2. Techni cal System Description
1 / 20 20 0 3
Table of Contents
MT-MD MT-MD- DE08C M.KAY
2- 1
Page
General Information
2- 7
2.1
Overview Sub- Racks
2- 8
2.1.1
Legend Overview Sub- Racks
2- 8
2.2
Top Level Sub- Rack
2- 9
2.2.1
Legend Top Level Sub- Rack
2- 9
2.3
UF Sub- Rack
2- 10
2.3.1
Legend UF Sub- Rack
2- 10
2.4
DF Sub- Rack
2- 11
2.4.1
Legend DF Sub- Rack
2- 11
2.5
Water Inlet Sub- Rack
2- 12
2.5.1
Legend Water Inlet Sub- Rack
2- 12
2.6
Rinsing Bridge
2- 13
2.6.1
Legend Rinsing Bridge
2- 13
2.7
Rear Door
2- 14
2.7.1
Legend Rear Door
2- 14
2.8
TFT Monitor
2- 15
2.8.1
Legend TFT Monitor
2- 15
2.9
Power Board 1 PB1
2- 16
2.9.1
Legend Power Board 1 PB1
2- 16
2.10
Power Board 2 PB2
2- 17
2.10.1
Legend Power Board 2 PB2
2- 17
2.11
Supervisor Board SB
2- 18
2.11.1
Legend Supervisor Board SB
2- 18
2.12
Controller Board CB
2- 19
2.12.1
Legend Controller Board CB
2- 19
2.13
Switch Mode Power Supply SMPS (Benning)
2- 20
2.13.1
Legend Switch Mode Power Supply SMPS
2- 20
2.14
Front Door
2- 21
2.14.1
Legend Front Door
2- 21
2.15
Flow Diagram
2- 22
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
B. Braun Medizintechnol zintechnologie ogie GmbH
2. Techni cal System Description
MT-MD MT-MD- DE08C M.KAY
1 / 20 20 0 3
2- 2
2.15.1
Legend Flow Diagram
2- 23
2.16
Description Flow Diagram
2- 25
2.16.1
Water Inlet Section with Upline Tank
2- 25
2.16.2
Degassing Circuit with Temperature System
2- 26
2.16.3
Dialysate Processing
2- 27
2.16.4
Central Bicarbonate and Concentrate Supply
2- 28
2.16.5
BIC Cartridge Holder
2- 28
2.16.6
Balance Chamber System
2- 29
2.16.7
Working Principle Balance Chamber System
2- 29
2.16.8
Ultrafiltration and Rinsing Bridge
2- 31
2.16.9
Disinfection and Cleaning Program
2- 31
2.17
Block Diagram
2- 32
2.17.1
Legend Block Diagram
2- 33
2.18
TFT Monitor
2- 35
2.18.1
Description TFT Monitor
2- 35
2.19
ABPM Option ion (Non-Inv -Invasive ive Bloo lood Pressure Measurement)
2-37 -37
2.19.1
Wiring DiagramABPM Option
2- 37
2.20
bioL ioLogic RR Option tion (Autom tomatic Bloo lood Pressure Stab tabilisa ilisation tion)
2- 38
2.21
Supervisor Board SB
2- 39
2.21.1
Supervisor
2- 39
2.21.1.1
Block DiagramSupervisor
2- 39
2.21.1.2
Description Supervisor
2- 39
2.22
Bicarbonate Conductivity Measurement
2- 41
2.22.1
Bloc lock Diag iagramBica icarbonate Conductivity ity Measurement
2-41 -41
2.22.2
Description Bicarbonate Conductivity Measurement
2-41 -41
2.23
END Conductivity Measurement Controller
2- 42
2.23.1
Description ion END Conductivity ity Measurement Controlle ller
2-42 -42
2.23.2
END Conductivity Measurement Supervisor
2- 42
2.23.3
Bloc lock Diag iagramEND Conductivity Measurement Su Superviso isor
2- 42
2.23.4
Descripti iptio on END Conductivit ivity y Measurement Superviso isor
2-42 -42
2.24
Temperature Measurement
2- 43
2.24.1
Bloc lock DiagramDegassing ing Temperature Measurement
2-43 -43
2.24.2
Design Degassing Temperature Measurement
2-43 -43
2.24.3
Description Bicarbonate Temperature Measurement
2-43 -43
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
B. Braun Medizintechnol zintechnologie ogie GmbH
2. Techni cal System Description
MT-MD MT-MD- DE08C M.KAY
1 / 20 20 0 3
2- 2
2.15.1
Legend Flow Diagram
2- 23
2.16
Description Flow Diagram
2- 25
2.16.1
Water Inlet Section with Upline Tank
2- 25
2.16.2
Degassing Circuit with Temperature System
2- 26
2.16.3
Dialysate Processing
2- 27
2.16.4
Central Bicarbonate and Concentrate Supply
2- 28
2.16.5
BIC Cartridge Holder
2- 28
2.16.6
Balance Chamber System
2- 29
2.16.7
Working Principle Balance Chamber System
2- 29
2.16.8
Ultrafiltration and Rinsing Bridge
2- 31
2.16.9
Disinfection and Cleaning Program
2- 31
2.17
Block Diagram
2- 32
2.17.1
Legend Block Diagram
2- 33
2.18
TFT Monitor
2- 35
2.18.1
Description TFT Monitor
2- 35
2.19
ABPM Option ion (Non-Inv -Invasive ive Bloo lood Pressure Measurement)
2-37 -37
2.19.1
Wiring DiagramABPM Option
2- 37
2.20
bioL ioLogic RR Option tion (Autom tomatic Bloo lood Pressure Stab tabilisa ilisation tion)
2- 38
2.21
Supervisor Board SB
2- 39
2.21.1
Supervisor
2- 39
2.21.1.1
Block DiagramSupervisor
2- 39
2.21.1.2
Description Supervisor
2- 39
2.22
Bicarbonate Conductivity Measurement
2- 41
2.22.1
Bloc lock Diag iagramBica icarbonate Conductivity ity Measurement
2-41 -41
2.22.2
Description Bicarbonate Conductivity Measurement
2-41 -41
2.23
END Conductivity Measurement Controller
2- 42
2.23.1
Description ion END Conductivity ity Measurement Controlle ller
2-42 -42
2.23.2
END Conductivity Measurement Supervisor
2- 42
2.23.3
Bloc lock Diag iagramEND Conductivity Measurement Su Superviso isor
2- 42
2.23.4
Descripti iptio on END Conductivit ivity y Measurement Superviso isor
2-42 -42
2.24
Temperature Measurement
2- 43
2.24.1
Bloc lock DiagramDegassing ing Temperature Measurement
2-43 -43
2.24.2
Design Degassing Temperature Measurement
2-43 -43
2.24.3
Description Bicarbonate Temperature Measurement
2-43 -43
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
B. Braun Medizintechnol zintechnologie ogie GmbH
2. Techni cal System Description
MT-MD MT-MD- DE08C M.KAY
1 / 20 20 0 3
2- 3
2.24.4
Descriptio iption n Dialy ialys sate Contro troller ller Temperatur ture Measurement
2- 43
2.24.5 2.24.5
Block DiagramDialysate Supe Supervi rvisor sor Temperature rature Measurement
2-44
2.24.6
Descriptio iption n Temperature tureMe Mea asurement Dialy ialys sate Superviso isor
2- 44
2.25
Level Measurement
2- 44
2.25.1
Block DiagramLevel Measurement Upline Tank
2-44 -44
2.25.2
Description Level Measurement Upline Tank
2- 44
2.26
Reed Contacts
2- 45
2.26.1
Block DiagramCoupling Status
2- 45
2.26.2
Description Coupling Status
2- 45
2.27
Pressure Measurement
2- 45
2.27.1
Block DiagramVenous Pressure Measurement
2-45 -45
2.27.2
Description Venous Pressure Measurement
2- 45
2.27.3
Block DiagramArterial Pressure Measurement
2-46 -46
2.27.4
Description Arterial Pressure Measurement
2- 46
2.28
Blood Inlet Pressure Measurement
2- 46
2.28.1
Block DiagramBlood Inlet Pressure Measurement
2-46 -46
2.28.2
Description Blood Inlet Pressure Measurement
2-46 -46
2.29
Level Sensors
2- 47
2.29.1
Block DiagramLevel Sensors Air Separator
2- 47
2.29.2
Description Level Sensors Air Separator
2- 47
2.30
Red Detector
2- 47
2.30.1
Block DiagramRed Detector
2- 47
2.30.2
Description Red Detector
2- 47
2.31 2.31
Degassing Pre Pres ssure ure Me Mea asure urement and and Dialysa ialysate Pre Pres ssure ure Measurement
2-48
2.31.1
Block DiagramPressure Measurement
2- 48
2.31.2
Description Pressure Measurement
2- 48
2.32
Monitoring Analogue 12 V Voltage Supply
2- 48
2.32.1
Bloc lock Diag iagramMo Mon nitor itoring ing of An Analog logue 12 V Voltag ltage Supply
2- 48
2.32.2
Descripti iptio on Monito itoring ing of An Analog logue 12 V Volta ltage Supply
2-48 -48
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
B. Braun Medizintechnol zintechnologie ogie GmbH
2. Technical System Description
MT-MD-DE08C M.KAY
1 / 20 0 3
2- 4
2.33
Safety Air Detector SAD and Venous Red Detector RDV
2-49
2.33.1
Block DiagramSAD/RDV
2-49
2.33.2
General Information SAD
2-49
2.33.3
Description Safety Air Detector SAD
2-50
2.33.4
Description Venous Red Detector RDV
2-52
2.34
Blood Leak Detector
2-53
2.34.1
Block DiagramBlood Leak Detector
2-53
2.34.2
Description Blood Leak Detector
2-53
2.35
Controller Board CB
2-54
2.35.1
Block DiagramController Board
2-54
2.35.2
Description Controller Board
2-55
2.36
Power Board 1 PB1
2-57
2.36.1
Block DiagramPower Board 1
2-57
2.36.2
Description Power Board 1
2-57
2.37
Power Board 2 PB2
2-59
2.37.1
Block DiagramPower Board 2
2-59
2.37.2
Description Power Board 2
2-60
2.38
Heparin Pump Compact
2-62
2.38.1
Block DiagramHeparin Pump Compact
2-62
2.38.2
Description Heparin Pump Compact
2-62
2.39
Single Needle Cross Over
2-64
2.39.1
Block DiagramSN Cross Over
2-64
2.39.2
Description SN Cross Over
2-65
2.40
Staff Call (Option)
2-66
2.40.1
Block DiagramStaff Call
2-66
2.40.2
Description Staff Call
2-66
2.40.3
Operating Modes Staff Call System
2-67
2.40.4
Block DiagramAlarmMonitoring
2-67
2.40.5
Pin Assignment
2-67
2.41
Switch Mode Power Supply SMPS (Benning)
2-68
2.41.1
Block DiagramSwitch Mode Power Supply
2-68
2.41.2
SystemIntegration
2-69
2.41.3
Layout Switch Mode Power Supply SMPS
2-70
2.41.4
Wiring DiagramSwitch Mode Power Supply SMPS with Battery Option
2-71
2.41.5
Description Switch Mode Power Supply
2-72
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
MT-MD-DE08C M.KAY
1 / 20 0 3
2- 5
2.41.6
Pin Assignment Switch Mode Power Supply
2-72
2.42
Safety Concept
2-77
2.42.1
Block DiagramSafety Concept
2-77
2.42.2
Description Safety Concept
2-78
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 20 0 3
2- 6
with the options:
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
•
Double Pump
•
BIC Cartridge Holder
•
Central Concentrate Supply
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 20 0 3
2- 7
General Information Operation is accomplished via a touch screen (TFT monitor). Two microprocessor systems control and monitor the machine. The hardware concept consists of the following systems:
Top Level System
•
Top Level System
•
Low Level System
The top level systemconsists of the following components: • •
Communication module Top level controller TLC (motherboard)
•
Hard disk drive
•
Floppy disk drive
•
ABPM (option)
The communication between the user and the machine is performed via the top level. Example data exchange to communication module: •
Entry via input mask of the touch screen or keyboard
•
Output via the output mask of the TFT monitor
Example data exchange to low level: •
Low Level System
Transmitting and receiving of data from/to low level controller
The low level systemconsistsof the following components: •
Low level controller LLC
•
Supervisor SB
•
Power board PB1 and power board PB2
The low level controls and monitors all functions. Example data exchange to top level controller: •
Transmitting and receiving of data from/to low level controller
Example data exchange low level controller to supervisor: •
Transmitting and receiving messages, data and commands from/to supervisor
All sensors are connected to the processor systemvia the supervisor board. The actuators, motors and valves are driven via the power boards 1 and 2.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.1
1 / 20 0 3
2- 8
Overview Sub-Racks
Fig. : Overview Sub-RacksRear View Dialog+
2.1.1
Legend Overview Sub-Racks 1 2
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Top Level Sub-Rack UF Sub-Rack
3 4
DF Sub-Rack Water Inlet Sub-Rack
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.2
1 / 20 0 3
2- 9
Top Level Sub-Rack
1
2
Fig. : Top Level Sub-Rack
2.2.1
Legend Top Level Sub-Rack 1 2
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Hard Disk Drive Motherboard
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.3
1 / 200 3
2 - 10
UF Sub-Rack
Fig. : UF Sub-Rack
2.3.1
Legend UF Sub-Rack Air Separator LA Air Separator Valve VLA Balance Chamber BK1/2 Bypass Valve VBP Dialyser Inlet Throttle DD DDE Dialyser Inlet Valve VDEBK1 VDEBK1 Dialyser Inlet Valve VDEBK2 VDEBK2 Dialyser Outlet Valve Balance Chamber VABK1
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Dialyser Outlet Valve Balance Chamber VABK2 Dialyser Dialyser Outlet Valve Balance Chamber VD VDABK1 ABK1 Dialyser Dialyser Outlet Valve Balance Chamber VD VDABK2 ABK2 Inlet Valve Balance Chamber VEBK VEBK1 1 Inlet Valve Balance Chamber VEBK VEBK2 2 Membrane Position Sensor Balance Chamber MSBK1/2 Outlet Flow Pump FPA FPA Outlet Flow Pump Throttle RVFPA RVFPA
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.4
1 / 200 3
2 - 11
DF Sub-Rack
Fig. : DF Sub-Rack
2.4.1
Legend DF Sub-Rack BIC Pump BICP Bicarbonate Conductivity BICLF Bicarbonate Temperature Sensor T TSBIC SBIC Bicarbonate Throttle RVB Concentrate Pump KP Concentrate Throttle RVK RVK Degassing Chamber EK EK Degassing Control Valve RVE
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Degassing Pressure Sensor PE Degassing Pump EP Dialysate Temperature Sensor Supervisor T TS SD D-S -S Dialysate Temperature Sensor T TSD SD ENDConductivity/Supervisor ENDLF/ENDLF-S LF/ENDLF-S Inlet Flow Pump FPE Inlet Flow Pump Throttle RVFPE RVFPE UF Pump UFP
