MANUAL
1 to 999 seconds (in 1-second increments)
Number of process cycles
1 to 9
to be preset Process time per cycle
60 seconds (automatic centrifugation time: 35 seconds)
Driving motor
Series motor, 300 W
Process indication
4 processes
Manual-mode rotor speed
3,000 rpm (High) and 1,200 rpm (Low)
End buzzer
Selectable from seven kinds of sound including five tunes, electronic beep and no sound
Safety devices
Door interlock, dual-overspeed detector and saline level sensor CE marking (★)
Applicable standard Power requirements
AC115V ±10% 10A 50/60Hz, AC220V ±10% 5A 50/60Hz
Dimensions
370 (W)×450 (D)×410 (H) mm
Weight
29 kg
The MC450 centrifuges satisfy CE marking requirements. The CE marking is an international symbol, which shows that the product conforms to EC directives. Standards concerning these directives are as follows: • Product Safety (EN61010-1 and EN61010-2-020) • Electromagnetic compatibility (EN61326-1)
2
3. List of Standard Accessories and Consumable Part The standard accessories of the MC450 are listed below.
Check that all the standard accessories are
supplied before installation. 3.1 List of standard accessories Bowl ass’y 1 pc.
Rotor ass’y 1 pc.
Distributor ass’y 1 pc.
Tank (5L) 1 pc.
Tubing (2.5m) 2 pcs.
Part No. S402776B
Part No. S307543A
Part No. S307457A
Part No. S402777B
Part No. 4744342
Drain pipe 2m
D10 adapter 25 pcs.
Test tubes 1 carton (50 pcs.)
Silicone grease 1 pc.
Hex. wrench 1 pc.
Part No. 660557
Part No. 8046004
(12 mm in dia. and 75 mm in length)
Part No. S401103
Part No. 480879
Part No. 480270
Instruction manual 1 pc.
Installation procedure 1 pc.
(Small)2 pcs., (Big)1 pc.
1 pc.
Part No.
Part No.
Part No.
S999727
S998002
Part No. 4666354 (Small) 4666355 (Big)
Hose band
3.2 List of consumable part Bellows pump set
Part No. S404323
3
Motor guard plate (2)
S411107
Tube connecter 1 pc.
Part No. 84610903
4. Location and Function of Parts ■ Front view ① Door
② Nozzle
③ Drain cover ④ Distributor ⑥ Bowl ⑤ Rotor ⑧ Door lock lever ⑦ Inner chamber Control panel
⑨POWER switch
Condition-setting panel
■ Rear view
⑬Drain connector
⑫ Pump inlet connector
⑭ Pump The pump cover is not mounted in this figure.
⑩Power connector
⑪ Sensor connector
4
⑮ Auxiliary drain tube
No.
Part
Description
①
Door
Test tubes are set on the rotor after opening the door. and nozzle are installed on the back of the door.
②
Nozzle
Through this nozzle, saline is supplied from the pump to the distributor.
③
Drain cover
Receives the waste liquid decanted from the rotor and discharges it to the outside.
④
Distributor
Distributes the saline supplied from the pump to each test tube held on the rotor.
⑤
Rotor
A 24-sample or a 12-sample rotor.
⑥
Bowl
Rotates together with the rotor. tube holders of the rotor swing.
⑦
Inner chamber
Designed to keep saline, waste liquid, etc. from leaking to the internal mechanism of the centrifuge. It is removable.
⑧
Door lock lever
Keeps the door locked while the rotor is running.
⑨
POWER switch
Turns on and off the centrifuge power. It doubles as a circuit protector. It automatically turns off the centrifuge power if abnormal current passes.
⑩
Power connector
To be connected to a power supply.
⑪
Sensor connector
To be connected to the sensor which detects the level of saline in the tank.
⑫
Pump inlet connector
To be connected to the saline tank by tubing.
⑬
Drain connector
The overflowed waste liquid is drained through this connector.
⑭
Pump
Provided for supplying saline into the centrifuge.
⑮
Auxiliary drain tube
The chamber is drained through this tube if the drain connector is clogged.
5
The drain cover
Its test tube holders swing.
It determines the angle to which the test-
5. Control Panel ② ①
③ ④
⑧
⑤ ⑥ ⑦
⑨
No.
