Mobile Stim & Combo Therapy Systems
SERVICE MANUAL Model- 2778 Mobile Combo Therapy System Applies to Serial numbers 1000 and above
Model- 2777 Mobile Stim Therapy System Applies to Serial numbers 1000 and above
© 2005 Encore Medical, L.P.
TABLE OF CONTENTS FOREWORD . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 1 1 SAFETY PRECAUTIONS .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 2 1.1 PRECAUTIONARY SYMBOL DEFINITIONS . .. 2 1.2 SAFETY PRECAUTIONS.. .. .. .. .. .. .. .. .. .. .. .. .. 2 2 THEORY OF OPERATION .. .. .. .. .. .. .. .. .. .. .. .. .. .. 5 2.1 OVERVIEW . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 5 2.2 POWER SUPPLY CIRCUIT.. .. .. .. .. .. .. .. .. .. .. .. 5 2.3 CONTROL BOARD .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 5 2.4 STIM BOARDS . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 5 2.5 ULTRASOUND BOARD AND APPLICATOR COMBINATION SYSTEMS ONLY . .. .. .. .. .. .. 5 2.6 USER INTERFACE AND ACCESSORIES .. .. .. .. 5 2.7 NIMH BATTERY . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 5 3 NOMENCLATURE . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 6 3.1 COMPONENT AND CONTROLS .. .. .. .. .. .. .. .. 6 3.2 HARDWARE AND SOFTWARE SYMBOL DEFINITIONS .. .. .. .. .. .. .. .. .. .. .. .. .. 8 4 SPECIFICATIONS .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 9 4.1 INTELECT MOBILE COMBO SYSTEM.. .. .. .. .. 9 4.2 INTELECT MOBILE STIM SYSTEM . .. .. .. .. .. 10 4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS . .. .. .. .. .. .. .. 11 4.4 INTELECT MOBILE COMBO ULTRASOUND SPECIFICATIONS .. .. .. .. .. .. 17 5 TROUBLESHOOTING . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 18 5.1 INTELECT MOBILE STIM AND COMBO ERROR MESSAGES . .. .. .. .. .. .. .. .. 18 5.2 INTELECT MOBILE STIM AND COMBO SYSTEM TESTING .. .. .. .. .. .. .. .. .. 21 5.3 VISUAL INSPECTION . .. .. .. .. .. .. .. .. .. .. .. .. 22 5.4 LEAKAGE TESTS . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 22 5.5 UNIT STARTUP AND FAN TESTING .. .. .. .. 22 5.6 ELECTRICAL STIMULATOR TEST SYSTEM SETUP .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 23 5.7 VMS™ MODE TEST .. .. .. .. .. .. .. .. .. .. .. .. .. .. 23 5.8 INTERFERENTIAL MODE TEST.. .. .. .. .. .. .. 24 5.9 PREMODULATED MODE TEST.. .. .. .. .. .. .. 24 5.10 RUSSIAN MODE TEST .. .. .. .. .. .. .. .. .. .. .. .. 25 5.11 MICROCURRENT MODE TEST .. .. .. .. .. .. .. 26 5.12 HIGH VOLTAGE PULSED CURRENT HVPC MODE TEST .. .. .. .. .. .. .. .. .. .. .. .. .. 27 5.13 ULTRASOUND TESTS. .. .. .. .. .. .. .. .. .. .. .. .. 28 5.14 ULTRASOUND APPLICATOR IDENTIFICATION TEST . .. .. .. .. .. .. .. .. .. .. .. 28 5.15 ULTRASOUND APPLICATOR OUTPUT TEST.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 29 5.16 ULTRASOUND DUTY CYCLE TEST . .. .. .. .. 30 5.17 COMBO OPERATION TEST .. .. .. .. .. .. .. .. .. 31
Intelect® Mobile Stim and Combo Therapy Systems
6 REMOVAL & REPLACEMENT .. .. .. .. .. .. .. .. .. .. .. 32 6.1 SEPARATING TOP & BOTTOM .. .. .. .. .. .. .. .. 32 6.2 THERAPY SYSTEM FAN .. .. .. .. .. .. .. .. .. .. .. .. 33 6.3 POWER SUPPLY .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 34 6.4 CHANNEL 1 STIM BOARD .. .. .. .. .. .. .. .. .. .. . 36 6.5 CHANNEL 2 STIM BOARD .. .. .. .. .. .. .. .. .. .. .. 37 6.6 ULTRASOUND BOARD COMBO SYSTEMS ONLY . .. .. .. .. .. .. .. .. .. .. .. 38 6.7 CONTROL BOARD ASSEMBLY .. .. .. .. .. .. .. .. 39 6.8 KEYMAT ASSEMBLY AND ON/OFF BUTTON KEYMAT .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 40 7 GENERAL MAINTENANCE .. .. .. .. .. .. .. .. .. .. .. .. 41 7.1 CLEANING THE SYSTEM .. .. .. .. .. .. .. .. .. .. .. 41 7.2 CALIBRATION REQUIREMENTS .. .. .. .. .. .. .. 41 7.3 FIELD SERVICE .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 41 7.4 FACTORY SERVICE.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 41 8 ULTRASOUND APPLICATOR CALIBRATION . .. 42 8.1 GENERAL PROCEDURES .. .. .. .. .. .. .. .. .. .. .. 42 9 PARTS . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 43 10 SCHEMATICS . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 51 11 WARRANTY. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 68
FOREWORD
Intelect® Mobile Stim and Combo Therapy Systems
Read, understand, and follow the Safety Precautions and all other information contained in this manual. This manual contains the necessary safety and field service information for those field service technicians, certified by Chattanooga Group, to perform field service on the Intelect Mobile Stim or Combo Therapy Systems. At the time of publication, the information contained herein was current and up-to-date. However, due to continual technological improvements and increased clinical knowledge in the field of electrotherapy, as well as Chattanooga Group’s policy of continual improvement, Chattanooga Group reserves the right to make periodic changes and improvements to their equipment and documentation without any obligation on the part of Chattanooga Group.
©2005 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent from the Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by the Chattanooga Group of Encore Medical, L.P.
1
1 SAFETY PRECAUTIONS
Intelect® Mobile Stim and Combo Therapy Systems
1.1 PRECAUTIONARY SYMBOL DEFINITIONS
1.2 SAFETY PRECAUTIONS
The precautionary instructions found in this manual are indicated by specific symbols. Understand these symbols and their definitions before operating or servicing this equipment. The definitions of these symbols are as follows:
Read, understand, and follow all safety precautions found in this manual. Below are general safety precautions that must be read and understood before attempting any service techniques on these systems.
A. CAUTION
Text with a “CAUTION” indicator will explain possible safety infractions that have the potential to cause minor to moderate injury or damage to equipment.
• Read, understand, and practice the precautionary and operating instructions. Know the limitations and hazards associated with using any electrical stimulation or ultrasound device. Observe the precautionary and operational decals placed on the unit. • DO NOT operate the Intelect Stim or Combo System when connected to any unit other than Chattanooga Group devices. Do not operate the unit in an environment of short-waveform diathermy use. • The Ultrasound modality should be routinely checked before each use to determine that all controls function normally; especially that the intensity control properly adjusts the intensity of the ultrasonic power output in a stable manner. Also, determine that the treatment time control actually terminates ultrasonic power output when the timer reaches zero. • Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous exposure to ultrasonic energy. • DO NOT use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the control panel as damage may result. • Operate, transport, and store this unit in temperatures between 59 °F and 104 °F (15 °C and 40 °C), with Relative Humidity ranging from 30%-60%. • Inappropriate handling of, and subjecting the ultrasound applicator to physical abuse, may adversely affect its characteristics. • Inspect Sound Head and Applicator handle for cracks, which may allow the ingress of conductive fluid before each use. • Inspect all cables, leads, and associated connectors before each use. • Never disconnect Applicator Cable, Lead Wires, Patient Switches, and Mains Power Cord from the system by pulling the cable or wire. Pulling cable or wire may cause system or accessory damage and result in injury to patient and personnel.
B. WARNING
Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.
C. DANGER
Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
D. DANGEROUS VOLTAGE
Text with a “Dangerous Voltage” indicator serves to inform the user of possible hazards resulting in the electrical charge delivered to the patient in certain treatment configurations of TENS waveforms.
E. CORROSIVE HAZARD NIMH BATTERY
Text with a “Corrosive Hazard” indicator will explain possible safety infractions if the chemical components of this product are exposed to air, skin, or other materials.
F. NOTE:
Throughout this manual “NOTE” may be found. These Notes are helpful information to aid in the particular area or function being described.
2
1 SAFETY PRECAUTIONS
Intelect® Mobile Stim and Combo Therapy Systems
1.2 SAFETY PRECAUTIONS (continued)
• These devices are restricted to sale by, or on the order of, a physician or licensed practitioner. This device should be used only under the continued supervision of a physician or licensed practitioner. • For continued protection against fire hazard, replace fuses only with ones of the same type and rating. • Make certain the unit is electrically grounded by connecting only to a grounded electrical service receptacle conforming to the applicable national and local electrical codes. • Care must be taken when operating this equipment around other equipment. Potential electromagnetic or other interference could occur to this or to the other equipment. Try to minimize this interference by not using other equipment in conjunction with it. • The safety of TENS waveforms for use during pregnancy or birth has not been established. • TENS is not effective for pain of central origin. This includes headache. • TENS should be used only under the continued supervision of a physician or licensed practitioner. • TENS waveforms have no curative value. • TENS is a symptomatic treatment, and as such, suppresses the sensation of pain which would otherwise serve as a protective mechanism. • The user must keep the device out of the reach of children. • Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when TENS stimulation is in use. • Powered muscle stimulators should be used only with the leads and electrodes recommended for use by the manufacturer. • In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the system and contact the dealer or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the system that must be tested by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the system. Use of a system that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user or cause extensive internal damage to the system. • Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous exposure to ultrasonic energy. • Before administering any treatment to a patient you should become acquainted with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings, and precautions. Consult other resources for additional information regarding the application of Electrotherapy and Ultrasound. • To prevent electrical shock, disconnect the unit from the power source before attempting any maintenance procedures. • Keep electrodes separated during treatment. Electrodes in contact with each other could result in improper stimulation or skin burns. • Long term effects of chronic electrical stimulation are unknown. • Stimulation should not be applied over the anterior neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing.
• Stimulation should not be applied transthoracically in that • • •
• • • •
3
the introduction of electrical current into the heart may cause cardiac arrhythmia. Stimulation should not be applied over swollen, infected, and inflamed areas or skin eruptions, (e.g., phlebitis, thrombophlebitis, varicose veins, etc.). Stimulation should not be applied over, or in proximity to, cancerous lesions. Output current density is related to electrode size. Improper application may result in patient injury. If any question arises as to the proper electrode size, consult a licensed practitioner prior to therapy. Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. Use only Degassed Water in Power Meter for testing Ultrasound Applicators. Use of other types of water will cause false test results. Refer to page 21 for Degassed Water Recipes. Do not aerate water when filling Power Meter. Unit failing Dielectric Withstand Test or Leakage Test could indicate serious internal problems! Do not place unit back into service! Send unit to factory for repair! Do not attempt to repair.
1 SAFETY PRECAUTIONS
Intelect® Mobile Stim and Combo Therapy Systems
1.2 SAFETY PRECAUTIONS (continued)
Stimulus delivered by the TENS waveforms of this device, in certain configurations, will deliver a charge of 25 microcoulombs (μC) or greater per pulse and may be sufficient to cause electrocution. Electrical current of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia. Patients with an implanted neurostimulation device must not be treated with or be in close proximity to any shortwave diathermy, microwave diathermy, therapeutic ultrasound diathermy, or laser diathermy anywhere on their body. Energy from diathermy (shortwave, microwave, ultrasound, and laser) can be transferred through the implanted neurostimulation system, can cause tissue damage, and can result in severe injury or death. Injury, damage, or death can occur during diathermy therapy even if the implanted neurostimulation system is turned “off.”