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.5
1 / 200 3
2 - 12
Water Inlet Sub-Rack RVDA Outlet Ausgang
H PE
EP
Fig. : Water Inlet Sub-Rack(with and without Heat Exchanger WTOption)
2.5.1
Legend Water Inlet Sub-Rack Degassing Chamber EK EK Degassing Control Valve RVE Degassing Pressure Sensor PE Degassing Pump EP Degassing Temperature Sensor T TSE SE Heat Exchanger WT (Option)
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Heater H Heater Inlet Temperature Sensor T TSHE SHE Heater Temperature Sensor T TSH SH Pressure Reducer Valve D DMV MV Upline Tank Inlet Valve VVBE VVBE Upline Tank VB
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.6
1 / 200 3
2 - 13
Rinsing Bridge
1
2
3
4
VD PDA VDE BL
Fig. : Rinsing Bridge
2.6.1
Legend Rinsing Bridge 1 Disinfection Valve VD 2 Pressure Sensor Dialysate Outlet PDA PDA
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
3 Dialyser Inlet Valve VDE 4 Blood LeakDetector BL
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.7
1 / 200 3
2 - 14
Rear Door
1 HOP
BIC-KV
PB2
PB1 SB 2 CB
SMPS
Fig. : Rear Door
2.7.1
Legend Rear Door BIC Cartridge Holder Board BIC-KV (Option) Controller Board CB HDF Online Power Board HOP HOP (Option) HFS 2 Board HFS2 with TSHE (Option) (2 2) Fan (1 1) Power Board 1 PB1
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Power Board 2 PB2 SAKA Board SAKA with TSHE (Option) (2 2) Switch Mode Power Supply SMPS Supervisor Board SB Heater Inlet Temperature Sensor Board T TSHE SHE (2 2)
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.8
1 / 200 3
2 - 15
TFT Monitor
Fig. : TFTMonitor
2.8.1
Legend TFT Monitor Backlight Inverter Board BIB Front Panel Board FPB TFT Monitor TFT
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Optical Status Display Board OSDB Touch Controller Board TCB Touch Screen
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.9
1 / 200 3
2 - 16
Power Board 1 PB1 V 7
C 1 2 6 C 3 7
V 2 0 R 7 5
C 1 4 8
R 7 1 V 5
V 2 2
V 2 1
U 1 3
V 3 4
C 5 0 V 3 3
C 3 9
R 3 1 2
P5
V 3 2
V 1 0
C 1 2 7 C 6 2
C 9 3 V 2 3 R 1 3 1
R 1 2 7
V 2 4
V 2 5
C 6 5 V 3 6 V 3 5
V 1 3
C 1 2 8
C 1 4 6
C 8 6
V 2 6 R 1 8 6
C 1 1 7
V 1 1
V 2 7
R 1 8 2
V 2 8
P7 U 1 5
V 3 8
C 8 9 V 4 0
C 3 4
V 3 9
V 4
C 1 2 5
C 1 5
C 1 2 4
U 1 4
V 3 7
C 6 4
C 1 4 7
P6
V 8
V 1 7 R 1 4
V 1 9
R 1
P1
V 1 V 1 8
U 1 2
V 3 1
C 1 8
V 3 0 V 2 9
C 1 7 P 4
P4
P3
P2 Fig. : Power Board 1 PB1
2.9.1
Legend Power Board1 PB1 P1 P2 P3 P4
MT-MD-DE08C M.KAY
Arterial BloodPump BPA BPA Supervisor Board SB SB Voltage Supply SensorsBlood Pump Cover switch
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
P5 P6 P7
Inlet Flow Pump FPE FPE Outlet Flow Pump FPA FPA Degassing Pump EP EP
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.10
1 / 200 3
2 - 17
Power Board 2 PB2
P3
V 3 2
V 3 3
V 3 4
V 3 5
P1
V 3 6
V 3 7
U 2 4 C 7 8
V 3 8
V 3 9
V 4 0
V 4 1
U 2 5
V 4 2
V 4 3
V 4 4
V 4 5 V 3 V 4 6
U 2 6 V 4 7
P5
C 9 3
P4
C 8 3
P2 Fig. : Power Board 2 PB2
2.10.1
LegendPower Board 2 PB2 P1
MT-MD-DE08C M.KAY
for Valves andSAKV: Upline Tank Inlet Valve RVVB RVVB Degassing Control Valve RVE VenousTubing Clamp Currentles Closed SAKV-SG Dialyser Outlet Valve VD A VDA Dialyser Inlet Valve VD VDE E Balance Chamber Valves VABK1/2, VDABK1/2, VEBK1/2, VEBK1/2, VDEB VDEBK1/2 K1/2 Air Separator Valve VLA Circulation Valve VZ VZ Bypass Valve VBP
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
P2 P3
P4 P5
Supervisor Board SB SB Piston Pumps: Concentrate Pump KP KP, Bicarbonate Pump BICP, BICP Ultrafiltration Pump UFP UFP Voltage Supply Disinfection Valve VD
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.11
1 / 200 3
2 - 18
Supervisor Board SB
P11 P12
P8
P19 K1
P5
P2
P9 14
P6 P7
P3
P18
P10
P23
P15 P17 P13
P1 P4 P16
Fig. : Supervisor Board SB
2.11.1
LegendSupervisor Board SB P1/P2/P3 P4 P5 P6 P7 P8 P9 P10 P1 1 P11 P12
MT-MD-DE08C M.KAY
Controller Board CB BICSS/KSS/RDVLED/SAD RTS/TX/CTS/RX Signals Staff Call Signals BPV/PBS/SAKA Signals BL Signals REM/HOFF/HREL Signals SCB-RX/TX/CTS/RTS Signals BP/EP/FP Signals VVBE/KP/BICP/UFP/SKAV/VDE/VDA Signals Heparin Pump Signals
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
P13 P14 P15 P16 P17 P18 P19 K1
Pressure Sensor PBE PBE Voltages +5 V/+24 VL/+24 VGB/+12 VAN/-12 VAN ENDLF/ENDLF-S/BICLF/TSD/TSD-S/TSBIC/TSESignals Voltages +5 VREF/+12 VAN/-12 VAN Voltages +5 V/12 VD/-12 VD Extension Connector Communication Program Adapter Hardware Switch: Position 0: Therapy Mode Position 2: TSM Service ProgramMode Position 3: Software Update Mode
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.12
1 / 200 3
2 - 19
Controller Board CB
P2
P5
U9
U2
U5
P3
U11
P1
U3
U12
U10
U6
U8
U22
U25
U18
U26
U1
U51
U33
Fig. : Controller Board CB
2.12.1
Legend Controller Board CB P1/P2/P3
MT-MD-DE08C M.KAY
Supervisor Board SB SB
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
P5
ProgramAdapter with Memory Card
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.13
1 / 200 3
2 - 20
Switch Mode Power Supply SMPS (Benning) LED +12VD / F401 - T1.25A TR5 +12VD / F402 - T5.00A TR5 +12VAN / F403 - T1.25A TR5 +5VD / F303 - T3.15A TR5 +5VD / F302 - T5.00A TR5 +5VD/ F301 - T3.15A TR5 +5VD / F304 - T3.15A TR5 +24 VGB/ F602 - T3.15A TR5 +24VGB/ F601 - T3.15A TR5
PE3 N3 L3 N2 L2
X2 F4 X3 X1.3 X1.1 X1.2 X1.6 X1.5 X1.4 X1 F3 F5 F6 F2 F1 PE
L
N
PE L1 N1
1
P12
F401 F402
3
F403
1
P10
F303 F302
3
+24V
1
F301 F304
F1 T6.25A 6.3x32 F2 T6.25A 6.3x32 F3 T3.15A TR5 F4 T3.15A TR5 F5/F6 - 110/120V: F20A 6.3x32 - 230V: M10A 6.3x32
P4 3
F602
1
F601
P5 3
1
F301 T3.15A TR5 F302 T5.0A TR5 F303 T3.15A TR5 F304 T3.15A TR5 F401 T1.25A TR5 F402 T5.00A TR5 F403 T1.25A TR5 F500 T3.15A TR5 F600 M10.00A 6.3x32 F601 T3.15A TR5 F602 T3.15A TR5
+
2
P13
P3
1 1
P9
1
3
4
P7
1 2
P8 1
X100
3
1
P2
1
P101 3
1
Fig. : Switch Mode Power Supply SMPS (Benning)
2.13.1
LegendSwitch Mode Power Supply SMPS X1 X2 X3 P2 P3 P4 P5 P6 P7
MainsInput, Heater Fluid Warmer (via Relay) Supervisor/Watchdog, Service Board Battery Connection (Screw Terminal) Power Board1/2 Supervisor/Controller Board Floppy Disk Drive Options, Service Board (Service Tool)
F1/F2 6.25 AT (6.3x32), Mains Input F3/F4 3.15 AT (TR5), Fluid Warmer +Monitor F5/F6 10 AM (6.3x32), Heater 1800 W (240 V) 20 AF (6.3x32), Heater 1800 W (110/120 V) F3 01 3.15 AT(TR5), +5 VD F301 F3 02 5.00 AT(TR5), +5 VD F302 F303/ 3.15 AT(TR5), +5 VD F304
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
P8 ABPM P9 Fan, Mains Switch P10 PC P11 Hard Disk Drive P12 Options P13 P14 EXT ON P101 Service Watchdog X1 X100 00 Fan F401 F401 F402 F403 F6 00 F600 F60 1/ F601/ F602
1.25 AT (TR5), +12 VD 5.00 AT (TR5), +12 VD 1.25 AT (TR5), +12 VAN 10 AM (6.3x32), +24 VL 3.15 AT (TR5), +24 VGB
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.14
1 / 200 3
2 - 21
Front Door HP
PV SN
HP
PA
PBS/SN
PBE
SN
PV SN
PBA
PBS/SN
PA SN
PBV
PBA
SAD
PBE
PBV
SAD
1 SAKV-SG
SAKA
2
SAKV-SG
SAKA
2
Fig. : Front Door
2.14.1
Legend Front Door Arterial Blood Pump BPA Arterial Pressure Sensor PA Arterial Tubing Clamp SAKA-SG SAKA-SG Cover for Suction Rods 2 Heparin Pump Compact HP Pressure Sensor PBE
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Pressure Sensor PBS/SN Safety Air Detector SAD SAD/Venous Red Detector RD V RDV Substitution Port 1 Venous Blood Pump BPV Venous Pressure Sensor PV Venous Tubing Clamp Currentles Closed SAKV-SG SAKV-SG
B. Braun Medizintechnologie GmbH
DM i T a l o g M + _ D s mD _ E C 0 h 8 a p C t M e r 2 . _ K 1 A -2 Y 0 0 3 . d o c / p d f < 0 1 1 0 0 3 > d d m m y y
PV
TSH
TSE
T
T
PA
PBE
BPA
RVE EP
EK
BPV
PA
PBE
Double-Needle
BPA
HP
HP
PE
SAD RDV
Water Sub-Rack Wasser-Einschub
H
PV
PBS
SAD RDV SAKA
SAKV-SG
Single-Needle
DFSub-Rack DF-Einschub
C NSVB
A
UF Sub-Rack UF-Einschub
Pyrogen Filter Pyrogenfilter
Rinsing Bridge Spülbrücke
DDE VVB
RVK
RVB
VDE
TSD TSD-S BK1
VDEBK1
TSBIC BICLF
VABK1
FPE
ENDLF ENDLF-S
VBP FBK1
LAB1
Z.D.
VDABK1
VEBK1 MSBK1
Dialysator
VBKO B
VDEBK2
VABK2
VEBK2
VDABK2
VLA
V.D.
LAFS
BL
FVD
PDA
MSBK2
RVDA
WT VVBE
WasserZU Water Inlet Luft Air Wasser Water Bicarbonat Bicarbonate Konzentrat Concentrate Dialysierfluß 1 DialysateFlow 1 Dialysierfluß 2 DialysateFlow 2 Dialysierflüssigkeit Dialysate Desinfektionsmittel Disinfectant Optionen Options
LAB2
Dialysat AB Dialysate Outlet
C
VZ
Zentrale BicarbonatVersorgung Central VBKS Bicarbonate FBIC Supply
Luft Air SBS1
UFP
KSS
RVFPA
D e s c r i p t i o n
S1
VD
FD
FK
BE
KE
BIC-Konzentrat BIC Concentrate
Heparin
KVA Zentrale KonzentratVersorgung Central Concentrate Supply
Disinfectant (Rear Side of Unit) Desinfektionsmittel (Geräterückseite)
Konzentrat Säurekonzentrat Concentrate Acid Concentrate
1 / 200 3
2.15.1 LegendFlow Diagram Abbreviation
Description
BE
Bicarbonate Withdrawal Rod
BICLF
Bicarbonate Conductivity
BICP
Bicarbonate Pump
BICSS
Bicarbonate Rinsing Connection Sensor
BK1
Balance Chamber 1
BK2
Balance Chamber 2
BL
Blood Leak Detector
BPA
Arterial Blood Pump
BPV
Venous Blood Pump
BVA
Bicarbonate Supply Connection (Central Supply)
DBK
Throttle BicarbonateCartridge Holder
DDE
Throttle Dialyser Inlet
DMV
Pressure Reducer Valve
EK
Degassing Chamber
ENDLF
END Conductivity
ENDLF-S
END Conductivity Supervisor
EP
Degassing Pump
FB
Filter Bicarbonate
FBIC
Filter Bicarbonate Cartridge
FBK1
Filter Balance Chamber 1
FBK2
Filter Balance Chamber 2
S2
1 / 2 0 0 3
2 2 2
2. Technical System Description
LA
SBS2
BICSS
Blut Arteriell Blood Arterial Blut Venös Blood Venous
FPA
B
2 . T e c h n i c a l S y s t e m
LVD
FB
BVA
BKUS
*
FBK2
BK2 FM
VBICP
DMV
Dialyser
KP
BICP
B . B r a u n M e d i z i n t e c h n o l o g i e G m b H
F l o w D i a g r a m
SAKV-SG
VB TSHE T
DBK
2 .1 5
2 - 23
2. Technical System Description
1 / 200 3
2 - 23
2.15.1 LegendFlow Diagram
MT-MD-DE08C M.KAY
Abbreviation
Description
BE
Bicarbonate Withdrawal Rod
BICLF
Bicarbonate Conductivity
BICP
Bicarbonate Pump
BICSS
Bicarbonate Rinsing Connection Sensor
BK1
Balance Chamber 1
BK2
Balance Chamber 2
BL
Blood Leak Detector
BPA
Arterial Blood Pump
BPV
Venous Blood Pump
BVA
Bicarbonate Supply Connection (Central Supply)
DBK
Throttle BicarbonateCartridge Holder
DDE
Throttle Dialyser Inlet
DMV
Pressure Reducer Valve
EK
Degassing Chamber
ENDLF
END Conductivity
ENDLF-S
END Conductivity Supervisor
EP
Degassing Pump
FB
Filter Bicarbonate
FBIC
Filter Bicarbonate Cartridge
FBK1
Filter Balance Chamber 1
FBK2
Filter Balance Chamber 2
FK
Filter Concentrate
FM
Flowmeter
FPA
Outlet Flow Pump
FPE
Inlet Flow Pump
FVD
Filter from Dialysate
H
Heater
HP
Heparin Syringe Pump
KE
Concentrate Withdrawal Rod
KP
Concentrate Pump
KSS
Concentrate Rinsing Connector Sensor
KVA
Concentrate Supply Connector (Central Supply)
LA
Air Separator
LAB1
Air Separator BIC Cartrige Holder 1
LAB2
Air Separator BIC Cartrige Holder 2
LAFS
Air Separator Level Sensors
LVD
Light Barrier Disinfection Valve
MSBK1
Membrane Position Sensor Balance Chamber 1
MSBK2
Membrane Position Sensor Balance Chamber 2
NSVB
Level Sensor Upline Tank
PA
Arterial Pressure Sensor
PBE
Pressure Sensor Blood Inlet
PBS
Blood Pressure Control Sensor
PDA
Pressure Sensor Dialysate Outlet
PE
Degassing Pressure Sensor
PV
Venous Pressure Sensor
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description RDV
Venous Red Detector
RVB
Throttle Bicarbonate
RVDA
Throttle DialysateValve
RVE
MT-MD-DE08C M.KAY
Throttle Flow Pump Outlet
RVFPE
Throttle Flow Pump Inlet
RVK
Throttle Concentrate
RVUF
Throttle Ultrafiltration
SAD
Safety Air Detector
SAKA
Arterial Tubing Clamp
SAKV-SD
Venous Tubing Clamp Currentles Closed
SBS1
Rinsing Bridge Connector Sensor 1
SBS2
Rinsing Bridge Connector Sensor 2
TSBIC
Bicarbonate Temperature Sensor
TSD
Dialysate Temperature Sensor
TSD-S
Dialysate Temperature Sensor Supervisor
TSE
Degassing Temperarture Sensor Thermal Fuse Heater Element
TSHE
Heater Inlet Temperature Sensor
UFP
Ultrafiltration Pump
VABK1
Outlet Valve Balance Chamber 1
VABK2
Outlet Valve Balance Chamber 2
VB
Upline Tank
VBICP
Bicarbonate Pump Valve
VBKO
Bicarbonate Cartridge Holder Top Valve
VBKS
Bicarbonate Cartridge Holder Concentrate Rod Valve
VBP
Bypass Valve
V.D.