Name
①
PROCESS LED
②
TIME indicator
③
AUTO switch
④
MANUAL switch
⑤
FEED switch
⑥
START switch
⑦
STOP switch
⑧
CYCLE indicator
⑩
Function Shows the current process of blood cell washing. Also indicates the process setting in the case of manual mode. While the centrifuge is in operation, the LED for the process concerned blinks. Indicates the remaining time of centrifugation (in 3 digits). In case of an abnormality, it shows the error code concerned. It also shows the rotor speed (×10 rpm) when the PUSH SPEED switch on the condition-setting panel is pressed. To be pressed for automatic washing cycle in succession. Its LED lights when the automatic mode is set. To be pressed when executing a single process (WASH, CENTRIFUGE, DECANT or AGITATE) by manual control. Its LED lights when the manual mode is set. In the automatic mode, it causes the value of cycle setting to count down. In the manual mode, it is used for setting desired one of the processes. To be pressed for starting a series of processes or a process. To be pressed to stop a process in the middle. It is also used for stopping the end buzzer and alarm buzzer. When the operation is stopped in the automatic mode or in the agitation/centrifugation (see ⑨), the operation can be restarted at the stopped point by pressing the START switch. When the operation is stopped in the manual mode, the operation is restarted from the start. Its LED lights when the operation is stopped in the middle, and blinks for 10 seconds then lights when the operation is completed. When the operation is stopped in the middle of a process in the automatic mode, its LED blinks. Indicates the number of washing-process cycles. While the centrifuge is in operation, it shows the number of the remaining cycles.
6
No.
Name
Function Sets the agitation time (agitation of liquid in test tubes) and the centrifugation time (manual mode) (see the figure below). Its LED lights when it is set. Operation is started by pressing the START switch.
Manual-mode set speed
⑨
AGI/CENT switch
Agitation set time
Manual-mode centrifugation time
This centrifuge does not accept the input of the START switch when the agitation set time is 0 second.
⑩
DOOR switch
To be pressed to release the door interlock. If it is pressed continuously, the centrifuge does not accept any input of the other switches for 4 seconds (alarm buzzer sounds).
7
6. Condition-setting Panel
①
No.
① ② ③ ④
②
⑤
⑥
⑦
⑨
⑩
⑪
⑫
⑧
Name
Function Sets the number of the washing-process cycles to be repeated (9 CYCLE setting switch cycles at maximum). Sets the centrifugation time of automatic-mode operation (99 AUTO CENTRI TIME setting seconds at maximum). The standard centrifugation time is 35 switch seconds. MANUAL CENTRI TIME Sets the centrifugation time of manual-mode operation (999 setting switch seconds at maximum). Sets one of the following rotor speeds for manual-mode MANUAL SPEED setting centrifugation: switch 3,000 rpm (High) and 1,200 rpm (Low) Sets the rotor speed for automatic- or manual-mode decantation by the combination of the H/L switch and the dial setting. (If the dial is set between 4 and 9, the rotor speed is the same as that of dial setting 0.)
⑤
DECANT switch
⑥
PUSH SPEED switch
⑦
④
③
SPEED
setting
OVERFLOW setting switch
Dial setting (No setting more than 3) 0 1 2 3 H 350 rpm 400 rpm 450 rpm 500 rpm L 330 rpm 370 rpm 410 rpm 450 rpm (Decantation is completed in a short time. Note that the above value may not necessarily match the speed indicated by the PUSH SPEED switch.) Switch
As this switch is pressed, the running speed of the rotor is shown (×10 rpm) on the TIME indicator on the control panel. Sets the pumping operation for overflow. Setting 0 1 2 3 4 5 6 to 9
Pumping operation for overflow No operation Operates only the first cycle. Operates only the first and the second cycles. Operates only the first, second and the third cycles. Operates only the first, second, third and the fourth cycles. Operates only the first, second, third, fourth and the fifth cycles. Selectable up to 9 in the same manner.
8
No.
