BIOHIT
M-LINE
TECHNICAL
SERVICE MANUAL
Page 1 of 42
Ver.03 22.02.05
Table of Contents 1 2 3
4
5
6
7 8
Product Product Description................... Description.............................. ...................... ...................... ....................... ....................... ...................... ...................... ................ ..... 3 Protective Protective tip cone filters and filter ejection ejection ....................... .................................. ...................... ...................... ..................... .......... 4 Maintenanc Maintenance............ e....................... ...................... ....................... ....................... ...................... ....................... ....................... ...................... ...................... ........... 4 3.1 Daily cleaning of the outer surface of the pipettor ......................... ............ ......................... ........................ ............ 4 3.2 Cleaning and decontaminating the lower parts ....................... ........... ........................ ........................ .................. ...... 5 3.2.1 Single channel pipettors: disassembling, regreasing and cleaning ................ ........... ..... 5 3.2.2 Single channel channel pipettors pipettors:: reassemblin reassembling g ....................... .................................. ....................... ....................... ........... 7 3.2.3 Cleaning Cleaning stop surfaces surfaces ....................... ................................... ....................... ....................... ........................ ........................ ............ 8 3.2.4 Multichannel pipettors: disassembling, regreasing and cleaning ................... ............. ...... 9 3.2.5 Multichannels pipettors: reassembling ......................... ............. ......................... .......................... ................... ...... 12 3.3. Sterilization Sterilization of the pipettor pipettor ....................... .................................. ....................... ....................... ...................... ....................... .................. ...... 13 Testing the performance and recalibration ........................ ........... ......................... ......................... ......................... ................ .... 14 4.1 Testing Testing the performanc performance e ....................... ................................... ........................ ....................... ....................... ........................ ............... ... 14 4.2 Recalibration Recalibration............ ....................... ...................... ...................... ....................... ....................... ...................... ....................... ....................... ............. 14 Disassembling and reassembling the handle ......................... ............ ......................... ......................... ........................ ........... 15 5.1 Disassembling single channel models m10, m20, m100, m200 and m1000:...... m1000:. ..... 15 5.2 Reassembling single channel models m10, m20, m100, m200 and m1000 ....... 17 5.3 Disassemblin Disassembling g model m5000 m5000 ...................... ................................. ...................... ....................... ....................... ...................... ........... 18 5.4 Reassembling model m5000....................... m5000.......... .......................... .......................... .......................... .......................... ................. .... 20 5.5 Disassemblin Disassembling g MCP-models MCP-models ...................... ................................. ...................... ...................... ...................... ...................... ............. 22 5.6 Reassemblin Reassembling g MCP models models ...................... ................................. ....................... ....................... ....................... ....................... ........... 24 Spare parts................ parts........................... ....................... ....................... ....................... ....................... ....................... ....................... ....................... ................. ..... 27 6.1 Handle assembly: m10, m20, m100, m200 and m1000 single channel pipettors 27 6.2 Handle assembly m5000........................ m5000............ ......................... ......................... ......................... ......................... ....................... ........... 28 6.3 Model m10 ....................... .................................. ....................... ....................... ....................... ....................... ....................... ....................... .............. ... 29 6.4 Model m20 ....................... .................................. ....................... ....................... ....................... ....................... ....................... ....................... .............. ... 30 6.5 Model m100 ....................... .................................. ...................... ....................... ....................... ....................... ....................... ....................... .............. 31 6.6 Model m200 ....................... .................................. ...................... ....................... ....................... ....................... ....................... ....................... .............. 32 6.7 Model Model m1000 ....................... ................................... ....................... ....................... ........................ ....................... ....................... ..................... ......... 33 6.8 Model Model m5000 ....................... ................................... ....................... ....................... ........................ ....................... ....................... ..................... ......... 34 6.9 Handle assembly; MCP m10, MCP m100 and MCP m300 ....................... ........... .................... ........ 35 6.10 Final assembly MCP models m10, m100 and m300 ......................... ............. ......................... .................. ..... 36 6.11 Dispensing Dispensing Head Head,, MCP MCP m10............................... m10.......................................... ....................... ....................... ...................... .............. ... 37 6.12 Dispensing Head, MCP m100........... m100 ........................ .......................... .......................... .......................... ......................... .............. .. 38 6.13 Dispensing Head, MCP m300........... m300 ........................ .......................... .......................... .......................... ......................... .............. .. 39 6.14 List of tools and accessories accessories.......... ...................... ....................... ...................... ....................... ....................... ...................... ............. 40 Warranty Warranty information information........... ...................... ...................... ...................... ...................... ...................... ...................... ....................... ..................... ......... 41 Certificate Certificate of Conformity............. Conformity........................ ...................... ...................... ...................... ....................... ....................... ...................... ............. .. 42
Page 2 of 42
Ver.03 22.02.05
Table of Contents 1 2 3
4
5
6
7 8