4
2 THEORY OF OPERATION
Intelect® Mobile Stim and Combo Therapy Systems
2.1 OVERVIEW
The Intelect Mobile Therapy Systems are comprised of several PC board assemblies housed within a common enclosure. These assemblies each support a distinct function in the product. The basic elements are User Interface, Control Board, Stim Boards, Ultrasound Board (Combo only), Ultrasound Applicator (Combo only), and Power Supply Circuits. 2.2 POWER SUPPLY CIRCUIT
A universal 75 Watt input power supply provides each system with the required 24 volts DC. The supply is connected to the mains at all times when the Mains Power Cord is attached and plugged into an outlet supplying 100 - 240 VAC. The 24 V supply is regulated locally at each PC board as required. 2.3 CONTROL BOARD
The Control Board serves just as its name implies. It controls the operation of the Stim Boards, Ultrasound Board, User Interface, and Accessories. The Control Board communicates to the Stim Boards and Ultrasound Board through a proprietary bus. The Control Board drives the display. The Control Board reads the menu buttons. The Control Board also reads the amplitude and the Contrast Control on the systems. Sound output is generated by the Control Board and routed to an internal speaker. 2.4 STIM BOARDS
The Stim Boards create all muscle stimulation output. Communication to the Stim Boards is via a proprietary bus. A Processor on each Stim Board acts on messages passed to it by the Control Board to set up waveforms and adjust output amplitude. Information can likewise be passed from each Stim Board back to the Control Board for monitoring current, etc. If a Stim Board does not respond as expected to a command from the Control Board, output is stopped and an Error Message is generated. 2.5 ULTRASOUND BOARD AND APPLICATOR COMBINATION SYSTEMS ONLY
The Ultrasound Board generates the 1 or 3.3 MHz output to drive the Sound Head of the Applicator. The Ultrasound Board is accessed through the proprietary bus by the Control Board. It can provide current and voltage information about the ultrasound output of the board. The calibration data for the Sound Head is passed through the Ultrasound Board from the Applicator to the Control Board. By storing the calibration data in the Applicator, there is no calibration necessary for the Ultrasound Board and any calibrated Chattanooga Group Intelect Advanced or Intelect Mobile Ultrasound Applicator can be connected and operated to provide accurate coupling and output. 2.6 USER INTERFACE AND ACCESSORIES
The LCD display panel provides the operator visible feedback in the way of menu choices. Pressing the User Interface buttons makes selections from the menus. The Control Board interprets these user inputs and responds accordingly. Audible feedback is given for such events as key presses and end of treatment. 2.7 NIMH BATTERY
The NiMH Battery Module incorporates a Nickel Metal Hydride (NiMH) Battery Pack and a PC Board. The PC Board monitors the Battery Charge Level. The Battery Pack supplies the required 24 VDC to the system which is then distributed to the respective PCB’s through the Universal Power Supply. The Battery Pack is interfaced with the system via a Wire Harness that facilitates communication with the Control Board and delivery of power to an Electrotherapy or Combination Therapy System. When the Therapy System is connected to a Mains Power Supply via the Mains Power Cord, the NiMH Battery Pack will charge. Once the Battery Pack is fully charged, the software will stop the charging process, eliminating the possibility of overcharging. Battery power is used only when the Therapy System is not connected to a Mains Power Supply.
5
3 NOMENCLATURE
Intelect® Mobile Stim and Combo Therapy Systems
3.1 COMPONENT AND CONTROLS A. Intelect Mobile Combo System
The nomenclature graphics below, Figure 3.1, indicate the general locations of the exterior components of the Two Channel Intelect Mobile Combo System.
Know the components and their functions before performing any operation of or service to the Intelect Mobile Combo System.
CONTRAST CONTROL
FAN VENT
ON/OFF POWER SWITCH
LCD
USER INTERFACE
ULTRASOUND APPLICATOR
PLYNTH
ELECTROTHERAPY LEAD WIRE RECEPTACLES ULTRASOUND RECEPTACLE
FIGURE 3.1
6
3 NOMENCLATURE
Intelect® Mobile Stim and Combo Therapy Systems
3.1 COMPONENT AND CONTROLS LOCATION (continued) B. Intelect Mobile Stim System
The nomenclature graphics below, Figure 3.2, indicate the general locations of the exterior components of the Two Channel Intelect Mobile Electrotherapy System.
Know the components and their functions before performing any operation of or service to the Intelect Mobile Stim system.
CONTRAST CONTROL
FAN VENT
LCD
ON/OFF POWER SWITCH
USER INTERFACE
PLYNTH
ELECTROTHERAPY LEAD WIRE RECEPTACLES
FIGURE 3.2
7
3 NOMENCLATURE
Intelect® Mobile Stim and Combo Therapy Systems
3.2 HARDWARE AND SOFTWARE SYMBOL DEFINITIONS
The symbols below are found on the system as well as within the software. These symbols are defined for the purpose of recognition and functionality when operating or performing service on the Intelect Mobile Combo or Stim Systems.
Know the symbols and their definitions before performing any operation of or service to the Intelect Mobile Combo or Stim Systems.
A. Intelect Mobile Combo and Stim Therapy System Hardware Symbols TREATMENT TIME
CONTRAST CONTROL
PARAMETER DISPLAY/ENTER
ON/OFF SWITCH
INTENSITY
DATA PORT
UP ARROW
STOP TREATMENT
DOWN ARROW PAUSE TREATMENT START TREATMENT
INCREASE
DECREASE
CLINICAL RESOURCES BACK CHANNEL 1 LEAD WIRES CHANNEL 2 LEAD WIRES
ULTRASOUND APPLICATOR
CHARGE LEVEL BATTERY CHARGING
8
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.1 INTELECT MOBILE COMBO SYSTEM
Figure 4.1 below provides physical details of the Intelect Mobile Combo. This section also provides waveform specifications to aid in troubleshooting.
Refer to this section when performing troubleshooting, replacement, and repair of the Intelect Mobile Combo System.
A. Intelect Mobile Combination Therapy System Physical Specifications
DEPTH
HEIGHT
WIDTH
FIGURE 4.1
Dimensions Width.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..25.7 cm (10.125 in) Height . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 18.4 cm (7.250 in) Depth.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 29.2 cm (11.5 in) Weight Standard Weight (with base) .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 2.3 kg (5.07 lb) Battery Pack. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 0.85 kg (1.87 lb) Power Input. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 100 - 240 VAC, 1.0 A, 50/60 Hz 100 W Max Output. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. +24 V, 3.125 A Fuses .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 3.15 A Time Lag (not user serviceable) Electrical Class .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. CLASS I Electrical Type Ultrasound TYPE B.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .
Operating Environment Temperature .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Between 15° C and 40° C (59° F and 104° F) Relative Humidity .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 30%-60% Atmospheric Pressure .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 950-1,050 h Pa Complies with: UL/IEC/EN 60601-1 IEC/EN 60601-1-2 IEC 60601-2-10 IEC 60601-2-5 0413
Electrotherapy TYPE BF. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . Battery Type . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Nickel Metal Hydride (NiMH) (1.2 V x 20 size AA)
9
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.2 INTELECT MOBILE STIM SYSTEM
Figure 4.2 below provides the physical details of the Intelect Mobile Stim. This section also provides waveform specifications to aid in troubleshooting.
Refer to this section when performing troubleshooting, replacement, and repair of the Intelect Mobile Stim .
A. Intelect Mobile Stim Therapy System Physical Specifications
DEPTH
HEIGHT
WIDTH
FIGURE 4.2
Dimensions Width.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..25.7 cm (10.125 in) Height . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 16.8 cm (6.625 in) Depth.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 29.2 cm (11.5 in) Weight Standard Weight (with base) .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 2.3 kg (5.07 lb) Battery Pack. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 0.85 kg (1.87 lb) Power Input. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 100 - 240 VAC, 1.0 A, 50/60 Hz 100 W Max Output. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. +24 V, 3.125 A Fuses .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 3.15 A Time Lag (not user serviceable) Electrical Class .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. CLASS I Electrical Type Electrotherapy TYPE BF. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . Battery Type . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Nickel Metal Hydride (NiMH) (1.2 V x 20 size AA)
Operating Environment Temperature .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Between 15° C and 40° C (59° F and 104° F) Relative Humidity .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 30%-60% Atmospheric Pressure .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 950-1,050 h Pa Complies with: UL/IEC/EN 60601-1 IEC/EN 60601-1-2 IEC 60601-2-10 IEC 60601-2-5 0413
10
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS
NOTE: All waveforms except High Voltage Pulsed Current (HVPC) of the Intelect Mobile Therapy System have been designed with a 200 mA current limit. VMS™, and all TENS waveform output intensities are measured, specified, and listed to peak, not peak to peak. All Waveforms are available on all channels.
The specifications found in this section provide the necessary waveform specifications to aid in troubleshooting. A waveform graphic from an oscilloscope is also provided for clarification. Refer to this section when performing troubleshooting, replacement, and repair of the Intelect Mobile Stim and Combo Systems.
A. IFC (Interferential) Traditional (4 Pole)- Figure 4.3
Interferential Current is a medium frequency waveform. Current is distributed from two channels (four electrodes). The currents cross in the body within the area being treated. The two currents interfere with each other at this crossing point, resulting in a modulation of the intensity (the current intensity increases and decreases at a regular frequency). Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2,000-10,000 Hz Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 seconds Sweep Low Beat Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz Scan Percentage . . . . . . . . . . . . . . . . . . . . . . . . .Static, 40%, and 100% Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.3
B. TENS- Asymmetrical Biphasic- Figure 4.4
The Asymmetrical Biphasic waveform has a short pulse duration. It is capable of strong stimulation of the nerve fibers in the skin as well as of muscle tissue. This waveform is often used in TENS devices. Because of its short pulse, the patient typically tolerates the current well, even at relatively high intensities. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-110 mA Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-1,000 μsec Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-250 Hz Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV* Burst Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-25 bps Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-250 Hz Amplitude Modulation. . . . . . . . . . . Off, 40%, 60%, 80%, and 100% Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.4
Stimulus delivered by the TENS waveforms of this device, in certain configurations, will deliver a charge of 25 microcoulombs (μC) or greater per pulse and may be sufficient to cause electrocution. Electrical current of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia.
*CC= Constant Current CV= Constant Voltage
11
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS (continued) C. TENS- Symmetrical Biphasic- Figure 4.5
The Symmetrical Biphasic waveform has a short pulse duration and is capable of strong stimulation of nerve fibers in the skin and in muscle. This waveform is often used in portable muscle stimulation units and some TENS devices. Because of its short pulse duration, the patient typically tolerates the current well, even at relatively high intensities. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-1,000 μsec Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-250 Hz Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV* Burst Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-25 bps Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-250 Hz Amplitude Modulation. . . . . . . . . . . Off, 40%, 60%, 80%, and 100% Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.5
D. High Voltage Pulsed Current (HVPC)- Figure 4.6
The High Voltage Pulsed Current (HVPC) has a very brief pulse duration characterized by 2 distinct peaks delivered at high voltage. The waveform is monophasic (current flows in one direction only). The high voltage causes a decreased skin resistance, making the current comfortable and easy to tolerate. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes or Probe Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-500 V Polarity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Positive or Negative Ramp . . . . . . . . . . . . . . . .0.5 seconds, 1 seconds, 2 seconds, 5 seconds Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Peak Current or Volts Sweep . . . . . . . . . . . . . . Continuous, 80/120 pps, 1/120 pps, 1/10 pps Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-120 Hz Cycle Time . . . . . . . . . . . . . . . . . . . . . . . 5/5, 4/12, 10/10, 10/20, 10/30, 10/50, and Continuous Treatment Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-99 Minutes
FIGURE 4.6
E. VMS™- Figure 4.7
VMS is a symmetrical biphasic waveform with a 100 μsec interphase interval. Because the pulse is relatively short, the waveform has a low skin load, making it suitable for applications requiring high intensities such as in muscle strengthening protocols. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-200 mA Channel Mode . . . . . . . . . . . . . . . . . . . . Single, Reciprocal, Co-Contract Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-1000 μsec Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV* Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off or On Set Intensity . . .Individual Channel Intensity Setting in Reciprocal and Co-Contract modes Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50 Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 pps Ramp . . . . . . . . . . . . . . .0.5 seconds, 1 seconds, 2 seconds, 5 seconds Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.7
*CC= Constant Current CV= Constant Voltage
12
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS (continued) F. Diadynamic Waveforms- Figures 4.8 - 4.12
The Diadynamic waveforms are rectified alternating currents. The alternating current is modified (rectified) to allow the current to flow in one direction only. Output Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-80 mA Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes MF: (Monophasé Fixe)- Figure 4.8 Frequency of 50 Hz: phase duration of 10 ms followed by a pause of 10 ms. DF: (Diphasé Fixe)- Figure 4.9 Frequency of 100 Hz: phase duration of 10 ms followed immediately by another identical phase of 10 ms. CP: Modulé en Courtes Périodes- Figure 4.10 1 second of MF followed abruptly by 1 second of DF. LP: (Modulé en Longues Périodes)- Figure 4.11 Rhythmical fluctuation between 2 MF currents. CP-iso: (Courtes Periodes Isodynamic)- Figure 4.12 A combination of MF and DF waveforms.