⑧
⑨
Name SALINE PRIME switch
PROGRAM switch
Function To be used for pump deaeration at the time of initial operation. The pump for saline supply operates while this switch is kept depressed. (This switch is ineffective while the centrifuge is in operation.) Sets the operation program. Setting Program (No setting more than 1) 0 5-second centrifugation is not performed at the final process of the final cycle in the automatic mode. 1 5-second centrifugation is performed at the final process of the final cycle in the automatic mode. 2 You can add the conditions “3000 rpm for 15 seconds ”*1 and “1200rpm for 60 seconds ” *2 to the centrifugation process at manual-mode. See the following procedures. 1. Press the MANUAL switch. The condition displayed on the condition-setting panel is performed. 2. Press the FEED switch. The centrifugation (3000 rpm for 15 seconds) is performed. 3. Press the FEED switch again. The centrifugation (1200 rpm for 60 seconds) is performed. *1:3000 rpm for 15 seconds: “H” is displayed on the CYCLE indicator and “15” is displayed in the TIME indicator. *2:1200 rpm for 60 seconds: “L” is displayed on the CYCLE indicator and “60” is displayed in the TIME indicator. 5-second centrifugation is not performed at the final process of the final cycle in the automatic mode.
⑩
AGITATE TIME setting switch
⑪
MELODY setting switch
⑫
OPTION switch
(If the dial is set between 3 and 9, the operation program is the same as that of dial setting 0.) Sets the agitation time (99 seconds at maximum). Sets the end buzzer. Setting Tunes (No setting more than 6) 0 No sound 1 Hitachi-no-ki 2 Oh Susanna 3 My Bonnie 4 My Old Kentucky Home 5 De Camptown Races 6 Electronic beep (two beeps) (You can turn down the sound level by covering the hole above the ① CYCLE setting switch with an adhesive tape.) This is a push-button switch for optional functions.
9
7. Preparation for Operation 7.1 Preparation (1) Prepare the test tubes (10 or 12 mm in diameter and 75 mm in length). When using the test tubes 10 mm in diameter, set the attached D10 a dapter to the rotor holder. CAUTION: ● Be sure to use the specified test tube with a wall thickness of 1 mm or more. When using thin test tubes such as disposable test tubes, do not use them repeatedly. If any of scratch, crack, inside warp, etc. is found on the test tubes, do not use them because they cannot bear a centrifugal force. It is recommended to use the tempered hard glass test tubes made by NICHIDEN-RIKA GLASS CO., LTD. or the Pyrex test tubes made by ASAHI TECHNO GLASS CORPORATION for this centrifuge. (2) Prepare the saline tank (standard accessory) and a waste liquid tank. Connect the signal cord of the liquid level sensor of the saline tank to the connector at the rear of the centrifuge. (3) Perform piping as shown in the figure. ・Do not set the saline tank at a place higher than the centrifuge or behind the centrifuge. ・Set the waste liquid tank at a place lower than the centrifuge. ・Connect the drain pipe to the drain connector. ・Place the tips of the drain pipe in the waste liquid tank or a sink positioning them higher than the
liquid level. The gray connector at the rear center of the centrifuge is used for draining off waste liquid from the chamber. After blood cell washing, waste liquid keeps draining off through the connector. The auxiliary drain tube under the connector is an auxiliary tube to drain waste liquid flowed from the chamber into the vicinity of the driving motor. Although waste liquid may not always drain off through it, prepare a tank for it. CAUTION: ● Make sure that the tips of the drain pipes are above the waste liquid level in the tank. Otherwise, the chamber cannot be drained completely and waste liquid stagnates in the chamber due to poor drainage. Poor drainage may also be caused if the pipe is bent or flattened or if waste liquid stagnates in the middle of the pipe as shown by the dotted lines in the following figure. Set the pipes properly so that waste liquid drains off smoothly. Be careful not to set the drain pipes at a place higher than the drain outlet of the centrifuge. ●Connect the saline tank to the pump inlet connector of the centrifuge properly with the hose bands (standard accessories). Otherwise the liquid might leak from the tube and the liquid might seep into the inside of the centrifuge. If the liquid might have seeped into the inside of the centrifuge, contact a Hitachi Koki authorized sales/service representative for cleaning and drying the centrifuge. ・Connect the saline tank to the pump inlet connector of the centrifuge by tubing. Hose band (Big)
Pump inlet connector
MC450
Saline tank Drain pipe Waste liquid tank
Hose band (Small)
10
7.2
Injection Volume Adjustment and Pump Deaeration
(1) Adjustment of injection volume of saline The MC450 is factory-adjusted to the injection volume of saline for the 12-mm diameter test tubes. When using the 12-mm diameter test tubes with this centrifuge, there is no need to adjust the injection volume. When using the 10-mm diameter test tubes, adjust the flow rate of the pump as follows:
① Remove the protective cover of the pump at the rear of the centrifuge (it is secured with two screws). ② Loosen the two locking screws of the pump. ③ Turn the adjusting screw counterclockwise to set the indicator to “60%” position. ④ Retighten the two locking screws.