Product Product Description................... Description.............................. ...................... ...................... ....................... ....................... ...................... ...................... ................ ..... 3 Protective Protective tip cone filters and filter ejection ejection ....................... .................................. ...................... ...................... ..................... .......... 4 Maintenanc Maintenance............ e....................... ...................... ....................... ....................... ...................... ....................... ....................... ...................... ...................... ........... 4 3.1 Daily cleaning of the outer surface of the pipettor ......................... ............ ......................... ........................ ............ 4 3.2 Cleaning and decontaminating the lower parts ....................... ........... ........................ ........................ .................. ...... 5 3.2.1 Single channel pipettors: disassembling, regreasing and cleaning ................ ........... ..... 5 3.2.2 Single channel channel pipettors pipettors:: reassemblin reassembling g ....................... .................................. ....................... ....................... ........... 7 3.2.3 Cleaning Cleaning stop surfaces surfaces ....................... ................................... ....................... ....................... ........................ ........................ ............ 8 3.2.4 Multichannel pipettors: disassembling, regreasing and cleaning ................... ............. ...... 9 3.2.5 Multichannels pipettors: reassembling ......................... ............. ......................... .......................... ................... ...... 12 3.3. Sterilization Sterilization of the pipettor pipettor ....................... .................................. ....................... ....................... ...................... ....................... .................. ...... 13 Testing the performance and recalibration ........................ ........... ......................... ......................... ......................... ................ .... 14 4.1 Testing Testing the performanc performance e ....................... ................................... ........................ ....................... ....................... ........................ ............... ... 14 4.2 Recalibration Recalibration............ ....................... ...................... ...................... ....................... ....................... ...................... ....................... ....................... ............. 14 Disassembling and reassembling the handle ......................... ............ ......................... ......................... ........................ ........... 15 5.1 Disassembling single channel models m10, m20, m100, m200 and m1000:...... m1000:. ..... 15 5.2 Reassembling single channel models m10, m20, m100, m200 and m1000 ....... 17 5.3 Disassemblin Disassembling g model m5000 m5000 ...................... ................................. ...................... ....................... ....................... ...................... ........... 18 5.4 Reassembling model m5000....................... m5000.......... .......................... .......................... .......................... .......................... ................. .... 20 5.5 Disassemblin Disassembling g MCP-models MCP-models ...................... ................................. ...................... ...................... ...................... ...................... ............. 22 5.6 Reassemblin Reassembling g MCP models models ...................... ................................. ....................... ....................... ....................... ....................... ........... 24 Spare parts................ parts........................... ....................... ....................... ....................... ....................... ....................... ....................... ....................... ................. ..... 27 6.1 Handle assembly: m10, m20, m100, m200 and m1000 single channel pipettors 27 6.2 Handle assembly m5000........................ m5000............ ......................... ......................... ......................... ......................... ....................... ........... 28 6.3 Model m10 ....................... .................................. ....................... ....................... ....................... ....................... ....................... ....................... .............. ... 29 6.4 Model m20 ....................... .................................. ....................... ....................... ....................... ....................... ....................... ....................... .............. ... 30 6.5 Model m100 ....................... .................................. ...................... ....................... ....................... ....................... ....................... ....................... .............. 31 6.6 Model m200 ....................... .................................. ...................... ....................... ....................... ....................... ....................... ....................... .............. 32 6.7 Model Model m1000 ....................... ................................... ....................... ....................... ........................ ....................... ....................... ..................... ......... 33 6.8 Model Model m5000 ....................... ................................... ....................... ....................... ........................ ....................... ....................... ..................... ......... 34 6.9 Handle assembly; MCP m10, MCP m100 and MCP m300 ....................... ........... .................... ........ 35 6.10 Final assembly MCP models m10, m100 and m300 ......................... ............. ......................... .................. ..... 36 6.11 Dispensing Dispensing Head Head,, MCP MCP m10............................... m10.......................................... ....................... ....................... ...................... .............. ... 37 6.12 Dispensing Head, MCP m100........... m100 ........................ .......................... .......................... .......................... ......................... .............. .. 38 6.13 Dispensing Head, MCP m300........... m300 ........................ .......................... .......................... .......................... ......................... .............. .. 39 6.14 List of tools and accessories accessories.......... ...................... ....................... ...................... ....................... ....................... ...................... ............. 40 Warranty Warranty information information........... ...................... ...................... ...................... ...................... ...................... ...................... ....................... ..................... ......... 41 Certificate Certificate of Conformity............. Conformity........................ ...................... ...................... ...................... ....................... ....................... ...................... ............. .. 42
Page 2 of 42
Ver.03 22.02.05
1
Product Description
Your new Biohit mLINE is an autoclavable air displacement pipettor. The attractive and ergonomical design of the mLINE pipettor together with its very low pipetting forces reduce the risk of repetitive strain injuries (RSI). The pipettor is designed for both right- and lefthanded use. The special tip cone design of the pipettor offers the possibility to t o use replaceable SafeCone filters to help prevent the contamination and damage of the pipettor. The operating button of the mLINE allows safe and easy filter ejection. The pipettor utilizes ut ilizes disposable tips.
