DEUTSCHE NORM
June 2006
D
DIN 878
ICS 17.040.30
Together with DIN EN ISO 463 :2006-06,
supersedes DIN 878:1983-10
Geometrical product specifications (GPS) – Mechanical dial gauges – Limits for metrological characteristics Geometrische Produktspezifikation (GPS) – Mechanische Messuhren – Grenzwerte für messtechnische Merkmale
5 1 : 0 1 1 3 8 0 5 1 0 2 1 0 0 1 7 0 7 6 1 7 . r N f L 1 9 2 6 3 9 6 . r N d K i t S . d t L . c i T n o y s a r b i l a K t e m e g E h t u e B d a o l n w o D n e m r o N
Document comprises 3 pages
Translation by DIN-Sprachendienst. In case of doubt, the German-language original should be consulted as the authoritative text.
©
No part of this translation may be reproduced without prior permission of DIN Deutsches Institut für Normung e. V., Berlin. Beuth Verlag GmbH, 10772 Berlin, Germany, has the exclusive right of sale for German Standards (DIN-Normen).
English price group 5 www.din.de www.beuth.de
!,vFw" 06.07
9833584
DIN 878:2006-06 (E)
Foreword This standard has been prepared by Technical Committee NA 152-03-07-02 Eindimensionale Längenprüftechnik of the Normenausschuss Technische Grundlagen Grundlagen (Fundamentals in Technology Standards Committee). DIN 878:1983-10 has been fully revised to reflect the publication of DIN EN ISO 463:2006, the latter being a framework standard that defines the most important design and metrological characteristics of mechanical dial gauges but does not specify numerical values for the metrological characteristics; Germany had been against the decision to delete these values. This new edition of DIN 878 has been published to ensure German industry continues to have access to the limit values of metrological characteristics. Amendments This standard differs from DIN 878:1983-10 878:1983-10 as follows. a)
Only numerical numerical values for for limits of metrological metrological characteristics characteristics are specified. specified.
b)
The title has been changed to reflect the content.
c)
The standard has been editorially revised.
Previous editions DIN 878-1: 1941-02, 1943-08, 1959-06 5 1 : 0 1 1 3 8 0 5 1 0 2 1 0 0 1 7 0 7 6 1 7 . r N f L 1 9 2 6 3 9 6 . r N d K i t S . d t L . c i T n o y s a r b i l a K t e m e g E h t u e B d a o l n w o D n e m r o N
DIN 878-2: 1955-12 DIN 878: 1970-06, 1979-01, 1983-10
1
Scope
This standard applies to mechanical gauges as in DIN EN ISO 463 having scale intervals of 0,01 mm and a measuring range of 0,4 mm; 0,8 mm; 3 mm; 5 mm and 10 mm. It specifies the numerical values for limits of metrological metrological characteristics. characteristics. This standard does not cover waterproof or watertight mechanical gauges, which are subject to special requirements.
2
Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. DIN EN ISO 1, 1, Geometrical product specifications (GPS) — Standard reference temperature for geometrical product specifications specifications and and verification DIN EN ISO 463, Geometrical Product Specifications (GPS) — Dimensional measuring equipment — Design and metrological characteristics of mechanical dial gauges
2
DIN 878:2006-06 (E)
DIN EN ISO 14253-1, 14253-1, Geometrical product specifications (GPS) — Inspection by measurement of workpieces and measuring equipment — Part 1: Decision rules for proving conformance or non-conformance with specifications DIN EN ISO 14978, Geometrical product specifications (GPS) — General concepts and requirements for GPS measuring equipment Internationales Wörterbuch der Metrologie (VIM) (International Vocabulary of Basic and General Terms in Metrology) 1)
3
Concepts
For the purposes of this standard, the concepts defined in DIN EN ISO 463, DIN EN ISO 14253-1, 14253-1, DIN EN ISO 14978 and the “International vocabulary of basic and general terms in metrology” (VIM) apply.
4
Requirements
4.1
Reference temperature
20 °C as in DIN EN ISO 1.
4.2
Limit values for metrological characteristics characteristics
The limit values for metrological characteristics are to be taken from Table 1. They apply to all gauge orientations. Table 1 — Limit values for metrological characteristics 5 1 : 0 1 1 3 8 0 5 1 0 2 1 0 0 1 7 0 7 6 1 7 . r N f L 1 9 2 6 3 9 6 . r N d K i t S . d t L . c i T n o y s a r b i l a K t e m e g E h t u e B d a o l n w o D n e m r o N
Limit values for hysteresis of error of indication
repeatability of error of indication
any 1/10 revolution
any 1/2 revolution
one revolution
the measuring range
0,4 mm
3 µm
3 µm
5 µma
—
—
7 µm
0,8 mm
3 µm
3 µm
5 µma
—
—
7 µm
3 mm
3 µm
3 µm
5 µm
8 µm
9 µm
10 µm
5 mm
3 µm
3 µm
5 µm
9 µm
10 µm
12 µm
10 mm
3 µm
3 µm
5 µm
9 µm
10 µm
15 µm
a
1)
errors of indication over a range of
Measuring range
Refers to 1/10 of the measuring range.
Obtainable from Beuth Verlag GmbH , AuslandsNormen-Service (ANS), 10772 Berlin, Order No. 13086, ISBN 3-410-13086-1.