O hio Medic al Cor C or porat ion D esk Referenc e
Medical, Laboratory, Industrial Air Systems Vacuum Systems Medgas Pipeline
®
Ohio Medical Corporation® has been the leading supplier of Healthcair® medical air and vacuum pumping systems, Ohio Medical Corporation® and Amvex medical suction regulators, Ohio Medical™ and Amvex air and oxygen flow meters and Aeros® portable suction equipment since 1958. Ohio Medical Corporation is an ISO 9001 and ISO 13485 registered and FDA regulated medical device manufacturing facility. All Ohio Medical medgas products are produced to conform to NFPA 99 standards. With a leading research and development team, dedicated service department, and knowledgeable customer support teams, we will continue to provide the medical industry with leading technology and quality products as we have for the past five decades. Ohio Medical is the only name you need to know for your medical suction and oxygen therapy and medgas equipment needs. Ohio Medical has maintained a presence in the medical industry for well over 45 years under the Ohio Medical, Amvex, and Aeros-Instruments Aeros-Instruments & Squire-Cogswell names. We are a leading supplier of medical pumping systems, systems, with a sound reputation in hospital environments environments around the world. In the 1950 the first Ohio Medical Vacuum regulator was introduced under the name of Ohio Chemical , followed by the Intermittent Vacuum Vacuum regulator in 1951 and the HH Model in 1953. These regulators were manufactured manufactured by Squire-Cogswell (now Ohio Medical) and private labeled for Ohio Medical, Ohmeda and successive companies since their introduction over 45 years ago. In July of 2005, Ohio Medical Corporation was formed, combining the Ohmeda® Suction & Oxygen Therapy products, Aeros portable suction equipment, the Healthcair and Selectorr pumping systems and medgas pipeline products. Finally in November 2007 Ohio Medical acquired Amvex Corporation a leading supplier of medical gas devices. This union created a synergistic blend of products and technology with strong customer and patient safety focus.
Ohio Medical Corporation® has been the leading supplier of Healthcair® medical air and vacuum pumping systems, Ohio Medical Corporation® and Amvex medical suction regulators, Ohio Medical™ and Amvex air and oxygen flow meters and Aeros® portable suction equipment since 1958. Ohio Medical Corporation is an ISO 9001 and ISO 13485 registered and FDA regulated medical device manufacturing facility. All Ohio Medical medgas products are produced to conform to NFPA 99 standards. With a leading research and development team, dedicated service department, and knowledgeable customer support teams, we will continue to provide the medical industry with leading technology and quality products as we have for the past five decades. Ohio Medical is the only name you need to know for your medical suction and oxygen therapy and medgas equipment needs. Ohio Medical has maintained a presence in the medical industry for well over 45 years under the Ohio Medical, Amvex, and Aeros-Instruments Aeros-Instruments & Squire-Cogswell names. We are a leading supplier of medical pumping systems, systems, with a sound reputation in hospital environments environments around the world. In the 1950 the first Ohio Medical Vacuum regulator was introduced under the name of Ohio Chemical , followed by the Intermittent Vacuum Vacuum regulator in 1951 and the HH Model in 1953. These regulators were manufactured manufactured by Squire-Cogswell (now Ohio Medical) and private labeled for Ohio Medical, Ohmeda and successive companies since their introduction over 45 years ago. In July of 2005, Ohio Medical Corporation was formed, combining the Ohmeda® Suction & Oxygen Therapy products, Aeros portable suction equipment, the Healthcair and Selectorr pumping systems and medgas pipeline products. Finally in November 2007 Ohio Medical acquired Amvex Corporation a leading supplier of medical gas devices. This union created a synergistic blend of products and technology with strong customer and patient safety focus.
Ta b l e o f C o n t e n t s
PIPELINE Outlets ............ ........................ ........................ ........................ ........................ ........................ ........................ ........................ ........................ ....................2 ........2 Integrated Flowmeter........... ....................... ........................ ........................ ........................ ........................ ........................ .......................3 ...........3 Retrot Wall Outlets................................. .............................................. ......................... ........................ ........................ ........................ ..............4 ..4 Zone Valve Boxes.................................. ............................................... ......................... ........................ ........................ ........................ ................5 ....5 Valves Va lves with Extensions ............ ........................ ....................... ....................... ........................ ........................ ........................ ....................5 ........5 Alarms: Area, Master and Combo C ombo ........................... ....................................... ........................ ....................... ......................6 ...........6 Refrot Alarm ............ ........................ ........................ ........................ ........................ ........................ ........................ ........................ ......................7 ..........7 Hose Retractor ........... ....................... ........................ ........................ ........................ ........................ ........................ ........................ ....................7 ........7 Hose and Assemblies: Bulk, Spiral Hoses and Hose Assemblies ............ ....................... .............8 ..8 Gas Control Panel ........... ....................... ........................ ......................... ......................... ........................ ........................ ........................ ..............9 ..9 Ceiling Columns/Pedastals........... ....................... ........................ ........................ ........................ ........................ ......................... ..............9 .9 Manifolds ............ ........................ ........................ ........................ ........................ ........................ ........................ ........................ ........................ ..............10 ..10 Emergency Oxygen Inlet Station .......................... ...................................... ........................ ........................ ....................... ............10 .10
SOT PRODUCTS Vacuum Va cuum Regulators ........... ....................... ........................ ........................ ........................ ........................ ........................ .......................11 ...........11 Flowmeters ........... ....................... ........................ ........................ ........................ ........................ ........................ ........................ ........................11 ............11 Adapters ........... ....................... ........................ ........................ ........................ ........................ ........................ ........................ ........................ ................ .... 11 Portable Suction Equipment ................................. ............................................. ........................ ........................ ....................... ............. 11
VACUUM SYSTEMS Rotary Vane Vacuum Systems ................................. ............................................. ........................ ....................... ....................12 .........12 Simplex Rotary Vane Vacuum System .......................... ...................................... ........................ ........................ ...............15 ...15 Rotary Claw Vacuum Systems ........... ....................... ........................ ........................ ........................ ....................... ..................17 .......17 Oasis Vacuum Systems............................. ......................................... ........................ ........................ ........................ .......................20 ...........20 “Water Mister” Liquid Ring Vacuum Systems ...................... .................................. ........................ .....................24 .........24
AIR SYSTEMS Scroll Compressed Air Systems ................................. ............................................. ........................ ....................... ..................27 .......27 Oil-Less Recipriocating Piston Air Compressor Systems ................ ............................ .....................30 .........30 Simplex Air System............................ ........................................ ........................ ........................ ........................ ........................ ...................33 .......33 Instrument Air System ........... ....................... ........................ ........................ ........................ ......................... ......................... ..................35 ......35 Air Treatment Modules (For Oil-less Compressors with Desiccant Dryers) .........36
TECHNICAL INFORMATION Calculated Peak Demand for a Medical Vacuum System ............... ........................... .....................38 .........38 Calculated Peak Demand for an Medical Air Ai r System ............... .......................... ....................... .................39 .....39 Manifold Sizing ............ ........................ ........................ ........................ ........................ ....................... ....................... ........................ ..................40 ......40 Medical Gas G as Equipment Mounting Height Detail ............. ......................... ........................ ....................... .............41 ..41 Suggetsed Medical Gas Master Alarm Points for Level 1 Facilities .....................42 .....................42 Medical Alarm Panel Wiring ................ ............................ ........................ ........................ ........................ ........................ ..................44 ......44
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
1
Outlets Features: Accepts Chemetron/Allied ® , Ohio Diamond, Puritan Bennett and DISS speci c adapters Available, Wall/Celing, Console and Ceiling Column Pin indexed to prevent interchangeability of gas services Cleaned for medical gas service NFPA 99 Compli Compliant ant Inlet pipe can be rotated 360 degrees for ease of installation Gas specic back bodies can accept either Quick Connect or DISS front identi cation bodies Outlet can accommodate various nished wall thicknesses from 3/8” (10 mm) to 1-1/4” (32 mm).
• • • • • • • •
Ohio Diamond Compatible Wall Outlet
Chemetron Compatible Wall Outlet
Ohio Diamond Compatible Console or Celing Column Outlet
DESCRIPTION/ PRODUCT
Chemetron Compatible Console or Ceiling Ceiling Column Outlet
Puritan-Bennett Compatible Wall Outlet
Puritan-Bennett Compatible Console or Ceiling Column Outlet
DISS Wall Outlet DISS Ceiling Outlet
DISS Console or Column Outlet
Diamond Compatible Wall
Chemetron Compatible Wall
Puritan-Bennett Compatible Wall
DISS Wall
DISS Ceiling
Oxygen
261010-1
261020-1
261040-1
261000-1
261000-3
Vacuum
261010-5
261020-5
261040-5
261000-5
261000-7
Air
261010-13
261020-13
N/A
261000-13
261000-15
Nitrous Oxide
261010-9
261020-9
261040-9
261000-9
261000-11
Evacuation
261010-17
261020-17
261040-13
261000-21
261000-23
Nitrogen
N/A
N/A
N/A
261000-17
261000-19
Carbon Dioxide
N/A
N/A
N/A
261000-25
261000-27
261690
261690
261290
261690
N/A
Slide
Slides, blanks and electrical receptacle are also available. Rebuild kits available
Chemetron ® is a registered trademark of Allied Healthcare Products, Inc.
Adapters are available from Ohio Medical Corporation
2
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Integrated Flowmeter The Integrated Flowmeter outlet combines a Dial Flowmeter and a Medical Gas Outlet into a single compact design. The right side port connection has a control knob which allows the user to adjust the ow setting, while the left side port is a direct connection to the gas supply. A Flow Indicator displays whether or not ow has been activated by the Control Knob. Flow Ranges available: 0-8 LPM, 0-15 LPM, 0-25 LPM. A large color coded front plate is used for ease of gas identi cation and aesthetic appeal. The latch-valve assembly is gas speci c. All outlets are cleaned and degreased for medical gas service, factory assembled and tested. All Ohio Medical ® products comply with NFPA-99.
FEATURES:
•
Flow rate meets the requirements of NFPA and CSA
•
Each outlet is 100 % tested and cleaned for medical gas service
•
Indexed to eliminate interchangeability of gas services
•
Flow Indicator
•
Five year warranty
NOTE: Not all objects appear as shown
Part
Confgurator
261XXX-X-XXX Flowrate 08................ 8 LPM 15............... 15 LPM 25............... 25 LPM
Connection Style 000................DISS 010.............Diamond 020........... Chemetron
Flowmeter Style C ..................Click T...............Thorpe** Gas / Outlet Type 1 .......................Oxy / Wall 13 .......................Air / Wall 2 ....................Oxy / Console 14 ....................Air / Console
** Thorpe style not yet available as of the date of this publication. Please contact Ohio Medical Corporation for updates.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
3
R e t r o t Wa l l Outlets Quick Connect Diamond 2&3 to Diamond or DISS Compatible FEATURES
Accepts Diamond or DISS Gas Speci c Adapters.