Dialyser Coupling (from Dialysate)
VD
Disinfection Valve
VDA
Dialyser Outlet Valve
VDABK1
Dialyser Outlet Valve Balance Chamber 1
VDABK2
Dialyser Outlet Valve Balance Chamber 2
VDE
Dialyser Inlet Valve
VDEBK1
Dialyser Inlet Valve Balance Chamber 1
VDEBK2
Dialyser Inlet Valve Balance Chamber 2
VEBK1
Inlet Valve Balance Chamber 1
VEBK2
Inlet Valve Balance Chamber 2
VLA
Air Separator Valve
VVB
Upline Tank Valve
VVBE
Upline Tank Inlet Valve
VZ
Circulation Valve
WT
Heat Exchanger
Z.D.
Dialyser Coupling (to Dialysate)
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
2 - 24
Degassing Control Valve
RVFPA
TSH
1 / 200 3
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.16
1 / 200 3
2 - 25
Description Flow Diagram The flow diagramcan be divided into six sections: •
Water Inlet Section with Upline Tank
•
Degassing Circuit with Temperature System
•
Dialysate Processing
•
Balance Chamber
•
Ultrafiltration
•
Rinsing Bridge
2.16.1 Water Inlet Section with Upline Tank The water inlet section has the following components •
Pressure Reducer Valve DMV
•
Upline Tank Inlet ValveVVBE (2/2 way valve)
•
Upline Tank VB
•
Level Sensors Upline Tank NSVB
•
Heat Exchanger WT(Option)
Pressure Reducer Valve DMV
TSE
The pressure reducer valve DMV limits the pressure of the inlet water (e.g. osmosis water) KSS to a maximumof approx. 1.3 bar.
VB
NSVB
Upline Tank Inlet ValveVVBE
RVE
WT
VVBE
The valve VVBE is time-delayed controlled via the level sensor NSVB (top) in the upline tank VB. The delay time depends on the dialysate flow. Level Sensors Upline Tank NSVB
RVB VZ
The level sensors are mounted in the upline tank. NSVB top: closed
- VVBE is closed
NSV B bottom (monitoring low water level): NSVB closed (alarm) - Water inlet is disturbed - Heater is switched off Heat Exchanger WT(Option)
DMV RVDA
The cold inlet water can be warmed up via the optional heat exchanger WT. Thereby the heat consumption to heat up the water can be reduced.
Fig. : Water Inlet with Upline Tank
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 26
2.16.2 Degassing Circuit with Temperature System The degassing circuit with temperature systemhas the following components:
Degassing Control Valve RVE Degassing Pressure Sensor PE Degassing Pump E P EP
•
Degassing Control Valve RVE
•
Degassing Pressure Sensor PE
•
Degassing Chamber EK EK
•
Degassing Pump EP EP
•
Thermal Fuse Heater Element T TSH
•
Temperature Sensor Heater Inlet T TSHE
•
Heater Element H
•
Degassing Temperature Sensor T TS SE E
The control valve RVE, pressure sensor PE, degassing chamber EK and degassing pump EP produce and measure a negative pressure respectively. The negative pressure is produced to separate the dissolved gas fromthe water. The control valve RVE reduces the flow (throttle principle) depending on the measured pressure at the pressure sensor PE. Thereby the desired negative pressure is gained between the control valve RVE and the degassing pump EP. The value of the negative pressure is approx. -500 mmHg and thus always higher than the lower pressure of the dialysate behind the dialyser. The degassing pump works with constant speed, which is determined by the dialysate, unless the negative pressure is insufficient at the smallest opening of RVE. Then the speed of EP is increased.
Heater Element H Thermal Fuse Heater Element TSH Degassing Tem Temperature Sensor TS E TSE
The heater H has an integrated thermal fuse TSH as a thermal cut-off. The temperature sensor TSE measures the actual temperature posterior to the heater.
Temperature Control
The temperature of the water inlet determines the amount of heat which the heater must supply, to replace the amount of heat (dialysate flow and dialysate temperature) withdrawn by the drainage. The differential temperature between the heater inlet (TSHE) and the heater outlet (TSE) determines the controlled variable for the heater, depending on the dialysate flow and dialysate temperature.
TSE
TSH
T
T
VLA RVE EP
EK
PE
H
T zumVB/to VB
TSHE
vomVB/fromVB
Fig. : Degassing Circuit with Temperature System
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 27
2.16.3 Dialysate Processing The dialysateprocessing has the following components: •
Bicarbonate Concentrate Pump BICP
•
Bicarbonate Throttle RVB RVB
•
Bicarbonate Temperature Sensor T TSBIC
•
Bicarbonate Conductivity BICLF BICLF
•
Concentrate PumpKP
•
Concentrate Throttle RVK
•
ENDConductivity ENDLF ENDLF
•
ENDConductivity Supervisor ENDLFS ENDLFS
•
Dialysate Temperature Sensor T TS SD D
•
Dialysate Temperature Sensor Supervisor T TSD SDS
•
Inlet Flow Pump FPE
•
Inlet Flow Pump Throttle RVFPE RVFPE
The main components of the dialysate preparation are the bicarbonate concentrate pump BICP and the concentrate pump KP, with the conductivity cells BICLF and ENDLF and a flow pump FPE. The flow pump FPE delivers the dialysate. The bicarbonate concentrate, which is added via the bicarbonate pump BICP, is measured by the conductivity measurement cell BICLF. Thereby the pump can control the given conductivity set-point value. The concentrate or acid concentrate addition has the same working principle. The nonreturn valves RVB and RVK stablise the dosage of the bicarbonate and concentrate. The temperature sensors TSBIC and TSD are responsible for: •
the temperature compensation of the conductivity measurement and
•
temperature measurement TSD after the addition of cold concentrate (second measurement sensor for temperature system) and thus compensation of temperature loss.
The conductivity sensor ENDLFS is an independent monitoring unit (supervisor). The geometry of the ENDLFS sensor is different (but has the same cell constant) than the ENDLF sensor of the controller. Thereby a deposit on the sensor can be identified. The temperature compensation is carried out by the temperature sensor TSDS. The temperature sensor additionally monitors the dialysate flow temperature for the supervisor. The ENDLFS and TSDS sensors have no influence on the respective control. The throttle RVFPE prevents a high pressure build-up and thus a bursting of tubing if the flow path is blocked behind FPE. If the set pressure is reached RVFPE is opened and the fluid can circulate.
RVFPE
vomVB fromVB
ENDLF ENDLF-S
BICLF TSBIC
TSD TSD-S
T
T T
RVB
FPE
RVK
BICP
zu Bilanzierungskammern to Balance Chambers
KP vomBicarbonat fromBicarbonate
vomKonzentrat fromConcentrate
Fig. : Dialysate Processing
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 28
2.16.4 Central Bicarbonate and Concentrate Supply A canister or central supply can be selected via the bicarbonate and concentrate supply connetion BVA and KVA. Thesupply connection is an option. The flow pump FPE guarantees a continuous control of the desired dialysate flow into the balance chambers. The flow rate is determined by the filling time of the balancechamber. The flow pump FPE is controlled via the predetermined volume of the chamber and a continuous detection of the position of the membrane.
2.16.5 BIC Cartridge Holder DBK DBK DBK
LAB1
T Throttle Bicarbonate Bicarbonate Cartridge Holder DBK ensures a constant pressure (approx. 200 mmHg) during the filling of the bicarbonate cartridge. Air Separator Sep Separa arator tor BIC BIC Cartrige Cartrige Holder Holder 1 1 Cartrige LAB1 ensures that only fluid can enter the bicarbonate cartridge.
VBKO
Bicarbonate Bicarbonate Cartridge CartridgeHolde Top Valve Valve Holderr Top The bicarbonate cartridge is filled to the limit presure (200 mmHg) after VBKO opens.
VBKS
Bicarbonate Bicarbonate Cartridge Holder Holder Concentrate Concentrate Rod Valve Valve Concen trate Rod The bicarbonatecartridge is vented during preparation and in therapy, i.e. VBKO closes and VBKS opens for a short time. This is repeated in regular intervals during threapy. VBKS is opened after the end of the therapy to empty the bicarbonate cartridge.
VBICP
Bicarbonate BicarbonatePump Valve If VBICP is opened the liquid level in LAB2 is increased. VBICP switches the BIC pump in bypass after the end of the therapy to empty the bicarbonate cartridge.
LAB2
Air Separator Sep Separa arator tor BIC BIC Cartrige Cartrige Holder Holder 2 2 Cartrige LAB2 serves as a buffer chamber for the bicarbonate cartridge (and canister) to prevent conductivity malfunctions/deviations during therapy.
VVB
Upline Tank Va Valv lve e VVB cuts off the main flow after the end of the therapy to empty the bicarbonate cartridge via FPE (VBICP and VBKS are opened).
BICLF
ENDLF
TSBIC
vomVorlaufbehälter fromUplineTank
VVB
DBK
TSD RVK
RVB BICP
FPE
ENDLF-S TSD-S
KP VBICP
LAB1
LAB2
VBKO
VBKS BVA
FBIC
FB BE
Fig. : BIC Cartridge Holder (Option)
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 29
2.16.6 Balance Chamber System The balance chamber systemhas the following components: •
Balance Chamber BK1
•
Balance Chamber BK2
•
Balance Chamber Dialyser Inlet Valve VDEBK1 VDEBK1and VDEBK2
•
Balance Chamber Inlet ValveVEBK1 VEBK1and VEBK2
•
Balance Chamber Membrane Position Sensor MSBK1 MSBK1and MSBK2
•
Balance Chamber Dialyser Outlet ValveVDABK1 VDABK1and VDABK2
•
Balance Chamber Outlet ValveVABK1 VABK1 and VABK2
The measurement and control of the ultrafiltration rate is accomplished by the double balance chamber systemand the ultrafiltration pump UFP. Both balance chambers BK1 and BK2 are identical. The chambers have flexible membranes, which can be moved to both sides. The membranes devide the chambers into two sub-compartments. The flow direction is defined by the membranes and the eight solenoid valves. The position of the membranes is measured by inductive membrane position sensors MSBK1 and MSBK2. The membrane position sensors (ferrites) are connected to the membranes and each move in a respective coil MSBK1 and MSBK2.
2.16.7 WorkingPrinciple Balance Chamber System The balance chamber BK1 is filled with dialysate at the beginning of phase 1. The membrane is in right position. The valves VDEBK1 and VDABK1 are opened. The balance chamber BK1 is filled by the outlet flow pump FPA, via valve VDABK1. Simultaneously the dialysate is removed from the balance chamber BK1 via valve VDEBK1. Phase 1 is completed and the membrane is in left position (see figure).
Phase 1:
The balance chamber BK2 is filled with fresh dialysate during this period. The used dialysate from the previous phase 2 is drained (see description phase 2).
DDE
PE/RVE
VDE
FPE BK1
VDEBK1
VEBK1
MSBK1
VDABK1
VDEBK2
VABK2
VLA VDA
LA_FS
MSBK2
BL
VDABK2
VEBK2
BK2
RVDA
VBP
VABK1
VZ
FPA
UFP
PDA
LA RVFPA
open offen closed geschlossen
Fig. : Phase 1 Balance Chamber
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description Phase 2:
1 / 200 3
2 - 30
After phase 1 is completed there is an automatic switch to the filled balance chamber BK2 to obtain a constant flow in the dialyser. The complete cylce is repeated in phase 2, i.e. valves VDEBK2 and VDABK2 are opened. The balance chamber BK2 is filled via valve VDABK2. Simultaneously the dialysate is drained from the balance chamber BK2 via valve VDEBK2. Phase 2 is completed and the menbrane is in left position (see figure). Simultaneously the balance chamber BK1 is filled with fresh dialysate. Therefore valve VEBK1 is opened. Valve VABK1 is also opened, to initiate the flow path for the used dialysate to the drain. The membrane moves to the right position. The outlet fluid volume is equal to the returned fluid volume, due to the closed balance chamber system The fluid volume removed from the closed system via the ultrafiltration pump UFP is replaced from the blood in the dialyser and equals the precise ultrafiltration volume. The system is initialised in preparation, i.e. the membrane sensors are automatically calibrated and the speed of the flow pumps FPE and FPA are determined. Thus a synchronisation of the membranes is guaranteed, and the pump speeds for the desired flow are determined.
DDE
PE/RVE
VDE
FPE BK1
VDEBK1
VEBK1
MSBK1
VDABK1
VDEBK2
VABK2
VLA VDA
LA_FS
MSBK2
BL
VDABK2
VEBK2
BK2
RVDA
VBP
VABK1
VZ
FPA
UFP
PDA
LA RVFPA
open offen closed geschlossen
Fig. : Phase 2 Balance Chamber
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 31
2.16.8 Ultrafiltration and Rinsing Bridge The main flow path and bypass have the following components: •
Dialyser Inlet ThrottleDDE D DE DD
•
Dialyser Inlet Valve VDE VDE VD
•
Dialyser Outlet Valve VDA VDA
•
Bypass Valve VBP
•
Outlet Flow PumpFPA
The flow path for the main flow and bypass are determined by the valves VDE, VDA and VBP. The built up flow from the flow pump FPA is stabilised by the throttle DDE. Valves VDE and VBP are closed for sequential therapy (ultrafiltration without dialysate fluid flow). The ultrafiltrate removal is carried out by the ultrafiltration pump UFP. Further components are: •
Red sensitive blood leak detector BL
•
Pressure sensor PDA which monitors the dialysate pressure (also used to calculate TMP)
•
Air separator LA with built in level sensors LAFS and air separator valve VLA
•
Throttle RVDA functions as a resistance to stabilise the flow of FPE
The throttle RVFPE prevents a high pressure build-up and thus a bursting of tubing if the flow path is blocked behind FPE. If the set pressure is reached RVFPE is opened and the fluid can circulate. The fluid level is lowered in the air separator LA, due to air bubbles from the dialyser (degassing or possible leakages). The air separator valve VLA is opened if the fluid level is lower than the bottom level sensor. The fluid level is increased, due to the negative pressure for the degassing range, until the level reaches the upper level sensor of the air separator LA.
2.16.9 Disinfection and Cleaning Program The user can select a disinfection or cleaning program. The position of the couplings are checked by the sensors BICSS, KSS, SBS1 and SBS2. Then the UF pump UFP starts running and builds up a negative pressure against the closed disinfection valve VD. At approx. -200 mmHg VD opens and disinfectant is sucked in by the UFP. The circulation valve VZ is open and fluid flows into the upline tank VB, because the throttle RVDA acts as a forward resistance. Thereby a quicker heat-up in the hot cleaning program is achieved and thus a reduction of disinfectant. There is no flow of fluid to the drain during suction, heat-up and circulation.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.17
1 / 200 3
2 - 32
Block Diagram
Fig. : Block Diagram
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 33
2.17.1 LegendBlock Diagram Supervisor Board SB Controller Sensoren BICLF BICSS ENDLF KSS NSVB PBE PE
Bicarbonate Rinsing Connection Sensor END Conductivity Concentrate Rinsing Connector Sensor Level Sensor Upline Tank Pressure Sensor Degassing Pressure Sensor
SBS1
Rinsing Bridge Connector Sensor 1
SBS2
Rinsing Bridge Connector Sensor 2
TSBIC
Bicarbonate Temperature Sensor
TSD
Dialysate Temperature Sensor
TSE
Degassing Temperarture Sensor
Supervisor Sensors ENDLFS TSDS Controller/Supervisor Sensors/Actuators BKUS BL
END Conductivity Supervisor Dialysate Temperature Sensor Supervisor
BottomBicarbonate Sensor Blood Leak Detector
FEDFFS
DF Filter Detection Sensor
FEDHDFS
HD Filter Detection Sensor
LAFS
Air Separator Level Sensors
MSBK1
Membrane Position Sensor Balance Chamber 1
MSBK2
Membrane Position Sensor Balance Chamber 2
PA PDA
Arterial Pressure Sensor Pressure Sensor Dialysate Outlet
PSABFS
Port Substition Drain Sensor
PSAUS
Port Substition Outlet Sensor
PV
Venous Pressure Sensor
RDV
Venous Red Detector
SAD
Safety Air Detector
TSHE
Heater Inlet Temperature Sensor
VBE
Filter Vent Valve
VBICP
BIC Pump Valve
VBKS
BIC Concentrate Suction Rod Valve
VBKO
MT-MD-DE08C M.KAY
Bicarbonate Conductivity
Top BIC Cartridge Valve
VDFF
DF Filter Valve
VSAA
Substitution Connection Outlet Valve (drain)
VSAE
Substitution Connection Inlet Valve
VSB
Substitution Bypass Valve
VVB
Upline Tank Valve
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 34
Power Board 1 PB1 BPA EP
Arterial Blood Pump Degassing Pump
FPA
Outlet Flow Pump
FPE
Inlet Flow Pump
Power Board 2 PB2 BICP
Bicarbonate Pump
KP
Concentrate Pump
RVE SAKV-SG UFP
Degassing Control Valve Venous Tubing Clamp Currentles Closed Ultrafiltration Pump
VABK1
Outlet Valve Balance Chamber 1
VABK2
Outlet Valve Balance Chamber 2
VB
Upline Tank
VBP
Bypass Valve
VD VDA
Disinfection Valve Dialyser Outlet Valve
VDABK1
Dialyser Outlet Valve Balance Chamber 1
VDABK2
Dialyser Outlet Valve Balance Chamber 2
VDE
Dialyser Inlet Valve
VDEBK1
Dialyser Inlet Valve Balance Chamber 1
VDEBK2
Dialyser Inlet Valve Balance Chamber 2
VEBK1
Inlet Valve Balance Chamber 1
VEBK2
Inlet Valve Balance Chamber 2
VLA VVBE
Air Separator Valve Upline Tanke Inlet Valve
VZ
Circulation Valve
HP
Heparin Pump Compact
Heparin Pump Comfort Board SN-Crossover Board BPV
Venous Blood Pump
PBS
Pressure Single Needle
PBS-S
Pressure Single Needle Supervisor
SAKA
Arterial Tubing Clamp
Power Supply H TSH
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Heater Element Thermal Fuse Heater Element
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.18
1 / 200 3
2 - 35
TFT Monitor
2.18.1 Description TFT Monitor TFT Monitor The 15" TFT monitor (TFT=thin filmtransistor) has a resolution of 1024 x 768 XGA. The TFT housing can be swivelled. Keyboard The following settings can be performed via the keyboard: • Start/stop arterial blood pump BPA • Increase/decrease speed of arterial blood pump BPA • Acknowledge alarms • Acknowledge entries Display: +-
Battery Option Keys:
Decrease speed of arterial blood pump BPA Fig. : TFT Monitor witt Touch Screen
stop
+-
Start and stop arterial blood pump BPA
Increasespeed of arterial blood pump BPA
stop
Fig. : Keyboard Membrane
Acknowledgealarms
Acknowledge entries The
start stop
key has two integrated yellow LEDs. In both the
and keys two red LEDs are integrated. All keys switch a resistor on the supervisor communication board to ground GNDD.