Adjusting screw
Reference (Index of injection volume) Test tube Injection Indicator volume setting Twenty-four 100 ml 80% 12-mm dia. test tubes Twenty-four 75 ml 60% 10-mm dia. test tubes Twelve 12-mm 50 ml 45% dia. test tubes Twelve 10-mm 38 ml 30% dia. test tubes
Indicator
Locking screws
(2) Pump deaeration Fill the saline tank with saline. Hold a beaker or other container to the nozzle. Press the SALINE PRIME switch on the condition-setting panel to discharge saline from the nozzle until no bubbles come out of the nozzle.
11
7.3
Setting of Bowl, Rotor and Distributor
(1) Bowl setting
① Apply silicone grease (standard accessory, general vacuum grease is also usable) to the inside of the bowl (see the arrow mark in the figure) and the loading portion of the O-ring of the decantation coil lightly by hand. Perform grease application once or twice a month.
Bowl
②Put both hands on the inside of the bowl as shown in the figure and push down the bowl horizontally Bowl applying a force with the thumbs and the other fingers alternately.
Decantation coil
③When the bowl is pushed about one-third or a half of the bowl height, put both hands as shown in the figure and push down the bowl furthermore turning the bowl until the bottom of the decantation coil becomes flush with the bottom of the bowl. ④Check that the bowl can be turned smoothly by hand.
Hold the bowl with both hands as shown in the figure. Remove the bowl gradually by rotating it.
Flush
Bowl
(2) Rotor setting Properly place the rotor on the drive shaft. Turn the rotor about a quarter so that the coupling pins are inserted into the rotor shaft holes.
Rotor shaft
Rotor
Coupling pin
Bowl
CAUTION: ● If the rotor is not set correctly, the rotor holder may come off the bowl and overswing.
(3) Distributor setting Fit the protrusions at the bottom of the distributor in the holes of the rotor.
12
Decantation coil
7.4 Setting the Operating Conditions The operating conditions are factory-set as follows. Setting dials and switches
Factory-setting
① CYCLE
3 cycles
② AUTO CENTRI TIME
35 seconds
③ MANUAL CENTRI TIME
60 seconds
④ MANUAL SPEED
L (1,200 rpm)
⑤ DECANT SPEED
H-1 (400 rpm)
⑦ OVERFLOW
1 (Operates only the first cycle)
⑨ PROGRAM
1 (5-second centrifugation is performed at the final cycle in the automatic mode)
⑩ AGITATE TIME
5 seconds
⑪ MELODY
1 (Hitachi-no-ki)
Coagulation of blood cells is influenced by the centrifugal speed (centrifugal force) and the centrifugation time. If the centrifugal force and the time are insufficient, coagulation may also be insufficient causing false negative results. To the contrary, excessive centrifugation may harden the blood cells firmly causing false coagulation or difficult resuspension. Thus the weak reaction may disappear. In addition, centrifugal conditions are greatly influenced by the amount of blood cells, specific gravity of blood cells, etc. Note that the above setting is just as a guide. Determine the optimum conditions according to the following evaluation. ① The supernatant is transparent and no blood cell is suspended after centrifugation. ② The blood cells that are precipitated at the bottom of the test tubes show clear outlines after centrifugation. ③ The blood cells are easily removed from the bottom and disentangled by a light agitation after decantation. ④ Check the reaction using reagents whether negative or positive. (Reference: “Cross match test – The methods and the solutions” published by GENDAISHA PUBLISHING Co.)