FIGURE 4.9
FIGURE 4.10
FIGURE 4.8 FIGURE 4.11
*CC= Constant Current CV= Constant Voltage
FIGURE 4.12
13
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS (continued) G. IFC (Interferential) Premodulated (2p)Figure 4.13
Premodulated Current is a medium frequency waveform. Current is distributed from one channel (two electrodes). The current intensity is modulated: it increases and decreases at a regular frequency (the Amplitude Modulation Frequency). Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2,000-10,000 Hz Beat Fixed (Sweep Off) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-200 Hz Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-199 Hz Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .2-200 Hz Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, and 10/50 Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV* Treatment Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.13
H. Russian- Figure 4.14
Russian Current is a sinusoidal waveform delivered in bursts or series of pulses. This method was claimed by its author (Kots) to produce maximal muscle strengthening effects without significant discomfort to the patient. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA Channel Mode . . . . . . . . . . . . . . . . . . . . Single, Reciprocal, Co-Contract Duty Cycle . . . . . . . . . . . . . . . . . . . . . .10%, 20%, 30%, 40%, and 50% Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV* Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off or On Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5/5, 4/12, 10/10, 10/20, 10/30, 10/50, and Continuous Burst Frequency (Anti-Fatigue Off) . . . . . . . . . . . . . . . . . . . .20-100 pps Ramp . . . . . . . . . . . . . . . .0.5 seconds, 1 seconds, 2 seconds, 5 seconds Treatment Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.14
I. Microcurrent- Figure 4.15
Microcurrent is a monophasic waveform of very low intensity. The literature reports beneficial effects of this waveform in the treatment of wounds. The physiological working mechanism of this effect is as yet not clearly understood. It is thought to promote tissue healing by stimulating the "current of injury": a current which naturally occurs in healing tissue. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes or Probe Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1,000.0 μA Polarity . . . . . . . . . . . . . . . . . . . . . . . . .Positive, Negative, or Alternating Treatment Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
*CC= Constant Current CV= Constant Voltage
FIGURE 4.15
14
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS (continued) J. MONOPHASIC: Monophasic Rectangular Pulsed Figure 4.16
The Monophasic Rectangular Pulsed waveform is an interrupted unidirectional current with a rectangular pulse shape. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-80 mA Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1-500.0 ms Phase Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5,000 ms Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.16 K. MONOPHASIC: Monophasic Triangular Pulsed Figure 4.17
The Monophasic Triangular Pulsed waveform is an interrupted unidirectional current with a triangular pulse shape. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-80 mA Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1-500.0 ms Phase Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5,000 ms Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
FIGURE 4.17 L. GALVANIC: Continuous- Figure 4.18
Continuous Galvanic Current is a direct current flowing in one direction only. Output Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA Polarity Reversal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .On or Off With Polarity Reversal On, Polarity will change after 50% of treatment time. Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . . Continuous, 5/60, and 10/60 Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-60 Minutes
FIGURE 4.18
*CC= Constant Current CV= Constant Voltage
15
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.3 INTELECT ELECTROTHERAPY WAVEFORM SPECIFICATIONS (continued) M. GALVANIC: Interrupted- Figure 4.19
Interrupted Galvanic Current is a direct current flowing in one direction only. The current is delivered in pulses. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA Polarity Reversal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .On or Off With Polarity Reversal On, Polarity will change after 50% of treatment time Cycle Time. . . . . . . . . . . . . . . . . . . . . . . . . Continuous, 5/60, and 10/60 Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8,000 Hz Duty Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95% Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-60 Minutes FIGURE 4.19
N. Träbert (Ultrareiz)- Figure 4.20
Träbert is a monophasic waveform with a phase duration of 2 ms and a pause of 5 ms resulting in a frequency of approximately 143 Hz. Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA Polarity Reversal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .On or Off With Polarity Reversal On, Polarity will change after 50% of treatment time. Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-60 Minutes
FIGURE 4.20
16
4 SPECIFICATIONS
Intelect® Mobile Stim and Combo Therapy Systems
4.4 INTELECT MOBILE COMBO ULTRASOUND SPECIFICATIONS
This section provides the necessary Ultrasound Specifications to aid in troubleshooting. Refer to these specifications as necessary when troubleshooting the Ultrasound PC Board and Applicators. A. Ultrasound
Frequency. . . . . . . . . . . . . . . . . . . . . . . . 1 MHz, ± 5%; 3.3 MHz, ±5% Duty Cycles . . . . . . . . . . . . . . . . . . . . 10%, 20%, 50%, and Continuous Pulse Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .16 Hz, 48 Hz, 100 Hz Pulse Duration . . . . . . . . . . . . . . . . . . . 1 mSec, ±20%; 2 mSec, ±20% 5 mSec, ±20% Output Power 10 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-20 Watts at 1 MHz, 0-10 Watts at 3.3 MHz 2 5 cm Crystal . . . . . . . . . . . . . . . . . . . . . . . . 0-10 Watts, 1 and 3.3 MHz 2 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . .0-4 Watts, 1 and 3.3 MHz 1 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-2 Watts 3.3 MHz Only Amplitude . . . . . . . . . . . . . . . . . . . 0 - 2.5 w/cm2 in Continuous mode, 0-3 w/cm2 in Pulsed modes Output accuracy . . . . . . . . . . . . . . . . ± 20% above 10% of maximum Temporal Peak to Average Ratios: 2:1, ± 20%, at 50% Duty Cycle 5:1, ± 20%, at 20% Duty Cycle 9:1, ± 20%, at 10% Duty Cycle Beam Nonuniformity Ratio. . . . . . . . . . 5.0 : 1 maximum Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Collimating Effective Radiating Areas . . . . . . . . . . .10 cm2 Crystal - 8.5 cm2, ±1.5 5 cm2 Crystal - 4.0 cm2, ±1.0 2 cm2 Crystal - 1.8 cm2, +0.2/-0.4 1 cm2 Crystal - 0.8 cm2, +0.2/-0.4 Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30 Minutes B. Head Warming Feature Specifications
The Head Warming feature of an Intelect Combination Therapy System utilizes Ultrasound output resulting in warming of the Sound Head to increase patient comfort. With Head Warming enabled, ultrasound is emitted without pressing the START button. The Applicator LED will not illuminate during the Head Warming period. US Channel will indicate "Warming". Output .. .. .. .. .. .. .. .. .. .. .. .. .. ..0 - 50% Cycling of maximum power Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 Mhz
Do not apply the Ultrasound Applicator to the patient during the Head Warming period. Applicator must remain in Applicator Hook during the Head Warming period.
17
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.1 INTELECT MOBILE STIM AND COMBO ERROR MESSAGES
A. The following information is provided as an aid in defining the Software Error Messages of the Intelect Mobile Therapy System. Once a particular Error Message is defined, the information will also list probable causes and possible remedies. Once the problem area is determined, subsequent tests for verification will be necessary to determine a “Bad Board”.
ERROR CODE
ERROR TYPE
DEFINITION
All Troubleshooting and tests will be to validate a “Bad Board” only. No component level troubleshooting information is or will be provided by Chattanooga Group for field troubleshooting of board components. B. Once a particular PC Board has been determined as bad, refer to the appropriate Removal and Replacement Section for the board affected and follow the instructions for replacement of the board.
PROBABLE CAUSES
POSSIBLE REMEDY
USER CORRECTABLE WARNING MESSAGES 100
WARNING
Ultrasound Applicator became unplugged.
Ultrasound Applicator was unplugged while an Ultrasound treatment was running.
Plug Ultrasound Applicator into proper receptacle on unit making certain it is completely seated.
101
WARNING
Ultrasound Applicator unplugged.
User attempted to start an Ultrasound treatment, but no Ultrasound Applicator was plugged into unit.
Plug Ultrasound Applicator into proper receptacle on unit making certain it is completely seated.
102
WARNING
Ultrasound Applicator not calibrated.
The Ultrasound Applicator plugged into the unit needs to be calibrated.
Contact dealer or Chattanooga Group for service.
103
WARNING
Ultrasound Channel not available.
User attempted to select Combo treatment, but the Ultrasound Channel was already in use.
Wait until Ultrasound treatment is completed or stop Ultrasound treatment and try again.
104
WARNING
Stim Channel not available.
User attempted to select an Electrotherapy or Combo treatment, but all Stim Channels are in use.
Wait until Electrotherapy treatment is completed or stop Electrotherapy treatment and try again.
105
WARNING
Stim Channels not available.
User attempted to select a two channel Electrotherapy treatment, but at least one of the two stim channels were already in use.
Wait until Electrotherapy treatment is completed or stop Electrotherapy treatment and try again.
106
WARNING
Overcurrent
Stim channel has exceeded allowed current level and the treatment has been stopped.
Reset treatement parameters and attempt session again.
107
WARNING
Bad Contact Quality.
Electrode contact is poor.
Apply new electrodes to the treatment area.
108
WARNING
Shorted Lead Wires
Lead Wires are bad.
Replace with new lead wires.
109
WARNING
Power Supply current limit.
User attempted to start two channels of Electrotherapy while running an Ultrasound treatment with a 10 cm2 Ultrasound Applicator and Ultrasound Output is currently set to greater than 15 Watts..
Wait until Ultrasound treatment is completed or stop Ultrasound treatment and try again or decrease ultrasound output to less than 15 Watts.
18
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.1 AND COMBO ERROR MESSAGES (continued) 5.1 INTELECT INTELECT MOBILE MOBILE STIM SYSTEM SOFTWARE ERROR MESSAGES (continued)
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the unit. Use of a unit that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user, or may cause extensive internal damage to the unit. ERROR CODE
ERROR TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
ERROR MESSAGES (200-213) REQUIRING TECHNICAL ASSISTANCE 200
ERROR
Error while attempting to save Ultrasound Applicator Calibration Data.
Could not save the Calibration Data to the Ultrasound Applicator.
1. Replace the Ultrasound Applicator with a known good Ultrasound Applicator. 2. Replace Ultrasound Board. 3. Replace Control Board.
201
ERROR
Error Applicator not calibrated OK.
Could not calibrate Ultrasound Applicator.
1. Refer to 8.1 and attempt to calibrate again. 2. Replace the Ultrasound Applicator with a known good Ultrasound Applicator. 3. Replace Ultrasound Board. 4. Replace Control Board.
202
ERROR
Timed out while saving the Ultrasound Applicator Calibration Data.
Could not save the Calibration Data to the Ultrasound Applicator.
1. Replace the Ultrasound Applicator with a known good Ultrasound Applicator. 2. Replace Ultrasound Board. 3. Replace Control Board.
203
ERROR
Error reading Protocol.
Error reading a Protocol from the EEPROM.
Restore Factory Settings, restore Factory Protocols and rebuild all User Protocols.
204
ERROR
Main Software Flash Erase Error.
Stim Main Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
205
ERROR
Main Software Flash Echo.
Stim Main Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
206
ERROR
Main Software CRC Error.
Stim Main Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board..
207
ERROR
Main Software Program Flash Error.
Stim Main Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
208
ERROR
Main Software Acknowledge Error.
Stim Main Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
209
ERROR
Software CRC Acknowledge Error.
Software upgrade Error
Replace Control Board..
210
ERROR
Channel Software Flash Erase Error.
Stim Channel Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
211
ERROR
Channel Software CRC Error.
Stim Channel Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
212
ERROR
Channel Software Program Flash Error.
Stim Channel Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
213
ERROR
Channel Software Acknowledge Error.
Stim Channel Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
214
ERROR
Channel Software CRC Acknowledge Error.
Stim Channel Software upgrade Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
19
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.1 INTELECT MOBILE STIM AND COMBO ERROR MESSAGES (continued)
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the unit. Use of a unit that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user, or may cause extensive internal damage to the unit. ERROR CODE
ERROR TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
CRITICAL ERRORS (300-314) DEMANDING TECHNICAL SERVICE 300
CRITICAL ERROR
Unit CFG Critical Error.
Error communicating with Stim Board on Powerup.
1. Replace appropriate Stim Board. 2. Replace Control Board.
301
CRITICAL ERROR
No Stim Board Critical Error.
Error detecting Stim Board on Powerup.
1. Replace appropriate Stim Board. 2. Replace Control Board.
302
CRITICAL ERROR
No Ultrasound Board Critical Error.
Error detecting Ultrasound Board on Powerup.
1. Replace Ultrasound Board. 2. Replace Control Board.
303
CRITICAL ERROR
EEPROM Critical Error.
Error reading EEPROM on Powerup.
Replace Control Board.
304
CRITICAL ERROR
Ultrasound Board Critical Error.
Error communicating with the Ultrasound Board.
1. Replace Ultrasound Board. 2. Replace Control Board.
305
CRITICAL ERROR
Ultrasound Board Write Critical Error.
Error communicating with the Ultrasound Board.
1. Replace Ultrasound Board. 2. Replace Control Board.
306
CRITICAL ERROR
Ultrasound Board Read_Write Critical Error.
Error communicating with the Ultrasound Board.
1. Replace Ultrasound Board. 2. Replace Control Board.
307
CRITICAL ERROR
Ultrasound Board Reset Critical Error.
Ultrasound Board Reset Error.
1. Replace Ultrasound Board. 2. Replace Control Board.
308
CRITICAL ERROR
Ultrasound Board Read Critical Error.
Error communicating with the Ultrasound Board.
1. Replace Ultrasound Board. 2. Replace Control Board.
309
CRITICAL ERROR
Ultrasound Board Calibration Critical Error.
Error calibrating Ultrasound Board.
1. Replace Ultrasound Board. 2. Replace Control Board.
310
CRITICAL ERROR
Stim Board Write Critical Error.
Error communicating with Stim Board.
1. Replace appropriate Stim Board. 2. Replace Control Board.
311
CRITICAL ERROR
Stim Board Bad Data Read Critical Error.
Error communicating with Stim Board.
1. Replace appropriate Stim Board. 2. Replace Control Board.
312
CRITICAL ERROR
Stim Board Main UP Reset Critical Error.
Error communicating with Stim Board.
1. Replace appropriate Stim Board. 2. Replace Control Board.
313
CRITICAL ERROR
Stim Board Channel 1 UP Reset Critical Error.
Error communicating with Stim Board.
1. Replace appropriate Stim Board. 2. Replace Control Board.
314
CRITICAL ERROR
Stim Board Channel 2 UP Reset Critical Error.
Error communicating with Stim Board.
1. Replace appropriate Stim Board. 2. Replace Control Board.
315
CRITICAL ERROR
Stim Board Reset Critical Error.
Stim Board Reset Error.