Optical Status DisplaysOSD An optical status display OSD is integrated into the TFT housing (top left and right). The red LED is cyclically checked during therapy. The following operating statuses are displayed: Red: Alarm Red Yellow: Warning Yellow Green: Green Trouble-free operation
Touch Screen/Touch Controller Board TCB The touch screen has a resolution of 4096 x 4096 with a 4-wire to technique and has an RS 232 interface (9600 Baud).
Fig. : Rear TFT Monitor
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 36
Backlight Inverter Board BIB
The backlight inverter board drives four lamps for the TFT monitor.
Front Panel Board FPB
The front panel board has five keys. If a key is pressed a signal is generated for the TLC and LLS. Three LEDs are integrated in each key. Thus the signals for the TLC are generated. The alarm acknowledge key additionally generates a signal for the LLS. A charge LEDis integrated on the FPB for the battery option. The volume can be set with a potentiometer. A signal is generated for the TLC. The LLS monitors the signal via the current and the pulse. The signals are generated for the TLC and LLS for the optical status displays OSDs. The signals for the brightness of the LEDs (OSD) are generated for the TLC. The signal (brightness for the TFT) for the backlight inverter board BIB is generated for the TLC. The signal for the parallel port is generated for the TLC. The signal for the RS 232 interface is generated for the LLS.
Loudspeaker
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
A loudspeaker for audible alarms is either integrated in the basic housing. The volume can be set with a potentiometer on the FPB (TFT housing, rear bottom left).
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.19
1 / 200 3
2 - 37
ABPM Option (None-Invasive Blood Pressure Measurement) ABPM Option
A none-invasive blood pressure measurement is possible with the ABPM option (automatic blood pressure measurement). The ABPM option works on an oscillometrical basis for the automatical control of the symptomatic hypertonia during dialysis. The ABPM option can be retrofitted in the dialysis machine.
ABPM Module/ABPM Interface Board Dialog+
The ABPM option consistsof the ABPM module and the ABPM interface board. The ABPM option will be assembled in the basic housing (left side).
VoltageSupply
The ABPM module is connected to the switch mode power supply via the ABPM interface board (connector P8).
Multi I/O
The ABPM module is connected to the motherboard (COM3 port) via the ABPM interface board.
2.19.1 Wiring DiagramABPM Option
ABPM Module ABPM-Modul Floppy Disk Drive Diskettenlaufwerk
left sidebasic housing linke Seite Grundgehäuse
P2
ABPM Interface Board ABPM Interface-Board
P1
Motherboard GX1LCD Switch Mode Power Supply SMPS Schaltnetzteil SMPS
1 COM4 COM2
COM3 1 1 COM1
P8
Fig. : Wiring Diagram ABPM Option
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.20
1 / 200 3
2 - 38
bioLogic RR Option (Automatic Blood Pressure Stabilisation) The bioLogic RR option can be installed and activated subsequently. The ABPM option must be present in the Dialog+to run the bioLogic RR option. The software for the bioLogic RR option is installed via an installation diskette. The diskette is automatically marked (assigned to the machine) during installation and subsequently can only be used for this specific Dialog+ machine. •
Installation diskette for bioLogic RR for SW ≥ 6.20
Note If the TLC software has to be reinstalled or the hard disk drive has to be replaced: Activate the option bioLogic RR, i.e. use the bioLogic RR installation diskette which belongs to the respective machine and activate again the option.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.21
1 / 200 3
2 - 39
Supervisor Board SB
2.21.1 Supervisor 2.21.1.1 Block DiagramSupervisor
E2PROM
RAM
SpeicherKarte Memory Card
Digitale Eingänge
Daten/Adressbus Data/Address Bus
Digital Inputs Digitale Eingänge Digital Inputs
CPU 80C535
Analoge Eingänge Analog Inputs Digitale Ausgänge Digital Outputs
ACIA DIABUS
Serielle Schnittstelle Serial Interface
TLCB
Serielle Schnittstelle Serial Interface Datenaustausch Controller Data Communication Controller
LLCB
E2PROM Digitale Ausgänge
Seriell Serial
Digital Outputs
Reset Generator Fig. : Block DiagramSupervisor Board
2.21.1.2 Description Supervisor The supervisor board connects the low level controller LLC with the peripheral. Additionally the supervisor (monitoring microprocessor) is integrated on the board. The following components are assembled on the board. The sensors are directly connected: • • • • • • • • • • • • • • •
Bicarbonate Conductivity Measurement END Conductivity Measurement Controller ENDConductivity Measurement Supervisor DegassingTemperature Measurement Bicarbonate Temperature Measurement Dialysate Temperature Measurement Controller Dialysate Temperature Measurement Supervisor Level Sensors Upline Tank Reed Contacts Rinsing Bridge Venous Pressure Sensor Arterial Pressure Sensor Level Sensors in the Air Separator Red Sensors Degassing Pressure Measurement Blood Inlet Pressure Measurement
Additionally the following components are available for monitoring and control: • •
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Monitoring of analogue 12 V supply voltage Synchronisation of actual pump values
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 40
The signal processing for the low level controller and supervisor are on the supervisor board. All sensors are connected to this board via plugs. The power board BP1 and PB2 are also connected to this board. Signals for single sensors of the processor system are also processed separately.
Microprocessor
The supervisor has the following components: •
80C535 CPU
•
E2PROM
Connection for:
80C535 CPU
•
Memory Card
•
Digital Inputs and Outputs
•
Analogue Inputs
•
Serial Interface
•
Communication with the Controller
•
Reset Generator
The supervisor CPU (central processing unit) has the following data: •
8 Bit Processor
•
256 x 8 RAM (internal)
•
6 8 Bit I/O Ports
•
3 16 Bit Counter
•
1 Serial Interface with max. 9600 Baud
•
8 Bit AD Converter with 8 Multiplexer Inputs
Address Decoding Logic
The processor is equipped with a 32 kB external RAM and a 1 MB E2PROM. The programcode is stored in the E2PROM. The RAM is used to store data.
Serial E2PROM
The serial E2PROM stores the supervisor sensor calibration data. The data is stored during calibration in the TSM service program and loaded before a therapyis activated.
Digital Inputs andOutputs
Additional signal memory (latches) for the in- and outputs are available, because the implemented ports of the 535 processor are limited. The outputs haveopen colectors. All inputs have TTL level.
Analogue Inputs
The processor has an internal 8 bit AD converter. The input voltage range is 0 to 5 V. The supervisor monitors the analogue sensors conductivity, temperature and pressure, these are directly connected with the inputs.
Serial Interface
The serial interface is used for the communication with the front panel board FPB (via the SUPBUS). The transfer rate is 9600 baud. The interface works in full duplex mode with V 24 level. An additional ACIA (asynchronous communications interface adapter) is implemented for the communication with the top level controller via the DIABUS. The ACIA works in full duplex mode with 19200 baud and 24 V level.
Communication with the Controller
The communication with the controller is realised by a parallel interface. The interface has signal memory (latches).
Reset Generator
The reset generator resets the processor after the supply voltage is switched on. Thereby the programcan start at a predefind address.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.22
1 / 200 3
2 - 41
Bicarbonate Conductivity Measurement
2.22.1 Block DiagramBicarbonate Conductivity Measurement +5 VREF Referenzspannung Reference Voltage
BICLF-A Sendeelektrode Transmitter Electrode
CLK4 Takt 4000 Hz Cycle 4000 Hz
Tiefpaß Low-Pass
Geschalteter Gleichrichter Switch Rectifier
Verstärker Amplifier
BICLF-C Empfangselektrode ReceiverElectrode
BIC-LF Ausgangssignal Output Signal Fig. : Block Diagram Bicarbonate Conductivity Measurement
2.22.2 Description Bicarbonate Conductivity Measurement The conductivity of the dialysate is determined by a resistance measurement. The measurement is performed by an alternating current with approx. 4 kHz. The calibration is accomplished by the controller. The measurement cell has two transmitter electrodes and a receiver electrode with a fixed cell constant. •
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
A transmitter voltage for the transmitter electrode BICLF_A signal is generated from the +5 VREF reference voltage by the CLK4 signal (4000 Hz).
•
The signal runs through the fluid.
•
The received BICLF_C signal is amplified.
•
The switched rectifier converts the a.c. voltage to a d.c. voltage BIC_LF.
•
A d.c. voltage, which is proportional to the conductivity is fed to the AD converter.
•
The temperature compensation and linearisation of the conductivity is performed by the controller.
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.23
1 / 200 3
2 - 42
END Conductivity Measurement Controller
2.23.1 Description ENDConductivity Measurement Controller The design of the END conductivity measurement controller is identical with the bicarbonate conductivity measurement in paragraph 2.22.
2.23.2 ENDConductivity Measurement Supervisor 2.23.3 Block DiagramEND Conductivity Measurement Supervisor
Taktgenerator 4000 Hz
Referenzspannung
Pulse Generator 4000 Hz
Reference Voltage ENDLF-S-A Sendeelektrode
Transmitter Electrode
END-S-LF Ausgangssignal Output Signal
Geschalteter Gleichrichter Switch Rectifier
Verstärker Amplifier
ENDLF-S-C Empfangselektrode Receiver Electrode
Fig. : Block DiagramENDConductivity Measurement Supervisor
2.23.4 Description ENDConductivity Measurement Supervisor The END conductivity measurement of the dialysate by the supervisor is in principle identical with the controller. The differences are: • •
The independent generation of the reference voltage and the clock signal Dimension of the Measurement Cell
The calibration is performed by the supervisor software.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.24
1 / 200 3
2 - 43
Temperature Measurement
2.24.1 Block DiagramDegassing Temperature Measurement
+5 VREF Referenzspannung Reference Voltage
U/I Wandler U/I Converter
Offset
TSE, TSHE, TSBIC, TSD Ausgangssignal Output Signal
PTC
Verstärker Amplifier
Tiefpaß Low-Pass
Fig. : Block DiagramDegassing Temperature Measurement
2.24.2
Design Degassing Temperature Measurement The temperature of the fluid is measured by a PTC resistor (PTC positive temperature coefficient). •
The PTC has a constant current flow of <0.1 mA
•
The voltage drop is measured and amplified in a differential amplifier
•
An offset voltage is added to lift the zero point. Thereby the measurement range of the ADconverter has an optimal working condition.
•
The output voltage is fed to the AD converter on the supervisor board via a low-pass filter.
2.24.3 Description Bicarbonate Temperature Measurement The design is identical with the degassing temperature measurement in paragraph2.24.1.
2.24.4 Description Dialysate Controller Temperature Measurement The design is identical with the degassing temperature measurement in paragraph2.24.1.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 44
2.24.5 Block DiagramDialysate Supervisor Temperature Measurement
ReferenzSpannung Reference Voltage
U/I Wandler
PTC
U/I Converter
Offset
Verstärker Amplifier
Tiefpaß
TSD-S Ausgangssignal Output Signal
Low-Pass
Fig. : Block Diagram Dialysate Supervisor Temperature Measurement
2.24.6 Description Dialysate Supervisor Temperature Measurement The temperature measurement of the supervisor has an independent reference voltage source. Thus a cross-interference with the temperature sensors of the controller is excluded.
2.25
Level Measurement
2.25.1 Block DiagramLevel Measurement Upline Tank
+5 V
NSVBO NSVBU Fig. : Block Diagram Upline Tank
2.25.2 Description Level Measurement Upline Tank The level sensor in the upline tank has two reed contacts. The contacts are switched by an internal magnet in a float ball. Pull-up resistors are on the input. The query is carried out by the digital inputs of the controller.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.26
1 / 200 3
2 - 45
Reed Contacts
2.26.1 Block DiagramCoupling Status
+5 V
SBS1 SBS2 KSS BICSS Fig. : Block Diagram Query Reed Contacts
2.26.2 Description CouplingStatus Magnets are integrated in the dialysate couplings und concentrate couplings. If the couplings are connected the reed contacts are switched. The query is performed by the digital inputs of the controller and supervisor. The inputs have pull- up resistors.
2.27
Pressure Measurement
2.27.1 Block DiagramVenousPressure Measurement
+5 VREF
+1,2 VREF
Offsetspannung
+5 V
Offset Voltage
Differenz Verstärker Differential Amplifier
Begrenzer Ausgangssignal PV Limiter ReferenceVoltage
Druckaufnehmer Pressure Sensor Fig. : Block Diagram Venous Pressure Measurement
2.27.2 Description VenousPressure Measurement The pressure sensor has a resistance bridge. The resistance value changes in accordance with the present pressure value. A constant +5 VREF is connected to the bridge.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
•
The measurement signal is tapped and amplified in the differential amplifier.
•
An offset voltage is added to lift the zero point. Thereby the measurement range of the AD converter has an optimal working condition.
•
The voltage is limited to +5 V by a clamp circuit on the output, in the event of a fault condition. A damage of the following circuit components is thereby prevented.
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 46
2.27.3 Block DiagramArterial Pressure Measurement +5 VREF
+2,5 VREF
Offsetspannung Offset Voltage
Differenz Verstärker Differential Amplifier
+5 V
Begrenzer Limiter
Ausgangssignal Reference VoltagePA
Druckaufnehmer Pressure Sensor Fig. : Block Diagram Arterial Pressure Measurement
2.27.4 Description Arterial Pressure Measurement The design of the arterial pressure measurement is identical with the venous pressure measurement in paragraph 2.27.2, with the exception of the offset voltage.