Page 3 of 42
Ver.03 22.02.05
2
Protective tip cone filters and filter ejection
The tip cone design of the Biohit mLINE pipettors (> 10 µl) allows the use of Biohit SafeCone filters in the tip cones. These removable filters prevent liquids and aerosols from entering the pipettor and thus protects the pipettor from contamination and damage. The Safe-Cone filters are available as Standard or Plus versions. It is recommended to use the Standard filter for general applications and the Plus filter for f or more demanding applications such as cell culture, bacterial and virological work and molecular biology. Filters need to be changed regularly. The interval for a filter f ilter replacement is application dependant but the recommendation is to change the filter daily (after 50 – 250 pipetting cycles) and always in case of over-aspiration. The Safe-Cone filters can be removed with the unique filter ejector without touching the filters by hand. Remove the operating button cap and eject the filter by pressing the operating button down. Make sure that the filter is disposed into a suitable waste container. Clean the tip cone if needed and put the new filter filt er in place.
3 Maintenance Biohit mLINE pipettors have been designed for easy in-house service. If the pipettor is used daily it is recommended to clean/decontaminate and check the performance of the pipettor every three months. Biohit also provides complete repair and recalibration service for your pipettor including a service report and performance certificate. Note: Please Note: Please make sure that the pipettor has been decontaminated before you send it to us or your local representative. Please advise if any hazardous material has been used with your pipettor. Note: The Note: The use of the tip cone filters may prolong the interval of the service. Change the filters regularly. Note: It Note: It is recommended to always use gloves when cleaning the pipettor.
3.1
Daily cleaning of the the outer surface surface of the the pipettor
Your Biohit mLINE pipettor should be checked every day for cleanliness. To clean and decontaminate the outer surface of your pipettor use Biohit Proline‚ Biocontrol (Cat.no. 724004, 5 litres) and a soft lint-free cloth. It is also possible to use ethanol (70%), isopropanol (60%) or a mild detergent as a cleaning agent. Gently clean the surface of the pipettor with moistened cloth and wipe dry. Pay special attention to the tip cone. Change the tip cone filter.
Page 4 of 42
Ver.03 22.02.05
3.2
Cleaning and decontaminating the lower parts
If your pipettor is in daily use it is recommended to clean/decontaminate and grease the pipettor every three months. To clean and decontaminate the lower parts of the pipettor follow these steps: 3.2.1 Single channel pipettors: disassembling, regreasing and cleaning 1.
Eject the tip cone filter
2.
Unscrew the tip ejector collar counter clockwise and remove it.
3.
Pipettors m10, m20, m100, m200 and m1000: Unscrew the tip cone holder counter clockwise and carefully remove it with the tip cone.
Pipettor m5000: Unscrew the tip cone cylinder counter clockwise and remove it.
4.
Remove the piston (if needed to replace, otherwise it is not required to remove it during basic maintenance ). Turn the pipettor upside down, push it against a table, hold the white collar with one hand and unscrew the piston counter clockwise with another hand.
5.
Clean the tip ejector collar, the tip cone holder, the tip cone (cylinder) and the piston with Biohit Proline Biocontrol, ethanol (70%), isopropanol (60%) or mild detergent and soft lint-free cloth.
Page 5 of 42
Ver.03 22.02.05
6.
7. 8.
Clean the interior of the tip ejector collar and the tip cone (cylinder) with a cotton swab. Be careful with the pipettors m10, m20 and m100 so that the seal inside the tip cone will not damage. Rinse the parts with distilled water if needed and let the parts dry. Pipettors m10, m20, and m100: Put a thin layer of grease on the piston:
m10 and m20: Grease on this area only
m100: Grease on this area only
Pipettors m200 and 1000: Put a thin layer of grease around the seal:
m200: Grease on the small seal only
m1000: Grease on the seal only
Pipettor m5000: Put a thin layer of grease on the seal and the interior of the tip cone cylinder:
Grease on the seal
Grease on the inner surface of the tip cone
Page 6 of 42
Ver.03 22.02.05
Note: Avoid excess grease. Use only the grease provided with the pipettor (731140, 2g tube). Note: The spare part code for the grease is 731141 (tube 20g). The 731140 (2g tube) provided along with the pipettor package is not available as a spare part. Note: Before reassembling check that no lint or particles are on the surface of the piston. Decontamination: For complete decontamination of the lower parts place the tip ejector collar, tip cone holder, tip cone and tip cone cylinder (only the model m5000) into a beaker containing Biohit Proline‚ Biocontrol and leave for at least 30 minutes. Wipe the piston with Biohit Proline‚ Biocontrol and the lint-free cloth. Rinse the parts with distilled or sterile water. Let the parts dry. Grease the piston and seal according to the instructions given earlier in this chapter.