Pin indexed to prevent interchangeability of gas services.
Each outlet is cleaned for medical gas service and pressure tested.
Retrot kit for diamond 2/3 rough-in part numbers 261220 Rough-In RetroFit Kit Pressure Gas Retrot kit for diamond 2/3 latch valve part numbers 261170-1 Latch Valve Oxy 261170-2 Latch Valve Vac 261170-3 Latch Valve N 2O 261170-4 Latch Valve Air 261170-6 Latch Valve WAGD
4
261221 Vacuum Rough-In RetroFit Kit
261160-1 261160-2 261160-3 261160-4 261160-5 261160-6 261160-7
Latch Valve DISS Oxy Latch Valve DISS Vac Latch Valve DISS N 2O Latch valve DISS Air Latch Valve DISS N 2 Latch Valve DISS WAGD Latch Valve DISS CO 2
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Zone Valve Boxes Single Zone Valve Box For 1/2”, 2”, 2-1/2” and 3” Valves
Multiple Zone Valve Box 4” & 6” Deep Box
Features: Full port design Blow-out proof stem Cleaned for medical gas service Teon seats Dual gauge port on pipe extension
Features: Full port design Blow-out proof stem Cleaned for medical gas service Teon seats Dual gauge port on pipe extension
Part Numbers:
Part Numbers:
261901-05
1/2”
Single Zone Valve Box
261901-07
3/4”
Single Zone Valve Box
261901-10
1”
Single Zone Valve Box
261901-12
1-1/4”
Single Zone Valve Box
261901-15
1-1/2”
Single Zone Valve Box
261902-XXXX
Double Zone Valve Box
261903-XXXXXX
Triplex Zone Valve Box
261904-XXXXXXXX
Quadruplex Zone Valve Box
261905-XXXXXXXXXX
Quintuplex Zone Valve Box
Replace XX with: The following require a 6-1/2” deep box 261901-20
2”
Single Zone Valve Box
261901-25
2-1/2”
Single Zone Valve Box
261901-30
3”
Single Zone Valve Box
05 = 1/2” 07 = 3/4” 10 = 1” 12 = 1 1/4” 15 = 1 1/2” 20 = 2” (Requires a 6” deep box)
Va l ve s w i t h Ex t e n s io n s Features: • • • • •
Full port design Blow-out proof stem Sizes range from 1/2” (13mm) to 4” (76.2mm) Cleaned for medical gas service Teon seats
Valve with Extensions • 1/2” through 4” - Dual gauge ports on pipe extensions • 4” - Single gauge port on pipe extension
Non-Locking Valves Part Number
Locking Valves Part Number
261630-05 261630-07 261630-10 261630-12 261630-15 261630-20 261630-25 261630-30 261630-40
261600-05 261600-07 261600-10 261600-12 261600-15 261600-20 261600-25 261600-30 261600-40
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
5
Alarms A r e a , Ma s t e r a n d Co m b o Features:
Complies with NFPA 99 & FCC Part 15
High visibility LED readouts
Adjustable audible alarm repeat
Optional BMS modules available
Field adjustable pressure settings
UL Listed
Equipped with local or remote sensors (A rea)
Normally Open or Closed Contacts (Master)
Combo Alarm (Local or Remote)
Master Alarm
261890-XX 10 Point 20 Point 30 Point 40 Point 50 Point
10 20 30 40 50
If Dry Contacts are required, add “C” to the part number
26189X- XXX Local Remote
Replace X with gas 1 = Oxygen 2 = Vacuum 3 = Nitrous Oxide 4 = Air 5 = Nitrogen 6 = Evacuation 7 = Carbon Dioxide
6 5 Master Alarm Modules 10 = 1, 10 point 20 = 2, 10 point 30 = 3, 10 point 40 = 4, 10 point
Example: 261890-20C
Area Alarm Local
261861-X 261862-XX 261863-XXX 261864-XXXX 261865-XXXXX 261866-XXXXXX 261867-XXXXXXX 261868-XXXXXXXX
1 Gas Area Alarm 2 Gas Area Alarm 3 Gas Area Alarm 4 Gas Area Alarm 5 Gas Area Alarm 6 Gas Area Alarm 7 Gas Area Alarm 8 Gas Area Alarm
Remote
X = gas
6
1 = Oxygen 2 = Vacuum 3 = Nitrous Oxide 4 = Air 5 = Nitrogen 6 = Evacuation 7 = Carbon Dioxide
261851-X 261852-XX 261853-XXX 261854-XXXX 261855-XXXXX 261856-XXXXXX 261857-XXXXXXX 261858-XXXXXXXX
1 Gas Area Alarm 2 Gas Area Alarm 3 Gas Area Alarm 4 Gas Area Alarm 5 Gas Area Alarm 6 Gas Area Alarm 7 Gas Area Alarm 8 Gas Area Alarm
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
R e t r o t A l a r m •
•
•
•
FEATURES AND BENEFITS • • • • •
High visibility LED readouts Field adjustable pressure settings for each module Local or Remote sensors Adjustable audible repeat cycle Universal trim plate sized for any brand alarm
The Ohio Medical Retrot Alarm is designed to replace and/or update most manufacturer’s master alarms, area alarms and combination alarms without disturbing the wall.
•
•
•
•
The retrot alarm kit includes a front panel display, a frame to cover the existing box, a power supply assembly, and area sensors if required. The retrot alarm is available in two sizes, four and six bay. The four bay box can accommodate up to 6 gases, or a combination of 4 gases and a 10 point master alarm with overall frame size of 19”L x 11.5”H. The six bay can accommodate up to 8 gases and a 10 point master with an overall frame size of 26”L x 11.5”H. The retrot box can be tted in any combination with each vertical slot capable of housing one master module or two area gas modules.
H o s e Re t r a c t o r The Ohio Medical Healthcair ® Series hose assemblies are manufactured with FDA approved medical grade anti-static PVC tubing. The tubing is reinforced with polyester braiding. The maximum working pressure is 200 psi @ 70oF. The ttings have either the coupling or the grips color coded per CGA C-9 standards. The ttings are permanently crimped on each end to the tubing with ferrules. Part Numbers DISS Female to Male DISS Male - 5’ Hose Length 263400-1-05 263400A-2-05 263400-3-05 263400-4-05 263400-5-05 263400A-6-05 263400-7-05
Oxygen Hose Assembly Vacuum Hose Assembly Nitrous Oxide Hose Assembly Air Hose Assembly Nitrogen Hose Assembly WAGD Hose Assembly Carbon Dioxide Hose Assembly
DISS Female to Ohmeda Compatible Quick Connect - 5’ Hose Length 263401-1-05 263401A-2-05 263401-3-05 263401-4-05 263401A-6-05
DISS Female to Chemetron Compatible Quick Connect - 5’ Hose Length 263402-1-05 263402A-2-05 263402-3-05 263402-4-05 263402A-6-05
Oxygen Hose Assembly Vacuum Hose Assembly Nitrous Oxide Hose Assembly Air Hose Assembly WAGD Hose Assembly
Retractor 261746Retractor
Oxygen Hose Assembly Vacuum Hose Assembly Nitrous Oxide Hose Assembly Air Hose Assembly WAGD Hose Assely
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
7
Hose and Assemblies B u l k , Sp i r a l H o s e s a n d H o s e A s s e m b l i e s Product Features:
Bulk Hose:
Spiral Hose:
Spiral hose is available in many combination and colors
Single, double, triple, and quadruple
3 ft coiled hose expands to 10 ft working length
Standard working lengths: 10 ft, 15 ft and 20 ft.
Available in a variety of gases
UL Listed, oxygen cleaned and tested
Standard 300 ft. rolls for 1/4” hose
Standard 250 ft. rolls for 5/16” hose
Amvex hose is made of high quality PVC material
Conductive and Non-Conductive available
Available in USA and ISO colors
Hose Assemblies:
CE marked & CSA approved
Available in custom lengths and con gurations to suit your requirements
Available in a variety of gases
Available in USA and ISO colors
Part Congurator Color Code: USA .........U ISO ............I
Length in feet
HS-XXYZ-XXYYXX Gas Type: Oxygen ...................O Med Air ...................A Vacuum...................V N2O ........................2 Nitrogen..................N WAGD ................... W CO2 ........................C Instrument Air ..........I
Inlet and Outlet Connection:
For correct hose matrix follow these rules when selecting inlet and outlet connections: 1. Female ttings rst 2. Numbers before letters 3. In alphabetical order
8
Hose Type: Conductive 1/4” ..................C4 Conductive 5/16” ................ C5 Non-Conductive 1/4”...........N5 Non-Conductive 5/16”.........N5 General Purpose 1/4” .........G4
Ohmeda Female ........................OF Ohmeda Male ........................... OM Chemetron Male ....................... CM Chemetron Female ....................CF 1/8” FNPT ...................................F2 1/4” FNPT ...................................F4 1/8” MNPT .................................M2 1/4” MNPT .................................M4 DISS Male ................................ DM DISS Male w/demand Check.....DC DISS Handtight ..........................DH DISS Hex Nut ............................ DF Schrader Female ....................... SF Schrader Male ...........................SM Puritan Female .......................... PF
Puritan Male ..............................PM Oxequip Male ............................XM Medstar Male ............................MM Medstar Female.........................MF British Male................................BM British Female............................ BF French male...............................FM French Female ...........................FF NIST Male................................. NM NIST Female ............................. NF German Male ............................ GM German Female.........................GF Blank End .................................. XX NOTE: Not all con gurations are possible.
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Ga s C o n t r o l Panel
Nitrogen, Air, CO2
Pressure Range: 0-300 PSIG
Pull to Lock
Push to Adjust
One DISS Outlet on Unit
Part Numbers: 261760 261911 263280 261911-SH
Nitrogen Control Panel Air Control Panel CO2 Control Panel Air Control Panel with Schrader Fitting
Ce i l i n g C o l u m n s Pedastals
Available Stationary or Retractable (Electrical Powered or Manual)
Complies with NFPA 99
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
9
Manifolds
Digital LCD Pressure Display
NEMA 4x Enclosure
Automatic Leak Detection
Automatic Transducer Diagnostic Testing
2634XX-XX-X*
Part Number Matrix 120V Oxygen ...................32 N2O .........................34 N2 .........................36 Air .........................38 CO2 .........................40 All Other Voltages Oxygen ...................33 N2O .........................35 N2 .........................37 Air .........................39 CO2 .........................41
Heater Kit.................H *Add to the end of the part number if a heater kit is required
Control Unit Only ....00 1 x 1 .....................01 2 x 2 .....................02 3 x 3 .....................03 4 x 4 .....................04 5 x 5 .....................05 6 x 6 .....................06 7 x 7 .....................07 8 x 8 .....................08 9 x 9 .....................09 10 x 10 ...................10
Em e r g e n c y Ox y g e n I n le t S t a t i o n Part Numbers: 261825 261701 261702 261703 261704 261705 232600 232602
10
Emergency Oxygen Inlet Station Check Valve, 3/4” Check Valve, 1” Check Valve, 1-1/4” Check Valve, 1-1/2” Check Valve, 2” Relief Valve, 1/2” NPT, Set at 75 PSIG Relief Valve Pipeaway Adapter, 1/2” NPT
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Va c u u m Re g u l a t o r s : D i g i t a l , A n a l o g , Pe d i a t r i c a n d N e o n a t a l M o d e l s •
•
• •
Patented Push-To-SetTM Improves patient safety Prevents unregulated suction Simple, one handed operation Quick-to-Max 2 turns to full wall vacuum for emergency situations Intermittent Mode: No-Delay “ON” cycle Large, Easy to Read Gauge, Color coded ranges • • • • •
Analog: Improved glow-in-thedark gauge Digital: Accuracy is +/- 1% of full scale Analog: Accuracy of +/- 5% of full scale Modular Components Digital units (cases) available in many colors.