1. Replace appropriate Stim Board. 2. Replace Control Board.
316
CRITICAL ERROR
Stim Powerup Test Failed Critical Error.
Stim Board failed its Self Test on Powerup.
1. Replace appropriate Stim Board. 2. Replace Control Board.
20
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.2 INTELECT MOBILE STIM AND COMBO SYSTEM TESTING
11. Degassed Water (<5 ppm) for Ultrasound Power Meter. Recipe(s) for Degassed Water 1) Boil Distilled Water for 30 Minutes. Place water in a non-porous container and immediately cover with cellophane. Allow to cool to room temperature of approximately 70 °F (21 °C). May be refrigerated to aid cooling time. or 2) Bring Distilled Water to a boil. Place the container under vacuum for 5 to 10 Minutes.
A. General
1. The following information is intended to aid in troubleshooting the major components of the Intelect Stim and Combo Therapy Systems to “Board Level” only. These tests are FACTORY standard testing procedures and methods used at the factory before shipment of any Intelect Therapy System. 2. Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshooting techniques are to determine “Bad Board” and board replacement only. No board component level troubleshooting is recommended, nor will information or parts be supplied by Chattanooga Group. Any board component level troubleshooting performed will be at sole risk and liability of the Service Technician performing such troubleshooting techniques. 3. Once a particular PC Board has been determined as bad, refer to the appropriate Removal and Replacement Section of this Manual for proper replacement.
NOTE: Two liter soft drink bottles are ideal storage and transport containers for degassed water as they are designed to keep oxygen out. Do not allow aeration of degassed water during transport or filling of the power meter. Do not use Tap Water or Distilled water in the Ultrasound Power Meter. Use only Degassed Water in order to obtain correct test results. The chart below illustrates the oxygen content of Degassed, Tap, and Distilled Water.
B. Special Tools, Fixtures, & Materials Required
WATER TYPE Degassed (per Recipe 1 or 2) Tap Water Distilled Water
1. Certain tests require the use of special tools and fixtures. These will be listed at the particular test where they are required. Testing with any other special tool or fixture other than those stated could give erroneous readings or test results. Always perform the tests exactly as stated to ensure accurate results. 2. Any special tools or fixtures required can be obtained through the Chattanooga Group Service Department. 3. Scope and other standard test equipment settings will be listed for each test performed to aid in performing the test to FACTORY standards and ensure proper readings. 4. The troubleshooting and repair of the Intelect Therapy Systems and Accessories should be performed only by authorized technicians trained and certified by Chattanooga Group.
ppm of OXYGEN Less than 5 ppm Up to 35 ppm Up to 20 ppm
D. Full Functional Tests
Perform the tests found in this section to verify Full Functionality of new Therapy System and accessories. E. Required Handtools
1. # 1 Phillips Screwdriver 2. Insulated Needle Nose Pliers 3. 1/4 in. Wrench
C. Equipment Required
Oscilloscope and Probes ESTI-2 Load Test Fixture Digital Multimeter Microcurrent Probe (Accessory) Intelect Mobile or Advanced Ultrasound Applicators (Accessories) Dielectric Withstand (Hi-Pot) and ground resistance tester NOTE: Adjust Dielectric Withstand tester to indicate fault with 120 k Ohm Load across the output when at specified test voltage. 7. Carbon Electrodes 8. Milliohm Meter 9. Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter 10. Dissolved Oxygen Test Kit used to test oxygen level of degassed water. 1. 2. 3. 4. 5. 6.
21
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.3 VISUAL INSPECTION General
Visually inspect the Intelect Mobile Therapy System. A visual inspection can, to an experienced technician, indicate possible abuse of the unit and internal problems. 5.4 LEAKAGE TESTS
Conduct all necessary leakage tests as required per “Chapter 7 Electrical Equipment” of the 1999, or later, edition of the NFPA (National Fire Protection Association) “Health Care Facility” standards. See Figure 5.1.
Unit failing Dielectric Withstand Test or Leakage Test could indicate serious internal problems! Do not place unit back into service! Send unit to factory for repair! Do not attempt to repair. 5.5 UNIT STARTUP AND FAN TESTING A. Test
1. Place unit face up on work surface. 2. Connect power cord to unit and plug into proper power receptacle. 3. Turn system on. Press the Enter button. IFC-2p should be highlighted. Press the Enter button. 4. Place hand at the back of system, at Contrast Control, to verify fan is blowing out. See Figure 5.2.
FIGURE 5.1
B. Test Results
1. Unit will not start, unit failed test. a) Possible bad Main Power Switch. b) Possible bad Power Supply. c) Possible bad power outlet or Mains Power Cord. 2. Screen does not display, unit failed test. a) Contrast Control needs adjusting. b) Possible bad display. c) Possible bad Control Board. d) Possible bad Power Supply. e) Visually check power LED. LED should illuminate Blue. Turn system off with Power button. Power LED should flash Blue. If Power LED illuminates Blue with system On and flashes Blue with system Off, the Power Supply is good. Replace Control Board. 3. Fan not blowing outward= Unit Failed Test a) Fan blowing inward. Fan wired wrong. Rewire or replace Fan. b) Fan not blowing. 1) Possible bad Fan. 2) Possible bad Power Supply. 3) Possible bad Control Board.
FIGURE 5.2
22
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.6 ELECTRICAL STIMULATOR TEST SYSTEM SETUP
The following tests for Stimulator Outputs will be performed on Channels 1 and 2. A. Equipment Required
1. ESTI-2 Load Test Fixture 2. Calibrated Oscilloscope and Probes B. System Set Up
1. Install known good Lead Wires to Channels 1 and 2 on the system. See Figure 5.3. 2. Connect Lead Wires from the system to the ESTI-2 Load Test Fixture. Channel 1 to Channel 1 IN and Channel 2 to Channel 2 IN. See Figure 5.4. 3. Connect Scope Probes to the Channel 1 To SCOPE and Channel 2 To SCOPE Tabs on the ESTI 2 Load Test Fixture respectively. See Figure 5.4. 4. Place ESTI-2 Load Switch in the 1 K position. See Figure 5.4. 5. Install Power Cord into system and plug into proper Power Supply. Turn system On.
FIGURE 5.3 SCOPE TO ESTI2
5.7 VMS™ MODE TEST A. VMS™ Mode Test Procedures
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15.
LOAD SWITCH TO 1 K
Set Scope; Time- 100 μS, Channel- 50 V, and Trigger- DC. Press Enter button. Highlight and select VMS, then press Enter button. Highlight Channel Mode and press the Enter button until Co-Contract is displayed to the right of Channel Mode. Highlight Cycle Time and press the Enter button. Highlight Continuous and press the Enter button. Highlight Phase Duration and press the Enter Button. Press the Up or Down Arrow button until 300 is displayed. Press the Enter button. Increase unit Intensity until 200 is displayed. Press START. Compare waveform on scope to Figure 5.5. Press STOP button. Highlight Channel 2. Repeat steps 2 through 13.
THERAPY SYSTEM TO ESTI2
FIGURE 5.4
B. VMS™ Mode Test Results
1. Waveform is the same between scope and Figure 5.5. Unit passed test. 2. No waveform or considerably different waveform. Unit failed test. Replace appropriate Stim PC Board. FIGURE 5.5
23
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.8 INTERFERENTIAL MODE TEST
It is assumed that the unit is ready for tests as described in 5.6 parts A and B. If not, set up unit per 5.6 parts A and B prior to performing tests. A. Interferential Mode Test Procedures
1. 2. 3. 4. 5. 6. 7. 8.
Set Scope; Time- 100 μS, Channel- 20 V, and Trigger- DC. Highlight Stim Channel 1. Press the Enter button. Highlight IFC-4p. Press the Enter button. Increase Intensity until 50 is displayed. Press START button. Compare waveform on scope to Figure 5.6. Press PAUSE button. Verify that the amplitude displayed below timer drops to zero (0). 9. Verify that Paused is displayed below the displayed amplitude. 10. Press STOP button
FIGURE 5.6
B. Interferential Mode Test Results
1. Waveform is the same between scope and Figure 5.6, amplitude dropped to zero when paused and “Paused” displayed below timer. Unit passed test. 2. No waveform or considerably different waveform. Unit failed test. Replace appropriate Stim Board. 3. Amplitude failed to “zero” when paused. Unit failed test. Replace appropriate Stim Board. 4. “Paused” did not display when unit paused. Unit failed test. Replace appropriate Stim Board.
5.9 PREMODULATED MODE TEST
Set up System per 5.6 parts A and B prior to performing test. A. Premodulated Mode Test Procedures
1. 2. 3. 4. 5. 6. 7. 8.
Set Scope; Time- 2.50 mS, Channel- 20 V, and Trigger- DC Highlight Stim Channel 1. Press the Enter button. Highlight IFC-2p. Press the Enter button. Increase Intensity until 50 is displayed. Press START button. Compare waveform on scope to Figure 5.7. Press STOP button. Highlight Channel 2 and repeat steps 3 through 7.
B. Premodulated Mode Test Results
1. Waveform is the same between scope and Figure 5.7. Unit passed test. 2. No waveform or considerably different waveform. Unit failed test. Replace appropriate Stim Board.
FIGURE 5.7
24
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.10 RUSSIAN MODE TEST
Set up System per 5.6 parts A and B prior to performing test. A. Russian Mode Test Procedures
1. 3. 4. 5.
Set Scope; Time- 5 mS, Channel- 50 V, and Trigger- DC Highlight Stim Channel 1. Press Enter button. Highlight Russian. Press Enter button. Highlight Channel Mode. Press the Enter button until Co-Contract is displayed. 6. Highlight Cycle Time. Press the Enter button. 7. Highlight Continuous. Press the Enter button. 8. Increase Intensity until 100 is displayed. 9. Press START button. 10. Compare waveform on scope to Figure 5.8. 11. Verify that both Channels reach 100. 12. Press STOP button. 13. Highlight Channel 2 and repeat steps 4 through 12.
SPEC: 200V PEAK TO PEAK ±10%
FIGURE 5.8
B. Russian Mode Test Results
1. Waveform is the same between scope and Figure 5.9 and amplitude reached 100. Unit passed test. 2. No waveform or considerably different waveform. Unit failed test. Replace appropriate Stim Board. 3. Amplitude failed to reach 100 on both Channels. Unit failed test. Replace appropriate Stim Board.
25
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.11 MICROCURRENT MODE TEST
1. Set up System per 5.6 parts A and B prior to performing test. 2. Place ESTI-2 Load Switch in the 10 K Micro position only for the Microcurrent Mode Tests. See Figure 5.9. LOAD SWITCH TO 10K MICRO
A. Microcurrent Mode Test Procedures.
1. 2. 3. 4. 5. 6. 7.
Set Scope; Time- 250 μS, Channel- 5.0 V, and Trigger- DC Highlight Stim Channel 1. Press the Enter button. Highlight Microcurrent. Press the Enter button. Highlight Frequency. Press the Up or Down Arrow button until 1000.0 Hz is displayed. Press the Enter button. Highlight Polarity. Press the Enter button until Alternating is displayed beside Polarity.
NOTE: The Frequency value will continue to Ramp and rotate due to Alternating Polarity being selected. This is normal.
FIGURE 5.9 PROPER POSITIVE + “MICROCURRENT” WAVEFORM
8. Increase Intensity until 1000 uA is displayed. 9. Press START button. 10. Compare waveform on scope to Figure 5.10 and Figure 5.10A. NOTE: The output will alternate between positive and negative on the scope. 11. Press STOP button. 12. Highlight Stim Channel 2. Press the Enter button and repeat steps 3 through 11. B. MicroCurrent Mode Test Results
1. Waveform is the same between scope and Figure 5.10 and Figure 5.10A. Unit passed test. 2. No waveform or considerably different waveform. Unit failed test. Replace appropriate Stim Board.
SPEC: 10V PEAK TO PEAK ±10%
FIGURE 5.10 PROPER NEGATIVE “MICROCURRENT” WAVEFORM
SPEC: 10V PEAK TO PEAK ±10%
FIGURE 5.10A
26
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.12 HIGH VOLTAGE PULSED CURRENT HVPC MODE TEST
Set up unit per 5.6 parts A and B prior to performing tests. PROPER NEGATIVE HIGH VOLT WAVEFORM
A. High Voltage Pulsed Current (HVPC) Mode Test Procedures
1. 2. 3. 4. 5.
Set Scope; Time- 25 μS, Channel- 50 V, and Trigger- DC Highlight Stim Channel 1. Press the Enter button. Highlight High Volt. Press the Enter button. Increase Intensity until 250 V is displayed. Highlight Display and press the Enter button until Peak Current is displayed. Press the Enter button. 6. Press START button. 7. Compare waveform on scope to Figure 5.11. 8. Highlight Polarity. Press the Enter button until Positive is displayed. 9. Compare waveform form on scope to Figure 5.12. 10. The numbers displayed for amplitude must not exceed 1.5 Amps. See Figure 5.13. 11. Press STOP button. 12. Highlight Channel 2. 13. Press the Enter button and repeat steps 3 through 12.