7.5 Preparation of the sample CAUTION: ●When the sample or the liquid is spilled, it might seep into the inside of the centrifuge. If it might have seeped into the inside of the centrifuge, contact a Hitachi Koki authorized sales/service representative for cleaning and drying the centrifuge. Blood cell washing Use one or two drops (about 50μl) of 3-5% erythrocyte suspension in one test tube. •
CAUTION: ● When using precipitated erythrocyte layer, use one or two drops (about 50μl or less) per a test tube. Otherwise a series of blood cell washing processes might be carried out incompletely. Centrifugation Use the 80% or less of the test tube capacity as the sample volume for centrifugation. In the MC450 centrifuge, the tubes are rotated with the tubes leaned at about 45°in the direction of the drive shaft. •
Centrifugal force
45
°
13
8. Operation CAUTION: ●How to close the door Close the door by pushing down at the forward center of the door strongly until a short beep is heard. If the door is closed insufficiently, an alarm (E1: DOOR OPEN) is indicated and the centrifuge does not start operation. If this alarm is indicated, properly close the door and press the START switch again. ●How to open the door Open the door gently after blood cell washing. Otherwise the remaining waste liquid from the inside of the drain cover might splash.
8.1 Operating Procedure in Automatic Mode Step
Operation
Description
1
Turn on the AUTO switch.
The switch lights its LED. (The automatic mode is selected automatically when the power switch is turned on.)
2
Turn on the START switch.
The switch lights its LED. To stop the centrifuge temporarily in the middle of a process, press the STOP switch and the centrifuge stops at once. For restarting the process, press the START switch.
3
Upon finish of all processes
The end buzzer sounds and the START switch extinguishes its LED. The door is opened. The buzzer stops as the STOP switch is turned on.
NOTE: In the automatic mode, the number of cycles having been set can be decreased by pushing the FEED switch. For example, if three cycles have been set, the number changes 3→2→1→3→2→1 as the FEED switch is pressed.
8.2 Operating Procedure in Manual Mode Step
Operation
Description
1
Turn on the MANUAL switch.
The switch lights its LED.
2
Press the FEED switch once or more to light the PROCESS LED of the desired process.
The CENTRIFUGE process (centrifugation) is set preferentially. For example, when executing only the WASH process, press the FEED switch three times and the WASH LED of the PROCESS indicator lights to show that the WASH process is set.
3
Turn on the START switch.
The switch lights its LED.
Upon finish of the selected process
The end buzzer sounds and the door is opened. START switch extinguishes its LED.
The
NOTE: 1) The door is automatically opened after the rotor is stopped by pressing the STOP switch in the middle of operation. 2) When power fails during operation and then it is restored, or when the POWER switch is turned off during operation and then turned on again, the automatic-mode operation is set ( AUTO LED lights) and the centrifuge stops. In such cases, set the desired process again and perform operation in the manual mode.
14
8.3 Operating Procedure of Agitation-Centrifugation Step
Operation
Description
1
Turn on the AGI/CENT switch.
The switch lights its LED.
2
Turn on the START switch.
The switch lights its LED. To stop the centrifuge temporarily in the middle of a process, press the STOP switch. For restarting the process, press the START switch.
3
Upon finish of the process
The end buzzer sounds and the START switch extinguishes its LED. The AUTO LED lights and the door is opened.
15
9. Maintenance WARNING: ●If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic or radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods. ●If there is a possibility that the centrifuge, rotor, or an accessory is contaminated by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor, or the accessory properly before requesting repairs from a Hitachi Koki authorized sales/service representative. Note that Hitachi Koki cannot repair the centrifuge, rotor or the accessory unless sterilization or decontamination is completed. ●It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or parts properly before returning them to a Hitachi Koki authorized sales/service representative. In such cases, copy the decontamination sheet at the end of this manual and fill out the copied sheet, then attach it to the item to be returned. Hitachi Koki may ask you about the treatment for t he centrifuge, rotor or the part if the decontamination is checked and judged as insufficient by Hitachi Koki. It is your responsibility to bear the cost of sterilization or decontamination. If you have any question, please send e-mail to “
[email protected]”. Note that Hitachi Koki cannot repair or inspect the centrifuge, rotor or the accessory unless sterilization or decontamination is completed. CAUTION: ●Do not operate this centrifuge in any manner not described in this instruction manual. Should you have any troubles with this centrifuge, call a Hitachi Koki authorized sales/service representative.
(1) Storage period of service parts Service parts are kept in stock for seven years after the discontinuation of production. The term “service parts” means the parts that are necessary to ensure the correct functioning of the centrifuge.