Digital: • Easy to read display • No moving parts means no calibration required • Accuracy is +/- 1% of full scale Analog: • Analog gauge designed for clarity • Accuracy of +/- 3% of full scale Features: • Can easily upgrade from Continuous to Cont/ Intermittent, or Analog to Digital • Durable • Easily service and clean the unit with a cold sterilant • Color cases available for inventory management
Adapters Flowmeters • • • • • •
Amvex Corporation manufactures a wide variety of quick connect medical gas adapters. Amvex offers Ohmeda, Chemetron, Puritan Bennett, Medstar, Oxequip and Schrader styles to suit your needs. Available options include 1/8”, 1/4” MNPT and FNPT, 1/4” and 5/16” Hose Barb and much more.
Impact resistant ow tube and shroud D-Shape knob prevents striping Pressure compensated, accurate ow readings White background improves visibility Color coded to help prevent cross connections 5 Year Warranty (1 year on the 0-1 LPM & 0-200 CCM Flowmeter)
Portables Moblvac ® III is a multipurpose portable suction unit. It produces both constant & intermittent suction with vacuum & ow levels equaling that of wall suction.
IDEAL FOR: Hospitals and extended care facilities requiring wall vacuum performance for Nasogastric, Thoracic, Oral, and Tracheal suctioning procedures.
Care-e-vac ® 3 is a high vacuum & high ow AC/DC powered suction unit. It provides constant suction when wall outlets are unavailable. The unit is compact and very lightweight. An optional oor stand & crash cart mount are available.
IDEAL FOR: Code carts, patient transportation or wherever a powerful suction unit with battery power is required.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
11
Ro t a r y Va n e Va c u u m Systems
•
Continuous on demand
•
Lubricated oil recirculated or dry running
•
NFPA 99 compliant
•
U.L. Listed Electrical Control Panel
•
Available in Simplex, Duplex, Triplex or Quad
•
Congurations:
Discharge
•
Tank Mounted: Up to 10 Hp
•
Stack Mounted: Up to 25 Hp
Inlet Vanes Casing
•
Warranty: 30/24 Months
•
Vibration Free
•
Oil sealed systems feature synthetic oil, reducing maintenance and oil changes.
Rotor
Lubricated
Dry
Model No.
HP
SCFM @ 19” HG (each pump)
Model No.
HP
SCFM @ 19” Hg (each pump)
S200
2
7.7
D100
1
3.25
S300L
3
11.3
D200
2
8,27
S300
3
16.1
D300
3
12.49
S500L
5
25.6
D500L
5
15.37
S500
5
43.2
D500
5
22.1
S750
7.5
60
D750
7.5
30.42
S1000
10
71.1
D1000
10
55.4
S1500
15
103
S2000
20
112
S2500
25
172
Maximum Vacuum
Maximum Vacuum 1, 2, & 10 Hp................25” Hg 3 & 5 Hp.......................27” Hg 7.5 Hp ..........................22” Hg
29.1” Hg Design capacity values should correspond with the “Calculated Peak Demand”. This will allow for the reserve pump as is required per NFPA for level 1 installations.
12
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
To t a l Re c i r c u l a t e d Ro t a r y V an e Co n t i n u o u s On De m a n d M e d i c a l Va c u u m Sy s t e m Oil Sealed Rotary Vane Medical Vacuum System The Ohio Medical Corporation Lubricated Rotary Vane Medical Vacuum System shall be fully NFPA 99 compliant for use in medical vacuum and dual Medical/Surgical applications. The unit will consist of electric motor driven pumps, vacuum receiver, electrical control system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module The total recirculating oil sealed rotary vane vacuum pump shall be single stage air cooled, and capable of producing a maximum vacuum level of 29.1” Hg. The pump assembly shall include an integral anti-suck back valve, exhaust oil separator delivering 99.9% oil-free air, oil level sight glass, and an exhaust pressure gauge. Each pump is protected by a temperature switch, check valve, pump isolation valve, source isolation valve and exible connector.
Vacuum Receiver The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain valve, vacuum gauge and the National Board label.
System Controls The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting pads and touch-up paint.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
13
To t a l Re c i r c u l a t e d Ro t a r y V a n e Co n t i n u o u s On De m a n d Va c u u m Sy s t e m Se l e c t i o n Ch a r t y g t i H c f a " o p 9 p a 1 y i m C t u n @ c p g M a p h i F a s c c e C a D S / e
14
y g t i H c f a " o p 5 a 2 t y p i m C @ u n c p g M a p h i F a s c c e C a D A / e
n o i t a r u m g e t s n y o S C
P H
e ) z s i S n o k l n l a a g T (
) A ( h t g n e L
) B ( h t d i W
) C ( t h g i e H
t h g i e W
l e d o M r e m e b t s m y u S N
7.7 / 7.7
19.7 / 19.7
Vertical Tank Mounted Duplex
2
80
60
40
75
1000
S200B-T2V
11.3 / 11.3
30.8 / 30.8
Vertical Tank Mounted Duplex
3
80
60
40
75
1100
S300LB-T2V
16.1 / 16.1
44 / 44
Vertical Tank Mounted Duplex
3
80
60
40
80
110
S300B-T2V
25.6 / 25.6
70 / 70
Vertical Tank Mounted Duplex
5
80
60
43
80
1500
S500LB-T2V
43.2 / 43.2
114.7 / 114.7
Horizontal Tank Mounted Duplex
5
120
80
36
66
1300
S500B-T2
60 / 60
163.8 / 163.8
Horizontal Tank Mounted Duplex
7.5
120
90
42
77
1700
S750B-T2
71.1 / 71.1
194 / 194
Horizontal Tank Mounted Duplex
10
120
90
46
77
2000
S1000B-T2
101 / 101
271 / 271
Stack Mounted Duplex
15
240
83
80
90
3900
S1500B-ST2
112 / 112
306 / 306
Stack Mounted Duplex
20
240
80
85
90
4400
S2000B-ST2
120 / 60
327.6 / 163.8
Stack Mounted Triplex
7.5
240
107
66
69
3300
S750B-ST3
142.2 / 71.1
388 / 194
Stack Mounted Triplex
10
240
107
66
69
3800
S1000B-ST3
172 / 172
460 / 460
Stack Mounted Duplex
25
240
80
94
90
4900
S2500B-ST2
180 / 60
491.4 / 163.8
Stack Mounted Quarduplex
7.5
240
107
66
80
4000
S750B-ST4
202 / 101
542 / 271
Stack Mounted Triplex
15
240
137
72
90
5800
S1500B-ST3
213.3 / 71.1
582 / 194
Stack Mounted Quarduplex
10
240
107
66
80
4800
S1000B-ST4
224 / 112
612 / 306
Stack Mounted Triplex
20
240
131
74
90
5900
S2000B-ST3
303 / 101
813 / 271
Stack Mounted Quarduplex
15
240
137
72
90
7300
S1500B-ST4
336 / 112
918 / 306
Stack Mounted Quarduplex
20
240
131
74
90
7500
S2000B-ST4
344 / 172
920 / 460
Stack Mounted Triplex
25
240
131
82
90
7000
S2500B-ST3
516 / 172
1380 / 460
Stack Mounted Quarduplex
25
240
131
82
90
8800
S2500B-ST4
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Si m p l ex Ro t a r y V an e Va c u u m Sy s t e m S p e c i c a t i o n /Se l e c t i o n C h ar t Oil Sealed Rotary Vane Simplex Vacuum System
The Ohio Medical Lubricated Rotary Vane Vacuum System shall consist of an electric motor driven pump, vacuum receiver, electrical control system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) m onths from date of start-up. Vacuum Pump Module The total recirculating oil sealed rotary vane vacuum pump shall be single stage air cooled, and capable of producing a maximum vacuum level of 29.1” Hg. The pump assembly shall include an integral anti-suck back valve, exhaust oil separator delivering 99.9% oil-free air, oil level sight glass, and an exhaust pressure gauge. Each pump is protected by a temperature switch, check valve, pump isolation valve, source isolation valve and exible connector. Vacuum Receiver The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain valve, vacuum gauge and the National Board label. System Controls The Underwriters Laboratories Listed simplex electrical motor control system shall be of a fuseless design in a NEMA 12 (dustproof) enclosure. The “Continuous On Demand” feature shall stop the operation of the motor during periods of low or no demand. The controls include a self protected combination motor control with short circuit, single phase and thermal overload protection, 120 volt control circuit transformer with fuseless primary and secondary protection, pressure sensor and a temperature switch with reset button. The cabinet door shall have an HMI (Human Machine Interface) system status display to include accumulated time and fault conditions; lighted Hand-Off-Automatic selector switch and safety disconnect operating handle. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for high air discharge temperature and thermal overload. Accessories Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting pads and touch-up paint.