FIGURE 5.11 PROPER POSITIVE HIGH VOLT WAVEFORM
B. High Voltage Pulsed Current (HVPC) Mode Test Results
1. Waveforms on scope the same as Figures 5.11 and 5.12. Amps do not exceed 1.5. Unit passed test. 2. No waveform or considerably different waveforms. Unit failed test. Replace appropriate Stim Board. 3. Amps exceed 1.5. Unit failed test. Replace appropriate Stim Board. FIGURE 5.12
AMPS MUST NOT EXCEED 1.5
FIGURE 5.13
27
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.13 ULTRASOUND TESTS A. Equipment Required
1. Degassed Water. Refer to page 21 for Degassed Water Recipes. 2. Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter. 3. Dissolved Oxygen Test Kit. Used to test oxygen level of degassed water. 4. Intelect Mobile or Advanced Ultrasound Applicator. 5.14 ULTRASOUND APPLICATOR IDENTIFICATION TEST
NOTE: Use any Intelect Mobile or Advanced Ultrasound Applicator for this test.
FIGURE 5.14
A. Ultrasound Applicator Identification Test Procedures
1. Without Ultrasound Applicator installed, turn unit on. 2. View the Ultrasound channel in the lower right corner of screen. It should read “Unplugged”. See Figure 5.14. 3. Connect Intelect Mobile or Advanced Applicator into Applicator receptacle. See Figure 5.15. Watch Applicator LED while connecting to system. The LED should flash Green five times. 4. Look at the Ultrasound channel. It should read Available. See Figure 5.15. 5. Highlight Ultrasound. Press the Enter button. 6. Highlight Warming. Press the Enter button until On is displayed beside Warming. 7. Press the Back button. Turn System Off and back On with Main Power Switch. After System boots, view the Ultrasound channel, Warming should be visible. See Figure 5.16. B. Ultrasound Applicator Identification Test Results
1. Unit operates as described in steps 2, 4, and 7. Unit passed test. 2. No Cal. displays in Ultrasound channel. a) Applicator not calibrated or needs recalibration. b) Possible bad Applicator. Retest with known good Applicator. 3. Unplugged displays after ten seconds of Applicator being connected to System. a) Possible bad applicator. Retest with known good Applicator. b) Possible bad internal connection at Ultrasound Board. c) Possible bad Ultrasound Board. d) Possible bad Control Board.
FIGURE 5.15
FIGURE 5.16
28
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.15 ULTRASOUND APPLICATOR OUTPUT TEST
Perform this test using all available Intelect Mobile or Advanced Ultrasound Applicators used with the System being tested. A. Ultrasound Applicator Output Test Procedures
1. Set up Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter per Operator's Instructions and fill test reservoir with Degassed Water.
Use only Degassed Water in Power Meter for testing Ultrasound Applicators. Use of other types of water will cause false test results. Refer to page 21 for Degassed Water Recipes. Do not aerate water when filling Power Meter. 2. Place an Applicator into the Power Meter retainer. Make certain the Sound Head is completely submerged in the degassed water and centered directly over the Stainless Steel Cone. See Figure 5.17. 3. Zero or Tare meter. 4. Highlight Ultrasound. Press the Enter button. 5. Highlight Duty Cycle. Press the Enter button. Highlight Continuous and press the Enter button. 6. Highlight Display. Press the Enter button until Watts displays. 7. Press START button. 8. Increase Intensity until the appropriate Watts is displayed per Figure 5.18. 9. Compare Power Meter readings to Figure 5.18 to all settings for the respective Applicator being tested as shown in Figure 5.18. 10. Press Frequency button until 3.3 MHz is displayed within the Frequency icon. Repeat test and compare readings to Figure 5.18.
FIGURE 5.17
APPLICATOR OUTPUT SPECIFICATIONS APPLICATOR SIZE
POWER SETTING WATTS
OUTPUT RANGE
1 cm2
1* 2* 1 2 4
0.8 - 1.2 1.6 - 2.4 0.8 - 1.2 1.6 - 2.4 3.2 - 4.8
5 cm2
1 2 5 10
0.8 - 1.2 1.6 - 2.4 4.0 - 6.0 8.0 - 12.0
10 cm2
1 5 10 15** 20**
0.8 - 1.2 4.0 - 6.0 8.0 - 12.0 12.0 - 18.0 16.0 - 24.0
2 cm2
NOTE: The Applicator LED should constantly illuminate green during the Applicator Output tests. B. Ultrasound Applicator Output Test Results
1. Output ranges fall within the specified ranges as listed in Figure 5.18. Unit passed test. 2. Readings fall outside specified ranges of Figure 5.18. a) Possible bad Degassed Water in Power Meter. b) Possible use of Power Meter other than Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter. c) Possible bad or out of calibration Applicator. d) Use known good Applicator. e) Check Ultrasound Board internal connections. f) Replace Ultrasound Board. g) Replace Control Board
* 3.3 MHz Only **1 MHz Only FIGURE 5.18
29
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.16 ULTRASOUND DUTY CYCLE TEST
This test is performed using only the 5 cm2 Intelect Mobile or Advanced Ultrasound Applicator. A. Ultrasound Duty Cycle Test Procedures
1. Set up Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter per Operator’s Instructions and fill test reservoir with Degassed Water.
Use only Degassed Water in Power Meter for testing Ultrasound Applicators. Use of other types of water will cause false test results. Refer to page 21 for Degassed Water Recipes. Do not aerate water when filling Power Meter. 2. Place the 5 cm2 Applicator into the Power Meter retainer. Make certain the Sound Head is completely submerged in the degassed water and centered directly over the Stainless Steel Cone. See Figure 5.19. 3. Zero or Tare meter. 4. Highlight Ultrasound on system. Press the Enter button. 5. Highlight Duty Cycle. Highlight Continuous and press the Enter button. 6. Highlight Display. Press the Enter button until Watts appears beside Display. 7. Press START button. 8. Increase Intensity until the appropriate Watts is displayed. See Figure 5.20. 9. Compare Power Meter reading to Figure 5.20. 10. Press the STOP button. 11. Highlight Duty Cycle and press the Enter button. Highlight the next level of Duty Cycle and repeat steps 6 through 10. Repeat for remaining Duty Cycle levels. 12. Highlight Frequency. Press the Enter button until 3.3 MHz is displayed beside Frequency. Repeat steps 4 through 11.
FIGURE 5.19
DUTY CYCLE SPECIFICATIONS APPLICATOR SIZE
DUTY CYCLE
OUTPUT RANGE
5 cm2
10% 20% 50% 100% (Continuous)
0.8 - 1.2 1.6 - 2.4 4.0 - 6.0 8.0 - 12.0
FIGURE 5.20
B. Ultrasound Duty Cycle Test Results
1. Duty Cycles fall within the specified ranges as listed in Figure 5.20. Unit passed test. 2. Readings fall outside specified ranges of Figure 5.20. a) Possible bad degassed water in Power Meter. b) Possible use of Power Meter other than Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter. c) Possible bad or out of calibration Applicator. Retest with known good Intelect Mobile or Advanced Applicator. d) Possible bad internal connection at Ultrasound Board. e) Replace possible bad Ultrasound Board. f) Replace possible bad Control Board.
30
5 TROUBLESHOOTING
Intelect® Mobile Stim and Combo Therapy Systems
5.17 COMBO OPERATION TEST
This test is performed using the 5 cm2 Applicator. Highlight Channel 1 and set up system per 5.6 parts A and B prior to performing tests. Connect the Intelect Mobile or Advanced 5 cm2 Applicator to the System. See Figure 5.21. Applicator LED will flash green five times. A. Combo Operation Test Procedures
1. Set Scope; Time- 50 μS, Channel- 20 V, and Trigger- DC. 2. Highlight Combo. Press the Enter button. 3. Highlight Display. Press the Enter button until Watts is displayed beside Display. 4. Highlight Waveform. Press the Enter button. 5. Press the Up or Down Arrow button until IFC-4p is highlighted. Press the Enter button. 6. Highlight Edit Stim. Press the Enter button. Increase Intensity until Channel 2 reads 50 mA. 7. Press START button. 8. Touch the Ultrasound Applicator to the Combo Contact on the ESTI-2 Load Test Fixture. The Combo Indicator on the ESTI-2 should illuminate. See Figure 5.22. 9. Compare waveform on scope to Figure 5.23.
FIGURE 5.21
B. Combo Operation Test Results
1. Waveform on scope the same as Figure 5.23 and the Combo Indicator illuminates. Unit passed test. 2. No waveform or considerably different waveform. Unit failed test. Replace Channel 1 Stim Board.
COMBO INDICATOR ILLUMINATED
FIGURE 5.22
FIGURE 5.23
31
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.1 SEPARATING TOP & BOTTOM
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. FIGURE 6.1 A. Tools and Equipment Required
1. #1 Phillips Screwdriver 2. Flat Blade Screwdriver
REMOVE 4 SCREWS
B. Removing Top from Bottom
1. Place system face down on a soft work surface. 2. Remove Lower Front Feet and Rear Fan Grill. See Figure 6.1. 3. Remove the four mounting screws securing the top and bottom together. See Figure 6.2. 4. Turn system over on its feet and carefully separate the System Top from the Bottom Housing. 5. Raise the system top and disconnect the Fan, Power Supply, and Battery Harnesses from the Control Board. See Figure 6.3. C. Replacing Top to Bottom
Replace System Top by reversing the above steps. NOTE: Do not over tighten the screws. Over tightening will damage the threads of the brass inserts.
FIGURE 6.2 POWER SUPPLY
FAN BATTERY
FIGURE 6.3
32
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.2 THERAPY SYSTEM FAN
REMOVE SCREWS
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. A. Tools and Equipment Required
#1 Phillips Screwdriver B. System Fan Removal
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Using a #1 Phillips Screwdriver, remove the two Fan Retaining Screws securing the Fan to the system Bottom. See Figure 6.4.
FIGURE 6.4
REMOVE BAFFLE
3. Remove the Fan Baffle from the Fan Housing. See Figure 6.5. C. Replacing System Fan
1. Replace new Fan, part number 27158, by reversing the steps above. NOTE: Do not over tighten the screws. Over tightening will damage the threads of the brass standoffs. 2. Reassemble Top and Bottom of system. Refer to 6.1, part C.
FIGURE 6.5
33 33
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.3 POWER SUPPLY
REMOVE SCREWS
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. A. Tools and Equipment Required
1. #1 Phillips Screwdriver 2. Insulated Needle Nose Pliers 3. Digital Multimeter B. Power Supply Removal
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Using the # 1 Phillips Screwdriver, remove the two screws securing the Power Supply to the system Bottom. See Figure 6.6.
FIGURE 6.6
POWER SUPPLIES RETAIN HIGH VOLTAGE! WHEN REMOVING FROM SYSTEM, HANDLE POWER SUPPLIES BY MOUNTING BRACKETS ONLY. 3. Lift Power Supply Assembly up to remove from mounting tabs. See Figure 6.7. 4. Using the Digital Multimeter, discharge Capacitor C4. See Figure 6.8.
MOUNTING TABS
FIGURE 6.7 CAPACITOR C4
FIGURE 6.8
34
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.3 POWER SUPPLY (continued)
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock.
REMOVE HARNESSES
5. Using Insulated Needle Nose Pliers, disconnect the Power Supply Harnesses from the Mains Connector. See Figure 6.9. 6. Remove Power Supply from system. PLIERS NOT SHOWN FOR CLARITY
C. Replacing Power Supply
1. Replace new Power Supply in reverse order of preceding steps. 2. Reassamble system Top to Bottom. Refer to 6.1, part C. FIGURE 6.9
NOTE: Do not over tighten the screws. Over tightening will damage the threads of the brass inserts.
35
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.4 CHANNEL 1 STIM BOARD
REMOVE SCREWS
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. A. Tools and Equipment Required
#1 Phillips Screwdriver B. Channel 1 Stim Board Removal
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Using the # 1 Phillips Screwdriver, remove the four screws securing Channel 1 Stim Board to the Stand Offs. See Figure 6.10. 3. Carefully lift the Channel 1 Stim Board from the unit. Make certain not to bend any of the Header Connector Pins on the board below during removal. See Figure 6.11.
FIGURE 6.10
C. Replacing Channel 1 Stim Board
HEADER CONNECTOR PINS
1. Replace new Channel 1 Stim Board in reverse order of preceding steps. Make certain all Header Connector Pins are properly engaged. See Figure 6.11. 2. Re-assamble system Top to Bottom. Refer to 6.1, part C. NOTE: Do not over tighten the screws. Over tightening will damage the threads of the brass inserts and Stand Offs.
FIGURE 6.11
36
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.5 CHANNEL 2 STIM BOARD
REMOVE STANDOFFS
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. A. Tools and Equipment Required
1. #1 Phillips Screwdriver 2. 1/4 in Wrench B. Channel 2 Stim Board Removal
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Remove Channel 1 Stim Board. Refer to 6.4, part B. 3. Using the 1/4 in Wrench, remove the four Stand Offs securing Channel 2 Stim Board in place. See Figure 6.12. 4. COMBO SYSTEM ONLY- Remove the 40 Pin Header from the back of the Channel 2 Stim Board. See Figure 6.13.