3.2.2 Single channel pipettors: reassembling 1.
Reinstall the piston (if it has been removed, removal of the piston is not necessarily needed during normal maintenance). Turn the pipettor upside down, push it against a table, hold the white collar with one hand and screw the piston clockwise with another hand.
2.
Pipettors m10, m20, m100, m200, m1000: Carefully place the tip cone on the piston and attach it by screwing the tip cone holder clockwise.
Pipettor m5000: Carefully place the tip cone cylinder on the piston and screw clockwise. Make sure that the tip cone (cylinder) is properly tightened. Avoid over tightening.
3.
Attach the tip ejector collar by screwing it clockwise.
4.
Place the new tip cone filter on its place with forceps
Page 7 of 42
Ver.03 22.02.05
5.
Press the operating button several times to ensure that the grease has spread evenly.
6.
Check the performance of the pipettor.
Note: It is always necessary to check the performance of the pipettor after in-house service or maintenance.
3.2.3 Cleaning stop surfaces If there is a problem with the accuracy of the pipettor or the pipettor is very dirty, the cleaning of the critical stop surfaces might be required. 1. Remove tip ejector collar, tip cone holder, tip cone and piston as described above 2. Turn the volume to the maximum. 3. Insert a 1000ul tip into the machinery: push upwards and simultaneously turn clockwise. The gray plastic part is connected to the machinery with a bayonet type of connection. Piston spring comes out with a metallic sleeve that acts as a stop surface.
4. Clean the stop surface thoroughly with ethanol and apply some compressed air inside the machinery. Stop surface
5. Insert the piston spring and metallic sleeve into the machinery with the help of a tip. Push upwards and turn counter clockwise.
Page 8 of 42
Ver.03 22.02.05
6. Reassembly the rest of the parts as described above.
3.2.4 Multichannel pipettors: disassembling, regreasing and cleaning 1. Unscrew the dispensing head by turning it counter clockwise and remove it.
2. Push down the piston support plate as far as it goes (for example with a pen) Note: It is very important that the pistons are at the lowest position during service. Otherwise the reassembly of the pistons and tip cones is tricky.
3. Push down the tip ejector bar locking piece (731145) with one hand and simultaneously hold the tip ejector bar at one end with another hand. Remove the tip ejector bar by pulling it out.
4. Attach a tip onto the tip cone. Push the tip cone upwards (it is spring loaded) and turn it counter clockwise. Pull the tip cone out of the tip cone housing
Page 9 of 42
Ver.03 22.02.05
5. Pull out the piston. It is connected to the piston support plate with a magnet. Please insert the piston into the tip cone if you store them on the table. That way piston won’t get damaged and collect dirt. Remove all tip cones and pistons.
6. Clean the tip cone housing, the tip ejector bar, the tip cones and the pistons with Biohit
Proline Biocontrol, ethanol (70%), isopropanol (60%) or mild detergent and soft lint-free cloth. 7. Disassembling tip cones: Models m10 and m100: Unscrew the o-ring support nut (m10, 731156) or seal support nut (m100, 726167).
Model m300: The tip cone can not be disassembled.
726148
726151
The assemblies include the following parts: 726170
726169
726168
726167
726147
m100
731159
Page 10 of 42
731158
731157
731156
731155
Ver.03 22.02.05
Clean the parts with Biohit Proline Biocontrol, ethanol (70%), isopropanol (60%) or mild detergent if needed. Rinse the parts with distilled water if needed and let the parts dry 8. Regreasing. Models m10 and m100: Place the parts in an order as indicated in the pictures below. It makes regreasing and reassembly easier. Apply a thin layer of grease on the seals and pistons only.
Model m300: Apply a thin layer of grease on the sealing only.
Note: Avoid excess grease. Use only the grease provided with the pipettor (731140, 2g tube). Note: The spare part code for the grease is 731141 (tube 20g). The 731140 (2g tube) provided with the pipettor package is not available as a spare part. Note: Before reassembling check that no lint or particles are on the surface of the piston. 9. Reassembling Models m10 and m100: Tighten the o-ring support nut (m10, 731156) or seal support nut (m100, 726167). Please make sure the nut is properly tightened but avoid over tightening. Keep the piston inserted all the time.
Page 11 of 42
Ver.03 22.02.05
Please move the piston up and down to spread grease evenly. Model m300: Insert the piston into the tip cone.
Please move the piston up and down to spread grease evenly. Decontamination: For complete decontamination of the lower parts place the tip cone housing, tip cones, pistons and tip ejector bar into a beaker containing Biohit Proline‚ Biocontrol and leave for at least 30 minutes. Rinse the parts with distilled or sterile water. Let the parts dry. Grease the seal and piston as described above.