g H " ) 9 1 R R @ O M T F 7 C 7 A 2 (
) R m R u m O T m u / i g x u c a a H " M V (
n o i t a r m u g e t s n y o S C
P H
e ) z s i S n o k l n l a a g T (
) A ( h t g n e L
) B ( h t d i W
) C ( t h g i e H
t h g i e W
l e d o m r e m e b t s m y u S N
21
29.1/20
Vertical Tank Mounted
2
60
30
43
74
290
S200B-T1V
30.8
29.1/20
Vertical Tank Mounted
3
60
30
45
74
310
S300LB-T1V
44
29.1/20
Vertical Tank Mounted
3
60
30
46
75
350
S300B-T1V
70
29.1/20
Vertical Tank Mounted
5
60
30
46
75
450
S500LB-T1V
118
29.1/20
Vertical Tank Mounted
5
90
31
50
80
550
S500B-T1V
163.8
29.1/20
Vertical Tank Mounted
7.5
90
31
50
81
600
S750B-T1V
194
29.1/20
Vertical Tank Mounted
10
90
31
54
81
1050
S1000B-T1V
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
15
Dr y Ru n n i n g Ro t a r y V a n e Va c u u m Sy s t e m S p e c i c a t i o n s /Se l e c t i o n C h ar t Oil-less Rotary Vane Vacuum System Ohio Medical Corporation Oil-less Rotary Vane, NFPA 99 compliant vacuum system shall consist of electric motor driven pumps, vacuum receiver, and an electrical control system. The components shall be modularly assembled to accommodate most existing doorways. The system shall include interconnecting piping and wiring to provide a functional operating package with applicable electrical and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) m onths from date of start-up. Vacuum Pump Module The dry running (oil-less) rotary vane vacuum pumps shall be single stage, air cooled, with self lubricating graphite vanes. The vacuum pump assembly shall include an integral inlet lter, vacuum regulating valve, check valve, pump isolation valve, a source isolation valve and exible connector. Vacuum Receiver The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain valve, vacuum gauge and the National Board label. System Controls The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals. Accessories Included for job site installation are inlet and discharge touch-up paint.
p " t y m 9 i 1 c u y a p t @ p n i c M a h c g i a F c a s p / e e a C g f D C S H o
16
" y p m 5 t 2 i c u y @ a p t p h n i a c g c a M c a i s p F / e e a C g f D C A H o
exible
n o i t a r u m g e t s n y o S C
connectors, source isolation valve, vibration mounting pads, and
P H
e ) z s i S n o k l n l a a g T (
) A ( h t g n e L
) B ( h t d i W
) C ( t h g i e H
t h g i e W
l e d o m r e m e b t s m y u S N
3.25 / 3.25
5.8 / 5.8
Vertical Tank Mounted Duplex
1
80
63
30
78
740
D100-T2V
8.27 / 8.27
13.8 / 13.8
Vertical Tank Mounted Duplex
2
80
63
30
80
810
D200B-T2V
12.49 / 12.49
18 / 18
Vertical Tank Mounted Duplex
3
80
63
32
83
940
D300B-T2V
15.37 / 15.37
20 / 20
Vertical Tank Mounted Duplex
5
80
63
32
83
980
D500LB-T2V
22.1 / 22.1
30 / 30
Vertical Tank Mounted Duplex
5
80
63
34
83
1000
D500B-T2V
30.42 / 30.42
42 / 42
Horizontal Tank Mounted Duplex
7.5
120
90
42
77
1670
D750B-T2
55.4 / 55.4
114 / 114
Horizontal Tank Mounted Duplex
10
120
102
60
77
2820
D1000B-T2
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Ro t a r y Cl a w Va c u u m Systems
•
Dry, non-contacting design
•
Low cost of ownership
•
No friction (or wearing parts) in pumping chamber
•
Clean air
•
NFPA 99 compliant
•
Low maintenance
•
Warranty: 30/24
•
Long life
•
Reliability
•
High efciency (cfm/hp) Duplex Tank Mounted
Model Number
H.P. Each
Maximum Vacuum Level ("Hg)
SCFM @ 19" Hg Each
Design Capacity SCFM @ 19" Hg
C300-T2
3
25.5
23.8
23.8
C500-T2
5
24
38.7
38.7
C750L-T2
7.5
24
52.6
52.6
C750-T2
7.5
22.5
68.2
68.2
Model Number
H.P. Each
Maximum Vacuum Level ("Hg)
SCFM @ 19" Hg Each
Design Capacity SCFM @ 19" Hg
C500-ST3
5
24
38.7
77.4
C750L-T2
7.5
24
52.6
105.2
C750-T2
7.5
22.5
68.2
136.4
C1500-ST3
15
22.5
120.1
222.6
Triplex Stack Mounted
Quadruplex Stack Mounted Model Number
H.P. Each
Maximum Vacuum Level ("Hg)
SCFM @ 19" Hg Each
Design Capacity SCFM @ 19" Hg
C500-ST4
5
24
38.7
116.1
C750L-T2
7.5
24
52.6
157.8
C750-T2
7.5
22.5
68.2
204.6
C1500-ST4
15
22.5
120.1
333.9
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
17
Dr y Ru n n i n g Co n t a c t l e s s Ro t a r y C la w M e d i c a l Va c u u m Sy s t e m Dry Running Contactless Rotary Claw Medical Vacuum System
The Ohio Medical Corporation ® Dry Running Rotary Claw Medical Vacuum System shall be fully NFPA 99 compliant for use in medical vacuum and WAGD applications. The unit will consist of electric motor driven pumps, vacuum receiver, electrical control system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module The dry running contactless rotary claw vacuum pump shall be single stage, and air cooled. The pump design utilizes two ductile iron claw type rotors rotating in opposite direction inside a cast iron housing. The timing of the rotors shall be maintained by a precision gear set which is separated from the oil free pumping chamber by a combination of lip seals. The vacuum pump assembly shall also include a 5 micron particulate inlet vacuum lter, vacuum regulator/relief valve, and exhaust silencer. The assembly shall also include check valves, pump isolation valves, a source isolation valve and exible connectors.
Vacuum Receiver The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain valve, vacuum gauge and the National Board label.
System Controls The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories Included for job site installation are inlet and discharge touch-up paint.
18
exible
connectors, source isolation valve, vibration mounting pads, and
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Dr y Ru n n i n g Co n t a c t l e s s Ro t a r y Cl a w M e d i c a l Va c u u m S y s t e m S e l e c t i o n Ch a r t y t i c a p a c / g y t i H c " p a 9 p 1 m a u C @ p n h g M c i s F a e e C f D S o
n o i t a r m u g e t s n y o S C
P H
e ) z s i S n o k l n l a a g T (
) A ( h t g n e L
) B ( h t d i W
) C ( t h g i e H
t h g i e W
l e d o m r e m e b t s m y u S N
23 / 23
Tank Mounted Duplex
3
120
105
30
68
710
C300B-T2
38.7 / 38.7
Tank Mounted Duplex
5
120
106
43
76
1930
C500B-T2
52.6 / 52.6
Tank Mounted Duplex
7.5
120
106
43
76
2060
C750B-T2
77.4 / 38.7
Stack Mounted Triplex
5
240
114
62
96
3920
C500B-ST3
68.2 / 68.2
Tank Mounted Duplex
10
120
120
48
80
2600
C1000B-T2
116.1 / 38.7
Stack Mounted Quadruplex
5
204
114
62
96
4990
C500B-ST4
105.2 / 52.6
Stack Mounted Triplex
7.5
240
114
62
96
4110
C750B-ST3
136.4 / 52.6
Stack Mounted Triplex
10
240
114
62
96
4600
C1000B-ST3
157.8 / 52.6
Stack Mounted Quadruplex
7.5
240
114
62
96
5250
C750B-ST4
222.6 / 111.3
Stack Mounted Triplex
15
240
114
62
96
4640
C1500B-ST3
204.6 / 68.2
Stack Mounted Quadruplex
10
240
114
62
96
5800
C1000B-ST4
333.9 / 111.3
Stack Mounted Quadruplex
15
240
114
62
96
5930
C1500B-ST4
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
19
Oa s i s V a c u u m Pu m p s Pr i n c i p l e o f Op e r a t i o n Air and Sealant Discharge
Air Inlet
Impeller Casing
Sealant Ring (oil or water) Trapped Air
Figure 2
Figure 1
The Oasis is a total recirculation liquid ring vacuum pump module. All components are contained within a 304 Stainless Steel enclosure. The above gure shows a cross section of a liquid ring vacuum pump. The impeller is cylindrical in shape with several ns and is located eccentrically in the impeller housing, which is partially lled with water. A port plate with suction and discharge openings is located on each side of the impeller cell. As the impeller rotates, the centrifugal force pushes the sealing liquid outward where it forms a liquid ring which follows the contour of the impeller housing. At the start, the impeller cell is completely lled with the sealing liquid. As the impeller advances, the liquid in the cell recedes and creates a vacuum in the empty space, which draws in the air or gas through the suction port. Continued rotation increases the gas volume in the impeller cell until the end of the suction cycle. From that , the liquid is gradually forced back into the impeller cell, compressing the gas to near atmosphere, until the compressed gas is expelled through the discharge port together with a portion of the sealing liquid. The complete cycle can be compared with a piston in a cylinder where the piston is the liquid ring and the impeller cell is the cylinder. The heat generated by the pump during the compression cycle is dissipated into the sealing liquid. The discharge is piped to a separation and recirculation tank. The air is discharged and the seal water is piped to an on board, high ef ciency, fan driven heat exchanger where the seal water is cooled and recirculated via a return line back to use in the pump. Any water lost due to evaporation is automatically added.
20
Oasis Model
H.P.
SCFM @ 19”Hg
Capacity (ACFM)
Make-Up Water Per Hour (gal)
OL150
1.5
6.6
18
0.04
OL300
3
12.1
33
0.08
OL400
4
17.6
48
0.15
OL500
5
26.5
72
0.20
OL750
7.5
31.6
86
0.22
OL1000
10
53
144
0.38
OL1500
15
69
187
0.53
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Estimated water usage in a typical medical application
Oa s i s V a c u u m Systems
•
NFPA Complaint
•
U.L. Listed Electrical Control Panel
•
Water Sealed/Air cooled
•
Available in simplex, duplex, triplex or quad
•
Modular Stack Mounted
•
Medical or Laboratory Applications
•
Minimal Maintenance, Maximum Durability
•
No Contacting Parts
•
Quiet Operation
•
Low Operating Temperature
•
Warranty 30/24 Months
•
Up to 48 hours of operation without make up water.* * System can run up to 48 hours without need for make up water in emergency situations based on typical medical vacuum application.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
21
Oa s i s V a c u u m System
Oasis Liquid Ring Water Sealed/Air Cooled Total Recirculating Vacuum System
Ohio Medical Corporation Oasis Liquid Ring, NFPA 99 compliant vacuum system shall consist of electric motor driven positive displacement non-pulsating liquid ring vacuum pumps, ASME vacuum receiver, electrical control system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module Each liquid ring vacuum pump shall have a cast iron body, stainless shaft, bronze impeller, and mechanical seals to be direct driven by a TEFC motor. Pump suction accessories shall include an inlet check valve, isolation valve and inlet exible connector. Pump discharge accessories shall include a water reservoir constructed of stainless steel. The recirculation reservoir shall have a level switch and sight glass to maintain proper liquid level. The recirculation lines shall include a water/air radiator. The total recirculation service liquid system, after initial priming, the pump shall require no more than 3 gallons of water per hour.
Vacuum Receiver The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain valve, sight glass, vacuum gauge and the National Board label. System Controls The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories Accessories included for job site installation are inl et and discharge exible connectors, source isolation valve, vibration mounting pads and touch-up paint.