FIGURE 6.12
C. Replacing Channel 2 Stim Board
1. COMBO SYSTEM ONLY- Install 40 Pin Header to back of new Channel 2 Stim Board. Make certain it is completely seated against board. See Figure 6.13. 2. Position the Channel 2 Stim Board over the Ultrasound Board (Combo Systems) or Control Board (2 Channel Stim Systems) aligning the 40 Pin Header with the 40 Pin Connector. See Figure 6.14. 3. Press the Channel 2 Stim Board into position until the board rests against the Ultrasound Board (Combo System) or Control Board (2 Channel Stim System) Stand Offs. 4. Secure the Channel 2 Stim Board using the Stand Offs removed in part B, step 3 above. 5. Install the Channel 1 Stim Board. Refer to 6.4, part C. 6. Re-assemble Top to Bottom. Refer to 6.1, part C.
REMOVE 40 PIN HEADER
FIGURE 6.13
NOTE: Do not over tighten the Stand Offs or screws. Over tightening will damage the threads of the brass inserts and Stand Offs.
ALIGN HEADER TO CONNECTOR
FIGURE 6.14
37
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.6 ULTRASOUND BOARD COMBO SYSTEMS ONLY
REMOVE STANDOFFS
A. Tools and
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. Equipment Required
1. #1 Phillips Screwdriver 2. 1/4 in Wrench B. Ultrasound Board Removal
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Remove Channel 1 Stim Board. Refer to 6.4, part B. 3. Remove Channel 2 Stim Board. Refer to 6.5, part B. 4. Using the 1/4 in Wrench, remove the four Stand Offs securing Ultrasound Board in place. See Figure 6.15. 5. Carefully remove the Ultrasound Board from the 40 Pin Control Board Header. See Figure 6.16. 6. Remove the 40 Pin Header from the Ultrasound Board. See Figure 6.17.
FIGURE 6.15
CONTROL BOARD HEADER
C. Replacing Ultrasound Board
1. Install 40 Pin Header to back of new Ultrasound Board. Make certain it is completely seated against board. See Figure 6.17. 2. Position the Ultrasound Board over the Control Board aligning the 40 Pin Header with the 40 Pin Connector on Control Board. See Figure 6.16. 3. Press the Ultrasound Board into position until the board rests against the Control Board Stand Offs. 4. Secure the Ultrasound Board using the Stand Offs removed in part B, step 4 above. 5. Install the Channel 2 Stim Board. Refer to 6.5, part C. 6. Install the Channel 1 Stim Board. Refer to 6.4, part C. 7. Re-assemble Top to Bottom. Refer to 6.1, part C.
FIGURE 6.16
NOTE: Do not over tighten the Stand Offs or screws. Over tightening will damage the threads of the brass inserts and Stand Offs.
INSTALL HEADER TO BACK OF ULTRASOUND BOARD
FIGURE 6.17
38
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.7 CONTROL BOARD ASSEMBLY
REMOVE CONTRAST KNOB
A. Tools and
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. Equipment Required
1. #1 Phillips Screwdriver 2. 1/4 in Wrench
FIGURE 6.18
B. Control Board Assembly Removal REMOVE SCREWS
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Remove Channel 1 Stim Board. Refer to 6.4, part B. 3. Remove Channel 2 Stim Board. Refer to 6.5, part B. 4. Remove Ultrasound Board (Combo Systems Only). Refer to 6.6, part B. 5. Remove the Contrast Control Knob. See Figure 6.18. 6. Remove the seven screws securing the Control Board Assembly in position. See Figure 6.19. 7. Remove the four LCD Bracket Mounting Screws. See Figure 6.19. 8. While lifting on the lower area of the Control Board Assembly with one hand, release the Alignment Tabs with the other to remove the Control Board from the system Top. See Figure 6.20. C. Replacing Control Board Assembly
1. Position the new Control Board Assembly over the Alignment Tabs of the system Top. Press the Control Board Assembly until the Alignment Tabs lock the Control Board into position. See Figure 6.20. 3. Install the four LCD Bracket Mounting Screws. See Figure 6.19. 4. Install the seven Control Board Mounting Screws. See Figure 6.19. 5. Install the Ultrasound Board (Combo Systems only). Refer to 6.6, part C. 6. Install the Channel 2 Stim Board. Refer to 6.5, part C. 7. Install the Channel 1 Stim Board. Refer to 6.4, part C. 8. Re-assemble Top to Bottom. Refer to 6.1, part C.
REMOVE SCREWS
FIGURE 6.19 RELEASE TABS LIFT HERE
NOTE: Do not over tighten the Stand Offs or screws. Over tightening will damage the threads of the brass inserts and Stand Offs.
FIGURE 6.20
39
6 REMOVAL & REPLACEMENT
Intelect® Mobile Stim and Combo Therapy Systems
6.8 KEYMAT ASSEMBLY AND ON/OFF BUTTON KEYMAT KEYMAT ON/OFF BUTTON
Unplug the unit from the power source before attempting removal or replacement procedures to prevent electrical shock. A. Tools and Equipment Required
1. #1 Phillips Screwdriver 2. 1/4 in Wrench B. Keymat Assembly Removal
1. Separate Top from Bottom. Refer to 6.1, part B. 2. Remove Channel 1 Stim Board. Refer to 6.4, part B. 3. Remove Channel 2 Stim Board. Refer to 6.5, part B. 4. Remove Ultrasound Board (Combo Systems only). Refer to 6.6, part B. 5. Remove Control Board Assembly. Refer to 6.7, part B. 6. Lift out Keymat Assembly and On/Off Button Keymat. See Figure 6.21.
FIGURE 6.21
C. Replacing Keymat Assembly
1. Place the new Keymat Assembly and On/Off Button Keymat into position. See Figure 6.21. 2. Install the Control Board Assembly. Refer to 6.7, part C. 3. Install the Ultrasound Board (Combo Systems only). Refer to 6.6, part C. 4. Install the Channel 2 Stim Board. Refer to 6.5, part C. 5. Install the Channel 1 Stim Board. Refer to 6.4, part C. 6. Re-assemble Top to Bottom. Refer to 6.1, part C. NOTE: Do not over tighten the Stand Offs or screws. Over tightening will damage the threads of the brass inserts and Stand Offs.
40
7 GENERAL MAINTENANCE
Intelect® Mobile Stim and Combo Therapy Systems
7.1 CLEANING THE SYSTEM A. Cleaning the Therapy System
With the system disconnected from the power source, clean the system with a clean, lint free cloth moistened with water and mild antibacterial soap. If a more sterile cleaning is needed, use a cloth moistened with an antimicrobial cleaner. Do not submerse the system in liquids. Should the unit accidentally become submersed, contact the dealer or Chattanooga Group Service Department immediately. Do not attempt to use a system that has been wet inside until inspected and tested by a Service Technician Certified by Chattanooga Group. Do not allow liquids to enter the ventilation holes in the optional modules. This could permanently damage the modules. B. Cleaning Therapy System Lens
Clean the Therapy System Lens with the NOVUS® Plastic Polishing System. NOVUS can be purchased by going to www.novuspolish.com on the internet. Follow the instructions as given by NOVUS on their product. Do Not Use alcohol or chlorine based solvents as this may damage the lens. 7.2 CALIBRATION REQUIREMENTS Ultrasound Applicators:
Annual calibration is required for all Ultrasound Applicators. Only the Applicators should be sent to the factory or a Field Technician certified by Chattanooga Group for this procedure. 7.3 FIELD SERVICE
A. All field service procedures as described in this Service Manual must be performed by a Service Technician certified by Chattanooga Group. B. Any attempted outside the scope of this Service Manual is the sole responsibility and liability of the Field Technician performing such procedures. C. After the performance of any Field Service, perform the tests as described in 5.3 through 5.17 to verify the system operates properly and within specifications prior to placing the unit back into operation. 7.4 FACTORY SERVICE
When the Intelect Combo or Stim Therapy System requires factory service, contact the dealer or Chattanooga Group Service Department.
41
8 ULTRASOUND APPLICATOR CALIBRATION
Intelect® Mobile Stim and Combo Therapy Systems
8.1 GENERAL PROCEDURES A. Tools and Equipment Required
1. All Ultrasound Applicators for the unit being serviced. 2. Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter, set to "Watts". 3. Degassed Water. Refer to page 21 for Degassed Water Recipes.
FIGURE 8.1
Use only Degassed Water in Power Meter for calibrating Ultrasound Applicators. Use of other types of water will cause false readings and bad test results. See page 21 for Degassed Water Recipes. Use of other brands or types of tools, equipment, fixtures, materials, and supplies other than those specifically listed on page 21 will give bad test and calibration results. If proper equipment is not available or cannot be obtained, send the Ultrasound Applicators to the factory for calibration. B. Ultrasound Applicator Calibration Procedures
1. Perform the following on all Ultrasound Applicators for the unit being serviced at least annually. 2. With the system on, press the Clinical Resources button once. See Figure 8.1. 3. Simultaneously press and hold the Treatment Time and Intensity Decrease buttons for approximately 2 seconds. See Figure 8.2. The calibration procedures screen should display. 4. Press the Down Arrow button until US Applicator Calibration is highlighted. 5. Press the Enter button. 6. Place the Ultrasound Applicator being calibrated into the Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter and set meter to "Watts”. See Figure 8.3. 7. Follow the instructions on the LCD Display. 8. When calibration is complete, place another applicator on the Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter and repeat steps 1 through 7 above.
FIGURE 8.2
SET METER TO WATTS
FIGURE 8.3
42
Intelect® Mobile Stim and Combo Therapy Systems
9 PARTS Combo Assembly
COMBO TOP ASSEMBLY See Pages 45 10 1 9 COMBO BASE ASSEMBLY See Page 46 2
2
3 8 4
5
6
7
ITEM
PART NO.
DESCRIPTION
1
27263
Rear Vent
1
2
27140
Screw, M3 x 8 mm
4
3
27365
Right Foot
1
4
27267
Battery Pack (Optional)
1
5
27410
Battery Compartment Cover
1
6
27253
Plynth
1
7
27150
Plynth Feet
2
8
27363
Left Foot
1
9
27969
Battery Harness
1
10
28018
Baffle
1
43
QTY
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Combo Ultrasound and Stim Boards
COMBO TOP ASSEMBLY See Page 45
1
2
3
4
5
6 7
8
ITEM
PART NO.
1
28019
DESCRIPTION
QTY
Header (Ultrasound Board to Control Board)
1
2
27269
Ultrasound Board
1
3
28020
Header (Channel 2 Stim Board to Ultrasound Board)
1
4
28017
Stand Off (Ultrasound Board to Channel 2 Stim Board) M3 x 19 mm
4
5
27498
Channel 2 Stim Board
1
6
27770
Stand Off (Channl 2 Stim Board to Channel 1 Stim Board) M3 x 16 mm
4
7
27419
Channel 1 Stim Board
1
8
27142
Screw, M3 x 6 mm SS- Pozi Pan Head
4
44
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Combo Top Assembly 1
2
3
4
ITEM
PART NO.
DESCRIPTION
1
27553
Top Assembly
1
2
27561
Keymat Assembly
1
3
28125
Control Board
1
4
27142
Screw, M3 x 6 mm
11
45
QTY
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Combo Base Assembly
9
10
1 8 2
3
4
7 5
6
ITEM
PART NO.
DESCRIPTION
1
27265
Power Supply
1
2
27142
Screw, M3 x 6 mm
2
3
27592
Power Supply Shield
1
4
27983
Combo Base
1
5
27152
Universal IEC Socket
1
6
27256
Ultrasound Applicator Hanger
1
7
27984
Combo Infill Panel
1
8
27158
Fan
1
9
27136
Screw, M4 x 35 mm
2
10
27367
Fan Baffle
1
46
QTY
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Stim Assembly STIM TOP ASSEMBLY See Pages 49
1 9
STIM BASE ASSEMBLY See Page 50
2
2
3 8 4 5
6
7
ITEM
PART NO.
1
27263
2 3
DESCRIPTION
QTY
Rear Vent
1
27142
Screw, M3 x 6 mm
4
27365
Right Foot
1
4
27267
Battery Pack (Optional)
1
5
27410
Battery Compartment Cover
1
6
27253
Plynth
1
7
27150
Plynth Feet
2
8
27363
Left Foot
1
9
27575
Baffle
1
47
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Stim Boards
STIM TOP ASSEMBLY See Page 49
1
2
3
4
5
ITEM
PART NO.
1
27580
Header (Stim Board to Control Board)
DESCRIPTION
1
2
27498
Channel 2 Stim Board
1
3
27770
Stand Off (Channel 2 Stim Board to Channel 1 Stim Board) M3 x 16 mm
4
4
27419
Channel 1 Stim Board
1
5
27142
Screw, M3 x 6 mm SS- Pozi Pan Head
4
48
QTY
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Stim Top Assembly 1
2
3
4
ITEM
PART NO.
1
27553
Top Assembly
DESCRIPTION
1
2
27561
Keymat Assembly
1
3
28125
Control Board
1
4
27142
Screw, M3 x 6 mm
11
49
QTY
9 PARTS
Intelect® Mobile Stim and Combo Therapy Systems
Stim Base Assembly 10 9 1 8
2
3 4
7 5
6
ITEM
PART NO.