3.2.5 Multichannels pipettors: reassembling 1. Please make sure that the piston support plate is at its lowest position.
2. Attach a tip on the tip cone and insert the piston assembly complete (piston+tip cone) into the tip cone housing. Push it upwards as far as it goes and turn the tip cone clockwise. Make sure that the tip cone is properly placed. appr.20m
Page 12 of 42
Ver.03 22.02.05
3. Reassemble the tip ejector bar. Push down the tip ejector bar locking piece (731145) with one hand and simultaneously hold the tip ejector bar with another hand and push it upwards. Make sure the snap fit will connect properly.
4.Insert tip cone filters. Attach the dispensing head to the handle and tighten it properly. Please avoid over tightening. Please make sure that the dispensing head can rotate freely.
3.3. Sterilization of the pipettor The entire mLINE pipettor can be sterilized by steam autoclaving at 121 oC, (252 oF), 1 atm for 20 minutes (please ref. instruction manual for details). Remove the tip cone filter (if fitted, Chapter 7.3). Put the pipettor into the sterilization bag and place it into the autoclave. After autoclaving the pipettor must be cooled down at least for two hours before use. It is recommended to grease the piston/seal after every 10 autoclaving cycle. It is recommended to check the performance of the pipettor after every autoclaving cycle.
Page 13 of 42
Ver.03 22.02.05
4
Testing the performance and recalibration
It is recommended to check the performance of your Biohit mLINE pipettors regularly (e.g. every 3 months) and always after in-house maintenance. However, the user should establish a regular testing routine for their pipettors with regard to accuracy requirements of the application, frequency of use, number of operators using the pipettor, nature of the liquid dispensed and the acceptable maximum permissible errors established by the user. (ISO 8655-1.) 4.1
Testing the performance
Performance testing should take place in a draught-free room at 15 – 30 OC, constant to +/- 0.5 OC and humidity above 50%. The pipettor, tips and the test water should have stood in the test room a sufficient time (at least 2 hours) to reach equilibrium with the room conditions. Use distilled or deionised water (grade 3, ISO 3696). Use an analytical balance with a readability of 0.01 mgs. (ISO 8655-6.) If the results fall within the specifications, the pipettor is ready for use. Otherwise check both systematic and random errors and, when necessary, proceed to the recalibration procedure . Note: Systematic error (inaccuracy) is the difference between the dispensed volume and the selected test volume. Random error (imprecision) is the scatter of the dispensed volumes around the mean of the dispensed volume. (ISO 8655-1) Note: Biohit specifications are achieved in strictly controlled conditions (ISO 8655-6). The user should establish own specifications based on the field of use and the accuracy requirements placed on the pipettor (ISO8655-1). Note: This method is based on ISO 8655. 4.2
Recalibration
The calibration of your mLINE pipettor has been factory checked and certified at 22 OC using grade 3 distilled water according to ISO 3696. The calibration is based on ISO 86556: Gravimetric test method for volumetric instruments. The pipettor’s specifications are guaranteed only with genuine Biohit tips. If you find the pipettor to be inaccurate after performance testing, please follow the instructions below: 1. 2. 3. 4.
Remove the lid of the calibration nut, located at the backside of the handle, with the aid of the calibration tool. Place the hexagonal head of the calibration tool into the hole of the calibration nut. Turn the adjustment lock counter clockwise to decrease and clockwise to increase the volume. Repeat testing the performance procedure. Continue until the results are correct.
Note: Biohit offers accredited calibration service. Please contact your local Biohit representative for further information. Note: Please do not use the plastic mLine calibration tool provided with the pipettor package for calibration purposes in technical service. The plastic tool has not been designed for continuous professional use. Please use a standard metallic tool.
Page 14 of 42
Ver.03 22.02.05
5
Disassembling and reassembling the handle
The upper part of the pipettor (handle) must be disassembled if, for example, there is a problem with the counter or any other inner part. The disassembly of the upper part is not a normal service routine! 5.1
Disassembling single channel models m10, m20, m100, m200 and m1000:
1) Remove the tip ejector collar, tip cone holder and tip cone 2) Next remove the tip ejector and handle collar by pushing the tip ejector upwards with your thumb.
3) Turn the pipettor upside down, push it against a table, hold the white collar with one hand and unscrew the piston counter clockwise with another hand.
4) Now the following parts have been removed: Tip ejector
Handle collar
Cap Piston
Tip cone holder
Tip cone
Page 15 of 42
Tip ejector collar
Ver.03 22.02.05
5) Remove spring lock, primary and secondary springs and tip cone locking ring with the mLine tool (726201). Push the tool upwards and simultaneously turn counter clockwise. The spring lock is locked with a bayonet type of connection.
Tip cone locking ring
Primary spring
Secondary spring
Spring lock
6) Pull out the machinery assembly (counter and operating button). Place your fore finger into the handle. Remove the display by pushing it with the fore finger.