22
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Oa s i s Va c u u m S y s t e m s Se l e c t i o n Ch a r t
y g t i c H a " p 9 a 1 C @ n g M i s F e C D S
M F C S p g m u H " P 9 r 1 e P @
12.1
12.1
17.6
l e d o M r e m e b t s m y u S N
P H
e ) z s i S n o l k l n a a g T (
" A " h t g n e L
" B " h t d i W
" C " t h g i e H
r e t a * W p f m o u H P P r e G P
Stack Mounted Duplex
3
120
79"
70"
92"
0.08
LO300B-ST2
17.6
Stack Mounted Duplex
5
120
79"
70"
92"
0.2
LO500B-ST2
31.6
31.6
Stack Mounted Duplex
7.5
120
79"
70"
92"
0.22
LO750B-ST2
35.2
17.6
Stack Mounted Triplex
5
240
116"
70"
92"
0.2
LO500B-ST3
52.8
17.6
Stack Mounted Quarduplex
5
240
154"
70"
92"
0.2
LO500B-ST4
53
53
Stack Mounted Duplex
10
120
79"
70"
92"
0.38
LO1000B-ST2
63.2
31.6
Stack Mounted Triplex
7.5
240
116"
70"
92"
0.22
LO750B-ST3
69
69
Stack Mounted Duplex
15
120
79"
70"
92"
0.53
LO1500B-ST2
94.8
31.6
Stack Mounted Quarduplex
7.5
240
154"
70"
92"
0.22
LO750B-ST4
106
53
Stack Mounted Triplex
10
240
116"
70"
92"
0.38
LO1000B-ST3
138
69
Stack Mounted Triplex
15
240
116"
70"
92"
0.53
LO1500B-ST3
159
53
Stack Mounted Quarduplex
10
240
154"
70"
92"
0.38
LO1000B-ST4
207
69
Stack Mounted Quarduplex
15
240
154"
70"
92"
0.53
LO1500B-ST4
n o i t a r m u g e t s n y o S C
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
23
“ Wa t e r M i s e r ” L i q u i d Ri n g Va c u u m Sy s t e m s
•
NFPA 99 compliant
•
U.L. Listed Electrical Control panel
•
Model No.
HP
SCFM @ 19" Hg (each pump)
Water Sealed
LR-3-1
3
11.7
•
Available in Simplex, Duplex, Triplex or Quad
LR-5-1
5
21.6
•
Congurations: Tank, Base or Stack Mounted
LR-75-1
7.5
33.7
•
Medical or Laboratory Applications
LR-10-1
10
47.5
LR-15-1
15
75.5
•
Minimal maintenance, Maximum Durability
LR-20-1
20
100.7
•
No contacting parts in pumping chamber
LR-30-1
30
137.4
•
Cool running and long wear
LR-40-1
40
199.6
•
Optional Heat Exchanger shown below for optimal seal conservation
•
24
Warranty: 30/24 Months
uid
Design capacity values should correspond with the “Calculated Peak Demand”. This will allow for the reserve pump as is required per NFPA for level 1 installations.
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
L i q u i d R i n g Co n t i n u o u s O n De m a n d Va c u u m S y s t e m , Pa r t i a l Re c i r c u l a t i o n Liquid Ring Water Sealed Partial Recirculating Vacuum System Ohio Medical Corporation ® partial recirculating, NFPA 99 compliant vacuum system shall consist of electric motor driven positive displacement non-pulsating liquid ring vacuum pumps, ASME vacuum receiver, electrical control system, and interconnection piping and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for serviceability. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Vacuum Pump Module Each liquid ring vacuum pump shall have cast iron body, stainless steel impeller, and mechanical seals to be direct driven by a TEFC motor. Pump suction accessories shall include an inlet check valve, isolation valve and inlet exible connector. Pump discharge accessories shall include a discharge air/water separator constructed of high density polyethylene that also serves as the partial recirculation reservoir tank. The recirculation reservoir shall have a two station, combination level sensor/sight glass to maintain proper liquid level. The pump discharge separator vents shall be manifolded, and a exible connector shall be provided for job site installation to facilitate connection of vent piping to the exterior of the building. Recirculation liquid control will be interactive through a temperature sensor, with a range of 75°F to 90°F. Each recirculation liquid line shall include a strainer and isolation/ ow regulation valve. The make-up water lines shall include a strainer, regulating valve, solenoid valve with manual priming valve and Each pump will require 59 oF cooling water.
exible connector.
Vacuum Receiver The vacuum receiver shall be constructed to ASME standards, rated for full vacuum, and include a valved by-pass, manual drain valve, sight glass, vacuum gauge and the National Board label.
System Controls The Underwriters Laboratories Listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, and an electronic controller to automatically switch the operating sequence of the vacuum pumps. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, pump operation, accumulated time, maintenance interval, fault conditions and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the audible signal can be silenced and the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for reserve pump in use (lag alarm). Dry contacts are furnished for remote (master) alarm signals.
Accessories Accessories included for job site installation are inlet and discharge exible connectors, source isolation valve, vibration mounting pads and touch-up paint.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
25
L i q u i d R i n g Co n t i n u o u s O n De m a n d V a c u u m Sy s t e m s Se l e c t i o n Ch a r t , Pa r t i a l Re c i r c u l a t i o n
26
p m u / p g y t i H h c " c a 9 a e p 1 f a o C @ y i n t g M c i F s a e C p a D S c
p m / u p y g t i H h c " c a 5 a e p 2 f a o C @ y i n M t g F c i s C a e p a D A c
11.7 / 11.7
27.5 / 27.5
21.6 / 21.6
n o i t a r u g n o C m e t s y S
) r M e P t a G w ( g p H u - " e k 9 a 1 m @ e d g e r a r i e u v q A e r
) A ( h t g n e L
) B ( h t d i W
) C ( t h g i e H
t h g i e W
l e d o M r e m b e t s m y u S N
P H
e ) z s i S n o l k l n a a g T (
Tank Mounted Duplex
3
120
0.7
88
40
86
1370
LP3B-T2
69.5 / 69.5
Tank Mounted Duplex
5
120
1.1
92
46
70
1600
LP5B-T2
33.7 / 33.7
85 / 85
Tank Mounted Duplex
7.5
200
1.6
92
46
76
2050
LP75B-T2
47.5 / 47.5
122 / 122
Tank Mounted Duplex
10
200
2.0
92
56
76
2340
LP10B-T2
75.5 / 75.5
200 / 200
Tank Mounted Duplex
15
200
2
92
56
76
2630
LP15B-T2
95 / 47.5
244 / 122
Stack Mounted Triplex
10
240
2.0
114
64
94
3940
LP10B-ST3
100.7 / 100.7
248 / 248
Base Mounted Duplex
20
240
3.3
114
64
86
3300
LP20B-B2
101.1 / 33.7
255 / 85
Stack Mounted Quadruplex
7.5
240
1.6
114
64
92
4200
LP75B-ST4
137.4 / 137.4
360 / 360
Base Mounted Duplex
30
240
4.9
130
70
100
4300
LP30B-B2
142.6 / 47.5
366 / 122
Stack Mounted Quadruplex
10
240
2.0
114
64
92
4900
LP10B-ST4
151 / 75.5
400 / 200
Stack Mounted Triplex
15
240
2.9
114
64
94
4370
LP15B-ST3
201.4 / 100.7
496 / 248
Base Mounted Triplex
20
240
3.3
150
64
86
4600
LP20B-B3
226.5 / 75.5
600 / 200
Stack Mounted Quadruplex
15
240
2.9
114
64
94
5500
LP15B-ST4
274.8 / 137.4
720 / 360
Base Mounted Triplex
30
240
4.9
176
70
100
6260
LP30B-B3
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Sc r o l l Co m p r e s s e d A i r Systems
•
Ohio Medical Scroll Systems features a dedicated pump to motor design. This ensures optimal design safety and energy efciency by cycling on fewer motors during low demand in lieu of starting multi pump modules.
•
Quiet
•
Low vibration
•
10,000 hour maintenance interval for tip seals
•
5,000 hour maintenance interval for shaft bearing
•
Warranty: 30/24
Base Mounted Modules (with motor) Part Number
Description
H.P.
RPM
SCFM @ 100 PSIG
SCFM @ 140 PSIG
Max Pressure
Voltage
230072
Scroll Module Low Pressure
3
2200
8.5
-
115
208/230/460/3/60
230073
Scroll Module Low Pressure
5
3150
14.5
-
115
208/230/460/3/60
230082
Scroll Module High Pressure
3
2000
-
7.2
145
208/230/460/3/60
230083
Scroll Module High Pressure
5
2900
-
12
145
208/230/460/3/60
Bare Compressor Air Ends (without motor) Part Number
Description
H.P.
Max Pressure
230074
Scroll Air End Low Pressure
3 or 5
115
230075
Scroll Air End High Pressure
3 or 5
145
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
27
Oi l -L e s s Ro t a r y Sc r o l l A i r Sy s t e m s w i t h De s i c c a n t D r y e r s Oil-Less Rotary Scroll Compressed Air System with Desiccant Dryers Ohio Medical Corporation ® Oil-less Rotary Scroll Compressed Air System shall be fully NFPA 99 compliant. The unit will consist of electric motor driven single stage oil-less rotary scroll air compressor modules, electrical control system, ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters, air treatment module, interconnection copper/brass piping, and wiring. The components shall be modularly assembled to accommodate most existing doorways and designed for serviceabili ty. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Air Compressor Modules The compressors shall have PTFE composite tip seals and be rated for 10,000 hours of operation. Compressors bearings shall be external to the compression chamber and shall be serviceable. Compressors with bearings that are not accessible for service and have limited life span shall not be accepted. Each belt driven compressor module inlet shall include an inline lter with particle retention of 10 microns, inlet isolation valve and ex connector. The compressor discharge piping includes air cooled after cooler with moisture separator and automatic drain, isolation valves, ASME safety pressure relief valve, thermal malfunction shutdown devise. Air Receiver The corrosion resistant coated receiver is equipped with an ASME safety pressure relief valve, sight glass pressure gauge, automatic drain, three-valve by-pass and source isolation valve. Air Treatment Module The air treatment module shall include dual dryers, dual ltration system, and a CO and Dew Point monitoring devise with local audible and visual signals and dry contacts for remote monitoring. The components shall be mounted on a common base with interconnecting copper/brass piping and upstream and downstream isolation valves. The isolation valves shall allow either set of components to be serviced without shutting down the system. The dryer system shall be duplexed and provide a pressure dew point of 10oF. Dryers shall be heatless desiccant design selected and sized to provide for the peak-calculated demand. The desiccant dryer shall be equipped with a Dewpoint Dependent Switching feature to minimize the need for purge air. The dual ltration system is designed to remove liquids and particulate matter, and consists of one micron coalescing lters with differential pressure indicators and automatic drain, air line pressure regulators with gauges, nal pressure relief valve, and sampling valve.