2.28
Blood Inlet Pressure Measurement
2.28.1 Block DiagramBlood Inlet Pressure Measurement
+5 VREF
+1,8 VREF
Offsetspannung Offset Voltage
Differenz Verstärker Differential Amplifier
+5 V
Begrenzer Limiter
Ausgangssignal Reference VoltagePBE
Druckaufnehmer Pressure Sensor Fig. : Block Diagram Blood Inlet Pressure Measurement
2.28.2 Description Blood Inlet Pressure Measurement The design of the blood inlet pressure measurement is identical with the venous pressure measurement in paragraph 2.27.2, with the exception of the offset voltage.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.29
1 / 200 3
2 - 47
Level Sensors
2.29.1 Block DiagramLevel SensorsAir Separator
Hochpaß High-Pass
Hochpaß
LAFSO-EL
Gleichrichter Rectifier
Tiefpaß Low-Pass
Komparator Comparator
LAFSO
LAFSU-EL
Gleichrichter Rectifier
Tiefpaß Low-Pass
Komparator Comparator
LAFSU
Oszillator Oscillator High-Pass
Fig. : Block Diagram Level Sensors
2.29.2 Description Level Sensors Air Separator The query of the fluid level in the air separator behind the balance chamber is accomplished by the conductivity of the dialysate. •
The oscillator generates an a.c. voltage with a frequnecy of approx. 8 kHz.
•
The d.c. voltage part is removed by a high-pass.
•
The output voltage is fed to the electrode.
•
2.30
If the electrode immerses into the fluid due to an increase of the fluid level, the voltage at the electrode is decreased.
•
The voltage is rectified for evaluation and smothed by a low-pass
•
This d.c. voltage is compared with a reference voltage by a comparator.
Red Detector
2.30.1 Block DiagramRed Detector
LEDAnsteuerung
CLK4
Grün Green
LEDDrive
RDV
Ausgangssignal Output Signal
Tiefpaß Low-Pass
Verstärker
Empfänger Receiver
Amplifier
Fig. : Block DiagramRed Detector
2.30.2 Description Red Detector The red detector works as a light barrier with a triggered 4 kHz green light. •
The received signal fromthe photo transistor is amplified.
•
The signal is only evaluated, if the green LED is also driven.
•
External interferences are thereby prevented.
The venous red detector RDV has an additional preamplifier.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.31
Degassing Pressure Measurement and Dialysate Pressure Measurement
2.31.1
Block DiagramPressure Measurement +5 VREF
+2.5 VREF
Offsetspannung Offset Voltage
Differenz Verstärker Differential Amplifier
1 / 200 3
2 - 48
+5 V
Begrenzer Limiter
Ausgangssignal Reference VoltagePE/PDA
Druckaufnehmer Pressure Sensor Fig. : Block Diagram Pressure Measurement
2.31.2
Description Pressure Measurement The respective pressure sensor has a resistance bridge. The resistance value changes in accordance with the present pressure value. A constant +5 VREF is connected to the bridge.
2.32
•
The measurement signal is tapped and amplified in the differential amplifier.
•
An offset voltage is added to lift the zero point. Thereby the measurement range of the AD converter has an optimal working condition.
•
The voltage is limited to +5 V by a clamp circuit on the output, in the event of a fault condition. Thereby a damage of the following circuit components is prevented.
Monitoring Analogue 12 V VoltageSupply
2.32.1 Block DiagramMonitoring of Analogue 12 V VoltageSupply
+12 VAN
Überwachung
P12ANOK
Monitoring -12 VAN
M12ANOK
Fig. : Block Diagram Analogue 12 V Voltage Supply
2.32.2 Description Monitoring Analogue 12 V VoltageSupply The analogue +12 VAN and -12 VAN supply voltages are monitored by two voltage monitoring components against deviation. The components have an internal reference voltage source. Thereby the connected voltage can be constantly monitored. The P12ANOK and M12ANOK signal are switched at the output, if the voltage is lower than 10.8 V.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.33
1 / 200 3
2 - 49
Safety Air Detector SAD and Venous Red Detector RDV
2.33.1 Block DiagramSAD/RDV
Fig.: Block Diagram Safety Air Detector SAD/RDV
2.33.2 General Information SAD Decrementation of Detected Air Volume
Flow Dependent AlarmLimits
If an alarm was not previously activated, the detected and added air volume from both the LLC and LLS systemis decremented with a rate of 1 µl s-1 Thus air bubbles which may have gathered in the venous bubble catcher and are accumulated in larger time intervals, do not automatically lead to an alarm, because the detected air volume is continuously subtracted by 1 µl s-1. The alarmlimits for the air alarmdepend upon the flow through the detector: • • •
50 ..... 200 ml min-1 blood flow =0.2 ml 200 ... 400 ml min-1 blood flow =0.3 ml 400 ... 850 ml min-1 blood flow =0.5 ml
Example -1 • set blood flow =250 ml min • no substitution -1 − Flow through the sensor =270 ml min : -1 -1 (270 ml min =250 ml min set blood flow +20 ml min-1 additional flow possible by external infusion devices, e.g. HDF − Alarm limit value =0.3 ml air
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
Event Chart
1 / 200 3
2 - 50
Example 200 ... 400 ml min-1 blood flow =0.3 ml: • • •
at t =0 an air bubble with 0.2 ml is detected after t =3 min a second air bubble with 0.2 ml is detected after time t =4 min a third air bubble with 0.2 ml is detected
Ergebnis: t [min]
Detected Air Volumein Software Counter [ml]
AlarmLimit Value [ml}
SADAlarm yes/ no
-1 0 1 2 3 4 6
0 0 +0,2 air 0,14 0,08 0,02 +0,2 air 0,16 +0,2 air 0 after acknowledgement
0,3 0,3 0,3 0,3 0,3 0,3 0,3
no no no no no yes no
After t = 3 min a second air bubble with 0.2 ml is detected. An alarm is not activated, because the first detected air volume is decremented with a rate of 1 µl s-1, during t =3 min =180 s (i.e. 0.18 ml). After t =4 min an SAD alarm is activated, because the limit of 0.3 ml is exceeded..
2.33.3 Description Safety Air Detector SAD Principles of Air Monitoring
The monitoring of air is performed by ultrasonic transmitting between two piezo elements. The piezo elements work as trasmitter and receiver. The transmitter piezo element transmits pulses. These pulses are transmitted through the tubing, which is filled with blood (air), to the receiver piezo element. The filled tubing is the transmitting path. If the tubing is filled with blood the transmitting signal is only slightly attenuated (so-called coupling resistance). If the tubing is filled with air the transmitting signal is attenuated very strongly. The received amplitude is evaluated. Thus the condition (attenuation) of the transmitting path can be derived. This received amplitude is compared with an alarmthreshold: •
Blood in tubing (no attenuation): Amplitude of receiver signal >alarm threshold.
•
Air in tubing (large attenuation): Amplitude of receiver signal
If air is in the tubing the alarm threshold is not reached and the SAD indicates air in system.
SAD Function
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
The ultrasonic transmitting pulse is triggered cyclically. The transmitting pulse is variably attenuated, depending on the air concentration in the tubing. The received voltage amplitude is compared with an alarm threshold. If the voltage drops below the alarm threshold the SAD and SAD_S signals of the flip-flops are set. The SAD signal is checked cyclically by the low level controller LLC and then reset by the SADRESET signal. Simultaneously the low level supervisor LLS checks cyclically the SAD_S signal and is then reset by the SADRESET_S signal. The LLC checks cyclically the function of the SAD during operation by switching from the alarm threshold to the test threshold. LLC and LLS expect an air signal after the activation of the test threshold. An air signal is present at the flip-flop SAD_S, due to the cyclic test, until the SADRESET_S signal resets the air signal. The LLS monitors cyclically if at least air was detected once in 1.5 s. If air was not detected an SAD alarm is activated.
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 51
SADCalibration
The test threshold and alarm threshold for the SAD are calibrated in the TSM service program. In dialysis the test threshold is cyclically checked to guarantee that the sensitivity of the SAD has not been reduced compared to calibration. Thus a reduction of sensitivity is detected caused by operation and ageing. The alarm threshold is cyclically measured (250 ms) by the LLS during therapy and compared with the fixed alarm threshold. The calibration threshold is the maximum sensitivity which can be reached. An individual calibration value is set in the TSM service programfor the calibration ex works.
Control Logic
In the control logic the signals are linked for programming the shift register and to generate the process control for the SAD. Additionally the transmitter pulses are triggered and the receiver window is closed or opened.
Divider and Oscillator 3.6864 MHz
A quartz oscillator with a divider generates the necessary clock signals.
Transmitter
The tramsmitter piezo element oscillates via an induction voltage (piezo effect) and transmits ultrasonic waves.
Receiver
The transmitted pulses from the transmitter are converted into a voltage in the receiver piezo element (piezo effect). The receiver voltage is fed to the comparator input.
Programming of Shift Registers 1 and2
The shift registers are selected by the SADSEL signal to enable dataloading. The USDI signal is loaded into the shift register by the clock pulse signal USCLK. The outputs of the shift registers are fed to the low-passes.
Shift Register 1 andLow-Pass1 (AlarmThreshold)
The output of the shift register 1 controls the control input of the multiplexer. A pulse width modulated voltage PWM is present at the output of the shift register 1. The low-pass 1 smooths the voltage which is fed to the multiplexer. This voltage is the alarmthreshold for the comparator.
Shift Register 2 andLow-Pass2 (Test Threshold)
The output of the shift register 2 controls the control input of the multiplexer. A pulse width modulated voltage PWM is present at the output of the shift register 2. The low-pass 2 smooths the voltage which is fed to the multiplexer. This voltage is fed to the comparator, during the cyclic SAD test.
Comparator
The comparator compares the receiver voltage with the alarm threshold. The receiver voltage is smaller than the alarm threshold if the transmitter signal is attenuated, due to air in the tubing. Together with the function of the receiver window the output signal of the comparator (to the flip-flop) becomes logic 1. The receiver signal is larger than the alarm threshold if blood is in the tubing. The output signal of the comparator (to the flip-flop) becomes logic 0.
SADSignal for LLC
The SD signal is fed to the flip-flop and sent to the LLC as SAD signal. The SAD signal is checked cyclically by the LLC. The LLC resets the flip-flop via the SADRESET signal. If air is detected at any time during a cycle the SAD output changes to logic 0. Thereby the LLC evaluates the complete last cycle as air.
SAD_S Signal for LLS
The SD signal is fed to the flip-flop and sent to the LLS as SAD_S signal. The SAD_S signal is checked cyclically bythe LLS. The LLS resets the flip-flop via the SADRESET_S signal. If air is detected at any time during a cycle the SAD_S output changes to logic 0. Thereby the LLS evluates the complete last cycle as air.
Test SAD and SAD_S
The voltage of the low-pass 2 is switched cyclically to the comparator output by the USTEST signal. Thus the comparator, the coupling and the dynamic of the circuit is checked. The alarm threshold and the test threshold (voltage of low-pass 2) is switched by the multiplexer.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 52
SADREF Signal
The amplifier is a differential amplifier with follow-up impedance. The difference between the voltage at the comparator input and the voltage at the voltage divider is amplified. LLS ckecks cycliccally (250 ms) the alarm threshold with the calibrated alarm threshold during therapy. If the value is out of limits a reference alarmis activated.
TIMECONTRSignal
The quartz oscillator monitors the SAD sampling frequency. If the control logic and quartz oscillator work correctly the divider is reset by the SADX signal. The counter output becomes logic 1 if the SADX signal is not present or the time between two reset pulses is too long. Thereby a drop below the SAD sampling frequencyis reported to the LLS. The TIMECONTR signal becomes logic 1 and inhibits the CLK input of the counter. Thereby an overflow of the counter and a removal of the TIMECONTR alarmis prevented. Furthermore the SADRESET signal of the LLC is monitored. If the SADRESET signal becomes logic 0, due to an error, the SADX signal cannot reset the counter. This leads to a TIMECONTR alarm.
2.33.4 Description Venous Red Detector RDV Assignment and Function RedDetector
The venous red detector RDV detects blood with a certain concentration. The RDV assists the user during the connection of the venous line to the patient and after therapy during disconnection with sodium chloride bags. The amplified RDV signal is processed on the supervisor board and sent to the TLC via the DIABUS.
Principle of Venous RedDetector RDV
The RDV consistsof a transmitter (green LED) and a receiver (phototransistor). A tubing is between transmitter and receiver. Blood, colourless turbid fluid or air can be in the tubing. The RDV generates a signal to detect red fluid or colourless fluid. The function principle of the RDV is based on the fact that blood absorbes all colours with the exception of red. The green light is attenuated if blood enters the RDV. Blood or colourless fluid can be distinguished with the phototransistor and the comparator.
RDV Signal
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
The receiver signal of the phototransistor is amplified. The RDV signal is fed to the supervisor board SB for further signal processing. The signal is fed to the low level controller LLC which provides the signal to the top level controller TLC.
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.34
1 / 200 3
2 - 53
Blood Leak Detector
2.34.1 Block DiagramBlood Leak Detector
BLRG
Auswahl rote oder grüne LED Selection red or green LED
Pulsweitensignal BLPWM Pulse-Width Signal
LEDAnsteuerung LED Drive
Ausgangssignal BLREC Output Signal
Tiefpaß Low-Pass
Rot Red
Grün Green
Verstärker Amplifier
Fig. : Block DiagramBlood Leak Detector
2.34.2 Description Blood Leak Detector The drive and evaluation circuit is mounted on the sensor head. The amplified signal of the receiver is fed to the supervisor board. The blood leak detector works with red and green light. The transmitter LEDs are mounted opposite to the receiver diodes (transmitted light). The drive and evaluation is carried out by the controller. The actual value is also fed to the supervisor.
MT-MD-DE08C M.KAY
Drive Principle
The drive of the LEDs is performed alternately by the BLPWM signal. The red LEDis used to level any turbidity.
Calibration
For calibration the red and green LED are driven by the BLRG signal, until both have identical output voltage. If a turbidity occurs, both drive signals are increased. If the dialysate is stained by blood only the green LED signal is attenuated. Theoutput signal is reduced and blood is thereby detected.
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.35
1 / 200 3
2 - 54
Controller Board CB
2.35.1 Block DiagramController Board
3
EEPROM
RTC A/D
MFP
n
o i ß t u a l n h i c s m r b e a T s u s B u
+5 VG P / REM 2 P
MUX
B n o i ß t u l a n h i c s m r b e a T s u s u B B
50 MHz
/2 K L C U P C
s u B a t a D t i B 6 1
K L C U P C
SCC
CIO
/ 1 P 16 Analoge Leitungen (P1) g u l 16 Analog Lines(P1) P l a 8 Leitungen (P3) n 8 Lines(P3) g i S 3 P / 2 P / 1 P r e k c e t s l a n g i 20 Leitungen/Baustein (P2) S 20 Lines/Component (P2)
s l e l a a n n g i g i S s l r e o r u t e t n o S C
Steuerlogik Control Logic
Puffer Buffer
RAM 512 K
CPU
Flash 521 K n
r r e e f f f f u u B P
24 Bit Address Bus
o i ß t u l a n h i c s m r b e a T s u s u B B n
32 Bit Data Bus
o i ß t u l a n h i c s m r b e a T s u s B u B
Erweiterungsstecker P4 Extension Plug P4
Fig. : Block Diagram Controller Board CB
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 55
2.35.2 Description Controller Board Microprocessor
CPU
The low level controller board has the following components: •
CPU
•
Clock Generator
•
Control Logic
•
Memory (RAM/Flash EPROM)
•
Counter/Timer and Parallel I/O Ports CIO
•
Serial Communication SCC SCC
•
AD Converter AD
•
Multiplexer MUX
•
System Clock Generation MFP
•
E2PROM
•
Real Time Clock RT RTC C
•
Extension Connection, i.e. for Second Microprocessor
The 68020 CPU (central processing unit) has the following data: •
32 Bit Processor
•
16 MB Address Range
•
32 Bit Data Bus (external)
•
24 Bit Address Bus
The 32 bit data bus is connected unbuffered to all memories and is shielded against the remaining peripheral by drive components. The data bus is terminated by special terminator components. Thereby the signals are terminated and in case of a tristate condition the last level is held. The 24 bit address bus is buffered by drive components and is terminated by diodes, to prevent reflections. All control and clock lines that are connected to several consumers are terminated by RC elements.
Clock Generation
The clock generation for the microprocessor and control logic is generated from a 50 MHz oscillator. A D flip-flop generates two opposite phase clock signals of 25 MHz. Thereby the necessary phase accuracy and rise times for the microprocessor is achieved. For peripheral components a 4 MHz and a 2 MHz are generated by a second 8 MHz oscillator. All clock lines are terminated.
Control Logic
The control logic is responsible for decoding and access control. All signals required for decoding are unbuffered and come directly from the CPU and from an other microprocessor. The access control implements wait cycles depending on the addressed component.