7) Now the following parts have been removed:
8) Clean, repair and replace parts if needed.
Page 16 of 42
Ver.03 22.02.05
5.2
Reassembling single channel models m10, m20, m100, m200 and m1000 1) Please note that there are two small slots on the machinery body and two small ribs on the calibration wheel inside the handle (please see the pics below).
Two slots Two ribs
2) Insert the machinery into the handle
3) Make sure that the two ribs on the calibration wheel go into the slots on the machinery body.
4) The machinery has been reassembled in a right way when the lines of the upper most counter wheel are visible
The lines must be visible this way
Page 17 of 42
Ver.03 22.02.05
5) Reassemble the tip cone locking ring, primary spring, secondary spring and spring lock in a right order. Lock the spring lock with the mLine tool by pushing the spring lock upwards and simultaneously turning it clockwise. Please place your thumb on the counter to keep the machinery body still. Otherwise it may slip out of the handle.
Note! Please make sure that the counter lines are clearly visible. 6) Reassemble the window and counter release nut; the upper snap fit must go first. If the window is not sitting properly, then the calibration wheel might not be sitting well. In that case push the wheel back into its groove. Replace the display sticker (used for attaching the display), if needed.
7) Reassemble the tip ejector and handle cover. Screw the piston
9) Reassemble the tip ejector collar, tip cone holder and tip cone. 10) Make sure the operating button and counter works properly 5.3
Disassembling model m5000
1) Remove the tip ejector collar, tip cone holder and tip cone 2) Next remove the tip ejector and handle collar by pushing the tip ejector upwards with your thumb.
Page 18 of 42
Ver.03 22.02.05
3) Turn the pipettor upside down, push it against a table, hold the white collar with one hand and unscrew the piston counter clockwise with another hand.
4) Now the following parts have been removed:
Handle collar
Tip ejector
Cap
Piston
Tip ejector collar
Tip cone
5) Remove the tip cone holder with mLine 5000 tool (726202) by pushing the tool upwards continuously and simultaneously turning counter clockwise
6) Remove spring lock, primary and secondary springs and tip cone locking ring with the mLine tool. Push the tool upwards and simultaneously turn counter clockwise. The spring lock is locked with a bayonet type of connection.
Tip cone locking ring
Primary spring
Page 19 of 42
Secondary spring
Spring lock
Tip cone holder
Ver.03 22.02.05
7) Pull out the machinery assembly (counter and operating button). Place your fore finger into the handle. Remove the display by pushing it with the fore finger.
9) Now the following parts have been removed:
10) Clean, repair and replace parts if needed.
5.4
Reassembling model m5000 1) Please note that there are two small slots on the machinery body and two small ribs on the calibration wheel inside the handle (please see the pics below).
Two ribs
Two slots
Page 20 of 42
Ver.03 22.02.05
2) Insert the machinery into the handle
3) Make sure that the two ribs on the calibration wheel go into the slots on the machinery body.
4) The machinery has been reassembled in a right way when the lines of the upper most counter wheel are visible
The lines must be visible this way
5) Reassemble the tip cone locking ring, primary spring, secondary spring and spring lock in a right order. Lock the spring lock with the mLIne tool by pushing the spring lock upwards and simultaneously turning it clockwise. Please place your thumb on the counter to keep the machinery body still. Otherwise it may slip out of the handle.
Note! Please make sure that the counter lines are clearly visible.
Page 21 of 42
Ver.03 22.02.05
6) Reassemble the tip cone locking nut by screwing it counter clockwise by hand until it “clicks”
7) Reassemble the window and counter release nut; the upper snap fit must go first. If the window is not sitting properly, then the calibration wheel might not be sitting well. In that case push the wheel back into its groove. Replace the display sticker (used for attaching the display), if needed.
8) Reassemble the tip ejector and handle cover. Screw the piston
9) Reassemble the tip ejector collar and tip cone. 10) Make sure the operating button and counter works properly 5.5
Disassembling MCP-models
1) Remove dispensing head 2) Next remove the tip ejector and handle collar by pushing the tip ejector upwards with your thumb.
Page 22 of 42
Ver.03 22.02.05
3) Turn the pipettor upside down, push it against a table, hold the white collar with one hand and unscrew the magnet holder (726166) counter clockwise with another hand. Note: Do not hold the white part with forceps; it might damage the part! 4) Unscrew the locking nut (726165) with the mLine MCP tool (726204). Insert the tool and turn counter clockwise. Remove all parts and pull out the machinery
5) Place your fore finger into the handle. Remove the display by pushing it with the fore finger.
6) Now the following parts have been removed:
7) Clean, repair and replace parts if needed. Clean the piston spring (726071) and stop surface (metallic sleeve) thoroughly with ethanol and apply some compressed air inside the machinery (ref chapter 3.2.3).