System Controls The Underwriters Laboratories listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, temperature switches with reset button, and an electronic controller to automatically change the operating sequence of the compressors. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, dewpoint, pump operation, accumulated time, maintenance interval, fault conditions, and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. Accessories Accessories included for job site installation are inlet and discharge exible connectors, vibration mounting pads, and source isolation valve.
28
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Oi l -L e s s Ro t a r y Sc r o l l A i r Sy s t e m s w i t h De s i c c a n t Dr y e r s Se l e c t i o n Ch a r t
G y I t i S c P a 0 p 5 a C @ n g M i s F e C D S
e l u d o m r G e I p S P M F 0 5 C S @
n o i t a r m u g e t s n y o S C
p m u p r e p P H
p m u P f o . o N
) A ( h t g n e l
) B ( h t d i W
) C ( t h g i e H
l e d o m r e m e b t s m y u S N
9.8
9.8
Tank Mounted Duplex
3
2
74
32
71
AS300-T2-DD
16.3
16.3
Tank Mounted Duplex
5
2
74
32
71
AS500-T2-DD
32.6
16.3
Tank Mounted Triplex
5
3
84
64
72
AS500-T3-DD
48.9
16.3
Stack Mounted Quadruplex
5
4
84
64
72
AS500-ST4-DD
65.2
16.3
Stack Mounted Pentaplex
5
5
119
64
72
AS500-ST5-DD
81.5
16.3
Stack Mounted Hexaplex
5
6
116
64
72
AS500-ST6-DD
97.8
16.3
Stack Mounted Heptaplex
5
7
119
64
72
AS500-ST7-DD
114.1
16.3
Stack Mounted Octoplex
5
8
119
64
72
AS500-ST8-DD
130.4
16.3
Stack Mounted Enneaplex
5
9
155
64
72
AS500-ST9-DD
146.7
16.3
Stack Mounted Decaplex
5
10
155
64
72
AS500-ST10-DD
163
16.3
Stack Mounted Hendecaplex
5
11
155
64
72
AS500-ST11-DD
179.3
16.3
Stack Mounted Dodecaplex
5
12
155
64
72
AS500-ST12-DD
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
29
Oi l -L e s s Re c i p r o c a t i n g Pi s t o n A i r Co m p r e s s o r Sy s t e m s
•
Oil-Less Air Compressors
•
NFPA 99 compliant
•
U.L. Listed Electrical Control panel
•
Available in Simplex, Duplex, Triplex and Quadruplex
•
Congurations: Tank, Base or Stack Mounted
•
Medical and Laboratory Con gurations
•
Warranty: 30/24 months
30
Model #
HP
SCFM @ 50 PSIG
SCFM @ 100 PSIG
A200
2
8.2
5.1
A300
3
12
10.1
A500
5
20.9
17
A750
7.5
32
26.8
A1000
10
41.3
36
A1500
15
58
51
A2000
20
71.5
62
A2500
25
109
94
A3000
30
118
102
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Re c i p r o c a t i n g Oi l -L e s s C o n t i n u o u s O n De m a n d Co m p r e s s e d A ir Sy s t e m w i t h D es i c c a n t D r y e r s Oil-Less Reciprocating Continuous on Demand Compressed Air System with Desiccant Dryers Ohio Medical Corporation reciprocating NFPA 99 compliant air compressor system shall consist of electric motor driven oil-less air compressors, an Underwriters Laboratories listed electrical control mounted in a NEMA 12 enclosure, ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters and an air treatment module. The components shall be modularly assembled to accommodate most existing doorways. The system shall include interconnecting copper/brass piping and wiring to provide a functional operating package with applicable electrical and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Compressor Modules The compressors shall consist of a crankcase, connecting rods, integral counterweights for smooth operation, and cylinders and heads designed for ef cient heat dissipation. Piston rings shall be provided to reduce wear and have a life span of 10,000 hours. Each compressor cylinder is protected by a temperature switch, which will stop the drive motor and provide an alarm signal in the event of abnormal discharge air temperature. Each belt driven compressor module shall include an inline lter with a particle retention of 10 microns, inlet isolation valve discharge isolation valve, and ASME safety pressure relief valve. Air Receiver The corrosion resistant coated receiver is equipped with an ASME safety pressure relief valve, sight glass pressure gauge, automatic drain, three-valve by-pass and source isolation valve. Air Treatment Module The air treatment module shall include dual dryers, dual ltration system, and a CO and Dew Point monitoring devise with local audible and visual signals and dry contacts for remote monitoring. The components shall be mounted on a common base with interconnecting copper/brass piping and upstream and downstream isolation valves. The isolation valves shall allow either set of components to be serviced without shutting down the system. The dryer system shall be duplexed and provide a pressure dew point of 10oF. Dryers shall be heatless desiccant design selected and sized to provide for the peak-calculated demand. The desiccant dryer shall be equipped with a Dewpoint Dependent Switching feature to minimize the need for purge air. The dual ltration system is designed to remove liquids and particulate matter, and consists of one micron coalescing lters with differential pressure indicators and automatic drain, air line pressure regulators with gauges, nal pressure relief valve, and sampling valve.
System Controls The Underwriters Laboratories listed electrical motor control system shall be of a fuseless design in a NEMA 12 enclosure. The “Continuous On Demand” feature shall stop the operation of the motors during periods of low or no demand. The controls include individual self protected combination motor controls with short circuit, single phase and thermal overload protection, individual 120 volt control circuit transformers with fuseless primary and secondary protection, pressure sensors, temperature switches with reset button, and an electronic controller to automatically change the operating sequence of the compressors. The cabinet door shall have an HMI (Human Machine Interface) system status display to include system pressure, dewpoint, pump operation, accumulated time, maintenance interval, fault conditions, and silence button; lighted Hand-Off-Automatic selector switches and safety disconnect operating handles. All required local alarm functions shall be integrated into the packaged system. Accessories Included for job site installation are inlet and discharge exible connectors, vibration mounting pads, source isolation valve, and touch-up paint.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
31
Re c i p r o c a t i n g Oi l -L e s s Co n t i n u o u s On De m a n d Co m p r e s s e d A i r Sy s t e m s w /De s i c c a n t D r y e r s Se l e c t i o n Ch a r t c h a c p @ a a e C M G I n F S g C i s S P e y 0 D t i 5
32
e v r f p e x o s e p ) y m e r l i t t s i u n l c p i m a h e u e p c n m t a a o n s y C e ( o s
n o i t a r m u g e t s n y o S C
P H
e ) z s i S n o k l n l a a g T (
f s r o t o y t o Q M
) A ( h t g n e l
) B ( h t d i W
) C ( t h g i e H
t h g i e W
l e d o m r e m b e t s m y u S N
8.2
8.2
Tank Mounted Duplex
2
120
2
78
47
69
1900
A200B-T2-DD
12
12
Tank Mounted Duplex
3
120
2
78
47
69
2000
A300B-T2-DD
20.9
20.9
Tank Mounted Duplex
5
120
2
80
50
77
2300
A500B-T2-DD
32
32
Tank Mounted Duplex
7.5
120
2
81
52
77
3000
A750B-T2-DD
41.3
41.3
Tank Mounted Duplex
10
120
2
67.5
105
85
3900
A1000B-T2-DD
58
58
Stack Mounted Duplex
15
120
2
69.5
126
92
5400
A1500B-ST2-DD
64
32
Stack Mounted Triplex
7.5
120
3
107.5
112
85
5100
A750B-ST3-DD
71.5
71.5
Stack Mounted Duplex
20
120
2
69.5
126
92
5900
A2000B-ST2-DD
82.6
41.3
Stack Mounted Triplex
10
120
3
107.5
112
85
5350
A1000B-ST3-DD
96
32
Stack Mounted Quadruplex
7.5
240
4
116
116
85
7000
A750B-ST4-DD
109
109
Base Mounted Duplex
25
120
2
139.5
125
80
6300
A2500B-B2-DD
116
58
Stack Mounted Triplex
15
240
3
120
128
90
7600
A1500B-ST3-DD
118
118
Base Mounted Duplex
30
120
2
139.5
125
80
6300
A3000B-B2-DD
123.9
41.3
Stack Mounted Quadruplex
10
240
4
116
116
85
7350
A1000B-ST4-DD
143
71.5
Stack Mounted Triplex
20
240
3
120
132
90
8200
A2000B-ST3-DD
174
58
Stack Mounted Quadruplex
15
240
4
120
145
90
10600
A1500B-ST4-DD
214.5
71.5
Stack Mounted Quadruplex
20
240
4
120
145
90
11500
A2000B-ST4-DD
218
109
Base Mounted Triplex
25
240
3
204.5
142
95
9000
A2500B-B3-DD
236
118
Base Mounted Triplex
30
240
3
204.5
142
95
9000
A3000B-B3-DD
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Simplex A i r Sy s t e m
Oil-Less Reciprocating Simplex Compressed Air System Ohio Medical Corporation air compressor system shall consist of electric motor driven oil-less air compressors, an Underwriters Laboratories listed electrical control mounted in a NEMA 12 enclosure, ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters and an air treatment module. The components shall be modularly assembled to accommodate most existing doorways. The system shall include interconnecting copper/ brass piping and wiring to provide a functional operating package wi th applicable electrical and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up. Compressor Modules The compressors shall consist of a crankcase, connecting rods, integral counterweights for smooth operation, and cylinders and heads designed for ef cient heat dissipation. Piston rings shall be provided to reduce wear and have a life span of 10,000 hours. Each compressor cylinder is protected by a temperature switch, which will stop the drive motor and provide an alarm signal in the event of abnormal discharge air temperature. Each belt driven compressor module shall include an inline lter with a particle retention of 10 microns, inlet isolation valve discharge isolation valve, and ASME safety pressure relief valve. Air Receiver The corrosion resistant coated receiver is equipped with an ASME safety pressure relief valve, sight glass pressure gauge, automatic drain, three-valve by-pass and source isolation valve. System Controls The Underwriters Laboratories Listed simplex electrical motor control system shall be of a fuseless design in a NEMA 12 (dustproof) enclosure. The “Continuous On Demand” feature shall stop the operation of the motor during periods of low or no demand. The controls include a individual self protected combination motor control with short circuit, single phase and thermal overload protection, 120 volt control circuit transformer with fuseless primary and secondary protection, pressure sensor and a temperature switch with reset button. The cabinet door shall have an HMI (Human Machine Interface) system status display to include accumulated time and fault conditions; lighted Hand-Off-Automatic selector switch and safety disconnect operating handle. All required local alarm functions shall be integrated into the packaged system. The circuitry shall be designed so the visual indicator will remain until the fault has been cleared and the reset button actuated. Local alarm functions shall be provided for high air discharge temperature and thermal overload. Ship Separate Items & Accessories (Field Installed) Ship Separate Items: Non-cycling refrigerated type air dryer, coalescing 1 micron gauge. Accessories: Inlet and discharge paint.