Memory (RAM/Flash EPROM)
The memory has four flash EPROMs and four static RAMs. The memories are connected parallel to the data bus. Both memory work with a 32 bit data bus.
Counter/Timer and Parallel I/O Ports (CIO)
Eight parallel ports are available as CIO components (counter/timer/parallel inputs and outputs). The CIO components have the following parts: •
two 8 Bit Ports
•
one 4 Bit Port
•
three 16 Bit Timers
The ports can be operated in different modes.
Serial Communication (SCC)
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
A SCC component (serial communication controller) is available for serial communication. The two channel controller can handle asynchronous and synchronous protocols.
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 56
AD Converter AD
The analogue signals in the range of 0 to 5 V are converted by a 2 MHz triggered AD converter. A 16 channel multiplexer MUX is connected prior to the AD converter. All 12 bits are transmitted simultaneously to the DIABUS by a read command.
Multiplexer MUX
The inputs of the multiplexer are connected to the signal plug P2. The input voltage range is 0 to 5 V. The inputs are protected against over voltages of ±20 V.
SystemClock Pulse Generator MFP
A multi function peripheral component MFP is available for the internal system clock generation. The MFP has the following parts: •
three 8 bit timers
•
one serial channel
•
eight in/out ports
A timer is pulsed with 3.68 MHz for the baud rate of the serial channels. Two timers are in series with the time basis of the software operating system. The serial interface is reserved for debugging information. The real time clock RTC and the serial E2PROM are connected to the port pins.
E2PROM
MT-MD-DE08C M.KAY
Specific parameters of the boards are stored in the serial E2PROM.
Real TimeClock RTC
The alarm output of the real time clock RTC is directly connected to plug P3. The mains power supply can thereby be switched on automatically. The voltage supply is buffered by +5 VG in the mains power supply.
Extension Port
The available extension port is related to the signals parallel to the microprocessor. A second microprocessor could have access to the board via this port.
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.36
1 / 200 3
2 - 57
Power Board 1 PB1
2.36.1 Block DiagramPower Board 1
EC-Motor Ansteuerung EC Motor Drive
Sollwert
Set Value
UND-Glied AND Element
MotorRegler Motor Controller
Tiefpaß Low-Pass
UND-Glied AND Element
Stop
Istwert Actual Value
BPADS
Treiber Drive
Komparator Comparator
UND-Glied AND Element
Meßwiderstand
EC-Motor EC Motor
PGND
Measurement Resistor
Tiefpaß Low-Pass NICHT-Glied NOT Element nur BPA only BPA NICHT-Glied NOT Element
NICHT-Glied NOT Element Komparator Comparator
BPAIMP
MotorAdapter Motor Adapter
Hallsensoren
Hall Sensors
PumpenklappenSchalter Pump Cover Switch GabellichtSchranke Light Barrier
Fig. : Block DiagramPower Board 1
2.36.2 Description Power Board 1 Drive Circuit for Pumps
The following drive circuits for the d.c. motors are on power board 1: •
Arterial Blood Pump BPA
•
Degassing Pump EP EP
•
Outlet Flow PumpFPA
•
Inlet Flow Pump FPE
The drives are controlled by the controller board via the supervisor board. The output signals are evaluated by the supervisor and controller board.
EC Motor with Hall Sensors
The electronic commutator d.c. motor has three hall sensors. The position of the magnetic rotors are detected with the hall sensors. The three phase motionless coils are fed with current according to the position of the rotors. The motor adapter detects the revolution. The speed dependent output signal is pulse width modulated. The sensitive degree and frequency are speed proportional. The signal is fed to the motor controller via a passive and active low-pass filter.
Asynchronous Actual Speed BPAIST_US Signal of Arterial Blood Pump
The BPAIST_US signal is generated by a NOT element from the speed dependent output signal of the motor adapter.
Stop BPASTOP Signal of Arterial Blood Pump
The EC motor can be stopped by the BPASTOP signal or by opening the pump cover. If the BPASTOP signal has logic 1, the EC motor is stopped.
Cover Switch BPADS Signal of Arterial Blood Pump
A permanent magnet is integrated in the pump cover. If the pump cover is closed a reed contact is closed. This signal is linked with the BPASTOP signal after a comparator and fed to the motor controller. If the pump cover is opened the motor is switched off.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 58
BPAIMP Signal of Arterial Blood Pump
The supervisor monitors the rotation of the arterial blood pump BPA via a slot disc. Theslot disc runs in a fork light barrier. The post comparator generates the BPAIMP signal.
BPA Signal of Arterial Blood Pump
The set-point value for the speed of the EC motor is available as a pulse width modulated (PWM) BPA signal. The BPA signal is converted to an analogue voltage by a low-pass filter and fed to the motor.
Current Limitation of EC Motor
A voltage which is proportional to a coil of the EC motor is present at a measurement resistance. A comparator compares this voltage with a reference voltage. The reference voltage is equivalent with the maxium motor current. The voltage of the measurement resistance is fed to a comparator via a lowpass filter. The low-pass filter prior to the comparator prevents an activation of the current limitation during short period load peaks of the EC motor.
Signals Input Signals
Description
BPA
Set-point value for motor speed of arterial blood pump
PWM signal
FPE
Set-point value for motor speed of inlet flow pump
PWM signal
FPA
Set-point value for motor speed of outlet flow pump
PWM signal
EP
Set-point value for motor speed of degassing pump
PWM signal
PUMP STOP Start/stop signal
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Signal Type
Logic 1=Stop
HS1
Hall sensor signal (integrated in motor)
-
HS2
Hall sensor signal (integrated in motor)
-
HS3
Hall sensor signal (integrated in motor)
-
Output Signals
Description
-IST
Proportional frequency signal for motor speed
-
BPADS
Pump cover position of arterial blood pump (BPA)
Logic 1=cover closed
BPAIMP
Processed signal of forklight barrier
-
L1
Coil control
-
L2
Coil control
-
L3
Coil control
-
Signal Type
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.37
1 / 200 3
2 - 59
Power Board 2 PB2
2.37.1 Block DiagramPower Board 2
2-Wege-Ventile 2 Way Valve Ventil Valve
NICHT-Glied NOT Element
UND-Glied AND Element
VS
NICHT-Glied NOT Element
UND-Glied AND Element
UND-Glied AND Element
Vref Meßwiderstand Measurement Resistor
Regelventil Entgasung (RVE) Degassing Control Valve (RVE) NICHT-Glied NOT Element
Strombegrenzung Current Limiter Komparator Comparator
UND-Glied AND Element
NICHT-Glied NOT Element
VS
NICHT-Glied NOT Element
Strombegrenzung Current Limiter
UND-Glied AND Element
Treiber Drive
PGND Ventil Valve
Vref Meßwiderstand Measurement Resistor
Ventil-Vorlaufbehälter-Eingang VVBE (2-Wege-Ventil) Upline Tank Inlet Valve VVBE (2 Way Valve) VVBE
Ventil Valve
Strombegrenzung Current Limiter Komparator Comparator
RVE
Treiber Drive
UND-Glied AND Element
Treiber Drive
PGND
Ventil Valve
Strombegrenzung Current Limiter Komparator Comparator
Vref Meßwiderstand Measurement Resistor
Venöse Schlauchabsperrklemme Venous Tubing Clamp SAKV
NICHT-Glied NOT Element
Treiber Drive
SAKV_S
NICHT-Glied NOT Element
Treiber Drive
PGND
Ventil Valve
Schrittmotor KP/BICP/UFP Stepper Motor KP/BICP/UFP F-schritt Isel
NICHT-Glied NOT Element
GAL
NICHT-Glied NOT Element
MotorRegler Motor Controller
Meßwiderstand /Measurement Resistor Meßwiderstand /Measurement Resistor
Motor PGND PGND
Fig. : Block DiagramPower Board 2
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 60
2.37.2 Description Power Board 2 Circuit Power Board 2 Drives
Drive circuits are on the power board 2 for the following actuators: •
15 2 way solenoid valves
•
one proportional control valve
•
one tubing clamp
•
threebipolar stepper motors
The 2 way valves are used e.g.: •
To control the inlet/outlet of the balance chambers
•
To switch the dialysateflow between main and by-pass
The proportional valve controls the degassing pressure The stepper motors are used for: •
Concentrate pump
•
Ultrafiltration pump
The drives are controlled by logic signals from the supervisor and controller board.
Driver 2 Way Valve
Each valve has a current limitation (mono-flop) integrated in the driver. The current limitation can be activated/deactivated by the VS signal. The valve current is determined indirectly via a voltage drop over a measurement resistor. The measurement resistor is in series with the valve coils. The voltage is compared with a reference voltageat a comparator. If the valve current reaches a limit value the output signal of the comparator switches and triggers a mono-flop. The current is cut-off for a short time, thus preventing an over load of the valves.
Driver 2 Way Valve VVBE
The driver for the upline tank inlet valve VVBE has an additional watchdog. The watchdog is triggered by a negative edge pulse of the VS signal. If the signal is not switched the valve closes after approx. 1.5 s. Thus an over flow of the upline tank is prevented.
Driver Proportional Valve
The proportional valve is driven by a pulse width signal. The current in the valve coil is controlled. The design of the current limitation is identical with the 2 way valve.
Driver Tubing Clamp
The SAKV and SAKVS signals for the tubing clamp are fed to switch transistors via an inverter. The switch transistors can close the clamp independently.
Driver Stepper Motor
The drive for the stepper motor has a GAL (generic array logic) and an integrated motor controller. The GAL generates the logic signals for the motor controller fromthe frequency signal of the stepper motor and the control signal of the coil current. The motor controller has an internal phase current limiter. The motor coil is thereby constant.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 61
Signals Input Signals
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Signal Types
SAKV
Switch signal for tubing clamp (controller)
Logic 1=Clamp open
SAKVS
Switch signal for tubing clamp (supervisor)
Logic 1=Clamp open
VB1
Switch signal for 2 way valve
Logic 1=Valve closed
VB2
Switch signal for 2 way valve
Logic 1=Valve closed
VB3
Switch signal for 2 way valve
Logic 1=Valve closed
VLA
Switch signal for 2 way valve
Logic 1=Valve closed
DV
Switch signal for 2 way valve
Logic 1=Valve closed
DV1
Switch signal for 2 way valve
Logic 1=Valve closed
VE1
Switch signal for 2 way valve
Logic 1=Valve closed
VE2
Switch signal for 2 way valve
Logic 1=Valve closed
VA1
Switch signal for 2 way valve
Logic 1=Valve closed
VA2
Switch signal for 2 way valve
Logic 1=Valve closed
VDE1
Switch signal for 2 way valve
Logic 1=Valve closed
VDE2
Switch signal for 2 way valve
Logic 1=Valve closed
VDA1
Switch signal for 2 way valve
Logic 1=Valve closed
VDA2
Switch signal for 2 way valve
Logic 1=Valve closed
Current limitation for 2 way valve
Logic 1=Valve current not limited
RVVB
Control proportional valve
PWM signal
RVE
Control proportional valve
PWM signal
BICP
Control Stepper motor
Frequency signal
KP
Control Stepper motor
Frequency signal
UFP
Control Stepper motor
Frequency signal
IBICP
Control stepper motor
Logic 1=large Coil current
IKP
Control stepper motor
Logic 1=large Coil current
IUFP
Control stepper motor
Logic 1=large Coil current
VS
MT-MD-DE08C M.KAY
Description
Output Signals
Description
Signal Type
ventil-IST
Actual value switch condition of 2 way valve
Logic 1=Valve closed
schrittPOS
Reed contact signal on the axis of the stepper motor
-
-Q11
Connection for stepper motor coil
-
-Q12
Connection for stepper motor coil
-
-Q21
Connection for stepper motor coil
-
-Q22
Connection for stepper motor coil
-
ventil-
Connection for valves
-
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.38
1 / 200 3
2 - 62
Heparin Pump Compact
2.38.1 Block DiagramHeparin Pump Compact
+24 VGB HP HPMORI HPISEL HPHALT DREHS FORMS
SchrittMotor Stepper Motor
MotorRegler Motor Controller
GAL
HPIST HPR HPKRALO Lichtschranke Kolbenplatte Light Barrier Piston Panel
KOLBS
HPKOLB
Fig. : Block Diagram Heparin PumpCompact
2.38.2 Description Heparin Pump Compact Stepper Motor
The stepper motor is controlled by a motor controller and a GAL (generic array logic). The logic signals for the motor controller are generated from: •
Direction of rotation HPMORI signal
•
Step frequency HP signal
•
Coil current HPHALTand HPISEL signal
The stepper motor works in a half step mode. An internal phase current limiter of the motor controller is kept constant depending of the sensor resistors. The current of the stepper motor can be set by the HPISEL and HPHALT inputs. The internal stop signal of the motor is linked with the following signals: •
Positive locking signal FORMS
•
Direction of rotation signal HPR
•
Stop signal HPHALT
The motor control is switched off if HPHALT is active or the positive locking is opened and the direction of rotation is set to closing.
Speed Recognition
After processing the DREHS signal with a schmitt-trigger (signal of the light barrier for detection of rotation) it is fed to a GAL. The level is only transferred with a pulse from HP. Thus mechanical vibrations of the slot disc are filtered, and the output signal is HPIST.
Plunger Plate, PositiveLocking
The plunger plate signal KOLBS and the positive locking signal FORMS for the light barrier signals are processed with a schmitt-trigger and are available as output signals HPKOLB and HPKRALO.
Syringe Sizes
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Syringe sizes of 10, 20 and 30 ml can be used in the heparin pump Compact.
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 63
Signals Signal HP
Set-point frequency for stepper motor
HPR
Direction of rotation via slot disc of stepper motor Setting of the phase current for the stepper motor driver
HPHALT
HPIST HPKRALO
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Description
Signal of the light barrier at motor Status of claw
HPKOLB
Status of plunger plate
HPMORI
Direction of rotation of stepper motor
HPISEL
Setting of the phase current for the stepper motor driver
Signal Type Step is performed at falling edge Logic 1=Drive running Logic 1=Stop=Phase current stepper motor =0 (INH of TCA3727) Updated only with HP pulse Logic 1=Positive locking is open Logic 1 =Plunger plate ist pressed Logic 1 =Drive opening Logic 0=100% current Logic 1=66% current
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.39
1 / 200 3
2 - 64
Single Needle Cross Over
2.39.1 Block DiagramSN Cross Over
+24 VGB
BPV
MotorRegler Motor Controller
UND-Glied PWM Tiefpaß Low-Pass AND Element UND-Glied AND Element
BPVSTOP
Endstufe End Stage Meßwiderstand
Komparator mit Tiefpaß Comparator with Low-Pass
UND-Glied AND Element
Tiefpaß Low-Pass
GNDL
Measurement Resistor
V ref MotorAdapter Motor Adapter
PWM
Tiefpaß (passiv)
Low-Pass (passive)
NICHT-Glied NOT Element
BPVIST-US Komparator Comparator
Pumpenklappenschalter mit Reedkontakt Pump Cover Switch with ReedContact
Komparator Comparator
BPVDS
EC-Motor EC Motor Hallsensoren Hall Sensors
Schlitzscheibe Slot Disk
Komparator Comparator
BPVIMP
Offsetspannung
+5 VREF
+5 V
Offset Voltage
Differenz Verstärker Differential Amplifier
Begrenzer Limiter
Druckaufnehmer Pressure Sensor
PBS Offsetspannung
+5 VREF
+5 V
Offset Voltage
Differenz Verstärker Differential Amplifier
Begrenzer Limiter
Druckaufnehmer PressureSensor
PBS_S
+12 VAN 5 V Referenzdiode 5 V Referance Diode
Impedanzwandler Impedance Converter
+5 Vref
+24 V Arterielle SchlauchAbsperrklemme Arterial Tubing Clamp
SAKA
FET-LeistungsSchalter FETPower Switch
Fig. : Block Diagram SN Cross Over
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Techni cal System Description
1 / 200 3
2 - 65
2.39.2 2.39.2 Description SN Cro Cross Over EC Motor with with Hall Sensors
Th The electro tronic commutator tor d.c. motor tor has thr three hall sensors. The positio ition n of the the magnetic rotors rotors are detected wit with the hall hall sensors. The three phase motionl oti onles ess coils are fed with current according to the position of the rotors. The motor adapter detects the the revolut olutiion. The speed dependent output signal signal is is pulse width width modulated. The sensitive tive degree and freque frequency are speed proporti proportional. The sign ignal is fed to the the motor contro trolle ller via a passive and activ tive low low-pa -pass filt filte er.