16
9.1 Maintenance of Pump, Tank and Tubing ・
Wash the inside of the tank and the tubing once a week. If saline is contaminated by bacteria, it may cause hemolysis or poor result. If the centrifuge is not used for a long period, drain out saline from the tank, tubing and pump. Check the volume of saline inside the tank before operation. Check that the tube is connected properly. When the tube is not connected properly, connect the tube securely to the saline tank and the pump inlet connector with the hose bands (standard accessories). When th e tube is lax, replace it to the new one. Replace the drain pipe and the tube with new ones every 1 to 3 years depending on the degree of stains.
・ ・
・
<Washing> ① Fill up the saline tank with washing solution (2% Sodium Hypochlorite Solution). Fill up the fluid passage with washing solution according to the same procedure as pump ② deaeration (see “7.2 Injection Volume Adjustment and Pump Deaeration”) and leave it for 12 to 24 hours. ③ After 12 to 24 hours, drain out washing solution from the saline tank and fill up the tank with distilled water. ④ Flow distilled water 2 to 3 liters from the saline tank according to the same procedure as pump deaeration. ⑤ Exchange distilled water in the saline tank for saline. ⑥ Flow saline 2 to 3 liters from the saline tank in the same manner as ④. The pump is a consumable part. Liquid may leak from the bellows of the pump due to deterioration. Although deterioration depends on the installation environment such as ultraviolet rays and temperature, it is recommended to replace t he bellows of the pump with new one (part number S403754) once every three years. It is also recommended to use saline that includes antiseptics.
*
9.2 Maintenance of Rotor, Distributor and Bowl After operation, remove the rotor, distributor and bowl properly, then wash and dry them well. See the following table for sterilization. Part
Sterilization
Rotor
Soak in a 2% sodium hypochlorite solution. and dry.
Distributor, bowl
Soak in a 2% sodium hypochlorite solution (within 1 hour). Then wash and dry.
Then wash
CAUTION: ●The distributor is made of polycarbonate. For decontamination, be sure to use a 2% sodium hypochlorite solution, otherwise the distributor might deteriorate. Use neutral detergent (pH 5 to 9). Do not dip the distributor in the diluted detergent solution for a long time, otherwise the strength of the distributor might be decreased.
9.3 Maintenance of Chamber, Inner Chamber and Drain Cover Remove the inner chamber and the drain cover from the centrifuge properly, then wash and dry them well. A 2% sodium hypochlorite solution is used for sterilization. When washing the inside of the chamber, be careful not to pour washing solution on the decantation coil directly. Wipe the decantation coil with a cloth or paper moistened with washing solution.
17
9.3.1 Removing and Mounting the Drain Cover ①
Slide the knob of the lock fastener at both sides in the arrow direction.
②
Pull the drain cover forward to remove it.
③
Insert the drain cover into the drain cover holder to mount it.
④
Push the lock fastener in the door securely until a click is heard.
CAUTION: ●When mounting the drain cover, check that the nozzle at the center of the door is put through the hole of the drain cover. If it is not properly mounted, mount the drain cover properly.
Lock fastener knob
Nozzle
Drain cover
Drain cover holder
9.3.2 Removing and Mounting the Inner Chamber ①
Remove the drain cover.
②
Remove the distributor, rotor and bowl.
③
Remove the inner chamber by pulling it up.
④
Reverse the above steps to mount the inner chamber.
CAUTION: ●Do not store washing solution in the chamber. Do not hose on the decantation coil. Otherwise, liquid may enter the inside of the centrifuge causing a failure.
・If a test tube is broken, be sure to remove the fragments inside the chamber and the drain pipe of the drain cover thoroughly with a brush.
18
10.Troubleshooting 10.1 Error code If any abnormality occurs during operation, corresponding error code blinks on the TIME indicator and the alarm buzzer sounds, then the centrifuge stops. Take the action described in Table 10-1 or contact a Hitachi Koki authorized sales/service representative for repair.
Table 10-1 Error codes Error code E1 E2
E3
Definition The door is open. The saline tank needs to be refilled.
A power failure occurs.
E4
The power supply is abnormal.
E5
The rotor overspeeds.
E6
An abnormal current flows.
E7
The speed sensor is abnormal. The current sensor is abnormal. The RAM is abnormal.