exible
lter,
line pressure regulator with
connectors, vibration mounting pads, source isolation valve and touch-up
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
33
Simplex A i r Se l e c t i o n M at r i x n o i t a r m u g e t s n y o S C
0 5 @ M F G I C S S P
34
P H
e ) z s i S n o l k l n a a g T (
) A ( h t g n e l
) B ( h t d i W
) C ( t h g i e H
t h g i e W
r e m l b e e t d m s o u y S m N
6.7
Oil-Less Scroll
Tank Mounted Simplex
2
80
75
32
54
745
AS200-T1
8.2
Oil-Less Reciprocating
Tank Mounted Simplex
2
80
75
32
54
745
A200B-T1
9.9
Oil-Less Scroll
Tank Mounted Simplex
3
80
75
32
54
770
AS300-T1
12
Oil-Less Reciprocating
Tank Mounted Simplex
3
80
75
32
54
770
A300B-T1
16.4
Oil-Less Scroll
Tank Mounted Simplex
5
80
75
32
54
962
AS500-T1
20.9
Oil-Less Reciprocating
Tank Mounted Simplex
5
80
75
32
54
962
A500B-T1
32
Oil-Less Reciprocating
Tank Mounted Simplex
7.5
80
75
32
54
1045
A750B-T1
41.3
Oil-Less Reciprocating
Tank Mounted Simplex
10
80
75
32
54
1156
A1000B-T1
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
I n st r u m e n t A i r System High Pressure Reciprocating Compressed Instrument Air System with Desiccant Dryers Ohio Medical Corporation Instrument Air NFPA 99 compliant compressor system shall consist of electric motor driven pressure lubricated air compressors, an Underwriters Laboratories listed electrical control mounted in a NEMA 12 enclosure, ASME air receiver, air cooled aftercoolers with individual drains, inline intake air lters and an air treatment module. The components shall be modularly assembled to accommodate most existing doorways. The system shall include interconnecting copper/brass piping and wiring to provide a functional operating package with applicable electrical and plumbing connections at the installation site. The packaged unit shall be factory tested prior to shipment and warranted for a period of (30) months from date of shipment or (24) months from date of start-up.
Compressor Modules Each compressor is to be a Quincy Compressor model QR-25 pressure lubricated reciprocating design, manufactured in the United States of America. Each pressure lubricated compressor will be v belt driven, single-acting, two-stage, air-cooled reciprocating type. Cylinders will be cast iron, with deep cooling ns. Valve assemblies can be removed individually without removing the cast iron head and inlet or discharge piping. The compressor shall have hardened and lapped disc-type valves for continuous service. There will be two pieces to the connecting rods with cast in tubes for positive oil distribution to the piston pins. Low pressure pistons will be aluminum with three automotive type compression rings and one automotive type oil ring. The high pressure pistons will be cast iron with three automotive type compression rings and one automotive oil ring. Connecting rod bearings shall be steel backed Babbitt lined automotive type. Crankshafts shall have integral counterweights and be ri e drilled for pressure lubrication to ensure positive oil distribution. Adjustable anti-friction tapered roller bearings shall be mounted on each end of the crankshaft to take both radial and thrust loads. Compressor ywheels shall be dynamically balanced and of the fan type. Finned intercoolers shall be an integral part of all two-stage units, and located in the direct blast of air from the fan-type ywheel. Compressors shall be equipped with a crankshaft driven oil pump located in the bearing carrier for easy removal and inspection. There is also an oil lter located on the bearing carrier as well as a pre- lter screen in the crankcase to help keep the oil clean. Low oil protection shall be a standard feature of the compressor, along with loadless starting.
Air Receiver The ASME corrosion resistant coated air receiver is equipped w ith sight glass, pressure gauge, manual and electronic drains, three-valve by-pass and source isolation valve. The receiver shall be rated for 350 psig. Air Treatment Module Desiccant air dryers shall be modular adsorption design. Each desiccant dryer shall be capable - 40°F pressure dew point at 100 psig. The desiccant dryer shall be twin tower design with switching controls to change from one tower to the other during regeneration. The dryer will come standard with energy management feature to control and delay purge air requirement. Carbon monoxide/dew point monitor shall be combination type and shall be in NEMA 12 enclosure. Desiccant dryers will have integral coalescing and particulate lters with activated charcoal lters with bypass piping. Coalescing lter will be rated at 98% ef ciency at 0.01 micron. System Controls The air treatment module shall include a microprocessor based, UL/CUL-508 listed and NFPA-99 compliant alternating control panel with a NEMA type 12 enclosure. The control panel is equipped with through the door, rotary type fused disconnect switches and a through the door, rotary type, non-fused disconnect switch which controls incoming power to the transformer that powers the CO/Dew Point Monitor. There are independent controls for each compressor with primary and secondary fused, 120VAC control circuit transformers. The panel will include an ON/OFF selector switch, elapsed time meter, two pressure switches (lead and lag), two motor starters. Each display board has a POWER ON L.E.D., COMPRESSOR RUNNING L.E.D., 2 mode of operation L.E.D.’s (Continuous or Start/Stop), 2 starting sequence L.E.D.’s (Lead or Auto). Three pressure sensitive membrane style pushbuttons to select the mode of operation (Continuous or Start/Stop), starting sequence (Lead or Auto) and ALARM CANCEL that silences the audible alarm horn and deactivates the alarm contacts. Four ashing fault indicators (Overload, Hi Temp, Lag Comp and System). One fault code display which indicates the type of system fault that has occurred. Audible alarm horn that sounds when one of the four fault conditions has occurred. www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
35
I n st r u m e n t A i r Se l e c t i o n M a t r i x
h c ) a e e v r I e G s S y t e P i c r 0 n 5 a i p p 2 a C m @ n u g p M i F s e e n o C D ( S
36
p m u p h c a e f G o I S y t i P c 0 a 5 p 2 a C @
n o i t a r m u g e t s n y o S C
P H
e ) z s i S n o k l n l a a g T (
s r o t o M f o y t Q
) A ( h t g n e l
) B ( h t d i W
) C ( t h g i e H
l e d o m r e m e b t s m y u S N
23.4
23.4
Tank Mounted Duplex
7.5
120
2
68"
80"
83"
AI750B-T2
30.2
30.2
Tank Mounted Duplex
10
120
2
68"
80"
83"
AI1000B-T2
46.8
23.4
Stack Mounted Triplex
7.5
120
3
106"
80"
83"
AI750B-ST3
60.4
30.2
Stack Mounted Triplex
10
120
3
106"
80"
83"
AI1000B-ST3
70.2
23.4
Stack Mounted Quadruplex
7.5
120
4
106"
80"
83"
AI750B -ST4
90.6
30.2
Stack Mounted Quadruplex
10
120
4
106"
80"
83"
AI1000B-ST4
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
A i r Tr e a t m e n t M o d u l e s ( Fo r Oi l -L e s s Co m p r e s s o r s w i t h De s i c c a n t D r y e r s )
Ohio Medical compressor control panel provides highly efcient purge and switching control
Clean, dry and oil free compressed air
compact and lightweight design
Perfect solution for a eld upgrade on older existing air compressor systems
Model Number
Replacement Elements
SCFM @ 100 PSIG
Dimensions (LxWlH)
Pre-Filter
After Filter
HCTM-B-DDS-6
6
34x16x36
233722
233777
HCTM-B-DDS-10
10
34x16x36
233722
233777
HCTM-B-DDS-24
24
34x20x40
233722
233777
HCTM-B-DDS-32
32
34x30x49
233722
233777
HCTM-B-DDS-42
42
34x20x56
233722
233777
HCTM-B-DDS-53
53
34x20x61
233773
223778
HCTM-B-DDS-65
65
34x20x69
233774
233779
HCTM-B-DDS-88
88
34x20x79
233774
233779
HCTM-B-DDS-106
106
40x30x66
233774
233779
HCTM-B-DDS-130
130
40x30x72
223775
233780
HCTM-B-DDS-176
176
40x30x82
223775
233780
HCTM-B-DDS-240
240
50x40x72
223776
233781
HCTM-B-DDS-360
360
50x40x72
-
-
Combination CO / Dewpont Moni tor (-112F to +68F) Model Number
Description
233013
CO/Dewpoint Combination Monitor
233015
CO Calibration Kit
2610-001
Humidi er (for CO Monitoring)
261922
Gas Specict DISS Air Check Assembly 1/8" NPT
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
37
Ca l c u l a t e d Pe a k De m a n d f o r a M e d i c a l Su r g i c a l Va c u u m Sy s t e m i n a T y p i c a l S ho r t Te r m Ge n e r a l H o s p i t a l Qty
Medical Gas Vacuum Outlet Locations
Design Flow in SCFM (Free Air) Per Room
Per Bed
Simultaneous Use Factor % Per Outlet
Total
ANESTHETIZING LOCATIONS: Specialized Surgeries (Open Heart, Organ Transplant)
4.0
100
Major / Outpatient OR
3.5
100
Cystoscopy / Endoscopy
2.0
100
Delivery Room
1.0
100
Emergency Operating Room
3.0
100
Cardiac Catheterization
1.0
10
Other Anesthetizing Areas (Minor O.R., Induction Rooms, etc.)
1.0
50
Waste Anesthetizing Gas Evacuation
2.0
100
ACUTE CARE (NON-ANESTHETIZING LOCATIONS): Post Operative Recovery Room
3.0
50
O.B. Recovery Room
2.0
50
Intensive Care Units (Except Cardiac)
2.0
75
Emergency Room (Trauma, Cardiac)
1.0
100
Cardiac Intensive Care
2.0
50
Neonatal I.C.U.
1.0
50
Special Procedure (X-Ray,Dialysis, Radiology)
1.5
25
Surgical Excision Room
1.0
10
SUBACUTE PATIENT CARE AREAS: Normal Nursery
1.0
Premature Nursery
1.0
Respiratory Care
10 20
1.5
10
Labor/Birthing
1.0
10
Patient Room (Surgical)
1.5
50
Patient Room (Medical)
1.0
10
Exam & Treatment Rooms
1.0
10
ER (Cast Room, OB/GYN)
1.0
10
Pre-Op Holding
1.0
10
Stress Test EKG & EEG
1.0
10
OTHER AREAS: Autopsy / Morgue
2.0
20
Central Supply / Sterile
1.5
10
Equipment Repair,Calibration & Teaching
1.5
10
Medical Lab / Pharmacy
1.0
10
Anesthesia/Workroom
1.5
10
Total Estimated Peak Flow (In SCFM @ 19" Hg) Total Estimated Peak Flow (in SCFM @ 19” Hg)
For the actual worksheet go to www.ohiomedical.com, go to “Technical Assistance”, select Calculated Peak Demand for a Typical Hospital (vacuum). This will take you to an excel spreadsheet, which all that you have to do is ll in the quantity and it will calculate for you.