Stop BPVSTOPSignal Signal of Venous Blood Blood Pump
The EC motor tor can be stop topped by the the BPVSTOP sign ignal or or by opening ing the the pump cover. If If the BPVSTOP signal has logic logic 1, the EC motor is stopped.
Cover Switch BPVDS Signal Signal of Venous Blood Blood Pump
A permanent magnet is is integrated integrated in the the pump cover. If If the pump cover is is close closed a reed contact is closed. This his signal is is linked linked with with the the BPVSTOP signal after after a comparator and fed to the motor control controlller. If If the the pump cover is opened the motor is switched off.
BPVIMP VIMP Signal of VenousBlood Blood Pump
The superviso isor monito itors the the rotatio tion of th the venous bloo lood pump BPV via a slot lot disc. Theslot disc runs in in a fork fork li light barrier. barrier. The post post comparator parator generates the BPVIMP signal.
BPV Signal Signal of VenousBlood Blood Pump
The set-po t-point int value lue for for the the speed of th the EC motor tor is available as a pulse lse width idth modulated odulated (PWM) BPV signal. signal. The BPV signal is converted rted to an analogue voltage by by a low-pa low-pass ss fil filter and fed to the the motor.
Current Limitati Limitation of EC Motor
A voltage which is proportional to a coil of the EC motor is present at a measurement resist resistance ance. A comparator compares this this voltage voltage with with a reference voltage voltage. The reference reference voltage voltage is equiva quivalent wit with the maxium motor current. current. The volta ltage of the the measurement re resista istance is fed to a comparator tor via a low lowpass pass filter. The The low-pa low-pass ss filter prior prior to the the comparator parator prevents prevents an activa activation tion of the current limitation during short period load peaks of the EC motor.
PBS andPBSSSignal Signals s of Singl Single e Needle dle Pressure Sensors
Each pressure sensor has a resistance resistance bridge bridge. The resistance resistance value changes in in accordance wit with the the present pressure val value. A constant +5 VREF is is connected to the bridge. •
SAKA Signal of Arterial Arterial Tubing Clamp
MT-MD MT-MD- DE08C M.KAY
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
The measurement sign ignal is tapped and amplifie lified in the the diffe ifferentia tial amplifi lifie er
•
An off offset voltage voltage is added to li lift the the zero point. point. Thereby the measurement range of the the AD converter rter has an optim timal working working condit condition.
•
The volta ltage is limite ited to +5 V by a clam lamp circ ircuit on the the output, in the the event of a fault fault condi condition. tion. Thereby a damage age of the the foll following circuit circuit components is prevented.
The arterial ial tub tubing ing clam lamp is activ tivated by the the SAKA sign ignal via a FET tra transisto istor.
B. Braun Medizintechnol zintechnologie ogie GmbH
2. Techni cal System Description
2.40
1 / 200 3
2 - 66
Staff Call (O (Optio tion)
2.40.1 2.40.1 Block lock DiagramStaff taff Call
PERSR
>1
PERSR- S PERSR-HU SR-HUP
P5 BR STAT BR: Soldering ringBridg Bridge Lötbrücke
P6 BR
1 sec
DYN
Voltage Monitoring Spannungsüberwachung
>1
1 sec
5
P7 BR DYNAUS
1
K1
3
Fig. : Block Diagram Staff Call
2.40 2.40.2 .2 Descrip criptio tion n Staff taff Call Operati rating Modes
The operatin ting modes static tic with ithout OFF alar larm, dynamic with ithout OFF alarm or dynamic with OFF alarm are set with soldering bridges on the staff call board. The defau fault settin tting g ex works is static tic with ithout OFF alar larm.
P5BR:
static without OFF alarm (default)
P6BR:
dynamic without without OFF alarm
P7BR:
dynamic wit with OFF alarm
Static without Off Alarm
The relay lay K1 switc itches as lon long as one of th the thr three inp inputs PERSR-S, -S, PERSR-HU -HUP or PERSR are acti active ve.
Dynamic withou without Off Alarm
If at lea least one input input PERSR-S, R-S, PERSR-HUP SR-HUP or PERSR PERSR changes from from active ctive to inactive nactive the relay relay K1 is switched switched for 1 second. The mono-f ono-flop U1 generates a switch switch tim time of one second.
Dynamic with Off Off Alarm
If at lea least one input input PERSR-S, R-S, PERSR-HUP SR-HUP or PERSR PERSR changes from from active ctive to inacti nactive ve the relay K1 is swit switche ched d for 1 second second or if the +12 + 12 VD VD voltag oltage e drops more than 10 %. The mono-f ono-flop U1 generates a switch switch ti time of one second. The supply supply vol voltage for the staff call circuit circuit is decoupl coupled by the diode V1 from the +5 V volt voltage and buffered buffered by a 1 F capacit capacitor C1.
MT-MD MT-MD- DE08C M.KAY
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
B. Braun Medizintechnol zintechnologie ogie GmbH
2. Techni cal System Description
1 / 200 3
2 - 67
2.40.3 2.40.3 Operating Mod Modes Staff Call System tem
/ AQ key to reset the alarm in the Press the operati operating mode static static without OFF alarm.
Unit ON/G N/Gerät EIN EIN Alarm
static
without OFF alarm larm ohneAUS-Alarm AUS-Alarm
dynamic
without OFF alarm larm ohneAUS-Alarm AUS-Alarm
dynamic with OFF alarm mit AUS-Alarm AUS-Alarm
Alarm Operation Betrieb
1s
Alarm Operation Betrieb
1s
Alarm Operation Betrieb
1s
Fig. : Operating Modes Staff Call System
Signals: PERSR- S
Staff call input of low level supervisor LLS
PERSR
Staff call input of low level controller LLC Staff call input of power supply
PERSR-HUP SR-HUP
(This li line is also acti active ve if the mains power suppl supply buzzer is is switched on or iif f there is is no WDlevel change in 0.7 s.)
2.40.4 2.40.4 Block DiagramAlarm AlarmMonitor itoring Legend:
Top Level Controll Controller TLC A- S
PERSR-T
WD
A
AKAL
PERSR-S R-S
A-S A-S:
LLS alarms (by data transm transmission)
AKAL:
LLC audible alarm
AKAL-S: AL-S:
LLS audible audible alarm alarm
transmiss ission ion) PERSR-T: SR-T: TLC staff call (by data tra
Low Level Controller LLC
Low Level Supervisor LLS AKAL-S L-S
transmission) A: LLC alarms (by data transm
PERSR
PERSR:
LLC staff call (hardware (hardware))
PERSR- S:
LLS staff call (hardware (hardware)
Voltage suppl upply staff call (hardware) are) PERSR-HUP: Vol PERSR
Power Supply PS
PERSRHUP
WD:
Staff Call Board PR
LLS watchdog connection connection (hardware)
Fig. : Block Diagram Alarm Monitoring Staff Call
2.40.5 2.40.5 Pin Assign ignment
5 green/grün
5 1
1
1brown/braun
AlarmTa Alarm Table:
3
3 white/weiß
TopView View Draufsicht
Cable Kabel
3 Medical Device medizin. zin. Gerät
5
The pin assign ignment is shown in the the fig figure.
Alarm Operation
Connector
Cable
1- 3 3- 5
white-brown white-green
Fig. : Pin Pin Assignment Staff Call Connector Connector
MT-MD MT-MD- DE08C M.KAY
Dialog+ Dialog+_sm_Chapter 2_12_1-2003.doc/pdf 2003.doc/pdf <011003>ddm 003>ddmmyy
B. Braun Medizintechnol zintechnologie ogie GmbH
DM i T a l o g M + _ D s mD _ E C 0 h 8 a p C t M e r 2 . _ K 1 A -2 Y 0 0 3 . d o c / p d f < 0 1 1 0 0 3 > d d m m y y
F i g . : B l o c k D i a g r a m S w i t c h M o d e P o w e r S u p p l y S M P S
2 2 .4 .4 1 1 .1
F1 T6.25A
Filter (EMC) Filter
L Mains Netz
N
(EMV)
F2 T6.25A
PE
L1
F5 M10A (230V) F20A (110V)
Heater Heizung N1
F6 M10A (230V) F20A (110V)
F5
F6
K4
DC
K3
0V
F_bat Battery 24V Akku 24V (Option)
PE1 F3 T3.15A
DC
Power Factor Correction PFC
+5VD +5VD
20..30Vdc
+5VD
I_bat
F4 T3.15A
ForwardConverter Abwärtswandler
B l o c k D i a g r a m S w i t c h M o d e P o w e r S u p p l y
S w i t c h M o d e P o w e r S u p p l y S M P S ( B e n n i n g )
+5VD 0V
D e s c r i p t i o n
+12VD
GND
0V +12VD 0V
L3
B . B r a u n M e d i z i n t e c h n o l o g i e G m b H
DM i T a l o g M + _ D s mD _ E C 0 h a 8 p C t M e r 2 . _ K 1 A -2 Y 0 0 3 . d o c / p d f < 0 1 1 0 0 3 > d d m m y y
+12VA 0V
Fluid N3 Warmer
-12VA
PE3
Signals Signale 20
Primary Primär
1 / 2 0 0 3
F301 - Power Board 1/2 - Hard Disk Drive - Floppy Disk Drive - Ext. - P14 F302 - Supervisor/Controller F303 - Motherboard F304 - ABPM - P7 - P8
1
Secondary Sekundär
2 6 8
F i g .: S y s t e m
Battery Akku
LC Display
I n t e g r a t i o n
2 .4 1 .2 S y s t e m I n t e g r a t i o n
LVDS
BIB
Backlight Inverter Board
LowVoltageDifferential Signal Adapter
TCB Touch Controller Board
F
TLC
Battery Option Akku-Option
Top Level Controller (GX1LCD Motherboard)
FPB
Charge LED Lade LED
Front Panel Board
FluidWarmer
Fan Mains Switch Lüfter Netzschalter
L N PE
L N PE
X3
Mains Input Netzeingang X1
B . B r a u n M e d i z i n t e c h n o l o g i e G m b H
-12VD
0V
Logic Logik
2 . T e c h n i c a l S y s t e m
L N PE
X2
L
Heater Heizung
SMPS Power Board Primary/ Primär
N
SMPS Power Board Secondary/ Sekundär
W 0 0 8 1
Power
P3 P9
Power
T1 K1
24 V Control 24 V Steuerung
Logic Logik Aux. Voltage Hilfspannung
T4 Thermal Fuse Thermosicherung
24V
Message Output MeldungenAusgang
2 x 1F
Heater / Heizung
5V Buzzer Summer Uamax
U Control U-Regler
P2 Heater Control Input
G K O N DHeizungssteuerung Eingang I L C R D D _ _ P _ N G H H H D
Drive for Battery, Buzzer, Clock Ansteuerung für Akku, Summer, Uhr
PFail
P2
P101 Watchdog
+5VD +12VD -12VD +12VAN -12VAN 24VL 24VGB 24VGD
P2
Machine Monitoring Input Geräteüberwachung
N F E S F _ S F _ F O D L L U O _ 4 M N A A K 4 D E G K K 2 K 2 B R / D A A / D A W
B24OK D24OK /PERSR_N /AKKU_OK PF PGD +5VG
2 . T e c h n i c a l S y s t e m D e s c r i p t i o n
1 / 2 0 0 3
2 6 9
DM i T a l o g M + _ D s mD _ E C 0 h 8 a p C t M e r 2 . _ K 1 A -2 Y 0 0 3 . d o c / p d f < 0 1 1 0 0 3 > d d m m y y
F i g .: S y s t e m I n t e g r a t i o n
2 .4 1 .2
Battery Akku
LC Display LVDS
BIB
Backlight Inverter Board
LowVoltageDifferential Signal Adapter
TCB
F
TLC
Touch Controller Board
Battery Option Akku-Option
Top Level Controller (GX1LCD Motherboard)
FPB
Charge LED Lade LED
Front Panel Board
FluidWarmer
L N PE
X1 L N PE
B . B r a u n M e d i z i n t e c h n o l o g i e G m b H
B . B r a u n M e d i z i n t e c h n o l o g i e G m b H
X3
Mains Input Netzeingang
X2
L
Heater Heizung
SMPS Power Board Primary/ Primär
N
SMPS Power Board Secondary/ Sekundär
Aux. Voltage Hilfspannung
Heater / Heizung
U Control U-Regler
P2 Heater Control Input
Heater
PFail
X2 F4 X3 X1.3 X1.1 X1.2 X1.6 X1.5 X1.4 X1 F3 F5 F6 F2 F1 N
Machine Monitoring Input N F E F S F _ F _ S D Geräteüberwachung L L U O _ O 4 M N A A K 4 D 2 E G K K 2 K B R / D A A / D A W
Fluid Warmer
PE3 N3 L3 N2 L2
L
P2
2 6 9
2 .4 1 . 3
Pay Attention to Pin Assignment! Auf Pinbelegung Achten! PE
P2
Watchdog
1 / 2 0 0 3
B24OK D24OK /PERSR_N /AKKU_OK PF PGD +5VG
Drive for Battery, Buzzer, Clock Ansteuerung für Akku, Summer, Uhr
P101
G K L O N DHeizungssteuerung Eingang I C R D D _ _ P _ N G H H H D
D e s c r i p t i o n
Message Output MeldungenAusgang
5V
Uamax
+5VD +12VD -12VD +12VAN -12VAN 24VL 24VGB 24VGD
24 V Control 24 V Steuerung
Logic Logik
Buzzer Summer
Mains Input
Power
P3 P9
2 x 1F
W 0 0 8 1
F i g . : L a y o u t S w i t c h M o d e P o w e r S u p p l y S M P S
24V
Power
T1 K1
T4 Thermal Fuse Thermosicherung
2 . T e c h n i c a l S y s t e m
Fan Mains Switch Lüfter Netzschalter
L N PE
DM i T a l o g M + _ D s mD _ E C 0 h a 8 p C t M e r 2 . _ K 1 A -2 Y 0 0 3 . d o c / p d f < 0 1 1 0 0 3 > d d m m y y
S y s t e m I n t e g r a t i o n
PE L1 N1
Regulator Regler
+12VD/ F401 - T3.15A TR5 +12VD/ F402 - T5.00A TR5 +12VAN / F403 - T1.25A TR5 +5VD/ F303 - T3.15A TR5 +5VD/ F302 - T5.00A TR5 +5VD / F301 - T3.15A TR5 +5VD/ F304 - T3.15A TR5 +24 VGD/ F602 - T3.15A TR5 +24VGB/ F601 - T3.15A TR5
LED
F401 F402
3
F403 F303 F302
1
F301 T3.15A TR5 F302 T5.0A TR5 F303 T3.15A TR5 F304 T3.15A TR5 F401 T3.15A TR5 F402 T5.00A TR5 F403 T1.25A TR5 F500 T3.15A TR5 F600 M10.00A 6.3x32 F601 T3.15A TR5 F602 T3.15A TR5
P10
Voltages PC
P4
Voltages Power Board 1/2
P5
Voltages Sup./Contr. Board
3
+24VL
F1 T6.25A 6.3x32 F2 T6.25A 6.3x32 F3 T3.15A TR5 F4 T3.15A TR5 F5/F6 - 110/120V: F20A 6.3x32 - 230V: M10A 6.3x32
P12
+5V +12V -12V
1
1
F301 F304 3
F602
1
F601
3
1
+
2
P13
P3 1 1
4
P7
P8 1
X100 Fan
Mains Switch/ Fan Voltage FDD Signals Supervisor Board
P9
1
3
1 2
1
P2
31
P101 3
Battery
L a y o u t S w i t c h M o d e P o w e r S u p p l y S M P S
2 . T e c h n i c a l S y s t e m D e s c r i p t i o n
1 / 2 0 0 3
1
Option Option Watch- EXT_ON Voltages ABPM dog HDD 2 7 0
DM i T a l o g M + _ D s mD _ E C 0 h 8 a p C t M e r 2 . _ K 1 A -2 Y 0 0 3 . d o c / p d f < 0 1 1 0 0 3 > d d m m y y
F i g . : L a y o u t S w i t c h M o d e P o w e r S u p p l y S M P S
Mains Input
Heater
2 .4 1 . 3
Fluid Warmer Pay Attention to Pin Assignment! Auf Pinbelegung Achten! PE3 N3 L3 N2 L2
X2 F4 X3 X1.3 X1.1 X1.2 X1.6 X1.5 X1.4 X1 F3 F5 F6 F2 F1 PE
L
N
PE L1 N1
Regulator Regler
+12VD/ F401 - T3.15A TR5 +12VD/ F402 - T5.00A TR5 +12VAN / F403 - T1.25A TR5 +5VD/ F303 - T3.15A TR5 +5VD/ F302 - T5.00A TR5 +5VD / F301 - T3.15A TR5 +5VD/ F304 - T3.15A TR5 +24 VGD/ F602 - T3.15A TR5 +24VGB/ F601 - T3.15A TR5
LED
F401 F402
3
F403 F303 F302
1
P10
Voltages PC
P4
Voltages Power Board 1/2
P5
Voltages Sup./Contr. Board
3
+24VL
1
F301 F304
F1 T6.25A 6.3x32 F2 T6.25A 6.3x32 F3 T3.15A TR5 F4 T3.15A TR5 F5/F6 - 110/120V: F20A 6.3x32 - 230V: M10A 6.3x32
3
F602
1
F601
3
1
F301 T3.15A TR5 F302 T5.0A TR5 F303 T3.15A TR5 F304 T3.15A TR5 F401 T3.15A TR5 F402 T5.00A TR5 F403 T1.25A TR5 F500 T3.15A TR5 F600 M10.00A 6.3x32 F601 T3.15A TR5 F602 T3.15A TR5
B . B r a u n M e d i z i n t e c h n o l o g i e G m b H
P12
+5V +12V -12V
1
+
2
P13
P3 1 1
4
P7
1 2
P8 1
Fan
1
P2
31
P101 3
X100
Mains Switch/ Fan Voltage FDD Signals Supervisor Board
P9
1
3
Battery
L a y o u t S w i t c h M o d e P o w e r S u p p l y S M P S
2 . T e c h n i c a l S y s t e m D e s c r i p t i o n
1 / 2 0 0 3
1
Option Option Watch- EXT_ON Voltages ABPM dog HDD 2 7 0
2. Technical System Description
1 / 200 3
2.41.4 Wiring DiagramSwitch Mode Power Supply SMPS with Battery Option
LC Display
J1
J1
(left) (links)
(right) (rechts)
J1
J3
J5
J7
Charge LED Lade LED
J9
2 - 71
2. Technical System Description
1 / 200 3
2 - 71
2.41.4 Wiring DiagramSwitch Mode Power Supply SMPS with Battery Option
LC Display
J1
J1
(left) (links)
(right) (rechts)
J3
J1
J5
J7
J9
Charge LED Lade LED
LED
P12
F401
PE3 N3 L3 N2 L2
X2 PE L N PE L1 N1 F4 X3 X1.3 X1.1 X1.2 X1.6 X1.5 X1.4 X1 F3
F402 F403
P10
F303 F302
Battery
F301 F304
2 x 12V
P4
F602 F601
Switch Mode Power Supply SMPS
P13
P5 P3
P9 P7 P8
X100
- -=blue + +=brown Pin 1/2 Charge LED Lade LED
P2
P101
Fig. : Wiring Diagram Switch Mode Power Supply SMPS with Battery Option
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 72
2.41.5 Description Switch Mode Power Supply The switch mode power supply SMPS is assembled in the rear door. The battery option can be retrofitted in a Dialog+. 110/120/230/240 V 50/60 Hz
Rated Voltage: Fuses: F1/F2
6.25 AT (6.3x32)
Mains input
F3/F4
3.15 AT (TR5)
Fluid warmer
F5/F6
230/240 V: 10 AM (6.3x32) 110/120 V: F20 A (6.3x32)
Heater 1800 W/240 V Heater 1800 W/110/120 V
F301
3.15 AT (TR5)
+5 VD
F302
5.00 AT (TR5)
+5 VD
F303/F304
3.15 AT (TR5)
+5 VD
F401
1.25 AT (TR5)
+12 VD
F402
5.00 AT (TR5)
+12 VD
F403
1.25 AT (TR5)
+12 VAN
F600
10 AM (6,3x32)
+24 VL
3.15 AT (TR5)
+24 VGB
F601/F602
The filter is required for EMC measures (EMC =electromagnetic compatibility).