DESCRIPTION
1
27265
Power Supply
1
2
27142
Screw, M3 x 6 mm
2
3
27592
Power Supply Shield
1
4
27412
Stim Base
1
5
27152
Universal IEC Socket
1
6
27258
Applicator Infill
1
7
27415
Stim Infill Panel
1
8
27158
Fan
1
9
27136
Screw, M4 x 35 mm
2
10
27367
Fan Baffle
1
50
QTY
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
SCL_APPL
SDA_APPL
1
1
1
1
MCLR
R6
R1
10K
1K
C2 10nF
C1 10nF
R2 1K
R4 42.2K 1% U1
1
3
0.1uF
1.0uF/16V
GP1/AN1/CIN-/VREF/ICSPCLK
GP3/MCLR/VPP
GP2/AN2/T0CKI/INT/ COUT PIC12F675-I/SN
8 7 6 5
C5
RT1
0.1uF L2
Next to Pin 1 Thermal connect
100K
2
2
C6
GP0/AN0/CIN+/ICSP DAT
GP4/AN3/T1G/OSC2/CL KOUT
4
C3
VSS
GP5/T1CKI/OSC1/CLKIN
1 2
1
VDD
2
1
1
VCC_APPL
2
2
2
2
F1
2
DGND_APPL NOT Case Connected
2
R5
2
1
R7
1
220
2
10K
1
Q4
1
LED1 1
2
3
220uH
MMBT3904
C4
2
R3 10K
1
2
10nF
D1 BAS16_NL 1
K1
3
1- N/C
2
2,4- US+
8
3- MCLR
1
5- +5V
5
6,8- US-
1
6
7- IC2 DATA
7
1
9- IC2 CLOCK 10- Applicator Body
4
C8
3 2 JP1
10nF
T1
1
2
3
4
5
6
7
8
9
10
US+
1
PAD3
50PAD PAD2
50PAD PAD1
G6K-2F-Y5VDC
1
8
2
7 1
3
6
4
5
PIN1
C9 330pF/200V
1
US-
1
100PAD
100PAD
PIN2
PIN3
1
EFD15
1
10nF
1
1
PAD4
50PAD
1
C7
1
1
50PAD
1
Case Connected
Transducer
51
5 cm2 Ultrasound Applicator
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
VCC
VCC
R74
C6 RP1
VCC
100K
.1uf
/CS
1
16
RS
2
15
R/W
3
14
NMI
4
13
RxD2
/ST BY
5
12
TxD2
MD0
6
11
E
7
10
/RST
8
9
CN2 R47
33 1 2 3 4 CON4
10K
VCC +24V U1
59 4 VCC
37 36 R27 10K
NMI
7
/ST BY
8
MD0 C2
6
U4
.01uf
2
VCC1
P43
VCC1
P44
AVCC
P45
AVref
P46 P47
NMI /STBY
P10
MD0
P11 P12
Vdd RST
3
DSP1
1
1 100
GND DS1233-10(3)
P13 /RES
P14
RES0
P15 P16
Y1 3 C4
P17 EXTAL P20
C5
20M Hz
22pf
P21
22pf
P22 P23 2 MD1
5
TxD1
97
RxD1
98 99
BUS_DI
R38
100
RxD0
13
BUS_DO
R39
100
TxD0
14
BUS_CLK
R40
100
SCK0
12
P24 XTAL
P25
MD1
P26
TxD1
P27
RxD1 SCK1
P30 P31
RxD0
P32
TxD0
P33
SCK0
P34 P35
VCC
P36 P37
P60
R31 10K
/CS_STIM
R41
100
91
/CS_US
R42
100
90
/CS_ AUX
R43
100
81
/BUS_RST
80 69 Q4
68
2N7002
58 57
48 R6
47
100K
31 30 21 20 11 10
96 95 94 93 46
15 70 71 92
9
P61 PB0
P62
PB1
P63
PB2
P64
PB3
P65
PB4
P66
PB5
P67
52
/CS
53
RS
4
54
R/W
5
55
E
56
/RST
V CC
15
/CS
E
9 10
75
11
74
12
73
13
72
DB1
C39
C40
C41
C42
C43
C44
C45
D20
R59
10K
1N6263
22K
C22
R60
.01uf
3.01K
24V_MON
1K VEE R52
100
R53
100
DB5 DB6 DB7
Cathode
K 1
Q6
C46
2N7002
LCD Displ ay 22pf
AC_IN
64
17
DB4
DQ
65
VEE
DB3
VSS
C_INH
66
R35
R13
DB2
14
67
3
LCD 240 * 128
DB0
8
76
VO
/RST
7
77
18K Panasonic: EVUF3AF25B14
R/W
6
78
A
VCC
RS
16
79
Anode
R12
2
22pf
22pf
22pf
22pf
22pf
22pf
22pf
Connector: Sa mtec HLE-122-02- S-DV-BE-A
BACKLIGHT
SD_REQ
63
SHUTDOWN
62
24V_EN
61
Vee_EN
60
BEEPER
VCC
82
BATT_IN
R4 100K +24V VCC
83
J3 BP1
BACKLIGHT
84
FAN
85
/EEPROM_CS
86
R14
R15
R16
R17
R18
R19
R20
R21
1
1K
1K
1K
1K
1K
1K
1K
1K
2
DIP1 SW DI P-4
/BATT_EN
87
FAN BEEPER
88
1
8
89
2
7
3
6
4
5
26
LED1
LED2
LED3
LED4
IND_1
IND_2
IND_3
IND_4
Q7 FAN
Q5
ZVN4306GV
BEEPER
2N7002
27 28
R54
29
100K
R7
32
100K
33 34 35
PB6 PB7
TxD2 RxD2
PA0
SCK2
PA1 PA2 PA3
P70
PA4
P71
PA5
P72
PA6
P73
PA7
P74 P75
P83
P76
P82
P77
49
R66
1K
TxD2
50
R67
1K
RxD2
51
38
24V_MON
VCC
39 40 41 42 43 44
R49
R50
R51
45
10K
10K
10K
P81
VCC
P80 AVSS
P90 P91
VSS
P92
VSS
P93
VSS
P94
VSS
P95 P96
VCC2
P97
25
R56
10K
24
J1
23
VSS
22
RxD1
19 18
SW2
SW3
SW4
SW5
17
TxD1 MD1 VCC
16
1
2
3
4
5
6
7
8
9
10
MCLR
V CC
B6
B7
PUB BUS D14 C16
H8S2144AF
.1uf
SW6
SW7
R75
R76
R77
10K
10K
10K
SW8
1N6263 JP1 VSS
D15
VSS SW10
VSS
SW11
VSS
1N6263
VSS
VCC
D16
VSS SW12
SW13
+24V
SW14
VSS
1N6263
+24V VSS
D17 CC 1N6263
VSS VSS VSS VSS
VCC R57 10K
R22
R23
R24
R25
1K
1K
1K
1K
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
HEADER 20X2 Connector: Sa mtec TLE-120-01 -G-DV-A
BUS_DI BUS_DO BUS_CLK /CS_US /CS_STIM CC VCC
VSS
C47
C48
22pf
22pf
+24V VSS +24V VSS VCC
VSS VSS
/CS_ AUX /BUS_RST C49 22pf
U8 8 C19 .1uF
7 3 4
Vcc /HOLD /WP Vss 25AA080
52
/CS SCLK SI SO
1
/EEPROM_CS
6
SCK0
5
TxD0
2
RxD0
Control Board 1 of 2
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
VCC V CC
+
+
C1
C12
C13
C14
C17
C18
.01uf
.01uf
.01uf
.01uf
100uf
100uf
D1 1N6263
VCC VCC SD_REQ
R1
V CC
100K
+24V
R73 VCC
C34 .1uF
1K
D2
BATT_IN
1N4148 U6 1
IN
FB
2
OUT
C8 330uf / 35V
FMM T551
1K
680
24V_DC
R2
24V_B ATT
Q1
10K
2N7002 R8
L1 330mH
3
GND
.01uf
Q3
5
ON/OFF
C37
R72
/BATT_EN
R46
4
5V_B ATT
LM2575T5.0(5) D7
C9
LED6
330uF
5V OK
D10
1M
SW1 R9 R11 15K
1N5819
ON/OFF
1N4148 +
D11
C20 100uf
10M R26
D12
10K
1N4148
C15 .1uF
1N4148 VSS_B ATT
U2 3 SHUTDOWN
R36
5
DG
2
24V_B ATT
IN
4.7K D19
1
1N6263
VCC
R58 10K
R28
D13
D18
1N4148
5.1V
4
COM IPS521
10K
Q2 2N7002
+24V
VEE VeeD8
C10 C36
C7
4.7uF
.1uF
R3
TP1
L2
R33
10uH
10K .22uF
C21
VCC
D9
4
2 R34
1N5819
1
R48 270K
C27
LED7
100pf
C11
B560C-13
VCC
24V_DC
Vee- OK
10uf / 35V /SHDN
GND
R30 NFB
3
D5
10K VR1
LT1617
R44
Vee Cal
1N6263
VCC
18K D6
8
10K
SW
+
Vee_EN
Vin
D3
1N5819
U7 5
1M
1uf
R5
AC_IN
U5A
50K
100K
3 R10
1 2
15K
R29
B560C-13
1K 24V_DC 4
LM358D R70
R37
10K
C28 .01uf
2.7K
C31 .1uF
C3
LED5 CN1
.01uf
R45 1 7.5K
2
24V OK
75W
R79
15
R78
15
C38
.001uf L3
24V_DC
CC
10uH U9 D4
8
C_INH
7
2
1N6263 C25
5
.022uf
4
L4 10uH
D21
Vcc
B560C-13 DR
6
COMP
C23
C32
C30
C33
C26
.1uF
10uF / 35V
.1uF
10uF / 35V
100uF / 35V
ISEN
FB
CC
1
VSS_B ATT 5V_B ATT
3
LM3478
C24
10uF / 35V
1uf
R62
R63
39K
4.7K
VCC
C_INH
R68
100
DQ
R69
100
1 2 3 4 5 6 CON6
R80 R64
R65
26.1K
.033 OHM
U5B
100 8
C35
J2 24V_B ATT
IRFR4105
PGND Vss
R61 560K
Q8
FA/SD
C50
LM358D
100pf 5
VSS_B ATT
7 6 D22 4
1N6263
R32 .033 OHM
C29
.1uF
53
R55
R71
150K
3.9K
Control Board 2 of 2
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
SOUNDHEAD
CONNE CTOR
FB1 HZ1206
L3
6
GND J1
C12 6
8
0.1uF 2
4.7uF/25V
KMDG-8 S-BS 7
C11
7
2 8
V CC
HEAD_C LOCK
1
1
100uH 1
4
R156
R151
0
0
2
4
3
33uF/50V
C6
2
1
9
1000pF/250V 2
1
8
21
HEAD_ DATA
1
1 21
C13
C14
10nF
10nF
10 2
9 8
1
VIN
ENG00787
1
100
DRAIN
5
10
8
GND
3 4
2
2
1
2
C10
C2 100pF/100V
1
100
TX2-24V
4
NL
1
2
C9 3300pF/250V
5
0.1uF/50V
2
L1 11.5uH/2A
R8
2
7
1
6
2
2
1 1
3
C98
47
L2
12
2.8uH/2A
1 12
IRF
GND
1
2
FB9
K2
TX2-24V
1
1
1
0.1uF/50V
C4 R2
K1
22_NL
C5
R7
2
+40 VADJ
0
S3
R160
R1
R152
US_GND
5
S1
2
4
47
1
R157
1
0
1
100pF/100V
2
3
1
T1
S2
GND 2
2
C1
S3
2
3
S2
5
GND
S1
1
5
L4
U3
1
1
2
4
OUT
3
3
US_GND
C22 2
1
V CC
GND
0_NL
U4
7
GND
2
R23
C23 0.1uF_NL
0.1uF
R17 2
HEAD_D ETE CT
1
300K_NL
C19
2 2
1.0M
2
1
GND
10K
2
1.0M
2
1.5K_NL R20
R13
TPS2819D BV
2.2 R11
1
1
3
IN
R12
R10
R22
1
0.1uF
1
1
1_3MHZ
2
2
2.2
GND
1
1
TPS2819D BV
GND
1
4
OUT
GND 1
IN
1
2
GND
IRLR110
1.5K_NL
0
Q8
2
R19
2
IRLR110
HZ1206
Q7
C20 0.1uF_NL R112
2N7002
1
4
Q4
R9 3
1
VDD
2
GND
4
GND
GND
V CC
VCC
0.1uF
2
2
2
5
C18
2
US Ground
2
4.7uF/25V
1
TESTPOINT
3
C16
2
0.1uF 5
V CC
2
VDD
FB10
1
U2 1
4.7uF/25V
TP1
3
2
1
C15
BAS16
100uH_NL
0.56
D3
C3
1 2
1
1
R5
HZ1206
12
V CC
GND
14.0K 3 12
+
GND
6 2
VRF
-
R18
LMH6642 M A
1.00K
4
C21 1
-5V
1
2
GND GND
0.1uF 1
2 R21 1.00K
DU TY_C YCLE U1A
CLK_REF
C17
13 2 1 74HC27D
2
R27
1
VCC
47pF
GND
620
2
1
1 1
12
R25
R26
2.2K
2.2K
U1B U5
3 4
2
I2C_ DATA
3 7 5
SCL
CLKN
SDA
CLKP
ADDR0
FBK
ADDR1
REF
XIN
6
R28
10
620
14
C28
11
VSS
47pF
2
GND
74HC27D
12
9
VDD
VSS
1
13
LOCK/IPRG
VDD
4
6
5 2
15
EXTLF
XOUT
8
16
1
1
2
2
1
FS6131-01
GND
GND
1
GND
VCC C29
C27
0.1uF
0.1uF
2 TESTPOINT 0.1uF 1
2
2
TP14
1 14
C25
GND
GND FB2
GND 9
1
10
8
11
V CC
U1C HZ1206
C26 4.7uF/25V
7
I2C_C LOCK
74HC27D
2
GND
GND
Ultrasound Board 1 of 3 (AMP) 54
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
FREQ. D13
R115
LEDRD
270
1
12
2
1_3MHZ
V CC
TP13 I_S A MPLE
TESTPOINT GND
2
1
V_S AMPLE
V_SUPP LY
R158
R159
1.00K
1.00K
R117
DU T Y_C YCLE
180
1
1
1
4.096V
C76 0.1uF
D14
2
U16
R A0/AN0
5
1
33K
GND
33K
R127 1
3
1 BUS_CLK
PWM
BUS_DO
BUS_DI
4
1
2
1
R124
R126
33K
10K
U15
2 1
2
1
C85 C83
4
LMV321M R130 20.5K
2
GND
1
GND
GND
STATUS 2
2 R122
LEDGN
270
GND
C84
1
D18
1
2
2
PWR
-
.033uF/5%
.01uF/5%
1
1
1
+
3
LMV7239M5
270 2
GND
20.5K
2 1
-
R121
LEDRD
STATUS 1
R129
2.61K
1
U18 D17
PIC18F252-20/SO
R128
12.1K
+
1
2
5
R123
18
1
2