Page 23 of 42
Ver.03 22.02.05
5.6
Reassembling MCP models
1. Please note that there are two small slots on the machinery body and two small ribs on the calibration wheel inside the handle (please see the pics below).
Two ribs
2. Insert the machinery into the handle
Two slots
3. Make sure that the two ribs on the calibration wheel go into the slots on the machinery body.
4. The machinery has been reassembled in a right way when the lines of the upper most counter wheel are visible
The lines must be visible this way
Page 24 of 42
Ver.03 22.02.05
5. Please place your thumb on the counter to keep the machinery body still. Otherwise it may slip out of the handle. Insert the piston spring (726071).
6. Insert the tip ejection adapter (726163) so that the flat end goes first. Then insert tip ejection collar (726163) and connecting collar (726164)
7. Insert the locking nut (726165) and tighten it with the mLine MCP tool (726204). Push down the spring loaded locking nut and simultaneously turn the tool clockwise. Please make sure the locking nut is properly tightened but avoid over tightening. Make sure that the connecting collar can be rotated freely. If not, then the nut is over tightened.
8. Reassemble the window and counter release nut; the upper snap fit must go first. If the window is not sitting properly, then the calibration wheel might not be sitting well. In that case push the wheel back into its groove. Replace the display sticker (used for attaching the display), if needed.
Page 25 of 42
Ver.03 22.02.05
9. Reassemble the tip ejector and handle cover. Screw the magnet holder.
10. Attach the dispensing head.
Page 26 of 42
Ver.03 22.02.05
6 6.1
Spare parts Handle assembly: m10, m20, m100, m200 and m1000 single channel pipettors
Machinery, complete; Model 10; 726061 Model 20: 726085 Model 100; 726062 Model 200, 726063 Model 1000; 726064
Calibration wheel; 726066
Handle, mline; 726012
Tip cone locking ring; 726069 Display; 10; 726022 100; 726026 20: 726082 200; 726027 1000; 726028 Counter release nut; 726068 Display sticker, 5x20pcs 726121
Primary spring; 726073
Secondary spring; 726074 Spring lock; 726075
mLine opening tool: 726201
Page 27 of 42
Ver.03 22.02.05
6.2
Handle assembly m5000
Machinery, complete; 5000; 726065
Calibration wheel: 726066
Handle, mline; 726012
Tip cone locking ring 5000;
Display; 5000; 726029
Counter release nut; 726068
Primary spring; 726073
Display sticker, 5x20pcs 726121
Secondary spring; 726074
Spring lock; 726075
mLine 5000 opening tool: 726202
Page 28 of 42
Ver.03 22.02.05
6.3
Model m10
Cap, set of 5 colours; 726001
Handle collar, 726002
ID window; 726004
Tip ejector; 726005 ID tag; 726003
Calibration nut cover;
Tip cone complete; 10; 731084 Tip cone 10, upper part, 731075 Tip cone 10 , lower part, 731077 O-rings, 10pcs: 731085
Piston 10; 731076
Tip cone holder; 726041 Tip ejector collar; 726051
mLine autoclavable grease, 20g tube: 731141
Page 29 of 42
Ver.03 22.02.05
6.4
Model m20 Cap, set of 5 colours; 726001 Handle collar, 726002
ID window; 726004 Tip ejector; 726005
ID tag; 726003
Calibration nut cover;
Piston 20; 726084
Tip cone 20; 726083
Tip cone holder; 726041
Tip ejector collar; 726052
mLine autoclavable grease, 20g tube: 731141
Page 30 of 42
Ver.03 22.02.05
6.5
Model m100 Cap, set of 5 colours; 726001
Handle collar, 726002
ID window; 726004 Tip ejector; 726005 ID tag; 726003
Calibration nut cover;
Tip cone; 100; 726042
Piston 100; 726034
Tip cone holder; 726041
Tip ejector collar; 726052
mLine autoclavable grease, 20g tube: 731141
Page 31 of 42
Ver.03 22.02.05
Cap, set of 5 colours; 726001
6.6
Model m200 Handle collar, 726002
ID window; 726004 ID tag; 726003
Tip ejector; 726005
Calibration nut cover;
Tip cone; 200; 726039
Piston 200; 726035
Tip cone holder; 726041
Tip ejector collar; 726052
mLine autoclavable grease, 20g tube: 731141
Page 32 of 42
Ver.03 22.02.05
Cap, set of 5 colours; 726001
6.7
Model m1000 Handle collar, 726002
ID window; 726004 ID tag; 726003
Tip ejector; 726005
Calibration nut cover; 726006
Tip cone;
Piston 1000; 726036