E8 E9
E10
The rotor overspeeds (hardware detection).
E11
The triac is abnormal.
Action to be taken Close the door securely and press the START switch. Refill the tank and start the centrifuge. In the manual mode, CENTRIFUGE and DECANT processes can be executed even when this error code is shown. The liquid level sensor may be faulty if the alarm buzzer sounds even though there is enough saline. Contact a Hitachi Koki authorized sales/service representative for repair. The error code can be temporarily cleared by pressing MAUAL and STOP switches simultaneously for three seconds or more (operation in the automatic mode is possible). It returns to the original state by turning on and off the power switch. If a power failure occurs while the centrifuge is running, wait a minute as the centrifuge stops rotation after the power failure. Press any of the switches on the control panel after a lapse of one minute or more from when the error code is indicated. If the error persists, contact a Hitachi Koki authorized sales/service representative for repair. This error may be caused by voltage drop when the rotor accelerating if your centrifuge is connected to an outlet to which multiple appliances are connected. Turn off the centrifuge power and connect the centrifuge to a proper outlet that can supply stable voltage to the centrifuge. The frequency of the power supply is abnormal. Turn off the power switch and on again. If the error persists, contact a Hitachi Koki authorized sales/service representative for repair. This error code is shown when the rotor runs at a speed higher than the specified speed. Contact a Hitachi Koki authorized sales/service representative for repair. This error code is shown when an abnormal current flows in the Contact a Hitachi Koki authorized sales/service circuit. representative for repair. The sensor to detect the rotor speed is abnormal. Contact a Hitachi Koki authorized sales/service representative for repair. The sensor to detect the current is abnormal. Contact a Hitachi Koki authorized sales/service representative for repair. Turn off and on the power switch. If the error persists, the microcomputer of the centrifuge to control its operation is abnormal. Contact a Hitachi Koki authorized sales/service representative for repair. This error code is shown when the rotor runs at a speed higher than the specified speed. Contact a Hitachi Koki authorized sales/service representative for repair. The element on the circuit board is abnormal. Contact a Hitachi Koki authorized sales/service representative for repair.
19
10.2 Troubleshooting when no Error Code is indicated No. 1
Symptom POWER switch does not light its LED even if it is turned on.
2
The centrifuge does not start operation.
3
Injection of saline does not take place.
4
Saline (waste liquid) scatters in the chamber. ①The amount of blood cells after washing is less than before. ②No washing is executed. ③Washing is executed.
5
No decantation is executed. Abnormal noise is heard. ①Contacting noise ②Others
6
7
8
9
Possible cause ① A power failure occurs. ②The power cord is disconnected. ①The door is not closed. ②The door switch is faulty. ①No saline is in the tank. ②The pump is not filled with saline. ③The pump is faulty. ④The pump connector is loose. ①Injection volume is excessive and it causes solution to overflow the test tubes. ②The nozzle is out of position and fails to supply the distributor with solution. ③The drain cover is broken.
Remedy ①Wait until power is restored. ②Contact a Hitachi Koki authorized sales/service representative for repair. ①Close the door. ②Contact a Hitachi Koki authorized sales/service representative for repair. ①② Refill the tank with saline to fill the pump. ③ Contact a Hitachi Koki authorized sales/service representative for repair. ④Retighten it securely.
① Adjust the injection volume properly. ② Contact a Hitachi Koki authorized sales/service representative for repair. ③ Contact a Hitachi Koki authorized sales/service representative for repair.
The coil is disconnected.
Contact a Hitachi Koki authorized sales/service representative for repair. ①The bowl and the rotor ①Set the bowl and the rotor correctly. contacts the drain ② Contact a Hitachi Koki authorized cover. sales/service representative for repair. ② The bearing and the locking screw for decantation coil are faulty. The centrifuge intensely ①Imbalance operation. ①Balance properly. vibrates. ②The bowl, rotor and ②Set the bowl, rotor and distributor distributor are not set correctly. correctly. ③ Contact a Hitachi Koki authorized ③Others sales/service representative for repair. A test tube breaks. ①The test tube is not ①Use strong test tubes enough to bear the strong. centrifugal force (see “7.1 Preparation”). ②The test tube is different ②Use applicable-sized test tubes. in size. ③Contact a Hitachi Koki authorized ③The rotor holder is sales/service representative for repair. deformed. Hemolysis is found in the ①Density of saline is ① Exchange for saline of proper density blood cells. improper. (0.9%). ②It is contaminated by ②Clean the tank, tube and pump (especially the inside of the tank). bacteria. ③Remove glass fragments in the chamber ③Fragments of a test and the drain cover. tube are mixed.