Filter (EMC)
PFC circuits are used for switch mode power supplies and ensure that the line current is drawn sinusoidally and in phase with the sinusoidal line voltage.
Power Factor Correction PFC
The forward converter (Buck converter also called down switcher) converts an input voltage into a lower output voltage, i.e. the 24 V is converted into ±12 V and +5 V.
ForwardConverter
2.41.6 Pin Assignment Switch Mode Power Supply P2 – Signals to Supervisor Board Pin Signal 1a PF 2a /REM
Description Power Fail Remote control power supply
3a WD_S
Supervisor WATCHDOG
I/O Pin Signal O 1b B24OFF_S I 2b D24OFF_S I
3b AKAL
4a nc 5a nc
I
4b AKAL_S 5b +5VG
Data for shift register 6a H_DIN 7a EXT_STATE Statusfor external switch-on possibility Voltage monitoring blood side 8a B24OK
I 6b PERSR_N O 7b AKKU_EN O 8b D24OK
9a H_DCLK 10a /AKKU_OK
I 9b H_PROG O 10b GND
Clock to load shift register Battery status
Description I/O Switch-off 24 Vsupervisor bloodside I Switch-off 24 V supervisor dialysate I side Audible alarm I Audible alarmsupervisor I Voltage supply for clock on controller O board Signal to activate the staff call O Enable battery operation I Voltage monitoring dialysate O Programming mode for shift register
I
P3 - Battery Connection (ScrewTerminal) Pin Signal Type 1 +24VAKKU stabilised 2 GNDAKKU
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Description Tolerance Charge and dischargecurrent 21 V ... 28 V Ground power
Current [A] 15
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 73
P4 - Voltage Supply Power Board 1/2 Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Signal +5VD GNDD -12VAN +5VD GNDD GNDL +12VAN GNDL GNDL +24VL +24VGB +24VGD +24VL +24VGB +24VGD
Type stabilised
Pin 1 2 3 4 5 6 7 8 9 10 11 12
Signal +12VAN -12VAN GNDAN +12VD -12VD GNDD +24VGB +24VL GNDL +5VD GNDD GNDL
Type stabilised stabilised
Pin 1 2
Signal +5VD GNDD
Type stabilised
Description +5 V digital Ground digital
Pin 1 2 3 4 5 6 7 8 9
Signal +5VD nc GNDD +12VD -12VD GNDD +12VAN -12VAN GNDAN
Type stabilised
Description +5 V digital
Pin 1 2 3 4 5 6 7 8 9
Signal +5VD nc GNDD +12VD -12VD GNDD +12VAN -12VAN GNDAN
Type stabilised
stabilised stabilised
stabilised
rectified rectified rectified rectified rectified rectified
Description +5 V digital (PB1) Ground digital (PB1) -12 V analogue (PB1) +5 V digital (PB2) Ground digital (PB2) Ground power (PB1) +12 V analogue (PB1) Ground power (PB1) Ground power (PB2) +24 V power (PB1) +24 V switched blood (PB1) nc (PB1); +24 V switched dialysate (PB1) +24 V power (PB2) +24 V switched blood (PB2) +24 V switched dialysate (PB2)
P5 - VoltageSupply Supervisor/Controller Board
stabilised stabilised 20...30 V 20...30 V stabilised
Description +12 V analogue -12 V analogue Ground analogue +12 V digital -12 Vdigital Ground digital +24 V switched blood side +24 V power Ground power +5 V digital Ground digital Ground power
P6 - 3½" FloppyDisk Drive
P7 - VoltageSupply Options
stabilised stabilised stabilised stabilised
Ground digital +12 V digital -12 Vdigital Ground digital +12 V analogue -12 V analogue Ground analogue
P8 - VoltageSupply Option ABPM (Automatic Blood Pressure Measurement)
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
stabilised stabilised stabilised stabilised
Description +5 V digital Ground digital +12 V digital +12 V digital Ground digital +12 V analogue -12 V analogue Ground analogue
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 74
P9 - Mains Switch andVoltage Supply Fan Pin 1 2 3 4 5 6 7 8 9 10 11 12
Signal AKKU_Laden GND Mains Switch Mains Switch +12VD +12VD Mains Switch nc GND GND Mains Switch nc
Description Display "Charge Battery" Digital ground Logic (1st level) +12 VH (1st level) +12 V digital (fan) +12 V digital (fan) Logic (2nd level)
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Signal +5VD nc GNDD +5VD nc GNDD +5VD nc GNDD +12VD nc GNDD -12VD +12VD GNDD
Type stabilised
Pin 1 2 3 4
Signal +5VD +12VD GNDD GNDD
Type stabilised stabilised
Description +5 V digital +12 V digital Ground digital Ground digital
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Signal +5VD +12VD GNDD +12VAN -12VD GNDD -12VAN GNDAN GNDD +24VL nc GNDL +24VGB +24VGD GNDL
Type stabilised stabilised
Description +5 V digital +12 V digital Ground digital +12 V analogue -12 Vdigital Ground digital -12 V analogue Ground analogue Ground digital +24 V power Coding Ground power +24 V switched blood +24 V switched dialysate Ground power
Digital ground (fan) Digital ground (fan) GNDH (2nd level)
P10 - Voltage Supply PC Description +5 V digital
stabilised
Ground digital +5 V digital Coding Ground digital +5V digital
stabilised
Ground digital +12 V digital
stabilised
stabilised stabilised
Ground digital -12 Vdigital +12 V digital Ground digital
P11 - VoltageSupply Hard Disk Drive
P12 - VoltageSupply Options
stabilised stabilised stabilised
20 ... 30 V
20 ... 30 V 20 ... 30 V
P13 - Connection Monitor Booster Pin Signal + 24 V (20 .... 28 V 1 GNDD 2 /HSS_ON 3 GNDD 4
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Type
Description Ground digital Ground digital
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 75
P14 - Input Signal /EXT_ON (Switch-on machine fromexternal source) Pin 1 2 3 4 5 6
Signal GNDD /EXT_ON EXT_STATE
Type
Description Ground digital Open collector input Input
+5VD
+5 V digital
ScrewTerminal: X1 - Mains Input, Heater Pin 1.1 1.2 1.3 1.4 1.5 1.6
Voltage L N PE N1 L1 PE
X3 - Fluid Warmer, Relay Monitor Booster Pin Voltage 3.1 L3 3.2 N3 3.3 PE3
Inputs: Mains Switch
2nd level
H_DIN
Data for shift register and watchdog controller
WD_S
Watchdog supervisor
/REM
Remote (clock on controller board)
AKAL
Audible alarm (controller)
AKAL_S
Audible alarm (supervisor)
MSWITCH
Mains switch
B24OFF
+24 VGB ON/OFF (supervisor)
D24OFF
+24 VGD ON/OFF (supervisor)
H_DCLK
Clock for programming shift register
H_PROG
Programming mode shift register
AKKU_EN Rated Mains Voltage
Enable battery operation (only possible in therapy mode ) 110/230 V 50/60 Hz
/EXT_ON The machine can be switched on with this signal (see menu 1.26 Battery Option: external ON ). EXT_STATE
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Status for external switch-on possibility (e.g. central disinfection), simultaneously input and output; the signal is looped through (connector P14/3 =input).
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 76
Outputs: OK B24OK +24VGB
D24OK PERSR_N
+24VGD OK
Staff call
/AKKU_OK Load status of the battery PF Connections Voltages
Power fail (power failure) Fluid warmer, heater +5 VD: ±12 VD: ±12 VAN: +24 VL: +24 VGB: +24 VGD:
Digital Digital Analog Power (non-regulated) Switched blood side Switched dialysis side
EXT_STATE
External machine switch-on (information for TLC), simultaneously input and output; the signal is only looped through (connector P2/7a =output).
Mains Switch ON/OFF
If the Dialog+is switched off via the Automati c Swit ch-Off function in therapy the machine is in standby mode, e.g. the Dialog+can be switched on and off via the disinfection program. The Dialog+is only disconnected from mains if the mains plug is pulled out of the mains socket.
Standby
The power consumption is ≤ 10 W in standby mode.
H_PROG
The H_PROG signal is an open-collector signal and is generated by the LLC.
H_DIN
H_DCLK
0
No programming mode: Data rotate in shift register
1
Programming mode: Shift register can be written
The H_DIN signal is an open-collector signal and is generated by the LLC (data for the shift register to drive the heater).
0
Half-wave of mains voltage for heater switched off
1
Half-wave of mains voltage for heater switched on
The H_DCLK signal is an open-collector signal and is generated by the LLC. The pending data at H_DIN are shifted into the shift register with the H_DCLK signal to drive the heater.
0, 1 Watchdog
Rectangular signal to shift the data into the shift register
A watchdog is integrated to prevent a permanent unitentional drive of the heater in case of a LLC reset. The watchdog is retriggered by the LLC with <2 s. The H_DIN signal is used for triggering. After switching on the switch mode power supply the watchdog prevents a drive of the heater for t =10 s. The watchdog has no safety function and is therefore not tested before the therapy starts.
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
2.42
1 / 200 3
2 - 77
SafetyConcept
2.42.1 Block DiagramSafety Concept
Controller-Kanal
Supervisor-Kanal
Controller Channel
Supervisor Channel
LC Display
Touch Screen
Supervisor-Fenster Supervisor Window
Controller-Fenster Controller Window Touch Controller Board TCB
LowVoltage Differential Signal Adapter LVDS
Front Panel Board FPB Supervisor/Controller Tasten Supervisor/Controller Keys
AQ
EQ
stop
BP
BP
Video Controller RAM
Controller-Fenster
UI-SW
Diskettenlaufwerk Floppy Disk Drive Festplatte Hard Disk Drive
Supervisor Window
SW-SC
SUPBUS
Top Level Controller
TLC DIABUS Low-Level-Controller
BUS
Low Level Controller
LLC
Motoren Motors
Low Level Supervisor
LLS
S M V
Sensoren Sensors
Low-Level-Supervisor
Ventile Valves
S
Sensoren Sensors
S
Sensoren Sensors
SR
Sicherheitsrelais Safety Relay
24 V Versorgungsspannung 24 V Supply Voltage Fig. : Hardware Safety Concept
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
B. Braun Medizintechnologie GmbH
2. Technical System Description
1 / 200 3
2 - 78
2.42.2 Description Safety Concept The complete control and monitoring of the system is performed by the controller. The supervisor is an independent protection system and controls the safety relevant process parameters. The double channel design of the hard- and software guarantees the monitoring of the safety relevant input and output values. The input of the setpoint value, the fault free transmission and monitoring of the output values are thereby guaranteed. The safety concept has the following processor systems: •
TopLevel Controller
Top level controller on the top level controller board T TLC
•
Low level controller on the low level controller board LLC LLC
•
Supervisor on the supervisor board LLS
The top level controller TLC consists of: •
PC hardware
•
Operating system(QNX, multi-tasking real time operating system)
The top level controller TLC has the following functions: •
Communication with the user
•
Processing of the process parameters for the LLC
•
Control of the ABPM
•
Balance of ultrafiltration/substitution
The input of the set-point values is performed by: •
Low Level Controller
Touch screen (colour TFT)
•
Floppy disk drive
•
LAN (local area network)
The low level controller LLC has: •
68020 microprocessor system
The low level controller LLC has the following functions:
Software Supervisor Communication Module SW-SC
•
Control of motors and valves
•
Monitoring
The software supervisor communication module (SW-SC module) is a safety independent software module within the TLC. The SW-SC module logically belongs to the supervisor. It has the following functions: •
Front Panel Board FPB
Supervisor
Display of safety relevant data from the supervisor on the TFT
The front panel board FPB has the following functions: •
Interaction with the user via front panel keys and LEDs
•
Drive of loudspeaker
The supervisor consist of: •
80535 microprocessor system
The supervisor has the following functions:
Enter of Set-point Values
•
Signal processing of the sensors
•
Monitoring of the safety relevant process pararmeters
The acquisition of the set-point values (therapy parameters) is performed as follows: •
MT-MD-DE08C M.KAY
Dialog+_sm_Chapter 2_1-2003.doc/pdf <011003>ddmmyy
Set-point values can be entered by the TLC for preparation of a therapy. The data is transmitted to the LLC and supervisor via the DIABUS.
B. Braun Medizintechnologie GmbH