38
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Ca l c u l a t e d Pe a k De m a n d f o r a M e d i c a l A i r Sy s t e m i n a T y p i c a l Sh o r t Te r m Ge n e r a l H o s p i t a l Medical Air Outlet Locations
Qty
Design Flow in SCFM (Free Air) Per Unit
Per Room
Per Bed
Per Outlet
Simultaneous Use Factor %
Totals
ANESTHETIZING LOCATIONS: Special Surgery (Open Heart, Organ Transplant, Ortho)
0.5
100
Major / Outpatient Surgery
0.5
100
Minor Surgery
0.5
75
Emergency Surgery
0.5
50
Cardiac Catherization
0.5
50
Endoscopy / Cytoscopy
1.0
10
Delivery Room / C-Section
0.5
100
ACUTE CARE LOCATIONS: Recovery Room/Surgical
2.0
50
Recovery Room - OB
0.5
25
ICU/CCU & Pediatric I.C.U.
2.0
50
Emergency Room (Trauma, Cardiac)
2.0
50
Anesthesia Workroom
1.5
Neonatal ICU
10 1.5
75
Pre-Op Holding
1.5
10
Dialysis Unit, Radiology
0.5
10
Ventilators Adult*
3.5
50
Ventilators Infants*
3.5
50
2.0
50
SUBACUTE CARE LOCATIONS: Pediatric Croup Tents** Nursery Full Term & Isolation ER (Cast Room, OB/GYN)
0.5 1.0
Patient Rooms (Medical & Surgical)
10 0.5
EEG & EKG
10 1.0
Exam & Treatment
1.0
Respiratory Care
50 10
1.0
Pulmonary Function Lab
50 1.0
Birthing & LDRP
25
1.0
50 50
OTHER: Respiratory Care Workroom
1.5
10
Autopsy / Morgue
1.5
10
Equipment Repair
1.5
10
Total Estimated Peak Flow (I n SCFM @ 50 PSIG) * Important! You must consult with respiratory therapy personnel concerning the maximum number of ventilators that could be used at one time and the average ow rates for each ventilator. DO NOT CALCULATE your total peak requirements without this information. ** If powered by Medical Air For the actual worksheet go to www.ohiomedical.com, go to “Technical Assistance”, select Calculated Peak Demand for a Typical Hospital (air). This will take you to an excel spreadsheet, which all that you have to do is ll in the quantity and it will calculate for you.
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
39
Manifold Sizing Medical Air Manifold: Cylinders Per Bank
=
400 Ft3
x
No. of beds
x
(cubic ft of gas used per month per bed)
12
÷
(months per year)
52
÷
(# of weeks per year)
234
÷
(Cubic ft per cylinder)
2 (Cylinders per bank)
Surgery Center Oxygen Cylinder Manifolds: Where usage is not known Cylinders Per Bank
=
400 Ft3
x
No. of beds
x
(cubic ft of gas used per month per bed)
12
÷
(months per year)
52
÷
(# of weeks per year)
244
÷
(Cubic ft per cylinder)
2 (Cylinders per bank)
Hospital Oxygen Cylinder Manifolds: Where usage is not known Cylinders Per Bank
=
700 Ft3
x
No. of beds
x
(cubic ft of gas used per month per bed)
12 (months per year)
÷
52
÷
(# of weeks per year)
244
÷
(Cubic ft per cylinder)
2 (Cylinders per bank)
*With a monthly demand of 30,000 ft, bulk O 2 should be used. Where usage is known:
Duplex Manifold Size Divide by 2 for each Bank 6 10 14 18 22 26 30
Usage, Ft3 per Month 5850 9750 13650 17600 21500 25350 29250
N2O Cylinder Manifold:
40
N2 Cylinder Manifold:
CO2 Cylinder Manifold:
Divide total by 2 for each Bank
Divide total by 2 for each Bank
Divide total by 2 for each Bank
No. of Operating Rooms 1 to 4 5 to 8 9 to 12 13 to 16 17 to 20
No. of Operating Rooms 1 to 2 3 to 4 5 to 6 7 to 8 9 to 10 11 to 12 13 to 14
No. of Operating Rooms 1 to 8 9 to 16 17 to 32
Total No. of Cylinders 4 8 12 16 20
Total No. of Cylinders 4 8 12 16 20 24 28
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
Total No. of Cylinders 4 8 12
M e d i c a l G a s Eq u i p m e n t M o u n t i n g Height Detail N ’ I L S . ” L R 2 A . 6 O
S T E L T U O
” 0 6
L S E A N G A P
” 2 6 E V L A X V O E B N O Z ” 6 6
G N I L I E C D E H S I N I F
M R A L A R E T S A M ” 6 6
R P O O T R C D A E R S T O E H R &
M R A L A A E R A
R O O L F D E H S I N I F
” 6 6 ” 4 ’ 6
Y - R L E L L O A B G S N L C A T N I M A I C L I U U R E L N T A R C O A C T C M E L E E R
” 3 6
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
41
Mechanical Room
Su g g e s t e d M e d i c a l G a s M a s t e r
MASTER ALARM #1 MAY BE LOCATED IN SECORITY (Requires 24 Hourt/Day Surveillance)
Soiurce/Mainline Isolation Valve
Soiurce/Mainline Isolation Valve
SUGGESTED ALARM POINTS FOR: MEDICAL/SURGICAL VACUUM PUMPS RESERVE VACUUM PUMP RUNNING 5.1.3.6.8 / 5.1.9.2.4(9) / 5.1.9.5.2 / 5.1.9.5.4(4) (Termination point located in vacuum control panel) MAIN LINE VACUUM LOW 5.1.9.2.4(8) (Terminal point from main line vacuum switch) WASTE ANESTHETIC GAS VACUUM PUMPS RESERVE VACUUM PUMP RUNNING 5.1.9.5.2 / 5.1.3.7.4.1 / 5.1.9.2.4(9) / 5.1.9.5.4(5) (Termination point located in vacuum control panel) MAINLINE EVACUATION LOW 5.1.9.2.4(8) (Termination point from main line vacuum switch)
42
SUGGESTED ALARM POINTS FOR: MEDICAL AIR COMPRESSOR (All Types) RESERVE COMPRESSOR RUNNING 5.1.9.5.4(1) / 5.1.3.5.14.5 (Termination point located in compressor control panel) MAINLINE PRESSURE HIGH 5.1.9.2.4(7) (Termination point from mainline pressure switch) MAINLINE PRESSURE LOW 5.1.9.2.4(7) (Termination point f rom mainline pressure switch) CARBON MONOXIDE HIGH 5.1.9.4.5(2) / 5.1.3.5.15(2) (Termination point from CO monitor) DEW POINT HIGH 5.1.9.2.4(10) / 5.1.3.5.15(1) (Termination point located in compressor control panel) Also, For Reciprocating (Piston type) Compressors HIGH TEMPERATURE 5.1.3.5.14.3 / 5.1.9.5.4(9) / 5.1.3.5.14.4(1) (Termination point located in compressor control panel) RECEIVER FLOODED (see note #1) 5.1.3.5.14.1 / 5.1.9.5.4(7) (Termination point located in compressor control panel) Also, For Liquid Ring Compressors: RECEIVER FLOODED 5.1.3.5.14.1 / 5.1.9.5.4(7) (Termination point located in compressor control panel) HIGH WATER SEPARATOR 5.1.3.5.14.2 / 5.1.9.5.4(8) (Termination point located in compressor control panel)
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
A l a r m Po i n t s f o r L ev e l 1 Fa c i l i t i e s
Medical Gas Room
Typical Nitrogen (N2) Installation
Soiurce/Mainline Isolation Valve MASTER ALARM #2 MAY BE LOCATED IN ENGINEERING DEPARTMENT/AREA
SUGGESTED ALARM FUNCTIONS FOR: MANIFOLDS WITHOUT RESERVE CYLINDER BANK CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.10.6 (Termination point located in power supply) MAINLINE PRESSURE HIGH 5.1.9.2.4(7) (Termination point from mainline pressure switch) MAINLINE PRESSURE LOW 5.1.9.2.4(7) (Termination point from mainline pressure switch)
Typical Oxygen (O2), Medical Air (Air), Nitrous Oxide (N2O), Carbon Dioxide (CO2) Installation Source/Mainline Isolation Valve 261920* Actuating Pressure Switch (Reserve In Use, Activate HP Cylinders set at 85 PSIG)
Check Valve Located in Pigtail
SUGGESTED ALARM FUNCTIONS FOR BULK OXYGEN SYSTEM CRYOGENIC BULK GAS UNITS WITH CRYOGENIC RESERVE
SUGGESTED HEIGHT
Regulator SUGGESTED ALARM FUNCTIONS FOR: MANIFOLDS WITH RESERVE CYLINDER BANK CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.12.9(1) (Termination point located in power supply) RESERVE IN USE 5.1.9.2.4(3) / 5.1.3.4.15.5 / 5.1.3.4.12.9(3) (Terminal point from actuating pressure switch 261920*) RESERVE LOW 5.1.3.4.12.9(4) / 5.1.9.2.4(5) (Terminal point from actuating pressure switch 261921*) MAINLINE PRESSURE HIGH 5.1.9.2.4(7) (Termination point from mainline pressure switch) MAINLINE PRESSURE LOW 5.1.9.2.4(7) (Termination point from mainline pressure switch)
261921 Actuating Pressure Switch (“Reserve Low”, 500 PSIG)
LIQUID LEVEL LOW 5.1.9.2.4(2) / 5.1.3.4.13.6(1) RESERVE IN USE 5.1.9.2.4(3) / 5.1.3.4.13.6(2) MAIN LINE PRESSURE HIGH 5.1.9.2.4(7) MAIN LINE PRESSURE LOW 5.1.9.2.4(7) RESERVE LOW 5.1.9.2.4(5) / 5.1.3.4.11.6(3) CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.13.6(5) RESERVE PRESSURE LOW 5.1.9.2.4(6) / 5.1.3.4.11.6(4) (Not Functional) CRYOGENIC BULK GAS UNITS WITH CYLINDER RESERVE LIQUID LEVEL LOW 5.1.3.4.13.6(1) / 5.1.9.2.4(2) RESERVE IN USE 5.1.9.2.4(3) / 5.1.3.4.13.6(2) / 5.1.3.4.13.5 MAIN LINE PRESSURE HIGH 5.1.9.2.4(7) MAIN LINE PRESSURE LOW 5.1.9.2.4(7) CHANGEOVER 5.1.9.2.4(1) / 5.1.3.4.13.6(5)
www.ohiomedical.com Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010
43
44
For ordering information, please call 1-800-448-0770 or 847-855-0500 Information/products in this catalog subject to change without notice. 255322 (Rev.11) 06/2010