Tip cone holder 1000;
Tip ejector collar; 726053
mLine autoclavable grease, 20g tube: 731141
Page 33 of 42
Ver.03 22.02.05
6.8
Model m5000 Cap, set of 5 colours; 726001 Handle collar, 726002
ID window; 726004
ID tag; 726003
Tip ejector; 726005
Calibration nut cover;
Tip cone holder 5000; 726076
Tip cone;
Piston 5000; 726037
Tip ejector collar 5000; 726054
mLine autoclavable grease, 20g tube: 731141
Page 34 of 42
Ver.03 22.02.05
6.9
Handle assembly; MCP m10, MCP m100 and MCP m300
Machinery complete mline MCP: 0,5-10: 726160 5-100: 726161 30-300: 726162
Calibration Wheel: 726066
Handle: 726012
Display: 0,5-10: 726022 5-100: 726141 30-300: 726142
Tip ejection adapter
Piston spring 726071
Counter Release nut: 726068 Display sticker, 5x20pcs 726121 Connecting collar and friction o-ring Locking nut 726165 Magnet holder 726166 mLine opening tool MCP: 726205
Page 35 of 42
Ver.03 22.02.05
6.10 Final assembly MCP models m10, m100 and m300
Cap, set of 5 colours; 726001
Handle collar, 726002 ID window; 726004
ID tag; 726003 Tip ejector; 726005
Calibration nut cover;
Dispensing head, please see details in the following pages
Page 36 of 42
Ver.03 22.02.05
6.11 Dispensing Head, MCP m10 Tip ejector bar locking piece; 731145
Locking bar springs; 2pcs, 731144
Tip cone housing 8ch/10; 726131 12ch/10;726134
Tip ejector bar 8ch/10; 726137 12ch/10;726139
Piston 10; 731155
Piston assembly complete 10ul; 731187 (includes both piston and tip cone complete)
O-ring support nut ; 731156 O-rings, 2pcs; 731157 Spring support; 731158 Tip cone assembly complete 10ul; 731160
Spring; 731159
mLine autoclavable grease, 20g tube: 731141
Page 37 of 42
Ver.03 22.02.05
6.12 Dispensing Head, MCP m100
Tip ejector bar locking piece; 731145
Locking bar springs; 2pcs, 731144
Tip cone housing 8ch/100; 726132 12ch/100;726135
Tip ejector bar 8ch/100; 726138 12ch/100;726140
Piston 100; 726147
Seal support nut ; 726167 Piston assembly complete 100ul; 726153 (includes piston and tip cone complete)
Seal support spring, 8 pcs; 726168 Seal support, 8pcs; 726169 Tip cone assembly complete 100ul; 726150
Seal 100/120, 8pcs 726170
mLine autoclavable grease, 20g tube: 731141
Page 38 of 42
Ver.03 22.02.05
6.13 Dispensing Head, MCP m300
Tip ejector bar locking piece; 731145
Locking bar springs; 2pcs, 731144
Tip cone housing 8ch/300; 726133 12ch/300;726136
Tip ejector bar 8ch/300; 726138 12ch/300;726140
Piston 300; 726148
Piston assembly complete 300ul; 726154 (includes piston and tip cone complete) Tip cone assembly complete 300ul; 726151 (includes also the spring)
mLine autoclavable grease, 20g tube: 731141
Page 39 of 42
Ver.03 22.02.05
6.14 List of tools and accessories 721009 Biohit forceps for filter replacement 725160 Biohit Holder for mLINE pipettor 726081 Blue cap set, 5 pieces 726200 Service kit, mLine single channel pipettors 726201 mline tool 726202 mLine 5000 tool 726203 mLine calibration tool 726204 mLine MCP tool 726206 Service kit, mLine multichannel pipettors 731141 Grease, 20g tube
Note! Please check always the spare part codes from the official price lists when ordering
Page 40 of 42
Ver.03 22.02.05
7
Warranty information
The Biohit mLINE pipettors are covered by warranty for 3 years against defects in materials and workmanship. Should your mLINE pipettor fail to function at any time, please contact your local Biohit representative. ANY WARRANTY WILL, HOWEVER, BE DEEMED AS VOID IF FAULT IS FOUND TO HAVE BEEN CAUSED BY MALTREATMENT, MISUSE, UNAUTHORIZED MAINTENANCE OF SERVICE OR NEGLIGENCE OF REGULAR MAINTENANCE AND SERVICE, ACCIDENTAL DAMAGE, INCORRECT STORAGE OR USE OF THE PRODUCTS FOR OPERATIONS OUTSIDE THEIR SPECIFIED LIMITATIONS, OUTSIDE THEIR SPECIFICATIONS, CONTRARY TO THE INSTRUCTIONS GIVEN IN THIS MANUAL OR WITH OTHER THAN THE MANUFACTURER’S ORIGINAL TIPS. Each Biohit mLINE pipettor is tested before shipping by the manufacturer. The Biohit Quality Assurance Procedure guarantees that the Biohit mLINE pipettor you have purchased is ready for use. Each Biohit mLINE pipettor is CE marked.
Page 41 of 42
Ver.03 22.02.05