GND
GND
R125
2
GND
TP15
C AL
17
RC7/ R X/ DT
0.1uF 2
1
18pF
16
RC5/SDO RC6/ TX/CK
MCLR/ Vpp
Hz
0.1uF
2
18pF 2
C79
2
C80
OSC2/CL K OUT
1
I2C_C LOCK
15
RC4/SDI/S DA
10
47
C86 3pole B esse l, -3dB @ 1K Hz, -51dB@10K Hz, -111dB @100K
C82
1
2 1
1
20M Hz
3.3uF/16V
2
1
20MHz Clock Reference
C77
12
14
RC3/SCK/SCL
R119
11
13
RC2/ CCP1
2
RC1/T1OSI/ CCP2 OSC1/CLKIN
V CC
5
RC0/T1OSO/T1CKI 9
CLK_REF
28
RB7/PGD
2
2
GND
I2C_ DATA
1
2
GND
Vss
Precision PWMReference
27
RB6/PGC
19
HEAD_C LOCK
26
RB5
0.1uF
Y1
25
RB4
C81
3.3uF/16V
24
RB3/PGM Vdd
2
1
1
C100
23
RB2 20
GND
22
RB1
HZ1206
GND HEAD_ DATA
21
RB0/INT
FB11
/CS_US
6
2
GND
LM4040DIM3-4.1
4
RA3/AN3/ V ref+ R A4/ TOCKI/C1OUT
V CC
470pF
1
Vss
C97
470pF
3
R A1/AN1 R A2/AN2/ Vref-/C Vref
8
C96
2
RA5/AN4/SS
GND
2
HEAD_D ETE CT
2
1
7
V_REF
3
.068uF/5%
D16 BAS16 3
R120 10K
GND 2
2
1
J3
/BUS_RST
3
PGC
PGD D15
2
1
4
3
6
5
8
7
10
9
RX TX GND
MD1 VCC
Header 2x5
2
PUB
B ZX84C5V1
2
V CC
1
1
VIN
1
Ready GND VCC
2.21K
4.7K
D21 V CC
2
1
GND
3
4
U17A 1
Du ty Cy cle
2
J4
21
CON2_NL
1
DU TY_C YCLE
220K
2
/BUS_RST
R137
1
47 1
11
2
11
74HC04
BAS16
R135
33uF/25V
U17C
2
C87
V CC Du ty Cy cle
7
74HC04
R132
D22
1
330K
5
6
1
22
1
1.00K
GND_US GND
74HC04
2
2
GND
GND
D23
270
LEDRD
24V_US
GND
GND_US
V CC
U17F U17E 13
12
74HC04
37
36
35
34
33
32
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
GND
Digital Ground R162 DR AIN 0
24V_US
V CC
GND_US
GNDD_US GNDD_US GNDD_US GNDD_US GNDD_US
R147
R148
R149
R150
47
47
47
47
GND
Digital Ground
2
VCC
BUS_DI BUS_DO BUS_CLK /CS_US
55
39
38
HEADER 20X2 74HC04
2
74HC04
40
1
10
1
11
2
8
1
U17D 9
1
GND
R142
2
2
GND
J5
Shu td own Ac tive High
2.2K 2
2
2N7002
2
Q17
3
23
R133
R139
1
270
BAS16 LEDY W
33K
2.2K
1
D20
U17B
3
R136
14
2
R138 1
D19
2 R141
2 2
VIN
C89 0.1uF
11
LEDGN
2
R134
1
R131
Ultrasound Board 2 of 3 (Ports)
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
1
TP2 TESTPOINT
V_SAMPLE R32 VOU T- FULL-WAVE RECTIFIER
499
TP3 TESTPOINT
2
1
1
FB3
U7A 3
2
1
2
5
LMV824M
+
1
1
7
C42
-
C34 2
1
1
1
C41
33nF
D4 BAT54S
-
VCC
HZ1206 U9B
-
GND
LMH6644 MA
+
2
301
2
2
R37
499
6
499 1
R36
LMH6644 MA 14
13
4
R29
U7D
2
3
1
+
1
12
R34
R38
R39
2.00K
1.47K
1
2
0.1uF
0.1uF
2
0.1uF
1
C31
2
VCC
1
2
2.00K GND
C32
GND 2
1 2
2
C33
2
2
1
2
11
0.1uF
1
GND
18pF_NL
C35
R33
R30 1.00K
U8 1
IN
6
NO
1
2
1
1
3.3uF/16V
1
R31 2
TP4 TESTPOINT
C49
2.00K
1.00K
1
VRF
-5V SUPP LY
R35
1
GND
2
22nF -5V
50mA max.
R57
499 2
1
U9C
301
1
LMV824M
1
2
10
C51 3.3uF/16V
+
1
8
1
1 R56
LMH6644 MA
+
9
6
10
3
U7C 2
0.1uF
0.1uF
LV
MAX660M
1
499 1
2
C54 C53
GND
GND
MAX4644EUT
5
OUT
U12
TP5 TESTPOINT
R52 IOU T- FULL-WAVE RECTIFIER
-5V
2
V+
2
8
OSC
4
NC
CAP-
GND
CAP+
3
2
0.1uF
FB4 HZ1206
C45
CAL
1
7
5
COM
FC
V+
4
2
2
1
VCC 1.00K
GND
8
9
GND
-
GND
C44 2
33nF
R58
3
R54
R49
R59
1.00K
2
1
1
4.02K 2
1
2
2.00K
499
U7B
1
2
5
2
+
2
7 6
1
LMH6644 MA 1
-
D7
C43
BAT54S
18pF_NL
1
GND C46 4.7nF TP6
1
2
GND
2
40 VOLT ADJUS TABLE SUPP LY
L8
TESTPOINT
R55
1.00K
2.00K
+40 VADJ
C47 1nF/100V
VIN
GND D9
R60
R61
200
200
4
R63
1 1
1
1
2
2
2 2
0.1uF R96
2
2 TP12
CON2
TESTPOINT
FB6
FERRITE
FERRITE
1
1
VIN
FB5
R110
220K
9.09K
C75 C74
4.7uF/35V
0.1uF
2
2
1
20 - 25.5 VDC
2
0.1uF
R108
2
1
2
2
R107
2
24V_US
1
R109
1
LMV824M VCC
2
GND
1
0-0.300V (0-3 A)
0.100 Ohm
4
2
J6 GND
MMBT3906
C37
GND_US
GND
1
GND GND
2.2K
C73 100pF
2
GND
GND
2
1
R89
U9A
1
3.32K
GND
GND
1
1 3
3
4.096V GND
1 2
2
2 1
MM BT3906
511 1 2
5 2
4
1.0M 1
2
Q13
1
2
C68
2.00K
1 3
R75
1.78K 2
C99
LMV321M
0.1uF
R92
1.0M
11 1
R72
V_SUPP LY
U14
Q14
3
1
1
2
1
V_REF
2
1 R88
2
2
1.0M
4
2.2 2
U CC3808AN-2 2
+
NL
1
6.81K
2
NL
R99
NDT3055L
R87
5
2
-
2
6
GND
TP11 TESTPOINT
GND
-
1 1
1
3
1
1
OUTB
RC
GND
C55
33K
+
1
OU TA
CS
2
2N7002
C70
2
2
2
+ R74
1
R161 1
2.2
560pF 2
649
1
18.0K
R93
Q16
NDT3055L
7
2.21K R71
R64 Q9
R95
8
VDD
FB
C61 2
GND
3
4
1
VCC
1
4
COMP
2 3
R77
GND
2 1
LMV824M
2
C69
U13
1
1M_NL
47
D11 30BQ100
0.1uF
Q15
GND
R68
1
GND
150 ENG00789
1
GND
4.99K
1
1
2 2
4.99K 2
-
PWR
12
2
R94
GND
4.99K
14 2
2
1
2
2
1 2
R80
GND
1
2
1nF/100V
GND
2
1
13 R65
R84
10nF
2
U9D
1
C60
TESTPOINT
GND
0.1uF/50V
0.1uF
470pF
TESTPOINT
1.0M
29.4K
8
C67
1
R76
0- 4.1V
1
TP7 2
150
2
1
C58 1nF_NL
1
TP10
2 2
10nF_NL
R67
200
C64
C62
C57 1
TP8 TESTPOINT
7
2 1
1
1
1
2
2
4.99K
C63 4.7uF/25V
100_NL
1 2
1 1
6
2
C66
2N7002
R69
1 2
R91
200
3
47
1
C56 3.3uF/16V_NL
R66
2
1 Q10
2
R90
3
R73
3
GND
TP9 TESTPOINT
1nF/100V
1
1
1 1
1 1
MBRS1100
1
47
2
5
0.1uF/50V
680uF/35V
DU TY_CYCLE
0.1uF/100V
1
C65
FEEDBK R86
1
C50
1
D10
1
VCC
C52
150
2
C48
1
I_SAMPLE
R62
30BQ100
T3
2 MBRS1100
1.00K
1
D8
12
2
2
1
2
R51 2
2
1.00K
1
1
R50 1
IRF
2
1
2
R53
1
47uH
GND
GND Analog G round
Ground Links
56
Ultrasound Board 3 of 3 (V Supply)
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
57
Stim Board- Channel 1 1 of 6
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 1 2 of 6 58
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 1 3 of 6 59
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 1 4 of 6 60
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 1 5 of 6 61
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 1 6 of 6 62
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 2 1 of 5 63
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 2 2 of 5 64
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 2 3 of 5 65
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 2 4 of 5 66
10- Schematics
Intelect® Mobile Stim and Combo Therapy Systems
Stim Board- Channel 2 5 of 5 67
11 WARRANTY
Intelect® Mobile Stim and Combo Therapy Systems
Chattanooga Group ("Company") warrants that the Intelect Mobile Combo and Stim Therapy Systems ("Products") are free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If these Products fail to function during the two year warranty period due to a defect in material or workmanship, Company or the selling dealer will repair or replace the respective Product without charge within a period of thirty days from the date on which the Product is returned to the Company or the dealer. All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by unauthorized centers or groups will void this warranty. The warranty period for certain accessories is 180 days. Accessories consist of Lead Wires, Electrodes, and Nylatex®. The warranty period for the Battery and Ultrasound Applicators is one year (12 Months). This Warranty Does Not Cover: • Replacement parts or labor furnished by anyone other than the Company, the selling dealer, or a certified Company service technician. • Defects or damage caused by labor furnished by someone other than Company, the selling dealer, or a certified Company service technician. • Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and required maintenance or any use that is inconsistent with the Product User's Manual. COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES. Some locations do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. To obtain service from Company or the selling dealer under this warranty: 1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company should be sent to: Chattanooga Group 4717 Adams Road Hixson, TN 37343 USA Phone: +1-423-870-7200 FAX: +1-423-870-2046 and 2. The Product must be returned to the Company or the selling dealer by the owner. A Return Authorization (RA) Number must be obtained before returning any product to the Company. This warranty gives you specific legal rights and you may also have other rights which vary from location to location. The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale of the Product. Any representative or agreement not contained in the warranty shall be void and of no effect. THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
68
4717 Adams Road P.O. Box 489 Hixson, TN 37343 U.S.A. +1-423-870-7200 OUTSIDE U.S.A. +1 423-870-2046 OUTSIDE U.S.A. FAX www.chattgroup.com 0413
28106A
© 2005 Encore Medical, L.P.