WARNING: ●Unauthorized disassembly of the centrifuge except by a Hitachi Koki authorized service representative are strictly prohibited.
20
11.Installation 11.1 Location (1)Install the centrifuge in a room where the ambient temperature is 5 to 30℃. (2)Place the centrifuge on a sturdy desk or a wagon with wheel locks or a strongly structured
workbench in a horizontal position. (3) Avoid installing the centrifuge in a dusty place. Otherwise, dust may get into the sample. (4)Do not position this centrifuge in direct sunlight. Otherwise, the temperature rises excessively and it may cause a failure. (5)Use the centrifuge in a well-ventilated room. Keep spaces at least 10 cm from the walls of the room. ( 6 ) This centrifuge is not explosion-proof. Do not use the centrifuge in or near places where flammable gases are generated or chemicals are stored.
11.2 Installation Procedure Refer to the attached “Installation Procedure (Part No.S998002)” for installation.
11.3 Removing the Motor Holder and Mounting the Motor Guard Plate (2) (1)Lift the front bottom of the centrifuge. (2)Remove the wing nut.
Accessory Motor Guard Plate (2)
(3)Remove the motor holder. (4)Secure the motor guard plate (2) with
the wing nut. Note: Save the removed motor holder for future transportation.
Motor holder
Wing nut
11.4 Connection of the Power Cord WARNING: ●To avoid electrical shocks, do not touch the power cord with wet hands. ●Do not hold the cord when disconnecting the power cord from the outlet. Hold the plug when disconnecting it. ● This centrifuge must be grounded properly. The power cord plug has a three-prong plug. One of the three prongs is provided for grounding. This centrifuge can be grounded just by putting the three- prong plug into a three-prong outlet. Do not connect the three-prong plug to a two-prong outlet using an adapter.
CAUTION: ●For connection to a different outlet, the power cord may be needed to be replaced. Follow local electrical codes.
11.5 Removing the Motor Packing (1)Turn on the power switch and open the door. Remove the motor packing. (2)When the centrifuge is not connected to the power source yet, open the door according to
“12. Opening the Door during Power Failure”.
21
12. Opening the Door during Power Failure WARNING: ●Opening the door while the rotor is rotating is very dangerous. Never unlock the door while the rotor is rotating. ●In the event where the door is opened while the rotor is still rotating, close the door immediately. ●Do not run the centrifuge with the hex. wrench inserted in the hole of the centrifuge.
The door cannot be opened and closed except when the centrifuge is powered on (POWER switch is on) and the rotor is stopped. If the door cannot be opened due to a power failure, open the door according to the following procedure. (1) Check that the rotor is not rotating. ■Listen carefully to make sure that no rotating sound is heard. ■It takes about 30 seconds until the rotor rotating at 3,000 rpm (maximum speed)
stops completely.
Allow for a sufficient time.
(2) Insert the attached hex. wrench into the small hole to unlock the door. ■ A small hole is provided on both sides of the centrifuge. ■Insert the attached hex. wrench straight into the small hole and push it in until a click
is heard. Then insert the hex. wrench into the small hole of the other side in the same manner. Thus the door is unlocked and it is opened
22
It is requested that you return the faulty product with this Decontamination Sheet in order to repair it safely in our plant. Be sure to decontaminate the product according to good laboratory procedures and methods, and fill out this Decontamination Sheet and attach it to the product to be returned to Hitachi Koki for repair.
Attention: Hitachi Koki Co., Ltd.
Decontamination Sheet
Date:
Name: Name of company (organization) or school: Division or faculty/Subject of study: Telephone number: Address: I performed decontamination to remove biological or chemical contaminants (including radioactive isotope) from this product as follows.
Model of centrifuge:
Serial number:
Model of rotor:
Serial number:
Accessory:
Serial number:
Contaminants used: Decontamination method (conditions):
Date of decontamination:
Signature